Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R)
K260931 · Unimed Medical Supplies, Inc. · DQA · Apr 14, 2026 · Cardiovascular
Device Facts
Record ID
K260931
Device Name
Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R)
Applicant
Unimed Medical Supplies, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Apr 14, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Unimed Reusable Finger Clip SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only.
Device Story
Reusable finger clip sensor; monitors functional arterial oxygen saturation (SpO2) and pulse rate (PR). Input: optical signals via 660nm/905nm LEDs and photodetector; output: electrical signals transmitted to compatible host monitor (e.g., Masimo Radical 7). Principle: two-wavelength relative optical absorption. Used in clinical settings; operated by healthcare professionals. Output displayed on host monitor to assist clinical decision-making regarding patient oxygenation status. Benefits: non-invasive, continuous monitoring of patient vital signs.
Clinical Evidence
No new clinical data. Clinical accuracy testing was not repeated as electro-optical components are identical to the predicate. Prior clinical study conducted per ISO 80601-2-61 and FDA Pulse Oximeter Guidance demonstrated accuracy against CO-oximetry in adult volunteers, including darkly pigmented subjects. Bench testing only for connector change verification.
Technological Characteristics
Reusable finger clip sensor; ABS housing with biocompatible silicone pad. Dual-wavelength LED (660nm/905nm) and photodetector. Double-shielded copper cable with DB9 connector. Powered by host monitor. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 10993-1.
Indications for Use
Indicated for continuous non-invasive SpO2 and pulse rate monitoring in adult patients (>30 kg) and pediatric patients (10-50 kg). Prescription use only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 14, 2026
Unimed Medical Supplies, Inc.
% Huanyu Zeng
RA Specialist
Unimed Medical Supplies Inc.
Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, Baoan District
Shenzhen, CHN 518108
Re: K260931
Trade/Device Name:
Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R)
Regulation Number: 21 CFR 870.2700
Regulation Name: Oximeter
Regulatory Class: Class II
Product Code: DQA
Dated: March 20, 2026
Received: March 20, 2026
Dear Huanyu Zeng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260931 - Huanyu Zeng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K260931 - Huanyu Zeng
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260931 | ☐ |
| Please provide the device trade name(s). | | ☐ |
| Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R) (U403-49R);
Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R) (U103-49R) | | |
| Please provide your Indications for Use below. | | ☐ |
| Unimed Reusable Finger Clip SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☑ Children (2 years old to < 12 years old)
☑ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater) | ☐ |
{4}
A brand of Xinwell
unimed
medical supplies
# 510(K) Summary
## 1. Submitter
Date Prepared: Mar. 20, 2026
Submitter/Manufacturer: Unimed Medical Supplies Inc.
Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, Baoan District, Shenzhen, China 518108
FDA Establishment Number: 3007307487
Contact:
Zeng Huanyu
RA Specialist
Tel: +86-755 26695165
E-mail: zenghy@unimed.cn
510(k) Submission Type: This is a Special 510(k).
## 2. Proposed Device
Trade Name: Unimed Reusable Finger Clip SpO₂ Sensors (U403-49R and U103-49R)
Common Name: Oximeter Sensor
Classification:
Medical Specialty: Cardiovascular
Regulation: 21 CFR 870.2700 – Oximeter
Product Code: DQA
Class: II
## 3. Predicate Device
| 510(K) No. | Trade Name |
| --- | --- |
| K242580 | Unimed Reusable Finger Clip SpO2 Sensors |
## 4. Device description
The subject devices, Unimed Reusable Finger Clip SpO₂ Sensors, are fully compatible reusable sensors for use with monitor that incorporates Masimo technology. These sensors are supplied non-sterile.
The subject sensors consist of a plug/connector, a cable, and a patient-contacting (finger clip) with LED and photodetector.
The compatibility of the sensors is validated with specific monitor, as shown below:
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
{5}
A brand of Xinwell
unimed
medical supplies
| Subject device | Compatible monitor |
| --- | --- |
| U403-49R | Masimo Radical 7 |
| U103-49R | |
Compared with the predicate devices (U403-125 and U103-125, cleared under K242580), the subject devices (U403-49R and U103-49R) have the following change:
- Change in interface connector only. All other specifications including LED, photodetector, and patient populations are identical to the predicate.
The purpose of this Special 510(k) submission is to receive clearance to market U403-49R and U103-49R, which differ from the predicate devices (K242580) only in connector type.
# 5. Intended use/Indications for use
Unimed Reusable Finger Clip SpO₂ Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only.
# 6. Comparison to predicate device
The subject devices, U403-49R and U103-49R, and the predicate devices, Unimed Reusable Finger Clip SpO2 Sensors (K242580), are compared in the following substantial equivalence table.
