K260918 · Amann Girrbach AG · EIH · Jun 29, 2026 · Dental
Device Facts
Record ID
K260918
Device Name
Zolid Lunaris
Applicant
Amann Girrbach AG
Product Code
EIH · Dental
Decision Date
Jun 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Zolid Lunaris is a zirconia ceramic CAD/CAM blank (type II, class 5 according to ISO 6872) for manufacturing fixed dental prostheses to restore tooth anatomy, physiology and aesthetics in (partially) edentulous patients.
Device Story
Zolid Lunaris is a zirconia ceramic CAD/CAM blank used by dental technicians and dentists to fabricate fixed dental prostheses; restores tooth anatomy, physiology, and aesthetics. Device is a disc-shaped blank (3Y-TZP basal zone for stability; 4Y-TZP incisal zone for optics) processed via CAD/CAM milling and sintering (1500°C for 2 hours). Final restoration is placed intraorally to replace missing or damaged teeth. Benefits include high flexural strength (≥800 MPa) and fracture toughness (≥4.0MPa√m) for long-term intraoral use.
Clinical Evidence
Bench testing only. Performance testing per ISO 6872 confirmed flexural strength ≥800 MPa and chemical solubility <100 μg/cm². Biocompatibility testing per ISO 10993-5 (cytotoxicity) and ISO 10993-18 (extractables) passed. Packaging, stability, and shelf-life (4 years) were validated.
Technological Characteristics
Zirconia ceramic (Y-TZP/4Y-TZP) CAD/CAM blank. Type II, Class 5 per ISO 6872. Flexural strength ≥800 MPa; E-module ≥200 GPa; Vickers hardness 1300±200 HV10. Supplied non-sterile. Sintering: 1500°C for 2 hours. Dimensions: 98mm diameter, 12-30mm height. Biocompatibility verified per ISO 10993.
Indications for Use
Indicated for (partially) edentulous patients requiring replacement of missing or damaged teeth via single or multi-unit dental restorations. Contraindications: insufficient tooth structure, known material allergies/intolerances, periapical inflammation, periodontal disease, insufficient root fillings, or other non-listed indications.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 29, 2026
Amann Girrbach AG
% Oliver Eikenberg
Global QA/RA Manager
Pure Global
111 Town Sq. Pl.
Suite 1203
Jersey City, New Jersey 07310
Re: K260918
Trade/Device Name: Zolid Lunaris
Regulation Number: 21 CFR 872.6660
Regulation Name: Porcelain Powder For Clinical Use
Regulatory Class: Class II
Product Code: EIH
Dated: March 19, 2026
Received: March 19, 2026
Dear Oliver Eikenberg:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260918 - Oliver Eikenberg
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260918 - Oliver Eikenberg
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**MICHAEL E. ADJODHA -S**
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260918
Device Name
Zolid Lunaris
Indications for Use (Describe)
Zolid Lunaris is a zirconia ceramic CAD/CAM blank (type II, class 5 according to ISO 6872) for manufacturing fixed dental prostheses to restore tooth anatomy, physiology and aesthetics in (partially) edentulous patients.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
# 510(k) Summary (traditional)
**Zolid Lunaris**
**K260918**
**Date Prepared:** 26 June 2026
## 510(k) Owner / Applicant Submission Sponsor
Amann Girrbach AG
Gewerbestrasse 10
6841 Maeder , AUSTRIA
Contact: Lisa Fassing, Senior Manager Regulatory Affairs – Registrations
Phone: +49723195701782
## Submission Correspondent
Pure Vision AI, Inc (dba Pure Global)
111 Town Square Place, Suite 1203
Jersey City, NJ 07310, USA
Contact: Oliver Eikenberg, PhD, Senior Consultant, Global Quality & Regulatory Affairs Manager,
Phone +4906971047787
## Device Information
Trade Name: Zolid Lunaris
Common Name: Powder, Porcelain
Regulation Number: 21 CFR 872.6660
Regulation Name: Porcelain powder for clinical use
Product Code: EIH
Classification: Class II
Review Panel: Dental
## Legally Marketed Predicate Device
Primary Predicate Device Name Ceramill® Zolid HT+ white
510(k) Number: K171876
Manufacturer: Amann Girrbach AG
Regulation Number: 21 CFR 872.6660
Regulation Name: Porcelain powder for clinical use
Product Code: EIH
Classification: Class II
Review Panel: Dental
{5}
## Device Description
Zolid Lunaris is a zirconia ceramic CAD/CAM blank (type II, class 5 according to ISO 6872) for manufacturing fixed dental prostheses to restore tooth anatomy, physiology and aesthetics in (partially) edentulous patients. The fixed dental prosthesis is intended for single use.
