K260855 · CMT Health PTE., Ltd. · FMI · Jun 10, 2026 · General Hospital
Device Facts
Record ID
K260855
Device Name
Profoject™ Safety Pen Needles
Applicant
CMT Health PTE., Ltd.
Product Code
FMI · General Hospital
Decision Date
Jun 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Profoject™ Safety Pen Needles are intended for use with pen injector devices for the subcutaneous injection of drugs. The device is intended for use in patients weighing ≥ 3.5 kg, from neonates to adults.
Device Story
Profoject™ Safety Pen Needles are sterile, single-use, disposable needles for subcutaneous drug delivery via pen injectors. Device features a passive safety mechanism: a spring-loaded needle shield that retracts during skin insertion and automatically extends/locks upon needle withdrawal to prevent accidental needlestick injuries and reuse. Operated manually by patients or healthcare professionals in clinical or home settings. Output is the delivery of medication; safety mechanism provides physical protection against sharps injuries. Device comprises housing, needle shield, hub, tube, spring, and packaging components. Performance verified via bench testing for mechanical integrity, sterility, and sharps protection.
Clinical Evidence
No clinical data included. Substantial equivalence is supported by non-clinical bench testing, including ISO 11608-2 (pen needle requirements), ISO 9626 (needle tubing), ISO 23908 (sharps protection), and ISO 10993-1 (biocompatibility). Packaging integrity verified via ASTM F1886, F1140, F88, and F1929. Sterility and residuals validated per ISO 11135 and ISO 10993-7.
Indicated for subcutaneous drug injection in patients weighing ≥ 3.5 kg, including neonates, infants, children, adolescents, and adults.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 10, 2026
CMT Health PTE., Ltd.
Monica Ma
Official Correspondent
150 Beach Rd., #28-05, Gateway West
Singapore, 189720
Singapore
Re: K260855
Trade/Device Name: Profoject™ Safety Pen Needles
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: March 16, 2026
Received: May 22, 2026
Dear Monica Ma:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
{1}
K260855 - Monica Ma
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260855 - Monica Ma
Page 3
Sincerely,
**SHRUTI N. MISTRY -S**
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General
Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260855 | ? |
| Please provide the device trade name(s). | | ? |
| Profoject™ Safety Pen Needles | | |
| Please provide your Indications for Use below. | | ? |
| Profoject™ Safety Pen Needles are intended for use with pen injector devices for the subcutaneous injection of drugs. The device is intended for use in patients weighing ≥ 3.5 kg, from neonates to adults. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
# 510(k) Summary
[As required by 21 CFR 807.92]
## 1. Submission Information
510(k) Number: K260855
Date: May 22, 2026
Type of 510(k) Submission: Traditional
Submitter: CMT HEALTH PTE. LTD.
150 BEACH ROAD, #28-05, GATEWAY WEST, SINGAPORE, 189720
Contact Person: Monica Ma
E-mail: ra@cmthealth.com
Tel: +65 6846 1379
## 2. Device Description
Proprietary Name: Profoject™ Safety Pen Needles
Common Name: Hypodermic Single Lumen Needle
Regulation Name: Needle, hypodermic, Single Lumen
Product Code: FMI
Device Class: II
Regulation Number: 21 CFR 880.5570
Review Panel: General Hospital
Indications for use: Profoject™ Safety Pen Needles are intended for use with pen injector devices for the subcutaneous injection of drugs. The device is intended for use in patients weighing $\geq 3.5$ kg, from neonates to adults.
Device Description: Compared with traditional pen needles, the Profoject™ Safety Pen Needles incorporate a safety feature to help prevent accidental needle-stick injuries to patients and healthcare professionals. The needle tube is equipped with a needle shield that automatically covers and locks the needle tube after use. During injection, the needle shield retracts as the needle tube is inserted into the skin. Once the injection is completed and the needle tube is withdrawn, the needle shield automatically extends to cover the needle tube. After lockout, the safety pen needle cannot be reused. The Profoject™ Safety Pen Needles are single-use, sterile, disposable medical devices intended for use with compatible pen injectors. The device comprises nine primary components: (1) Housing, (2) Needle shield, (3) Needle hub, (4) Needle tube, (5) Outer container, (6) Spring, (7) Retaining ring, (8) Sliding clasp and (9) Sealed paper. The outer container and sealed paper form the unit packaging of the Profoject™ Safety Pen Needles, maintaining sterility until use. The needle hub is designed to screw onto compatible pen injectors.
