This product is intended for use in the withdrawal and injection of fluids for medical purposes. This product is compatible for use with standard luer slip and luer lock syringes. Additionally, after withdrawal of the needle from the body, the attached needle safety sheath can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
Device Story
Single-use sterile hypodermic needle; used for intradermal, subcutaneous, or intramuscular fluid injection/withdrawal. Device components: needle tube, hub, safety cover, hinge, protective sheath. Operation: user attaches needle to standard luer slip/lock syringe; manually pivots safety cover away for use; after procedure, manually pivots cover over needle until locked to shield tip. Used by healthcare professionals in clinical settings. Benefits: facilitates fluid delivery while providing integrated sharps injury prevention mechanism.
Clinical Evidence
No clinical data. Bench testing only. Evaluated physical/mechanical properties (ISO 7864, ISO 9626, ISO 6009, EN ISO 8536-4), safety mechanism performance (per FDA guidance), Luer connector compatibility (ISO 80369-7), sterility (EN ISO 11137-2), and biocompatibility (ISO 10993 series including cytotoxicity, sensitization, irritation, systemic toxicity, and hemocompatibility).
Technological Characteristics
Materials: medical-grade polypropylene (PP) and SU5304 stainless steel. 6% Luer female conical connector (ISO 80369-7). Sizes: 18G-25G, 16mm-51mm lengths, Regular/Thin Wall. Sterilization: electron beam irradiation (SAL 10^-6). Manual safety mechanism; no electronic or software components.
Indications for Use
Indicated for withdrawal and injection of fluids for medical purposes in neonates, infants, children, adolescents, and adults. Compatible with standard luer slip and luer lock syringes. Includes a manually activated safety sheath to minimize accidental needlestick risk.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 7, 2026
Hangzhou Qiantang Longyue Biotechnology Co., Ltd.
% Ren Jiangfeng
Regulatory Affairs Manager
Shanghai Lingfu Technology Co., Ltd.
Rm. 1115, Block A, Xuhui Vanke Centre, # 55 Ding'An Rd., Xuhui District
Shanghai, Shanghai 200235, China
Re: K260933
Trade/Device Name: Safety Sterile Hypodermic Needles (NR-S1816R, NR-S1916R, NR-S1925R, NR-S1938R, NR-S2016R, NR-S2025R, NR-S2038R, NR-S2051R, NR-S2116R, NR-S2125R, NR-S2138R, NR-S2151R, NR-S2216R, NR-S2225R, NR-S2238R, NR-S2251R, NR-S2316R, NR-S2325R, NR-S2338R, NR-S2351R, NR-S2516R, NR-S2525R, NR-S2538R, NR-S2551R, NR-S2516T, NR-S2525T, NR-S2538T, NR-S2551T)
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: July 8, 2026
Received: July 8, 2026
Dear Ren Jiangfeng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260933 - Ren Jiangfeng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260933 - Ren Jiangfeng
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# SHRUTI N. MISTRY -S
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K260933 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Safety Sterile Hypodermic Needles (NR-S1816R, NR-S1916R, NR-S1925R, NR-S1938R, NR-S2016R, NR-S2025R, NR-S2038R, NR-S2051R, NR-S2116R, NR-S2125R, NR-S2138R, NR-S2151R, NR-S2216R, NR-S2225R, NR-S2238R, NR-S2251R, NR-S2316R, NR-S2325R, NR-S2338R, NR-S2351R, NR-S2516R, NR-S2525R, NR-S2538R, NR-S2551R, NR-S2516T, NR-S2525T, NR-S2538T, NR-S2551T) | | | |
| Please provide your Indications for Use below. | | | ? |
| This product is intended for use in the withdrawal and injection of fluids for medical purposes. This product is compatible for use with standard luer slip and luer lock syringes. Additionally, after withdrawal of the needle from the body, the attached needle safety sheath can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
510(k) #: K260933
# 510(k) Summary
Prepared on: 2026-08-07
Contact Details
21 CFR 807.92(a)(1)
Applicant Name
Hangzhou Qiantang Longyue Biotechnology Co., LTD
Applicant Address
Room 104, 201, 301, 302, Building 12, Block 1, No. 619 Wangmei Road, Linping Subdistrict, Linping District Hangzhou Zhejiang 310000 China
Applicant Contact Telephone
86-17775381549
Applicant Contact
CHEN Meijun
Applicant Contact Email
meijun_chen@nextech-x.com
Correspondent Name
Shanghai Ling Fu Technology Co., Ltd.
Correspondent Address
Room 1115, Block A, Xuhui Vanke Centre, No. 55 Ding'an Road, Xuhui District Shanghai Shanghai 200235 China
Correspondent Contact Telephone
86-21-33561520
Correspondent Contact
Ren Jiangfeng
Correspondent Contact Email
jiangfeng.ren@llins-tech.com
Device Name
21 CFR 807.92(a)(2)
Device Trade Name
Safety Sterile Hypodermic Needles (NR-S1816R, NR-S1916R, NR-S1925R, NR-S1938R, NR-S2016R, NR-S2025R, NR-S2038R, NR-S2051R, NR-S2116R, NR-S2125R, NR-S2138R, NR-S2151R, NR-S2216R, NR-S2225R, NR-S2238R, NR-S2251R, NR-S2316R, NR-S2325R, NR-S2338R, NR-S2351R, NR-S2516R, NR-S2525R, NR-S2538R, NR-S2551R, NR-S2516T, NR-S2525T, NR-S2538T, NR-S2551T)
Common Name
Hypodermic single lumen needle
Classification Name
Needle, Hypodermic, Single Lumen
Regulation Number
880.5570
Product Code(s)
FMI
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K152362
Terumo SurGuard 3 Safety Hypodermic Needle
FMI
Device Description Summary
21 CFR 807.92(a)(4)
The Safety Sterile Hypodermic Needles are single-use medical devices intended for the injection of medicinal liquids into human intradermal, subcutaneous, or intramuscular tissues. The device primarily consists of a safety cover, needle hub, needle tube, protective
{5}
sheath, and hinge.
