The VERAFEYE Imaging and Guidance System is intended for cardiac applications. The system provides 2D & 3D images of the heart, cardiac valves, great vessels, and surrounding anatomical structures for the evaluation of the presence or absence of pathology. The VERAFEYE Imaging and Guidance System is also intended for catheter-based cardiac electrophysiological (EP) procedures in the Left and Right Atrium. The VERAFEYE Imaging and Guidance System provides the reconstruction of chamber geometry from ultrasound data, and visualization of the chamber anatomy and intracardiac catheter location during procedures.
Device Story
The VERAFEYE Imaging and Guidance System is a non-sterile, reusable ultrasound imaging and guidance system for use in cardiac catheterization labs. It takes ultrasound signals from an imaging catheter and electromagnetic (EM) tracking signals from pose sensors located on the imaging catheter and compatible third-party ablation catheters. The system console processes these inputs to generate real-time 2D (B-mode) and 3D volumetric images of cardiac structures; it reconstructs chamber geometry and displays the live pose of catheters relative to anatomical landmarks. Operated by clinicians, the system provides visual guidance during EP procedures. The output is displayed on monitors to assist in navigation and anatomical assessment. The device does not deliver therapy or perform autonomous clinical decisions; it benefits patients by providing real-time anatomical visualization and catheter tracking during minimally invasive cardiac interventions.
Clinical Evidence
No clinical data was utilized. Safety and effectiveness were demonstrated through bench testing, including imaging performance (resolution, accuracy), operating environment compliance (thermal/mechanical index, acoustic output), clinical workflow simulation, and software verification.
Technological Characteristics
System includes console, Catheter Interface Unit, EM tracking subsystem, and ECG module. Materials include TPU and medical-grade acrylic. Sensing uses ultrasound transducers and EM pose sensors (electrical coils). Connectivity includes USB for data export and hot-fix installation. Software performs signal acquisition, B-mode/3D reconstruction, and EM tracking. Sterilization: Not provided sterile.
Indications for Use
Indicated for adult patients undergoing cardiac applications, including evaluation of heart pathology and catheter-based cardiac electrophysiological (EP) procedures in the Left and Right Atrium.
{0}
![LOGO]
June 15, 2026
Luma Vision Limited
Marta Walker
Director of Quality and Regulatory at Luma Vision
Block C, Parkview House-Beech Hill Office Campus
Beech Hill Rd.
Dublin, D04 K5D0
Ireland
Re: K260885
Trade/Device Name: VERAFEYE Imaging and Guidance System (VF-VIS-002)
Regulation Number: 21 CFR 892.1560
Regulation Name: Ultrasonic pulsed echo imaging system
Regulatory Class: Class II
Product Code: IYO, DQK
Dated: March 17, 2026
Received: March 17, 2026
Dear Marta Walker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260885 - Marta Walker
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K260885 - Marta Walker
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for: **MARCO CANNELLA -S**
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K260885 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| VERAFEYE Imaging and Guidance System (VF-VIS-002) | | | |
| Please provide your Indications for Use below. | | | ? |
| The VERAFEYE Imaging and Guidance System is intended for cardiac applications. The system provides 2D & 3D images of the heart, cardiac valves, great vessels, and surrounding anatomical structures for the evaluation of the presence or absence of pathology. | | | |
| The VERAFEYE Imaging and Guidance System is also intended for catheter-based cardiac electrophysiological (EP) procedures in the Left and Right Atrium. The VERAFEYE Imaging and Guidance System provides the reconstruction of chamber geometry from ultrasound data, and visualization of the chamber anatomy and intracardiac catheter location during procedures. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}

Luma Vision Limited
Block C, Parkview House
Beech Hill Office Campus
Beech Hill Road
Dublin D04 K5D0, Ireland
www.lumavision.com
