K260850 · Osteomni, Inc. · ODP · May 12, 2026 · Orthopedic
Device Facts
Record ID
K260850
Device Name
OSTEOMNI SPINAL CAGES SYSTEM
Applicant
Osteomni, Inc.
Product Code
ODP · Orthopedic
Decision Date
May 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OSTEOMNI CERVICAL PEEK CAGES are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six (6) weeks of prior non-operative treatment. Cervical Peek Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and packed with autogenous bone. OSTEOMNI CERVICAL PEEK CAGES are to be used with supplemental fixation. OSTEOMNI CERVICAL Ti CAGES are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six (6) weeks of prior non-operative treatment. Cervical Ti- Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and packed with autogenous bone. OSTEOMNI CERVICAL Ti CAGES are to be used with supplemental fixation. OSTEOMNI LUMBAR PEEK CAGES are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region. OSTEOMNI LUMBAR Ti CAGES are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region. OSTEOMNI CERVICAL CORPECTOMY CAGE is intended for use in skeletally mature patients in the cervical spine (C2-T1) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders. OSTEOMNI LUMBAR CORPECTOMY CAGE is intended for use in skeletally mature patients in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders. OSTEOMNI CORPECTOMY CAGES are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. OSTEOMNI CORPECTOMY CAGES are intended to be used with supplemental spinal fixation systems cleared for use in the cervical, thoracic, and/or lumbar spine. The use of bone grafting material is optional.
Device Story
System of intervertebral body fusion (IBFD) and corpectomy cages; designed to restore spinal height and lordotic angle; maintains disc space stability until bony fusion occurs. Implants manufactured from PEEK and/or titanium alloy; inner chamber packed with autogenous bone graft. Used in surgical settings by surgeons; requires supplemental spinal fixation. Cervical cages placed via anterior approach. Corpectomy cages replace collapsed/unstable vertebral bodies due to tumor, trauma, or infection; can provide structural integrity in palliative tumor cases without fusion. Benefits include spinal stabilization and decompression of neural tissues.
Clinical Evidence
No clinical data; bench testing only. Substantial equivalence supported by design identity to predicate device.
Technological Characteristics
Materials: PEEK and titanium alloy. Form factor: Intervertebral body fusion cages and corpectomy cages. Principle: Mechanical stabilization of spinal segments/vertebral bodies. Supplemental fixation required. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature adults with degenerative disc disease (DDD) of cervical spine (C2-T1) with radicular symptoms (failed 6 weeks non-operative treatment) or lumbar spine (L2-S1) with up to Grade 1 spondylolisthesis/retrolisthesis (failed 6 months non-operative treatment). Also indicated for vertebral body replacement in cervical (C2-T1) or thoracolumbar (T1-L5) spine due to tumor, osteomyelitis, or trauma; or for spinal column integrity in advanced stage tumors where fusion is not feasible.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 12, 2026
Osteomni, Inc.
Gulten Shurdom
Director
9956 NW 89th Ct.
Medley, Florida 33178
Re: K260850
Trade/Device Name: Osteomni Spinal Cages System
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: ODP, OVE, MAX, MQP, PLR
Dated: March 16, 2026
Received: March 16, 2026
Dear Gulten Shurdom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260850 - Gulten Shurdom
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260850 - Gulten Shurdom
Page 3
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Page 1 of 2
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260850 | ? |
| Please provide the device trade name(s). | | ? |
| OSTEOMNI SPINAL CAGES SYSTEM | | |
| Please provide your Indications for Use below. | | ? |
| OSTEOMNI CERVICAL PEEK CAGES are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six (6) weeks of prior non-operative treatment. Cervical Peek Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and packed with autogenous bone. OSTEOMNI CERVICAL PEEK CAGES are to be used with supplemental fixation. OSTEOMNI CERVICAL Ti CAGES are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six (6) weeks of prior non-operative treatment. Cervical Ti- Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and packed with autogenous bone. OSTEOMNI CERVICAL Ti CAGES are to be used with supplemental fixation. OSTEOMNI LUMBAR PEEK CAGES are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region. OSTEOMNI LUMBAR Ti CAGES are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region. OSTEOMNI CERVICAL CORPECTOMY CAGE is intended for use in skeletally mature patients in the cervical spine (C2-T1) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders. OSTEOMNI LUMBAR CORPECTOMY CAGE is intended for use in skeletally mature patients in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders. OSTEOMNI CORPECTOMY CAGES are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. | | |
{4}
Page 2 of 2
OSTEOMNI CORPECTOMY CAGES are intended to be used with supplemental spinal fixation systems cleared for use in the cervical, thoracic, and/or lumbar spine. The use of bone grafting material is optional.