The needle hub is equipped with a 6% Luer female conical connector that conforms to ISO 80369-7. This connector allows the device to be securely connected to a syringe equipped with a compatible 6% Luer male conical connector. The device is manufactured primarily from medical-grade polypropylene (PP) and SU5304 stainless steel.
To reduce the risk of accidental needlestick injury after use, the device incorporates a manually activated needle safety feature. Prior to removing the protective sheath, the user must manually pivot the safety cover away from the needle to allow access for use. After completion of the injection, the user must again manually pivot the safety cover over the exposed needle until it engages and locks in place, thereby fully shielding the needle. The safety mechanism is not automatically or passively activated and requires deliberate user action to engage.
The Safety Sterile Hypodermic Needles are supplied sterile and are sterilized using electron beam irradiation. The devices have a shelf life of five years. They are available in various sizes to accommodate different clinical needs, including needle gauges ranging from 18G to 25G and nominal needle lengths ranging from 16 mm to 51 mm, in both Regular Wall (RW) and Thin Wall (TW) configurations.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
This product is intended for use in the withdrawal and injection of fluids for medical purposes. This product is compatible for use with standard luer slip and luer lock syringes. Additionally, after withdrawal of the needle from the body, the attached needle safety sheath can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The Indications for Use for the subject device, Safety Sterile Hypodermic Needles, and the predicate device, Terumo SurGuard 3 Safety Hypodermic Needle (K152362), are not identical in exact wording, but they share the exact same fundamental intended use. Both devices are intended to interface with standard Luer syringes for the fluid delivery into the body, and both incorporate a manually activated safety shield to minimize the risk of accidental needlestick injuries after the procedure is complete.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device, the Safety Sterile Hypodermic Needles, has the same basic technological characteristics, design, materials, and principle of operation as the legally marketed predicate device, the Terumo SurGuard 3 Safety Hypodermic Needle (K152362). The minor dimensional differences between the subject and predicate devices merely provide healthcare professionals with a broader range of options to accommodate varying clinical needs (e.g., deeper tissue injections or varying fluid viscosities). These size variations do not alter the fundamental scientific technology, principle of operation, or intended therapeutic effect of the device. Comprehensive bench testing has demonstrated that the subject device's various dimensions perform safely and effectively. Therefore, the technological characteristics of the subject device are substantially equivalent to the predicate device and do not raise any different questions regarding safety and effectiveness.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The subject device underwent a comprehensive suite of nonclinical tests to verify its safety and effectiveness for its intended use. The performance evaluation encompassed the following broad categories:
1) Physical, Mechanical, and Chemical Performance
Testing evaluated needle appearance, dimensions, stiffness, resistance to breakage, corrosion resistance, bond strength between the hub and needle tube, lumen patency, particulate contamination, color coding, and applicable chemical properties. The testing referenced EN ISO 7864:2016, ISO 9626:2016, ISO 6009:2016, and EN ISO 8536-4:2020.
2) Needlestick Prevention Mechanism
The safety mechanism was evaluated for activation force, resistance to premature activation, strength after activation, drop impact resistance, puncture resistance, and simulated clinical use. The simulated clinical evaluation referenced FDA's Medical Devices with Sharps Injury Prevention Features — Guidance for Industry and FDA Staff.
3) Luer Connector Compatibility
The 6% Luer female connector was evaluated for liquid leakage, air leakage, stress cracking, axial separation, torque resistance, and resistance to thread overriding in accordance with ISO 80369-7:2021.
4) Sterility and Bacterial Endotoxins
Electron-beam sterilization was evaluated in accordance with EN ISO 11137-2:2015 to support a Sterility Assurance Level of 10⁻⁶. Bacterial endotoxin testing was performed according to the applicable EP/USP requirements, with an acceptance criterion of less than 20 EU per device.
5) Biological Evaluation
The biological safety evaluation was conducted in accordance with the following standards:
• ISO 10993-1:2018 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
• ISO 10993-4:2017 Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood
• ISO 10993-5:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
• ISO 10993-10:2021 Biological evaluation of medical devices — Part 10: Tests for skin sensitization
• ISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
{6}
- ISO 10993-23:2021 Biological evaluation of medical devices — Part 23: Tests for irritation
The biological evaluation included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogen, and hemocompatibility/hemolysis evaluations using a representative worst-case model.
Clinical Data: Not Applicable
The Safety Sterile Hypodermic Needles are substantially equivalent to the predicate device (Terumo SurGuard 3 Safety Hypodermic Needle). The non-clinical testing demonstrates that the device is as safe and effective as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.