## 3. Predicate Device Identification
Predicate 510(k) Number: K210748
Marketing clearance date: May 25, 2022
Product name: Well-life™ Safety Pen Needles
Manufacturer: W. L. MED Co., Ltd.
1
{5}
#### **4. Non-Clinical Test Conclusion**
Non-clinical verification of the Profoject™ Safety Pen Needles have been conducted to evaluate their safety, performance, and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods, and ultimately support a substantial equivalence determination. Particularly, the following was conducted to adequately demonstrate the effectiveness of the proposed device in accordance with relevant test methods cited below:
##### **Performance Testing**
1. ISO 11608-2:2022 Needle-based injection systems for medical use - Requirements and test methods - Part 2: Double-ended Pen Needles.
2. ISO 9626 Second edition 2016-08-01 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
3. ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
4. ASTM F1140/F1140M-13 (2020) Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
5. ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
6. ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
7. USP - NF <71> Sterility Tests
8. ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)]
9. ISTA 3A: 2018 Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
10. ISO 11135 Second edition 2014-07-15 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
11. USP - NF <788> Particulate Matter in Injections
12. USP - NF <85> Bacterial Endotoxin Testing
13. ISO 23908: 2011 Sharps Injury Protection – Requirements and Test Methods – Sharps Protection Features for Single Use Hypodermic Needles, Introducers for Catheters and Needles Uses for Blood Sampling
##### **Usability Testing**
FDA guidance document *Medical Devices with Sharps Injury Prevention Features*.
##### **Biocompatibility Testing**
The biocompatibility evaluation for the Profoject™ Safety Pen Needles were conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1. Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed device is classified as an externally communicating with prolonged, blood path indirect contact.
1. ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2. ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
2
{6}
3. ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
4. ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
5. ISO 10993-4 Third edition 2017-04 Biological evaluation of medical devices--Part 4: Selection of tests for interactions with blood
# **5. Substantial Equivalent Based on Assessment of Clinical Performance Data**
Clinical data was not including in this submission
3
{7}
# 6. Substantially Equivalent Comparison Conclusion
| Parameters | | | Proposed Device | | Predicate Device | | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 510(k) Number | | K260855 | | K210748 | | -- |
| 2 | 510(k) Holder | | CMT HEALTH PTE. LTD. | | W. L. Med Co., Ltd. | | -- |
| 3 | Trade Name | | Profoject™ Safety Pen Needles | | Well-life™ Safety Pen Needles | | -- |
| 4 | Product code | | FMI | | FMI | | Same |
| 5 | Regulation Number | | 21 CFR 880.5570 | | 21 CFR 880.5570 | | Same |
| 6 | Common Name | | Hypodermic Single Lumen Needle | | Hypodermic Single Lumen Needle | | Same |
| 7 | Device Class | | II | | II | | Same |
| 8 | Indications for use | | Profoject™ Safety Pen Needles are intended for use with pen injector devices for the subcutaneous injection of drugs. The device is intended for use in patients weighing ≥ 3.5 kg, from neonates to adults. | | The Well-life safety pen needle is intended for use with pen injector devices for the subcutaneous injection of insulin. | | Different Discussion 1 |
| 9 | Prescription/OTC | | OTC | | OTC | | Same |
| 10 | Supplied sterile | | Yes | | Yes | | Same |
| 11 | Single use | | Yes | | Yes | | Same |
| 12 | Principle of operation | | Manual | | Manual | | Same |
| 13 | Method of attachment to pen injector | | Screw threads | | Screw threads | | Same |
| 14 | Models | Needle Gauge | 29G, 30G, 31G, 32G, 33G, 34G | | 29G, 30G, 31G, 32G, 33G, 34G | | Different Discussion 2 |
| | | Needle Length | 4mm, 5mm, 6mm, 8mm | | 4mm, 5mm, 6mm, 8mm, 10 mm, 12.7mm | | |
| 15 | Wall type | | Thin-walled Extra-thin-walled | | Regular walled Thin-walled | | Different Discussion 3 |
| 16 | Biocompatibility | | Conform ISO 10993-1 | | Conform ISO 10993-1 | | Same |
| 17 | Sterility | | SAL=10^{-6} | | SAL=10^{-6} | | Same |
| 18 | Sterilization method | | EO sterilized | | EO sterilized | | Same |
| 19 | Shelf Life | | 5 years | | 5 years | | Same |
| 20 | Unit Packaging | | Polypropylene container with seal made of dialysis paper | | Polypropylene container with seal made of medical grade paper | | Different Discussion 4 |
| 21 | Patient-contact Material | Needle tube | SUS 304 | Needle tube | STS 304 | Different Discussion 5 | |
| | | | Lubricant | Polydimethylsiloxane | Silicon | | Polydimethylsiloxane |
| 22 | Configuration | | Housing, Needle shield, Needle hub, Needle tube, Outer container, Spring, Retaining ring, Sliding clasp, and Sealed paper | | Needle, Hub, Needle Cap, Sterile Cap, Guide, Safety Cap, and Spring | | Different Discussion 6 |
