The Protaryx Transseptal Puncture Device is used to introduce various cardiovascular catheters to the heart, including the left side of the heart. The device enables left heart access through a puncture of the atrial septum during a transseptal catheterization procedure.
Device Story
Sterile, single-use introducer catheter system comprising sheath, dilator, and pigtail guidewire; used by electrophysiologists and interventional cardiologists in catheter labs. Device facilitates left heart access via atrial septal puncture during transseptal catheterization. Operates under fluoroscopic and echocardiographic guidance. Combines features of RF-based access systems to enable catheter introduction and angiography. Clinical benefit includes minimally invasive access to left heart chambers for diagnostic or therapeutic procedures.
Clinical Evidence
Bench testing only. Includes mechanical verification (ISO 10555-1), biological safety (ISO 10993-1), sterilization validation (ISO 11135-1), packaging/shelf-life validation (ANSI/AAMI/ISO 11607), and pyrogen testing. Usability was assessed via a GLP animal study.
Technological Characteristics
Sterile, single-use introducer catheter (sheath, dilator, pigtail guidewire). Materials evaluated per ISO 10993-1. Sterilization via Ethylene Oxide (SAL 10^-6) per ISO 11135-1. Mechanical compliance with ISO 10555-1. Packaging validated per ANSI/AAMI/ISO 11607.
Indications for Use
Indicated for patients requiring transseptal catheterization to access the left heart for cardiovascular procedures; used to introduce catheters to the heart chambers.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
SafeCross Transseptal Puncture Device and Introducer System (TSP/I) (K203459)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2026
Protaryx Medical, Inc
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Dr. Suite #510k
Saint Paul, Minnesota 55114
Re: K260839
Trade/Device Name: Protaryx Transseptal Puncture Device (PTX2-001)
Regulation Number: 21 CFR 870.1340
Regulation Name: Catheter introducer
Regulatory Class: Class II
Product Code: DYB, DXF
Dated: February 23, 2025
Received: March 13, 2026
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260839 - Prithul Bom
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260839 - Prithul Bom
Page 3
Sincerely,
Katherine N. Trivedi - S
Digitally signed by
Katherine N. Trivedi - S
Date: 2026.04.10
16:30:06 -06'00'
Katherine Trivedi
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260839 | |
| Device Name Protaryx Transseptal Puncture Device (PTX2-001) | |
| Indications for Use (Describe) The Protaryx Transseptal Puncture Device is used to introduce various cardiovascular catheters to the heart, including the left side of the heart. The device enables left heart access through a puncture of the atrial septum during a transseptal catheterization procedure. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Protaryx
# 510(K) SUMMARY
K260839
## Submitter Information
Company Name: Protaryx Medical, Inc.
Company Address: 323 West Camden Street, Suite 600, Baltimore, MD 21201, USA
Company Phone: +1 (514) 554-9628
Company Facsimile: N/A
Contact Person: David Mester, CEO
Summary Prepared on: April 10, 2026
## Device Identification
Device Trade Name: Protaryx Transseptal Puncture Device (PTX2-001)
Device Common Name: Introducer
Classification Name: Introducer, Catheter (21 CFR 870.1340)
Product Code: DYB and DXF
Review Panel: Cardiovascular
Device Class: Class II
## Identification of Legally Marketed Device(s)
Primary Predicate Device: SafeCross Transseptal Puncture Device and Introducer System (TSP/I)
Manufacturer: East End Medical Inc.
510(k): K203459
Indications for Use: The SafeCross Transseptal Puncture Device and Introducer (TSP/I) System is used to introduce various cardiovascular catheters to the heart, including the left side of the heart. The system enables left heart access through a puncture of the atrial septum during a transseptal catheterization procedure. In addition, the device can be used for monitoring intracardiac pressures, sampling blood, and infusing solutions.
Secondary Predicate Device: VersaCross RF Wire (ProTrack RF Anchor Wire)
Manufacturer: Boston Scientific Inc.
510(k): K242076
Indications for Use: The ProTrack RF Anchor Wire is indicated for the creation of an atrial septal defect in the heart.
Reference Device: VersaCross Transseptal Sheath
Manufacturer: Boston Scientific Inc.
510(k): K183655
Indications for Use: The VersaCross Transseptal Sheath is used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation /puncture.
## Indications for Use
The Protaryx Transseptal Puncture Device is used to introduce various cardiovascular catheters to the heart, including the left side of the heart. The device enables left heart access through a puncture of the atrial septum during a transseptal catheterization procedure.
