Multi-center registry study; Concurrent control population from routine clinical practice
The registry study evaluated the rate of catheter balloon injury (CBI) and device accuracy (sensitivity/specificity) in a real-world clinical setting, comparing the TUCSV device arm against a concurrent control population where the device was not used.
Post-market, 3-month prospective, multi-center, open label, single-arm registry study; Prospective, multi-center, open-label, single-arm registry study with a concurrent control population; Follow-up/Duration: 3 months
Adult male or female (Age ≥18) patients requiring placement of a urinary catheter; Sample Size: 1,014 patients (ITT population); Number of Sites: 3
Concurrent control population (patients from clinical areas where the TUCSV device was not made available)
Rate of catheter balloon injury (CBI); Device accuracy (sensitivity and specificity)
Indications for Use
The TUC Safety Valve is indicated for single use during inflation of a Foley catheter retaining balloon in adult patients. The TUCSV alerts the user of premature inflation of the balloon within the urethra and alleviates the pressure, to aid in the prevention of incorrect balloon placement.
Device Story
TUCSV is a sterile, single-use accessory placed between an inflation syringe and a Foley retention balloon catheter using standard Luer connections. It functions as a mechanical safety valve; it incorporates a flow restrictor to prevent rapid inflation and a pressure relief valve that expels inflation fluid if the balloon is inadvertently inflated within the urethra. The device is operated by clinicians during urological procedures. By venting fluid upon detecting high-pressure resistance characteristic of urethral placement, the device alerts the user to misplacement and prevents balloon inflation in the urethra. This mechanism reduces the risk of catheter balloon injury (CBI) and associated trauma, such as hematuria or severe pain, providing a safety safeguard during catheterization.
Clinical Evidence
Prospective, multi-center, open-label, single-arm registry study (n=1,014 ITT; 911 PP). Primary endpoint: Catheter Balloon Injury (CBI) rate was 0.0% in the TUCSV arm. Secondary endpoints: Sensitivity 100%, Specificity 99.7%, PPV 83.3%, NPV 100%, Accuracy 99.7%. Concurrent control group (n=4,648) showed 0.73% CBI rate. GLP animal study confirmed no significant urethral tissue response or histological abnormalities compared to controls.
Technological Characteristics
Materials: Medical grade plastic and silicone. Principle: Mechanical pressure relief valve and flow restrictor. Connectivity: None (standalone accessory). Sterilization: Radiation (ISO 11137-1/2). Form factor: Inline Luer-connected accessory. Standards: ISO 80369-7 (connectors), ASTM F1980-21 (aging), ASTM D4332-22 (conditioning), ASTM D4169-22 (distribution), ASTM F1929-15 (dye penetration), ASTM F88-23 (seal strength), ISO 11607-1:2019 (sterile barrier).
Indications for Use
Indicated for adult patients (male/female) undergoing Foley catheterization. Used to prevent urethral injury from premature/incorrect inflation of the retention balloon within the urethra.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}

August 5, 2026
Class Medical, Ltd.
% David Worrell
Regulatory Consultant
Rqm+
5000 Centregreen Way
Suite 100
Cary, North Carolina 27513
Re: K260919
Trade/Device Name: Transurethral Catheter Safety Valve (TUCSV) (80050)
Regulation Number: 21 CFR 876.5130
Regulation Name: Urological Catheter and accessories
Regulatory Class: II
Product Code: KNY
Dated: July 3, 2026
Received: July 6, 2026
Dear David Worrell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260919 - David Worrell
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260919 - David Worrell
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSICA K. NGUYEN - S
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260919 | ? |
| Please provide the device trade name(s). | | ? |
| Transurethral Catheter Safety Valve (TUCSV) (80050) | | |
| Please provide your Indications for Use below. | | ? |
| The TUC Safety Valve is indicated for single use during inflation of a Foley catheter retaining balloon in adult patients. The TUCSV alerts the user of premature inflation of the balloon within the urethra and alleviates the pressure, to aid in the prevention of incorrect balloon placement. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
CLASS MEDICAL
Simplicity that works...