| Feature | Subject Device
(Model: U403-49R, U103-49R) | Predicate Device
(K242580) | Comparison to Predicate Device |
| --- | --- | --- | --- |
| General Information | | | |
| Device name | Unimed Reusable Finger Clip SpO2 Sensors | Unimed Reusable Finger Clip SpO2 Sensors | Identical |
| Classification Regulation/ Product Code | 21 CFR 870.2700, Class II/DQA | 21 CFR 870.2700, Class II/DQA | Identical |
| Intended use/Indications for use | Unimed Reusable Finger Clip SpO2 Sensors are indicated for continuous | Unimed Reusable Finger Clip SpO2 Sensors are indicated for continuous | Identical |
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
{6}
A brand of Xinwell
unimed
medical supplies
| Feature | Subject Device
(Model: U403-49R, U103-49R) | Predicate Device
(K242580) | Comparison to Predicate Device |
| --- | --- | --- | --- |
| | non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only. | non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only. | |
| Principle of operation | Two-wavelength relative optical absorption | Two-wavelength relative optical absorption | Same |
| Key performance specifications/ characteristics | | | |
| Intended patient population | Adult (>30 kg) / Pediatric (10-50 kg) | Adult (>30 kg) / Pediatric (10-50 kg) | Identical |
| Intended application site | Finger | Finger | Identical |
| Prescription or OTC | Rx Only | Rx Only | Identical |
| Use type | Reusable | Reusable | Identical |
| Sensor Structure Composition | DB9 connector, cable, LED & photodetector, finger clip | Card-edge connector, cable, LED & photodetector, finger clip | Different connector type only (DB9 vs. card-edge). All other components identical. |
| Raw material | Connector: DB9 connector
Cable: Double-shielded copper conductor in TPU jacket
LED: 660nm/905nm
Finger clip: ABS Silicone pad: Biocompatible silicone | Connector: Card-edge connector
Cable: Double-shielded copper conductor in TPU jacket
LED: 660nm/905nm
Finger clip: ABS Silicone pad: Biocompatible silicone | Equivalent |
| Performance (Arms) | | | |
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn
E-mail: sales@unimed.cn
{7}
A brand of Xinwell
unimed
medical supplies
| Feature | Subject Device
(Model: U403-49R, U103-49R) | Predicate Device
(K242580) | Comparison to Predicate Device |
| --- | --- | --- | --- |
| Saturation Accuracy, No Motion (70-100%) | ±3% | ±3% | Identical |
| Pulse Rate Accuracy, No Motion | ±3 bpm (30-250 bpm) | ±3 bpm (30-250 bpm) | Identical |
| Environmental | | | |
| Operating Temperature | 5 to 40 °C | 5 to 40 °C | Identical |
| Operational/Storage Humidity | 10 to 85% | 10 to 85% | Identical |
| Biocompatibility | Pass ISO 10993 cytotoxicity, skin irritation and skin sensitivity tests | Pass ISO 10993 cytotoxicity, skin irritation and skin sensitivity tests | Identical |
| Energy source | Monitor power supply | Monitor power supply | Identical |
## 7. Verification and validation testing
### Non-clinical test data
This submission is a Special 510(k). The proposed device change consists of a connector modification; the key electro-optical components and patient-contacting materials remain unchanged from the predicate device. Accordingly, biocompatibility and clinical performance evaluations were not repeated, as these assessments remain valid and applicable to the subject devices.
Non-clinical verification testing was conducted to confirm that the proposed connector change does not adversely affect device safety or performance and that the Unimed Reusable Finger Clip $\mathrm{SpO}_2$ Sensors (models U403-49R and U103-49R) remain substantially equivalent to the predicate devices. Testing was performed in conformance with the following recognized standards:
- IEC 60601-1 Medical electrical equipment — Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020)
- IEC 60601-1-2 Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests (IEC 60601-1-2:2014+AMD1:2020)
- ISO 80601-2-61 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
{8}
A brand of Xinwell
unimed medical supplies
(ISO 80601-2-61:2017, COR1:2018)
Verification testing addressed risks associated with the connector change, including electromagnetic compatibility, non-clinical SpO₂ and pulse rate performance (simulated use), mechanical durability (bending and drop), mechanical integrity (tensile strength), and electrical safety (dielectric strength). All tests passed their respective acceptance criteria, confirming that device safety and performance are maintained.
## Biocompatibility
Biocompatibility evaluation was not repeated for this Special 510(k) submission. The patient-contacting components — the finger clip housing and silicone pad — are identical to those of the predicate device. These components involve contact with intact skin and are classified as prolonged exposure per ISO 10993-1:2018. Prior biocompatibility testing on the predicate device addressed all relevant endpoint assessments, including cytotoxicity (ISO 10993-5), skin sensitization (ISO 10993-10), and skin irritation (ISO 10993-23). As the patient-contacting materials are unchanged, this data remains applicable to the subject devices.
## Clinical Performance Data
Clinical accuracy testing was not repeated for this Special 510(k) submission. The electro-optical components governing SpO₂ measurement accuracy are unchanged from the predicate device. The prior clinical study, conducted under an approved protocol with subject informed consent per ISO 80601-2-61 and FDA's Pulse Oximeter Guidance, demonstrated the accuracy of the U403-49R against arterial oxygen saturation (SaO₂) measured by CO-oximetry in human adult volunteers including a sufficient proportion of darkly pigmented subjects. This clinical data remains applicable to the subject devices.
## 8. Substantial Equivalence Statement
Based on the comparison, analysis, and the submitted verification and validation data, Unimed believes that the Unimed Reusable Finger Clip SpO₂ Sensors are as safe and effective and are substantially equivalent to the predicate devices.
Add: Bld#8, Nangang 3rd Industrial Park, Tangtou, Shiyan, 518108 Shenzhen, People's Republic of China
http://www.unimed.cn E-mail: sales@unimed.cn
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.