info@lumavision.com
Department of Health and Human Services
Centre of Device and Radiological Health
Office of Device Evaluation
Traditional 510(k) section- K260885
# **510(k) Summary of Safety and Effectiveness Information**
**as required by section 21 CFR 807.92**
# **Submitter of 510(k)**
| **Manufacturer Name:** | Luma Vision Limited |
| --- | --- |
| **Address:** | Block C, Parkview House, Beech Hill Office Campus, Beech Hill Road, Dublin D04 K5D0, Ireland |
| **Contact Person:** | Marta Walker |
| **Phone:** | 0031680132668 |
| **Email:** | marta.walker@lumavision.com |
| **Job Title:** | Quality Assurance & Regulatory Affairs Director |
| **Date Prepared:** | 15^{th} June 2026 |
K260885
Page 1 of 13
{5}
# **Traditional 510(k) Submission for VERAFEYE Imaging and Guidance System**
# **Device Information:**
| **Trade / Proprietary Name:** | VERAFEYE Imaging and Guidance System (VF-VIS-002) |
| --- | --- |
| **Common / Usual Name:** | VERAFEYE Imaging and Guidance System |
| **Device:** | System, Imaging, Pulsed Echo, Ultrasonic |
| **Regulation Description:** | Ultrasonic Pulsed Echo Imaging System |
| **Regulation Number:** | 21 CFR §892.1560 |
| **Device Classification:** | Class II |
| **Product Code:** | IYO, DQK |
**Panel:** Division of Cardiovascular Devices
# **Predicate & Reference Devices:**
# ***Primary Predicate Device:*** (K242893)
*Trade / Device Name: VERAFEYE Imaging System (VF-VIS-001)*
*Regulation Number: 21 CFR §892.1560*
*Regulatory Name: Ultrasonic Pulsed Echo Imaging System*
*Regulatory Class: Class II*
*Product Code: IYO*
# ***Secondary Predicate Device:*** (K251234)
*Trade / Device Name: EnSite™ X EP System*
*Regulation Number: 21 CFR §870.1425*
*Regulatory Name: Programmable Diagnostic Computer*
*Regulatory Class: Class II*
*Product Code: DQK*
K260885
Page 2 of 13
{6}
# Device Description
The VERAFEYE Imaging and Guidance System is a non-sterile, reusable ultrasound imaging and guidance system intended for use in a cardiac catheterization laboratory for adult cardiac applications. The system consists of a System Console with integrated computing hardware and software, a Catheter Interface Unit (CIU), a three-dimensional electromagnetic tracking subsystem, a pulse phase (ECG) module, and an ablation interface for compatibility with third-party ablation catheters. An itemization of the components is provided in the following table.
| Component | Model No. | Sub-component | Patient Contact Material |
| --- | --- | --- | --- |
| Catheter Interface Unit | VF-CIU-002 | N/A | N/A |
| System Console | VF-SC-002 | Sensor Interface Unit | N/A |
| | | System Control Unit | N/A |
| | | Window Field Generator | N/A |
| | | Window Field Generator Bracket | N/A |
| | | Pulse Phase Module | N/A |
| | | Reference Sensor | N/A |
| | | Pulse Phase Module Leads | Thermoplastic Polyurethane (TPU) |
| | 3M™ Red Dot™ Model 2560 or 2570 | Pulse Phase Module Electrode, single use | Thermoset Polyurethane |
| | DE-009c | Reference Sensor Patch, single use | Medical grade acrylic |
| | 60-860 – CIV-Flex™ probe cover | Reference Sensor Sheath, single use | Medical grade acrylic |
| Ablation Interface | VF-ABI-001 | Ablation Interface | N/A |
No part of the VERAFEYE Imaging and Guidance System is provided sterile.
The system transmits ultrasound energy into adult patients to generate real-time two-dimensional (B-mode) and three-dimensional volumetric images of the heart, cardiac valves, great vessels, and surrounding anatomical structures for evaluation of the presence
K260885
Page 3 of 13
{7}
or absence of pathology. It provides scale indication with respect to anatomical structures to support clinical assessment in conjunction with other medical information.
The system utilizes compatible intracardiac imaging catheters intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices within the heart. The system receives ultrasound and tracking signals from the imaging catheter as well as supported third-party ablation catheters, converts analog signals to digital data, processes imaging information, and displays reconstructed chamber geometry and catheter position to assist clinicians during electrophysiology and other percutaneous cardiac procedures. The catheter is intended for imaging guidance only and not for treatment delivery.
