K260892 · Shenzhen HugeMed Medical Technical Development Co., Ltd. · BTA · May 14, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260892
Device Name
Suction Pump (SU-90)
Applicant
Shenzhen HugeMed Medical Technical Development Co., Ltd.
Product Code
BTA · General, Plastic Surgery
Decision Date
May 14, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SU-90 Suction Pump is indicated for vacuum extraction, aspiration during flexible endoscopy and aspiration and removal of surgical fluids, tissue(including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system. Intended use in hospital environments.
Device Story
Suction Pump (SU-90) is a portable, powered aspiration device for hospital use; operates via diaphragm pump powered by rechargeable lithium battery or AC adapter; generates vacuum pressure up to -90kPa and flow rate ≤24 L/min. Device components include diaphragm pump, battery, PCB, display, and enclosure. Used with external accessories (suction tube, air filters, relief valve, liquid storage bottle) to remove fluids/tissues from patients. Healthcare providers operate device to manage surgical or airway suctioning; output is physical aspiration of materials into storage containers. Benefits include effective fluid/tissue removal during procedures or bedside care. Device is non-sterile, reusable, and requires cleaning per instructions.
Clinical Evidence
Bench testing only. Performance validated against ISO 10079-1:2022 and ISO 10079-4:2021 (free air flow, vacuum accuracy, ingress protection, noise). Electrical safety and EMC tested per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and IEC TS 60601-4-2. Shelf-life validated via ASTM F1980-21; service life via IEC 62506:2023; packaging via ASTM D4169 DC13-2023.
Technological Characteristics
Powered suction pump; diaphragm pump mechanism; AC/DC power (100-240V input, 14.4V internal battery); Class II, Type BF electrical safety; IP22 ingress protection. Dimensions: 240*264*128 mm. Non-sterile, reusable pump unit. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10079-1, ASTM F1980-21, ASTM D4169 DC13-2023.
Indications for Use
Indicated for vacuum extraction, aspiration during flexible endoscopy, and aspiration/removal of surgical fluids, tissue (including bone), gases, bodily fluids, or infectious materials from wounds, airways, or respiratory support systems in hospital environments.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2026
Shenzhen HugeMed Medical Technical Development Co., Ltd.
Joyce Huang
Regulatory engineer
Contact Address
Re: K260892
Trade/Device Name: Suction Pump (SU-90)
Regulation Number: 21 CFR 878.4780
Regulation Name: Powered Suction Pump
Regulatory Class: Class II
Product Code: BTA
Dated: [NOTE: Use date of most recent supplement]
Received: March 19, 2026
Dear Joyce Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260892 - Joyce Huang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260892 - Joyce Huang
Page 3
Sincerely,
Colin K. Chen-S
Digitally signed by
Colin K. Chen-S
Date: 2026.05.14
09:51:52 -04'00'
Colin Kejing Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | ? |
| Please provide the device trade name(s). | | ? |
| Suction Pump (SU-90) | | |
| Please provide your Indications for Use below. | | ? |
| The SU-90 Suction Pump is indicated for vacuum extraction, aspiration during flexible endoscopy and aspiration and removal of surgical fluids, tissue(including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system. Intended use in hospital environments. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260892
Page 1 of 6
# 510(k) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
## I. Submitter
Shenzhen HugeMed Medical Technical Development Co., Ltd.
401, 501, Building 4, Haizhi Technology Park, Fortis, No. 17, Bulan Road, Xialilang Community, Nanwan Street, Longgang District, Shenzhen, Guangdong, 518112, China
Tel.: +86-15992143701
E-mail: huanghuoying@hugemed.cn
Contact Person: Joyce.Huang
Date Prepared: May 11, 2026
## II. Subject Device
Name of Device: Suction Pump
Common or Usual Name: Powered suction pump
Classification Name: Pump, Portable, Aspiration (Manual Or Powered)
Regulatory Class: II
Regulation number: 21 CFR 878.4780
Product Code: BTA
Review Panel: General & Plastic Surgery
## III. Predicate device
Predicate Device: Basic Suction Pump
510(k) Number: K130123
Product Code: BTA
Manufacturer: Medela AG
## IV. Device description
The Suction Pump (Model:SU-90) is equipped with a rechargeable battery and designed to be used in hospital environments. The Suction Pump produces a flow rate not
{5}
K260892
Page 2 of 6
exceeding 24 L/min and has a maximum vacuum pressure of -90kPa, which is marked "high vacuum - high flow".
The Suction Pump is mainly composed of diaphragm pump, rechargeable lithium battery, PCB board, display screen, and enclosure.
The Suction Pump is to be used in conjunction with accessories like suction tube, air filters, relief valve and liquid storage bottle (these accessories are outside the scope of this submission). The Suction Pump is a non-sterile device and can be used on multiple patients when following the proper cleaning techniques detailed in the Instructions For Use, but the accessories are all single patient use and should be disposed of properly and in accordance with their respective instructions for use.
## V. Indications for use
The SU-90 Suction Pump is indicated for vacuum extraction, aspiration during flexible endoscopy and aspiration and removal of surgical fluids, tissue(including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system. Intended use in hospital environments.