The indications are for Replacement of missing or damaged teeth by means of
- Single tooth dental restorations
- Multi-unit dental restorations
Zolid Lunaris disc-shaped blanks are available in 7 different sizes (Ø x height in mm): 98 x 12, 98 x 14, 98 x 16, 98 x 18, 98 x 20, 98 x 25, 98 x 30.
The zirconia ceramic CAD/CAM blank is supplied non-sterile.
The device achieves its intended medical purpose through its mechanical and material properties – namely high flexural strength (≥800 MPa) and fracture toughness (≥4.0MPa√m) according to ISO 6872 – allowing for long-term intraoral use. Due to its mechanical properties (Y-TZP ZrO₂), Zolid Lunaris is classified as Type II, Class 5 ceramic according to ISO 6872.
Once the dental restoration is fabricated and placed in the patient's mouth, it functions to restore the tooth anatomy, function, and aesthetics in patients with missing or damaged teeth while integrating harmoniously with the surrounding tissues. The 3Y-TZP basal zone provides mechanical stability, while the 4Y-TZP incisal zone enhances optical properties, offering an optimized balance between strength and aesthetics.
## Substantial Equivalence Discussion –
| FEATURE | SUBJECT DEVICE | PREDICATE DEVICE | COMPARISON / COMMENT |
| --- | --- | --- | --- |
| Manufacturer | Amann Girrbach AG | Amann Girrbach AG | |
| Trade Name | Zolid Lunaris | Ceramill® Zolid HT+ white | |
| INTENDED USE COMPARISON | | | |
| 510(k) Number | K260918 | K171876 | N/A |
| Product Code | EIH | EIH | same |
| Regulation Number | 21 CFR 872.6660 | 21 CFR 872.6660 | same |
| Regulation Name | Porcelain powder for clinical use | Porcelain powder for clinical use | same |
| Indications for Use | Zolid Lunaris is a zirconia ceramic CAD/CAM blank (type II, class 5 according to ISO 6872) for manufacturing fixed dental prostheses to restore tooth anatomy, physiology and aesthetics in (partially) edentulous patients. | Zirconium-oxide blanks for permanent and removable dental prosthetics. | same |
{6}
| FEATURE | SUBJECT DEVICE | PREDICATE DEVICE | COMPARISON / COMMENT |
| --- | --- | --- | --- |
| **Manufacturer** | **Amann Girrbach AG** | **Amann Girrbach AG** | |
| **Trade Name** | **Zolid Lunaris** | **Ceramill® Zolid HT+ white** | |
| Conditions of Use | Replacement of missing or damaged teeth by means of -Single tooth dental restorations -Multi-unit dental restorations | -Anatomically reduced crown and bridge frames in the anterior and posterior tooth range, and monolithic (fully anatomical) crowns and bridges; -Anatomically reduced four to multi-unit bridge frames with a maximum of three connected intermediate units in the anterior region and a maximum of two connected intermediate links in the posterior region; -Monolithic four to multi-unit bridges with a maximum of three connected intermediate units in the anterior region and a maximum of two connected intermediate links in the posterior region; -Cantilever frames and bridges with a maximum of one bridge unit (maximum one free-end pontic and no further than the second premolar). | same |
| Contraindications | -Insufficient tooth structure -Known allergies and intolerances to the material in the medical device -Periapical inflammation or existing periodontal disease -Insufficient root fillings -All indications that are not listed under 'Indications' | -Insufficient tooth-structure availability -Insufficient preparation results -Insufficient oral hygiene -More than two connected bridge units in the posterior region, more than three connected intermediate units in the anterior region -Known incompatibilities with respect to the components -Heavily discolored hard tooth structure -Provisional insertion | same |