The principle of operation for the imaging aspects of the VERAFEYE Imaging and Guidance System is the same as that for commonly used ultrasound imaging systems (both external and intra-operative). To localize the imaging information with respect to the anatomy and anatomical structures i.e. the navigational aspect of the VERAFEYE Imaging and Guidance System, magnetic pose sensors are incorporated adjacent to the ultrasound transducer of the VERAFEYE Imaging Catheter. The pose sensors utilize electrical coils in which signals are induced when placed in a moving electromagnetic (EM) field; the signals are indicative of the position and orientation of the coil within the moving EM field. The pose sensors allow for the pose of the tip, and therefore the imaging plane, to be known with respect to the EM field generator. The image data, tracking information and heartbeat signal information allow for the reconstruction of the relevant anatomical chambers, in order to provide global orientation for navigation along with live 2D and 4D ultrasound imaging. For compatible third-party ablation catheters, the same principle of EM tracking of the catheter tip is used to display the live pose of the device within the anatomical chamber data.
The primary responsibility of the software is to control the hardware for signal acquisition (ultrasound system, motor, EM tracking system and ECG interface), acquire the digitized signals into B-mode images and/or volumes and present them to the user. Based on that data the user can create semantic landmarks like the left/right atrium. The imaging catheter and, if connected, the third-party ablation catheter are shown in relation to the imaging, volumes and/or landmarks. Auxiliary functions are screenshot and video clip recording as well as recording export to external USB devices.
The tracking of the imaging catheter as well as the optional, supported third-party ablation catheter is done via EM tracking. The catheters contain sensor(s) which are within an EM field generated by a field generator. The EM tracking system digitizes the pose (3D position
K260885
Page 4 of 13
{8}
and orientation) which then is used by the software to show the catheters relative to the global anatomical orientation.
| Performance Characteristic | System Capability |
| --- | --- |
| 2D B-mode image resolution | 0.5mm |
| Tracking positional accuracy | 3mm |
| Tracking orientation accuracy | 5° |
| Intra-volume point to point accuracy | 5mm @ 2cm depth |
| Combined volume accuracy | 3mm |
| Image acquisition frequency | 20Hz |
| Tracking frequency | 30Hz |
The intended use environment for the VERAFEYE Imaging and Guidance System is a cardiac catheter laboratory (cath lab), a minimally invasive operating environment designed for cardiac procedures. The workspace is centered around an adjustable patient bed integrated with a C-arm X-ray system, enabling fluoroscopic imaging while requiring staff to wear protective lead equipment. The operating room includes ECG monitoring, ceiling-mounted display panels, and imaging systems such as fluoroscopy, echocardiography (including TEE and ICE), and advanced technologies like electro-anatomical mapping and ablation guidance. Ambient noise levels typically range from 50–65 dBA, while lighting is adjusted throughout procedures to support patient preparation, vascular access, and image-guided interventions. The cath lab is divided into sterile and non-sterile zones, with the sterile area extending from the patient bed to the patient's abdomen. Clinical staff, equipment, and workspaces within this boundary must maintain sterility to ensure patient safety and support efficient procedural workflows.
**Intended Use:** Cardiac
# **Indications for Use:**
The VERAFEYE Imaging and Guidance System is intended for cardiac applications. The system provides 2D & 3D images of the heart, cardiac valves, great vessels, and surrounding anatomical structures for the evaluation of the presence or absence of pathology.
The VERAFEYE Imaging and Guidance System is also intended for catheter-based cardiac electrophysiological (EP) procedures in the Left and Right Atrium. The VERAFEYE Imaging and Guidance System provides the reconstruction of chamber geometry from ultrasound data, and visualization of the chamber anatomy and intracardiac catheter location during procedures.
K260885
Page 5 of 13
{9}
## **Indications for Use Discussion**
The intended use of the VERAFEYE Imaging and Guidance System combines functionalities from the 2 predicate devices without introducing a new/modified clinical purpose, patient population, or therapeutic indication that could impact safety or effectiveness.
Similar to the Primary Predicate Device (VERAFEYE Imaging System), the system provides 2D and 3D imaging of cardiac anatomy for evaluation of the presence or absence of pathology. Similar to the Secondary Predicate Device (EnSite X EP System), it supports electrophysiology (EP) procedures by providing visualization of the chamber anatomy and intracardiac catheter location. The chamber geometry reconstruction function is derived from ultrasound imaging data and serves as an anatomical guidance tool during EP procedures.
The VERAFEYE Imaging and Guidance System remains limited to diagnostic imaging and procedural guidance within the heart and does not deliver therapy or perform autonomous clinical decision-making. The proposed intended use represents an integration of established imaging and electrophysiology guidance capabilities rather than a change in the clinical purpose or risk profile.