## Device Description
The Protaryx Transseptal Puncture device is a sterile, single-use introducer catheter device. The device is comprised of a sheath, a dilator, and a pigtail guidewire.
The Protaryx Transseptal Puncture device is designed for catheterization and angiography of specific heart chambers and locations. It is used in catheterization procedures primarily by Electrophysiologists and Interventional Cardiologists. Procedures using the devices are performed in fully equipped catheter labs with imaging equipment, including fluoroscopy and echocardiography under sterile technique.
{5}
Protaryx
# Comparison of Predicate Devices
The intended use for the Protaryx Transseptal Puncture device is substantially equivalent to the primary predicate device - East End Medical SafeCross device. The secondary predicate device and reference device - Boston Scientific VersaCross RF Wire (ProTrack RF Anchor Wire) and Boston Scientific VersaCross Transseptal Sheath were used to support the basis for substantial equivalence.
The Protaryx Transseptal Puncture Device combines the features/ functions of these devices into a single device. The proposed and predicate devices share the same fundamental scientific technology, including principles of operation and mechanism of action (see Table 1 below) using an RF Guidewire as the means of access. The differences in design and technological characteristics between the proposed and predicate devices do not raise different questions of safety and effectiveness. The results of verification and validation testing demonstrate substantial equivalence of the Protaryx Transseptal Puncture device with the combined predicate devices.
Table 1: Comparison of Subject and Predicate Device
| Characteristic | Comparison Results |
| --- | --- |
| Intended Use | Identical |
| Indications for Use | Similar |
| Fundamental Scientific Technology | Identical |
| Operating Principles | Identical |
| Mechanism of Action | Identical |
| Materials | Similar |
| Technological Aspects | Similar |
| Packaging and Sterilization | Similar |
# Performance Testing Summary
Performance Testing has been completed to demonstrate substantial equivalence of the subject device and predicate device. All test requirements were met as specified by applicable standards and the test protocols. The device was subjected to the following verification and validation activities.
- Performance and Physical Testing:
Mechanical verification testing was conducted for the subject Protaryx Transseptal Puncture Device to ensure compliance with the requirements of ISO 10555-1 and Protaryx Medical, LLC self-enforced requirements. The following tests were performed:
- Dimensional verification
- Simulated Use
- Catheter Bond Strength
- Tip Pull
- Tensile Strength
- Flexibility and Kink
- Torque Strength/Response
- Corrosion resistance
- Shapeability
- Surface Defects
- Hemostasis testing
- Usability testing (via a GLP (21CFR58) Animal study): Biological Safety Testing
The biological safety of the subject device was verified in accordance with the requirements of ISO 10993-1:2009/Cor.1:2010 and the September 2023 FDA guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".
- Cytotoxicity
- Sensitization
- Intracutaneous Reactivity
{6}
Protaryx
- Acute Systemic Toxicity
- Hemocompatibility (Hemolysis, Complement Activation)
- Thromboresistance
- Material Mediated Pyrogenicity
- Sterilization
Sterilization verification was completed for the subject device to the requirements of ISO 11135-1:2014. Sterilization is performed with Ethylene Oxide to a Sterility Assurance Level (SAL) of $10^{-6}$. Residual limits are in accordance with ISO 10993-7:2008/Cor.1:2009.
- Packaging Validation and Shelf Life
- Visual inspection, Bubble Leak and Seal Strength testing was used to evaluate the integrity of the packaging configuration. Testing was conducted after sterilization, environmental conditioning, and simulated shipping and distribution. Ship testing was performed to ensure the integrity of the device packaging through the rigors of shipping and handling. The seal strength and sterile barrier integrity was validated per ANSI/AAMI/ISO 11607:2006 (Parts 1 and 2) over the proposed shelf life of the device.
- Pyrogen Testing
- The subject device is supplied non-pyrogenic. LAL testing using the Kinetic Chromogenic method was conducted to ensure the device meets current FDA and USP pyrogen limit specifications.
- Bench-top Validation.
- Bench-top validation testing was conducted to assess valve durability of the sheath, radiopacity, compatibility with other devices and to evaluate the design and function.
# Conclusion
The intended use and fundamental scientific technology, including principles of operation and mechanism of action, of the Protaryx Transseptal Puncture device is equivalent in form, fit and function to that of the cited Predicates (Safe Cross Transseptal Puncture and Introducer (TSP/I), the VersaCross RF Wire (ProTrack RF Anchor Wire) and the VersaCross Transseptal Sheath). Differences in design and technological characteristics do not raise any different questions of safety and effectiveness. The results of verification and validation activities support the substantial equivalence of the Protaryx Transseptal Puncture Device to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.