4
{8}
| 23 | Performance Testing | Comply with ISO 9626, ISO 11608-2 and ISO 23908 | Comply with ISO 9626, ISO 11608-2 and ISO 23908 | Same |
| --- | --- | --- | --- | --- |
| 24 | Principle of operation for the sharps protection feature | The Profoject™ Safety Pen Needles incorporate a safety feature to help prevent accidental needle-stick injuries to patients and healthcare professionals. The needle tube is equipped with a needle shield that automatically covers and locks the needle tube after use. During injection, the needle shield retracts as the needle tube is inserted into the skin. Once the injection is completed and the needle tube is withdrawn, the needle shield automatically extends to cover the needle tube. After it is locked out, the safety pen needle cannot be reused. | The Well-life™ Safety Pen Needle (i.e., S type) is designed to reduce occurrence of accidental needle sticks from patient end of the needle by providing a shield that covers and locks the needle after use. As the user proceeds with inserting the needle into the skin, the Needle Cap will retract. After the injection is completed and needle is removed from the skin, the safety cap will automatically extend to cover the needle. Once the pen needle is locked out, it can no longer be used. | Different Discussion 7 |
| Differences between proposed device and predicate device: | | | | |
| Discussion 1: Indications for use The broader indication for use does not introduce new or different questions of safety or effectiveness when compared to the predicate device as the device’s mechanism remains unchanged. | | | | |
| Discussion 2: Models The needle length of the proposed device differs from those of the predicate devices. This difference in needle size does not affect device performance. All needle sizes of the proposed device have been tested, and the results demonstrate that the device meets the performance criteria. Therefore, this difference does not affect the substantial equivalence between the proposed and predicate devices. | | | | |
| Discussion 3: Wall type The wall thickness of the proposed device differs from that of the predicate device. Both thin-walled and extra-thin-walled needles of the proposed device have been tested, and the results demonstrate that the device meets the performance criteria. Therefore, this difference does not affect the substantial equivalence between the proposed and predicate devices. | | | | |
| Discussion 4: Unit Packaging While the sealing paper materials of the proposed device and the predicate device differ, performance testing has demonstrated that the proposed device maintains a SAL=10^{-6}. Consequently, the difference in unit packaging materials does not raise new or different questions of safety and effectiveness when compared to the predicate device. | | | | |
| Discussion 5: Patient-contact Material The needle tube material of the proposed device is designated as “SUS 304” (Japanese Industrial Standard), while that of the predicate device is designated as “STS 304” (Korean Standard). Although the nomenclature differs, both designations refer to the same underlying material: 304 austenitic stainless steel. The chemical composition and mechanical properties of SUS 304 and STS 304 are identical within the allowable tolerances of their respective | | | | |
5
{9}
standards. In addition, biocompatibility testing and ISO 9626 performance testing have been conducted on the proposed device’s needle tube material to demonstrate that it meets biocompatibility and ISO 9626 requirements. This difference does not raise any new or different questions of safety and effectiveness when compared to the predicate device.
#### Discussion 6: Configuration
The configuration of proposed device is different from predicate device, depending on the design and sales requirements of the product. Performance testing has confirmed that differences in design do not result in any new or different questions of safety or effectiveness when compared to the predicate device.
#### Discussion 7: Principle of operation for the sharps protection feature
The proposed device and the predicate device operate on the same fundamental principle of action, despite minor differences in component nomenclature (e.g., “needle shield” vs. “needle cap”). In both devices, a passive shield retracts during skin insertion, automatically extends upon needle withdrawal, and locks permanently after use, thereby preventing reuse and reducing the risk of accidental needlestick injuries. The claimed terminological variations do not affect the physical sequence of operation, the clinical functional outcome, or the safety mechanism.
#### **The Conclusions:**
The conclusions drawn from the non-clinical tests demonstrate that the devices are as safe, as effective, and performs as well as the legally marketed predicate device identified in the submission. Thus the subject device is substantially equivalent to the predicate device.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.