| Intended User | Professionally trained dental technicians and dentists. | The product may only be used by trained dental technicians. | same |
| Prescription Use | Yes | Yes | same |
| **TECHNICOLOGICAL CHARACTERISTIC COMPARISON** | | | |
{7}
| FEATURE | SUBJECT DEVICE | PREDICATE DEVICE | COMPARISON / COMMENT |
| --- | --- | --- | --- |
| Manufacturer | Amann Girrbach AG | Amann Girrbach AG | |
| Trade Name | Zolid Lunaris | Ceramill® Zolid HT+ white | |
| Physical Form / State | Solid | Solid | same |
| Chemical Composition | ZrO₂ + HfO₂ + Y₂O₃ ≥ 99.0 wt. % Y₂O₃: 6.3 – 7.2 wt. % HfO₂: ≤ 5.0 wt. % Al₂O₃: ≤ 0.01 wt. % Others oxides: ≤ 1.0 wt. % | ZrO₂ + HfO₂ + Y₂O₃ ≥ 99.0 wt. % Y₂O₃: 6.7 – 7.2 wt. % HfO₂: ≤ 5.0 wt. % Al₂O₃: ≤ 0.5 wt. % Others oxides: ≤ 1 wt. % | similar |
| Shapes | Disks (round-shape) | Disks (« U »-shape or round-shape) | similar |
| Dimensions | 98 mm of diameter, in seven different heights (12, 14, 16, 18, 20, 25, 30 mm). | “U”-shape: 90 x 72, in eight different heights (10, 12, 14, 16, 18, 20, 25, 30 mm). - Round-shape: 98 mm of diameter, in eight different heights (10, 12, 14, 16, 18, 20, 25, 30 mm). | similar |
| Shades | 16 VITA shades (A1 - D4) and four bleach shades (BL0, BL1, BL2, BL3) | white | different |
| Supplied sterile | Non-sterile. | Non-sterile. | same |
| Sintering Temperature and Duration | 1500 °C – 2 hours | 1450 °C – 2 hours | similar - Both meet ISO 6872:2024 |
| Shelf life | 4 years | 5 years | similar |
| Flexural Strength | ≥800 MPa acc. to ISO 6872 | 1100 ± 150 MPa acc. to ISO 6872 | similar - Both meet ISO 6872:2024 minimum average strength of 800 MPa thus, both are classified as dental ceramics of Type II, Class 5. |
| E-module | ≥200 GPa | ≥ 200 GPa | same |
| Thermal expansion coefficient (CTE) (25-500°C) | 10.5±0.5 (10⁻⁶/K) | 10.4 ± 0.5 (10⁻⁶/K) | similar |
| Chemical Solubility | <100 μg/cm² acc. to ISO 6872 | <100 μg/cm² acc. to ISO 6872 | same |
| Vickers hardness | 1300±200 HV10 | 1300 ± 200 HV10 | same |
| Final Density | ≥ 6.05 g/cm³ | ≥ 6.05 g/cm³ | same |
| Porosity (%) | 0 (no open porosity) | 0 (no open porosity) | same |
| Grain Size | ≤ 0.7 μm | ≤ 0.6 μm | similar |
| Radioactivity | ≤ 1.0 Bq/g | ≤ 1.0 Bq/g | same |
| Biocompatibility | Verified by cytotoxicity testing according to ISO 10993-5 and chemical analysis for organic and inorganic extractables according to ISO 10993-18. | Verified by cytotoxicity testing according to ISO 10993-5 and chemical analysis for organic and inorganic extractables according to ISO 10993-18. | same |
{8}
Any minor differences do not raise new questions regarding safety or effectiveness.
The following testing and evaluations demonstrate that the subject device performs safely and effectively:
- Performance Testing acc. to ISO 6872: Meets the ISO 6872 standard for type II, class 5 zirconia ceramics (flexural strength ≥800 MPa, chemical solubility <100 μg/cm²).
- Biocompatibility acc. to ISO 10993: Passed ISO 10993 tests with no cytotoxic effects, confirming its safety for use in dental applications.
- Packaging: Validated to ensure packaging stability, protecting the product during storage and transit for the defined shelf life.
- Stability and Shelf-Life: 4 years, with no degradation or adverse effects expected.
## Conclusion
Based on the comparison of intended use, technological characteristics, and performance data, the subject device Zolid Lunaris is substantially equivalent to the predicate device Ceramill® Zolid HT+ white (K171876). Differences in chemical composition, shapes, dimensions, shades, sintering temperature and duration, shelf life, flexural strength, thermal expansion coefficient (CTE) and grain size do not raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.