Page 1 of 6
K260919
Transurethral Catheter Safety Valve (TUCSV)
# 510(K) SUMMARY
# A. Applicant Name, Address, Telephone
Applicant name
Class Medical Ltd.
Applicant address
Unit 1D Annacotty Business Park
Annacotty Co. Limerick, Ireland
Applicant phone
+353 85 2823340
# B. Applicant Contact
Mr. Danny Forde
Chief Executive Officer
dforde@classmedical.ie
# C. Primary Correspondent/Consultant
David Worrell
Regulatory Consultant
RQM+
dworrell@rqmplus.com
# D. Date Prepared
August 5, 2026
# E. Device Name
Device trade name:
Transurethral Catheter Safety Valve (TUCSV)
(80050)
Device common name:
Catheter accessory
Classification name:
Urological catheters and accessories
Device class:
Class II
Classification regulation
21 CFR 876.5130
Product Codes:
KNY
# F. Predicate Device
Signal Catheter K182635. The predicate device has not been subject to a design-related recall.
# G. Device Description
The TUCSV Device is a sterile, single-use accessory that is used in conjunction with a Foley balloon-type catheter during urological procedures. The TUCSV is placed between the inflation syringe and the retention balloon catheter using standard Luer connections. The device contains a flow restrictor to minimize rapid inflation of the retention balloon. It incorporates a pressure relief valve to expel inflation fluid which indicates misplacement of the retention balloon within the urethra. This aids in the prevention of incorrect balloon placement (e.g., if the balloon is inadvertently inflated before entering the bladder).
{5}
CLASS MEDICAL
Simplicity that works...
Page 2 of 6
K260919
Transurethral Catheter Safety Valve (TUCSV)
# H. Indications for Use
The TUC Safety Valve is indicated for single use during inflation of a Foley catheter retaining balloon in adult patients. The TUCSV alerts the user of premature inflation of the balloon within the urethra and alleviates the pressure, to aid in the prevention of incorrect balloon placement.
# I. Comparison of Technological Characteristics
The subject device Transurethral Catheter Safety Valve is substantially equivalent to the predicate device with respect to same intended use, and similar indications for use, technology, materials of construction, and performance. Table 1 summarizes the comparison between the subject device and the predicate device.
Table 1. Comparison of Subject Device and Predicate Device
| Features | Class Medical Ltd. TUCSV Device | Safe Medical Design Signal Catheter K182635 |
| --- | --- | --- |
| | Subject device | Predicate device |
| Device Classification | 21 CFR 876.5130 | |
| Common Name | Urological Catheter and Accessories | |
| Indications for Use | The TUC Safety Valve is indicated for single use during inflation of a Foley catheter retaining balloon in adult patients. The TUCSV alerts the user of premature inflation of the balloon within the urethra and alleviates the pressure, to aid in prevention of incorrect balloon placement. | The Signal Catheter is indicated for urological bladder drainage, with a maximum patient indwelling time of 29 days. |
| Technology - Mechanism of Action for Pressure Relief | The TUCSV is positioned between the inflation syringe and the balloon catheter using standard Luer connections. It uses a pressure relief valve to indicate pressurization and misplacement of the retention balloon within the urethra. Inflation fluid is expelled from the valve to alleviate the pressure and inflation of the retention balloon. | The signal balloon is an integral component and positioned in the catheter proximal hub. It is designed to inflate during excessive pressure in the retention balloon when it is constricted or misplaced within the urethra. The signal balloon inflates to alleviate the fluid inflation and resulting pressure in the retention balloon. |
| Device Activation | Occurs when the retention balloon is constricted within the urethra when it cannot be safely inflated at the nominal inflation parameters of the catheter. | |
| Inflation Fluid Flow Restriction | Flow rate is regulated by integrated restrictor screw to minimize rapid inflation of the retention balloon. | Flow rate is user controlled via force of syringe depression for retention balloon inflation. |
{6}
CLASS MEDICAL
Simplicity that works...