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| --- | --- | --- |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
K260850 - Page 1 of 2
| 510(k) #: | 510(k) Summary | Prepared on: 2026-05-01 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | OSTEOMNI INC. | |
| Applicant Address | 9956 NW 89TH CT MEDLEY, FL, 33178 MEDLEY FL 33178 United States | |
| Applicant Contact Telephone | +1 7869993120 | |
| Applicant Contact | Mrs. GULTEN SHURDOM | |
| Applicant Contact Email | cy@osteomni.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | OSTEOMNI SPINAL CAGES SYSTEM | |
| Common Name | Intervertebral Body Fusion Device; Vertebral Body Replacement Device | |
| Classification Name | Intervertebral Body Fusion Device; Spinal Intervertebral Body Fixation Orthosis | |
| Regulation Number | 21 CFR 888.3080; 21 CFR 888.3060 | |
| Product Code(s) | ODP, OVE, MAX, MQP, PLR | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K252686 | EffortMed PEEK Cages & Corpectomy Cages | ODP, OVE, M |
| | | |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The OSTEOMNI SPINAL CAGES SYSTEM includes cervical intervertebral body fusion (IBFD), transforaminal lumbar interbody fusion (TLIF), and posterior lateral interbody fusion (PLIF) devices. The system was designed to restore height and lordotic angle in the spine. The main role of the cages is to help maintain the cleared disc space stable and intact, until a healthy bony fusion occurs between the adjoining vertebrae. To help achieve this, the inner chamber of the cage body is filled with bone graft before implantation. The system also includes cervical and thoracolumbar corpectomy cages used in cervical spine (C2-T1) and in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to, tumors, fractures, and infections. Implants are manufactured from PEEK and/or titanium alloy. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| OSTEOMNI CERVICAL PEEK CAGES are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six (6) weeks of prior non-operative treatment. Cervical Peek Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and packed with autogenous bone. OSTEOMNI CERVICAL PEEK CAGES are to be used with supplemental fixation. OSTEOMNI CERVICAL Ti CAGES are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six (6) weeks of prior non-operative | | |
{6}
K260850 - Page 2 of 2
treatment. Cervical Ti-Cages facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and packed with autogenous bone. OSTEOMNI CERVICAL Ti CAGES are to be used with supplemental fixation.
OSTEOMNI LUMBAR PEEK CAGES are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region.
OSTEOMNI LUMBAR Ti CAGES are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region.
OSTEOMNI CERVICAL CORPECTOMY CAGE is intended for use in skeletally mature patients in the cervical spine (C2-T1) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.
OSTEOMNI LUMBAR CORPECTOMY CAGE is intended for use in skeletally mature patients in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.
OSTEOMNI CORPECTOMY CAGES are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
OSTEOMNI CORPECTOMY CAGES are intended to be used with supplemental spinal fixation systems cleared for use in the cervical, thoracic, and/or lumbar spine. The use of bone grafting material is optional.
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| --- | --- |
| The indications for use for the OSTEOMNI SPINAL CAGES SYSTEM and the predicate device are the same. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| Design: The designs of the Cervical and Lumbar PEEK Cages and the Corpectomy Cages are identical to the previously cleared predicate devices (K252686), with the exception of additional heights being added to one cervical cage design. This submission adds titanium alloy versions the Cervical and Lumbar Cages. Sizes: Additional heights have been added to one of the cervical cage designs. Material: Titanium alloy versions of the Cervical and Lumbar Cages are introduced in this submission. Principles off operation: no difference | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| The designs of the Cervical and Lumbar PEEK Cages and the Corpectomy Cages are identical to the previously cleared predicate devices (K252686), with the exception of additional heights being added to one cervical cage design. This submission adds titanium alloy versions the Cervical and Lumbar Cages. The added designs do not represent new worst cases as compared to the previously cleared designs. Therefore, the non-clinical testing performed in support of the predicate device is applicable to the Osteomni devices. Clinical Testing - Not applicable Osteomni Spinal Cages System is substantially equivalent to the legally marketed predicate device identified above. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.