## **Functional and Technological Comparison**
The VERAFEYE Imaging and Guiding System (VF-VIS -002) adds the following functionalities to the VERAFEYE Imaging System (VF-VIS-001):
1. Creation of anatomical models of the left and right atria.
2. Real-time display of the position of the imaging catheter and supported ablation catheters. (Note: VF-VIS-002 supports the Abbott FlexAbility and TactiCath ablation catheters)
3. Allow the user to place markers on the anatomical models that signal locations that have been ablated (ablation markers).
Table 1 below provides a functional and technological comparison between the VERAFEYE Imaging and Guidance System VF-VIS-002 and the legally marketed predicate devices, VERAFEYE Imaging System VF-VIS-001 (K242893) and EnSite™ X EP System (K251234).
As shown in Table 1, the subject device has the same intended use, classification, and similar technological characteristics as the predicate devices.
Consistent with the VERAFEYE Imaging System predicate, the subject device is intended for cardiac applications and provides two-dimensional (B-mode) and three-dimensional ultrasound images of the heart, cardiac valves, great vessels, and surrounding anatomical
K260885
Page 6 of 13
{10}
structures to evaluate the presence or absence of pathology, and to provide scale information that may be used adjunctively with other medical data for clinical diagnosis.
Consistent with the EnSite™ X EP System predicate, the subject device is also intended for catheter-based cardiac electrophysiology procedures and provides visualization of intracardiac catheter location and reconstruction of chamber geometry to support electrophysiology diagnostics and monitoring.
Together, the predicate devices represent the full scope of ultrasound imaging and electrophysiology guidance functions of the VERAFEYE Imaging and Guidance System VF-VIS-002.
**Table 1. Predicate Device Comparison Table- VERAFEYE Imaging and Guidance System**
# **Substantial Equivalence Imaging and Guidance System**
| Description | Predicate Device 1 | Predicate Device 2 | | |
| --- | --- | --- | --- | --- |
| Manufacturer | VERAFEYE Imaging System VF-VIS-001 K# K242893 | EnSite™ X EP System K# K251234 | VERAFEYE Imaging and Guidance System VF-VIS-002 K# K260885 | Discussion of Equivalence & Differences |
| Classification | Class II | Class II | Class II | Same as both predicate devices |
| Regulation: | 21 CFR 892.1560 – IYO | 21 CFR 870.1425- DQK | 21 CFR 892.1560 – IYO 21 CFR 870.1425- DQK | Same as both predicate devices |
| Regulation Name | IYO - Ultrasonic pulsed echo imaging system. | DQK- Programmable diagnostic computer. | IYO - Ultrasonic pulsed echo imaging system. DQK- Programmable diagnostic computer. | Same as both predicate devices |
| Product Code | IYO | DQK | IYO, DQK | Same as both predicate devices |
| Intended Use Environment | Cardiac - Intracardiac Echocardiography | Cardiac Electrophysiology, Diagnostics, and Monitoring Devices | *Cardiac Electrophysiology - Intracardiac Echocardiography* | Applicable indications taken from both predicates; |
K260885
Page 7 of 13
{11}
| Description | Predicate Device 1 | Predicate Device 2 | | |
| --- | --- | --- | --- | --- |
| Manufacturer | VERAFEYE Imaging System VF-VIS-001 K# K242893 | EnSite™ X EP System K# K251234 | VERAFEYE Imaging and Guidance System VF-VIS-002 K# K260885 | Discussion of Equivalence & Differences |
| | | | | ICE from Device 1 & Electrophysiology from Device 2 |
| **Indications for Use** | The VERAFEYE Imaging System is intended for cardiac applications. The system transmits ultrasound energy into adult patients creating 2D (B-mode), & 3D images of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system also provides the ability to indicate scale with respect to anatomical structures, which provides information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The system utilizes catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other | The EnSite X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. The EnSite X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological procedures | The VERAFEYE Imaging and Guidance System is intended for cardiac applications. The system provides 2D & 3D images of the heart, cardiac valves, great vessels, and surrounding anatomical structures for the evaluation of the presence or absence of pathology. The VERAFEYE Imaging and Guidance System is also intended for catheter-based cardiac electrophysiological (EP) procedures in the Left and Right Atrium. The VERAFEYE Imaging and Guidance System provides the reconstruction of chamber geometry from ultrasound data, and visualization of the chamber anatomy and intracardiac catheter location during procedures. | Equivalent to features prescribed in both – see note *1 |
K260885
Page 8 of 13
{12}
| Description | Predicate Device 1 | Predicate Device 2 | | |
| --- | --- | --- | --- | --- |