Page 3 of 6
K260919
Transurethral Catheter Safety Valve (TUCSV)
| Features | Class Medical Ltd. TUCSV Device | Safe Medical Design Signal Catheter K182635 |
| --- | --- | --- |
| | Subject device | Predicate device |
| Patient Contact | No. TUCSV is placed between the syringe and catheter and is external to the patient. | Yes. Balloon catheter inserted into the urethra and bladder. The auxiliary signal balloon is external to the patient. |
| Material | Medical Grade Plastic and Silicone | Silicone |
| Sterile Single Use | Yes | Yes |
The subject device and predicate device have the same intended use. The subject device and predicate device are sterile, single-use devices intended to prevent injury from accidental inflation of the retention balloon in the urethra.
The subject device and predicate device use different technological characteristics to perform the same general purpose to alert the user to inadvertent retention balloon inflation when the balloon is misplaced within the urethra. The TUCSV device has a valve external to the balloon catheter that releases excess inflation fluid when the retention balloon is inflated in the urethra. The predicate device contains an external balloon within the catheter hub that inflates with fluid when the retention balloon is inflated in the urethra. When the subject device and predicate device are used as intended, they achieve the same general purpose and these differences in technological characteristics do not raise different questions of safety and effectiveness.
### J. Summary of Non-Clinical Testing
Bench testing of the subject device met all predetermined acceptance criteria, and included:
- Dimensional verification testing
- Compatibility with commercially available fluid-filled syringes and Foley-type catheters
- Pressure-release testing
- Retention balloon flow-rate testing
- Pull-tab force testing
- Valve repeatability testing
- Luer connection testing per FDA-recognized standard ISO 80369-7 Second edition 2021-05 – Small bore connectors for liquids and gases in healthcare applications
Shelf-life testing was conducted using the following FDA-recognized standards:
- Accelerated Aging: ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- Environmental Conditioning: ASTM D4332-22 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing
- Simulated Distribution ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
- Sterile Barrier Integrity: Dye Penetration ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
{7}
CLASS MEDICAL
Simplicity that works...
Page 4 of 6
K260919
Transurethral Catheter Safety Valve (TUCSV)
- Seal Strength Testing: ASTM F88-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
The TUCSV device is provided STERILE and for single use only. The sterilization method is radiation (R). A sterilization validation was performed according to the following FDA-recognized standards:
- ISO 11137-1:2006/(R)2015 Sterilization of healthcare products – Radiation Part 1
- ISO 11137-2 Third edition 2013-06 [Including AMD1:2022] – Sterilization of healthcare products – Radiation Part 2
Sterile barrier systems were evaluated for performance and stability in accordance with ISO 11607-1:2019 Requirements for materials, sterile barrier systems, and packaging systems.
Usability Engineering FDA-recognized standards IEC 62366-1:2015/A1:2020 Medical Devices – Part 1: Application of usability engineering to medical devices processes and FDA Guidance Applying Human Factors and Usability Engineering to Medical Devices, dated February 3, 2016, were used to plan and conduct a usability study for the TUCSV device. Specific use scenarios and tasks were identified, and an analysis of hazards and risks was conducted to determine safety risks associated with use of the subject device. All critical tasks were evaluated by the human factors study. The usability study was completed successfully. No changes to the device design or device labeling were considered necessary based on results of the usability factors study.
A GLP animal study compared urethral tissue response from inflation of the anchoring balloon of a urinary catheter, to demonstrate that the activation pressures of the device do not cause urethral damage. No macroscopic or histological abnormalities were observed between the two groups. Histomorphometry results showed urethral wall thickness was comparable for the two groups. Results from the animal study confirmed no significant urethral tissue response at anchoring balloon inflations in the urethra at clinically relevant high pressure compared to low pressure in the animal model.
