The DYNEX® External Fixation Systems are intended for external fixation with the following indications: Stabilization of Fractures & Osteotomy Rear and Mid-foot Arthrodesis Adult and Pediatric Leg Lengthening Correction of Bony or Soft Tissue Defects or Deformities in Upper & Lower Extremities Pseudoarthrosis or Non-Union of Long Bones The DYNEX® TRIAD - TTT Fixation System is intended for external fixation in the treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities via tibial bone transport.
Device Story
DYNEX External Fixation Systems comprise modular orthopedic hardware including rings, partial rings, footplates, rails, struts, half pins, transport pins, and guidewires; used for external skeletal fixation. System facilitates bone stabilization, lengthening, and transport. Operated by surgeons in clinical settings; frame attached to bone via pins and wires. TRIAD-TTT sub-system includes linear bone adjustment mechanism allowing up to 10mm transverse bone transport. Devices provide mechanical support for bone healing and deformity correction. Components are single-use; instruments are reusable and require sterilization.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing including static axial strength, slipping torque, four-point bending, and torsional strength per ASTM F1541-24, alongside biological evaluation per ISO 10993-1:2018 and steam sterilization validation.
Technological Characteristics
Metallic bone fixation system; components include rings, rails, struts, and pins. Materials are consistent with predicate devices. Mechanical fixation via tensioned wires and half pins. Features linear bone adjustment mechanism for transport. Sterilization via steam. Testing performed per ASTM F1541-24.
Indications for Use
Indicated for adult and pediatric patients requiring stabilization of fractures and osteotomies, rear and mid-foot arthrodesis, leg lengthening, correction of bony or soft tissue defects or deformities in upper and lower extremities, and treatment of pseudoarthrosis or non-union of long bones. The TRIAD-TTT system is specifically indicated for tibial bone transport.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Dynex Ring Fixation System. Diametrix Ring Fixation System (K202054)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 17, 2026
Vilex, LLC
Brock Johnson
President
111 Moffitt Street
McMinnville, TN 37110
Re: K260900
Trade/Device Name: DYNEX® External Fixation Systems
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Codes: KTT, JDW
Dated: March 17, 2026
Received: March 18, 2026
Dear Brock Johnson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260900 - Brock Johnson
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for
{2}
K260900 - Brock Johnson
Page 3
more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G. Allen
Digitally signed by Peter G. Allen
Date: 2026.04.17 13:31:42
04'00"
for Lixin Liu, Ph.D.
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260900 | ? |
| Please provide the device trade name(s). | | ? |
| DYNEX® External Fixation Systems | | |
| Please provide your Indications for Use below. | | ? |
| The DYNEX® External Fixation Systems are intended for external fixation with the following indications:
• Stabilization of Fractures & Osteotomy
• Rear and Mid-foot Arthrodesis
• Adult and Pediatric Leg Lengthening
• Correction of Bony or Soft Tissue Defects or Deformities in Upper & Lower Extremities
• Pseudoarthrosis or Non-Union of Long Bones
The DYNEX® TRIAD - TTT Fixation System is intended for external fixation in the treatment of non-union or
pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities via tibial bone
transport. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260900 (Page 1 of 2)
| 510(k) #: | 510(k) Summary | Prepared on: 2026-04-13 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Vilex LLC | |
| Applicant Address | 111 Moffitt Street McMinnville TN 37110 United States | |
| Applicant Contact Telephone | 8019164157 | |
| Applicant Contact | Mr. Brock Johnson | |
| Applicant Contact Email | brock.johnson@vilex.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | DYNEX External Fixation Systems | |
| Common Name | Single/multiple component metallic bone fixation appliances and accessories | |
| Classification Name | Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component | |
| Regulation Number | 888.3030 | |
| Product Code(s) | KTT, JDW (CLASS 2) - PIN, FIXATION, THREADED | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K202054 | Dynex Ring Fixation System. Diametrix Ring Fixation System | KTT |
| K242861 | TrueLok Elevate | KTT |
| K061493 | Apex Pin Fixation System | JDW |
| K231343 | Redemption Duo Hindfoot Nail System | HSB |
| K250304 | TITANEX Screw Systems | HWC |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The DYNEX® External Fixation Systems include the following sub-systems:
- DYNEX® External Fixation System
- DYNEX® Mini Rail System
- DYNEX® TRIAD - TTT Fixation System
The DYNEX® External Fixation System includes rings, partial rings, footplates, struts, hardware, and instruments necessary for the construction of an external fixation frame to which tensioned wires and half pins are attached to the bone and to the frame itself.
The DYNEX® Mini Rail Fixation System includes rails, half pins, guidewires, and instruments necessary for the construction of external fixation rails to which half pins and wires are attached to the bone and to rails. | | |
{5}
K260900 (Page 2 of 2)
The DYNEX® TRIAD - TTT Fixation System includes a fixation frame, transport pins, fixation pins, and instruments necessary for the attachment of the pins and frame to the patient. The frame is fixated to the tibia to facilitate transverse tibial bone transport. The DYNEX TRIAD - TTT Fixation System is intended for use with the DYNEX External Fixation System and is compatible with the DYNEX half pins, rings, and accessories.
DYNEX® External Fixation System components are for single-use only and not designated or sold for any use except as indicated.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The DYNEX® External Fixation Systems are intended for external fixation with the following indications:
- Stabilization of Fractures & Osteotomy
- Rear and Mid-foot Arthrodesis
- Adult and Pediatric Leg Lengthening
- Correction of Bony or Soft Tissue Defects or Deformities in Upper & Lower Extremities
- Pseudoarthrosis or Non-Union of Long Bones
The DYNEX® TRIAD - TTT Fixation System is intended for external fixation in the treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities via tibial bone transport.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The intended use of the subject device and predicate device is external fixation. The indications for use of the subject device and predicate device are similar. The subject device expands the indications for use of the predicate device to accommodate the line extension by including the following additional indications: correction of bony defects, correction of soft tissue defects and deformities, and pseudoarthrosis or non-union of long bones. The additional indications for use do not change the intended use of the subject device and therefore does not constitute a new intended use compared to the predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
This submission presents the DYNEX TRIAD - TTT Fixation System as a line extension of the DYNEX External Fixation Systems. The DYNEX TRIAD - TTT Fixation System adds a new fixation frame, transport pins, and the necessary instruments and accessories. The DYNEX TRIAD - TTT Fixation System is intended for use with the DYNEX External Fixation Systems. The transport pins are a similar design and are composed of the same material as the DYNEX half pins. The fixation frame is attached to bone using the DYNEX half pins and the transport pins. The fixation frame includes a linear bone adjustment mechanism that allows for up to 10mm of transverse bone transport. The fixation frame is composed of similar materials as the predicate rings, struts, and connectors. The instruments and accessories are composed of similar materials as the predicate instruments and accessories.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
No clinical tests were required to demonstrate substantial equivalence of the subject device to the predicate device.
The following non-clinical tests were performed to evaluate the technological differences of the subject compared to the predicate devices:
- Static Axial Strength per ASTM F1541-24
- Slipping Torque per ASTM F1541-24
- Four-Point Bending per ASTM F1541-24
- Torsional Strength per ASTM F1541-24
- Steam Sterilization and Reprocessing Adoption Analysis
- Biological Evaluation per FDA Guidance of ISO 10993-1:2018
The results of the non-clinical tests met all acceptance criteria. Therefore, the subject device is as safe and effective as the predicate device and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.