| Manufacturer | VERAFEYE Imaging System VF-VIS-001 **K#** K242893 | EnSite™ X EP System **K#** K251234 | VERAFEYE Imaging and Guidance System VF-VIS-002 **K#** K260885 | Discussion of Equivalence & Differences |
| | devices in the heart of adult patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. | | | |
| **Operating Frequency** | 5.25MHz | N/A | 5.25MHz | Same as predicate Device 1 |
| **Modes** | 2D (B-mode) and 3D | N/A | 2D (B-mode) and 3D | Same as predicate Device 1 |
| Transducers/catheters: | VERAFEYE ICE Catheter VF-VIC-001 | N/A | VERAFEYE ICE Catheter VF-VIC-002 | Same as predicate Device 1 & compatible catheter |
| **System Features** | | | | |
| Cardiac Measurements and Calculations | Yes: Length & area | N/A | Yes: Length & area | Same as predicate Device 1 |
| Cybersecurity Features | yes | N/A | yes | Same as Device 1 |
| Local user accounts with user authentication & authorization | yes | N/A | yes | Same as predicate Device 1 |
| Table side Remote Control: | The VERAFEYE ultrasound imaging system will provide control options through mouse and keyboard | N/A | The VERAFEYE ultrasound imaging system will provide control options through mouse and keyboard | Same as predicate Device 1 |
| Service Save logs: | yes | N/A | yes | Same as predicate Device 1 |
| Security Hot-fix Installation: | Yes: Through USB drive | N/A | Yes: Through USB drive | Same as predicate Device 1 |
K260885
Page 9 of 13
{13}
| Description | Predicate Device 1 | Predicate Device 2 | | |
| --- | --- | --- | --- | --- |
| Manufacturer | VERAFEYE Imaging System VF-VIS-001 **K# K242893** | EnSite™ X EP System **K# K251234** | VERAFEYE Imaging and Guidance System VF-VIS-002 **K# K260885** | Discussion of Equivalence & Differences |
| Study Back-up and Restore: | yes | N/A | yes | Same as predicate Device 1 |
| Output Display Standard (Track 3): | yes | N/A | yes | Same as predicate Device 1 |
| Patient Contact Materials: | yes | N/A | yes | Same as predicate Device 1 |
| Motor Unit: | Motor Unit – VERAFEYE system | N/A | Motor Unit – VERAFEYE system | Same as Device 1 |
| | Function: Rotates the inner part of the catheter NOTE: The motor unit, is equivalent of the motor unit of the Ultra ICE Plus catheter (# K181042) – we can take this part of that product as a reference device, as although it has a different intended use (for the entire Ultra ICE catheter than VERAFEYE, the “motor unit” part of the Ultra ICE catheter is equivalent operational principle (rotates) to the VERAFEYE motor unit as their intended use is to rotate the catheter and to | N/A | Function: Rotates the inner part of the catheter NOTE: The motor unit, is equivalent of the motor unit of the Ultra ICE Plus catheter (# K181042) – we can take this part of that product as a reference device, as although it has a different intended use (for the entire Ultra ICE catheter than VERAFEYE, the “motor unit” part of the Ultra ICE catheter is equivalent operational principle (rotates) to the VERAFEYE motor unit as their intended use is to rotate the catheter and to convert signals into imaging. | Same as predicate Device 1 |
K260885
Page 10 of 13
{14}
| Description | Predicate Device 1 | Predicate Device 2 | | |
| --- | --- | --- | --- | --- |
| Manufacturer | VERAFEYE Imaging System VF-VIS-001 K# K242893 | EnSite™ X EP System K# K251234 | VERAFEYE Imaging and Guidance System VF-VIS-002 K# K260885 | Discussion of Equivalence & Differences |
| | convert signals into imaging. | | | |
Note *1- The proposed intended use of the VERAFEYE Imaging and Guidance System is acceptable because it combines functionalities from the 2 predicate devices without introducing a new/modified clinical purpose, patient population, or therapeutic indication that could impact safety or effectiveness. Similar to the Primary Predicate Device (VERAFEYE Imaging System), the system provides 2D and 3D imaging of cardiac anatomy for evaluation of the presence or absence of pathology. Similar to the Secondary Predicate Device (EnSite X EP System), it supports electrophysiology (EP) procedures by providing visualization of the chamber anatomy and intracardiac catheter location. The chamber geometry reconstruction function is derived from ultrasound imaging data and serves as an anatomical guidance tool during EP procedures. The VERAFEYE Imaging and Guidance System remains limited to diagnostic imaging and procedural guidance within the heart and does not deliver therapy or perform autonomous clinical decision-making. The proposed intended use represents an integration of established imaging and electrophysiology guidance capabilities rather than a change in the clinical purpose or risk profile
## Non-clinical Performance testing
Non-clinical performance testing to substantiate the effectiveness of the VERAFEYE Imaging and Guidance System encompasses the requirements pertaining to Imaging & Guidance, Operating Environment, Clinical workflow and Lifetime. The specific tests are summarized in the table below.