### K. Summary of Clinical Testing
A post-market, 3-month prospective, multi-center, open label, single-arm registry study was conducted in Ireland and the United Kingdom. A concurrent control population was included in the study which included patients from clinical areas at three study sites where the TUCSV device was not made available. The objective of the study was to determine the safety and effectiveness of the TUCSV when used in conjunction with a urethral catheter. The patient population included adult male or female (Age ≥18) patients with no known pre-existing urethral anomaly requiring the placement of a urinary catheter who provided consent to participate in the study.
The intention to treat (ITT) population included 1,014 patients, including 839 males and 175 females. The per protocol (PP) population included 911 patients, including 739 males and 172 females. A post-hoc analysis was conducted to evaluate the following endpoints:
1. Primary Safety and Efficacy Endpoint: Rate of catheter balloon injury (CBI). Analysis was performed on the intent-to-treat population: 0.0%
Suspected catheter balloon injuries were confirmed by a Urologist in an examination immediately following the catheterization procedure.
{8}
CLASS MEDICAL
Simplicity that works...
Page 5 of 6
K260919
Transurethral Catheter Safety Valve (TUCSV)
CBI was defined as failure to complete urethral catheterization after balloon inflation with associated failure to drain the bladder normally and severe pain in awake patients, usually accompanied by urethral bleeding/hematuria.
2. Secondary Endpoint: Device Accuracy Performance. Device Accuracy was analyzed as sensitivity and specificity of the device in indicating to the user that the balloon is placed correctly or incorrectly. Analysis was performed on the per protocol population only. Accuracy was calculated as follows:
- Sensitivity of the device to incorrect catheter placement – Percent rate of TUCSV vents correctly when the balloon is inflated in the urethra: 100%
- Specificity of the device to correct placement – Percent rate of TUCSV not venting correctly when the balloon is inflated in the bladder: 99.7%
- Percent of inappropriate venting (false positive): 0.3%
- Percent of failed venting when balloon was inflated in the urethra (false negative): 0.0%
- Positive predictive value (PPV) - the proportion of venting events in which the balloon was in fact malpositioned. Calculated as true positives divided by all venting events (true positives plus false positives): 83.3%
- Negative predictive value (NPV) - the proportion of non-venting procedures in which the balloon was in fact correctly placed. Calculated as true negatives divided by all non-venting procedures (true negatives plus false negatives): 100%
- Accuracy - the proportion of all procedures in which the device correctly indicated balloon position, whether by venting or not venting. Calculated as true positives plus true negatives, divided by the total number of procedures: 99.7%
The study procedure achieved a 100% success rate for the TUCSV arm where catheterization was successful. The observed vent rate was 1.97% in the TUCSV arm. No CBIs were reported in the ITT population (TUCSV arm). No CBIs were reported in the PP population (TUCSV arm). In the concurrent control population, against an estimated 4,648 catheterizations over the 3-month monitoring period, the CBI was 1.24% in males, 0.22% in females and 0.73% overall.
No urethral injury occurred with the TUCSV. One adverse event was reported which met the definition of a serious adverse event but was classified unrelated to the device or procedure. Three pain events were reported at venting; all were deemed as normal procedural discomfort anticipated with successful catheterization.
24 protocol deviations were identified, all in the TUCSV arm, where the case report form incorrectly documented 'No' or 'NA' to the question "Was the TUCSV used as per the instructional video?". All 24 procedures were successful.
L. Conclusion
{9}
CLASS MEDICAL
Simplicity that works
Page 6 of 6
K260919
Transurethral Catheter Safety Valve (TUCSV)
Non-clinical performance testing results demonstrate the subject device meets established specifications, met pre-determined acceptance criteria and conforms with FDA-recognized standards for medical devices. The subject device performs as well as the predicate device for its intended use and supports a determination of substantial equivalence.
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.