## VI. Comparison of technological characteristics
The subject and predicate devices differ in flow rate, power source, electrical requirements, and recommended accessories. These differences do not raise new issues on safety or effectiveness of the subject device.
| Device Characteristic | Subject Device | Predicate Device |
| --- | --- | --- |
| Product Name | Suction Pump | Basic Suction Pump |
| 510(k) Number | K260892 | K130123 |
| Product Owner | Shenzhen HugeMed Medical Technical Development Co., Ltd. | Medela AG |
| Device Classification | 878.4780 Powered Suction Pump Class II | 878.4780 Powered Suction Pump Class II |
| Product Code | BTA | BTA |
| Intended Use | The SU-90 Suction Pump is indicated for vacuum extraction, aspiration during flexible endoscopy and aspiration and | The Basic Suction Pump is indicated for vacuum extraction, aspiration during flexible endoscopy |
{6}
K260892
Page 3 of 6
| | removal of surgical fluids, tissue(including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system. Intended use in hospital environments. | and aspiration and removal of surgical fluids, tissue (including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system either during surgery or at the bedside. |
| --- | --- | --- |
| Prescription vs. OTC | Prescription | Prescription |
| Maximum vacuum | -90 kPa | -90 kPa |
| Flow | ≤24 l/min | 30 l/min |
| Equipment Type | high vacuum and high flow | high vacuum and high flow |
| Working Mode | Constant | Constant |
| Environment of use | In hospital environments | In hospital, ambulatory surgery center, clinic, and doctors' practice. |
| Power | AC/DC | AC |
| Voltage | Power adapter input: 100–240 V~, 50/60 Hz; Internal power supply: DC 14.4 V; | 100-240 V, 50/60 Hz |
| Electrical Requirements | Class II, Type BF | Class I, Type CF |
| IP-Protection | IP22 | IP22 |
| Device Dimensions | 240*264*128 mm | Rack version: 305*210*375 mm Portable version: 305*285*375 mm Rack/trolley: 510*985*470 mm |
3
{7}
K260892
Page 4 of 6
| Accessories | Air filter; Relief valve and liquid storage bottle; Suction tube; adapter; these accessories are outside the scope of this submission. | Patient tubing; Connectors; Liners; Reusable lids; Jars; Locking clasp; Reusable sets; Vacuum tubing ; Filters; Holders; Accessories for high volume setups; Specimen cups; Wall holders; Vacuum Assisted Delivery cups; |
| --- | --- | --- |
| Sterile vs. Non-Sterile | Pump is non-sterile. Disposable components are sterile. | Pump is non-sterile. Disposable components are sterile. |
| Disposable vs. Non-Disposable | Pump is reusable. Disposable components are single patient use. | Pump is reusable. Disposable components are single patient use. |
| Operating Environment | Temperature: +5~40 °C Humidity: 30 ~ 80 % , non-condensing Atmospheric Pressure: 70 ~ 106 kPa | Temperature: +5~40 °C Humidity: 30~75 % Atmospheric Pressure: 70~106 kPa |
| Storage Environment | Temperature: -20~60 °C Humidity: 10 ~ 95 % , non-condensing Atmospheric Pressure: 50 ~ 106 kPa | Temperature: -20~55 °C Humidity: 20~95 % Atmospheric Pressure: 50~106 kPa |
4
{8}
K260892
Page 5 of 6
| Consensus Standards | IEC 60601-1 | IEC 60601-1 |
| --- | --- | --- |
| | IEC 60601-1-2 | IEC 60601-1-2 |
| | ISO 10079-1 | ISO 10079-1 |
## VII. Performance data
The following performance data were provided in support of the substantial equivalence determination.
### Bench testing
ISO 10079-1:2022 and ISO 10079-4:2021 were utilized for testing the Suction Pump, with the following parameters tested per the standard:
- Free air flow
- Vacuum level
- Accuracy of vacuum levels
- Protection against ingress of solid objects and liquids
- Noise level
Additionally, the battery use life of the Suction Pump was also tested.
### Electromagnetic Compatibility and Electrical Safety
The Electromagnetic Compatibility and Electrical Safety testing for the Suction Pump was conducted in accordance with:
- IEC 60601-1:2005+A1:2012+A2:2020
- IEC 60601-1-2:2020
- IEC 60601-2-18:2009
- IEC TS 60601-4-2:2024
### Software
The Suction Pump contains software. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions".
{9}
K260892
Page 6 of 6
- Shelf-Life and Service Life
The shelf life is determined based on product performance testing after accelerated aging test according to ASTM F1980-21.
The service life was validated via accelerated testing according to IEC 62506:2023 and mechanical life testing of key components.
- Packaging
Packaging performance was tested in accordance with ASTM D4169 DC13-2023.
VIII. Conclusions
The subject device has the same indication for use and has similar design features and technological characteristics as the predicate device. Performance testing data demonstrates that the subject device is as safe and effective as the predicate device. Accordingly, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.