Clinical data was not utilized to demonstrate safety or effectiveness of the VERAFEYE Imaging and Guidance system.
| Test Category | Performance trait evaluated | Enabled function |
| --- | --- | --- |
| Imaging and guidance | - Field of View - Imaging geometry - Image resolution - Dynamic range - Point to point accuracy - Roll accuracy - Device pose accuracy - Temporal accuracy | The system is able to provide images suitable for the visualization of 3^{rd} party devices & cardiac anatomy. It also provides continuous tracking data with respect to the position of the catheter and compatible 3^{rd} party therapy devices |
| Operating environment | - Thermal index - Mechanical index - Acoustic output - Surface temperature - Emissions & dielectric strength | 3D imaging is provided while remaining compliant with the standards for acoustic output for cardiac devices |
K260885
Page 11 of 13
{15}
| Clinical workflow | - Spatial map display & control, - Operational latency physiological signals monitoring - Third party tool representation | Operation of the system can be achieved within the normal flow of the electrophysiology procedure |
| --- | --- | --- |
| Lifetime | - System runtime - Expected service life | The function of the system is maintained consistent with the length of electrophysiology procedures |
| Software | Software verification at unit, integration and system level | Live imaging, creation and display of anatomical maps, tracking of the image & third party ablation catheter & export of recorded data. |
| Design validation | - Pre-clinical - Summative | Operation of the system conforms with the clinical workflow in a simulated use environment |
There are no technological difference between the subject device and predicate device 1. The technological differences between the subject device and predicate device 2 stem from the generation of the displayed cardiac maps. For the subject device, the maps are generated from application of tracking data to the image data such that a map is created depicting structures that are observable in the ultrasound image. For predicate device 2, the maps are generated from an application of the tracking data to the measured electro-potential data such that the maps that are generated are an illustration of electro-physiologic function. The ability of the subject device to create anatomical maps is a function of ultrasound imaging capability that the system has; predicate device 2 does not incorporate an ultrasound imaging capability.
### Conclusion:
The complimentary functions of predicate device 1 and device 2 are combined in the VERAFEYE Imaging and Guidance System. The subject device and predicate devices share common intended use, principles of operation and technological characteristics though all of the features are not present together in either one of the predicates. Predicate device 1 employs technical characteristics and functionality that supports ultrasound image acquisition; predicate device 2 employs technology and functionality that supports the combination of imaging data, tracking data and anatomical map creation. The combination of previously existing but dissimilar technologies into a single platform facilitates mitigation of hazards associated with lack of availability of one of
K260885
Page 12 of 13
{16}
the technologies. Any hazards arising from combination of normally separated technologies is addressed through compliance with the relevant standards and the appropriate particular standards.
After analyzing the results of the bench, laboratory and electrical safety tests, it can be concluded that the VERAFEYE Imaging and Guidance System and the predicate devices (K242893 and K251234) are substantially equivalent.
The subject device (VERAFEYE Imaging and Guiding System VF-VIS-002) and predicate devices are all intended for use in intravascular and/or intracardiac imaging. The results of the relevant performance data and compatibility testing support a determination that the proposed subject devices do not raise new questions of safety or effectiveness and is substantially equivalent to the predicate devices.
K260885
Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.