# Gastrointestinal Videoscope Olympus GIF-2T1500 (K260824)

_Olympus Medical Systems Corporation · FDS · Aug 27, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K260824

## Device Facts

- **Applicant:** Olympus Medical Systems Corporation
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** Aug 27, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-2T1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

## Device Story

Reusable gastrointestinal videoscope; utilizes CMOS RGB color filter image sensor; features two instrument channels (2.8 mm and 3.7 mm) for accessory insertion, fluid introduction, or suction (2.8 mm channel only). Used with EVIS X1 Video System Center (CV-1500), monitor, and ancillary equipment. Operated by clinicians in clinical settings for visualization and endoscopic surgery. Provides NBI, RDI, and WLI imaging modes; includes electronic zoom, pre-freeze, and auxiliary water feeding. Output displayed on monitor for real-time visualization, aiding diagnosis and therapeutic intervention. Reprocessing via MAJ-420 connection tube.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (durability, thermal safety, dimensional analysis, photobiological safety, fluid rates, optical performance, video latency) and an animal study evaluating NBI, TXI, RDI, and BAI-MAC imaging functions. Biocompatibility, electrical safety (IEC 60601 series), and software/cybersecurity validation were also performed.

## Technological Characteristics

Reusable gastrointestinal videoscope; CMOS RGB color filter sensor; two instrument channels (2.8 mm/3.7 mm). Compatible with EVIS X1 CV-1500. Features NBI, RDI, WLI, TXI, and BAI-MAC imaging. Electrical safety per IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, and IEC TR 60601-4-2. Biocompatibility per ISO 10993-1. Reprocessing via MAJ-420 tube.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- FUJIFILM Endoscope Model EI-740D/S ([K221551](/device/K221551.md))

## Reference Devices

- [K112680](/device/K112680.md)
- [K232997](/device/K232997.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG ADMINISTRATION

August 27, 2026

Olympus Medical Systems Corporation
% Brenda Geary
Director, Regulatory Affairs
Olympus Corporation of the Americas
800 W. Park Dr.
Westborough, Massachusetts 01581

Re: K260824

Trade/Device Name: Gastrointestinal Videoscope Olympus GIF-2T1500

Regulation Number: 21 CFR 876.1500

Regulation Name: Endoscope And Accessories

Regulatory Class: Class II

Product Code: FDS, NWB

Dated: July 28, 2026

Received: July 28, 2026

Dear Brenda Geary:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K260824 - Brenda Geary

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K260824 - Brenda Geary

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# SHANIL P. HAUGEN -S

Shanil P. Haugen, Ph.D.

Assistant Director

DHT3A: Division of Renal, Gastrointestinal,

Obesity and Transplant Devices

OHT3: Office of GastroRenal, ObGyn,

General Hospital and Urology Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260824 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Gastrointestinal Videoscope Olympus GIF-2T1500  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-2T1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

OLYMPUS®

Traditional 510(k)
Gastrointestinal Videoscope Olympus GIF-2T1500

K260824
Page 1 of 4

# 510(k) Summary

## I. Submitter

Olympus Medical Systems Corporation

2951 Ishikawa-cho

Hachioji-shi

Tokyo 192-8507

Japan

Contact Person: Shinichiro Kawachi

Phone: +81 42-642-2111

Date Prepared: August 5, 2026

## II. Device

|  Device Proprietary Names | Gastrointestinal Videoscope Olympus GIF-2T1500  |
| --- | --- |
|  Common or Usual Names | Endoscope  |
|  Classification Name | Endoscope and Accessories  |
|  Regulation Number | 876.1500  |
|  Product Codes | FDS, NWB  |
|  Device Classification | Class II  |

## III. Predicate Device

Substantial equivalence is claimed to the following device:

- FUJIFILM Endoscope Model EI-740D/S, K221551, Fujifilm Corporation

## IV. Device Description

The Gastrointestinal Videoscope Olympus GIF-2T1500 (GIF-2T1500) is a non-sterile, reusable device, intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equivalent for endoscopy and endoscopic surgery. The device is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

Page 1 of 4

{5}

K260824  
Page 2 of 4

**OLYMPUS®**

**Traditional 510(k)  
Gastrointestinal Videoscope Olympus GIF-2T1500**---

The GIF-2T1500 features a CMOS (RGB color filter) image sensor and two (2) instrument channels (Ø 2.8 mm and 3.7 mm). The Ø 2.8 mm channel runs from the distal end to the connector and is used for inserting endoscopic accessories, introducing liquid from a syringe attached to the biopsy valve, or aspirating fluids, debris, or air from the lumen. The Ø 3.7 mm channel runs from the distal end to the instrument channel port of the control section and is used for inserting endoscopic accessories or introducing liquid from a syringe attached to the biopsy valve; this channel is not connected to the suction channel, therefore, cannot be used to aspirate the lumen. The endoscope consists of three parts: the insertion section, the control section, and the connector section.

The GIF-2T1500 is compatible with the EVIS X1 Video System Center (CV-1500).

The MAJ-420 Channel Connection Tube is a reprocessing accessory that attaches to the GIF-2T1500 instrument channel ports and is used to inject reprocessing fluids into both instrument channels of the endoscope. The tube can also be used to flush air through both channels to expel fluids.

## V. Indications for Use

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-2T1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

## VI. Comparison of Technological Characteristics

The Gastrointestinal Videoscope Olympus GIF-2T1500 is substantially equivalent to the Endoscope Model EI-740D/S (K221551) as the subject and predicate devices have the same intended use and similar design and technological characteristics.

The subject device and predicate device are gastroscopes featuring the same direction of view and field of view, working length, and bending angulations. The devices include two instrument channels, and the distal ends are equipped with the same functions. Both endoscopes maintain records of device information and provide pre-freeze (anti-blur), electronic zoom, high frequency cauterization, and auxiliary water feeding functions.

Page 2 of 4

{6}

OLYMPUS®

Traditional 510(k)

Gastrointestinal Videoscope Olympus GIF-2T1500

K260824

Page 3 of 4

When compared to the predicate device, the subject device utilizes a different image sensor (CMOS/RGB Color Filer vs. CCD) and has a slightly larger depth of field (2 - 100 mm vs 3 - 100 mm), a smaller overall diameter, and a longer total length (1350 mm vs. 1330 mm). Although the subject device's distal end outer diameter, maximum insertion section outer diameter, and insertion tube outer diameter are slightly smaller than those of the predicate device, they fall within the range of outer diameters for other legally marketed gastroscopes indicated for use in the upper digestive tract (e.g., K112680 and K232997). The subject device's first instrument channel has a smaller diameter (2.8 mm vs 3.2 mm) and Endo Therapy accessories enter and exit the subject device's endoscopic image from this channel at an angle. Although the subject device's first instrument channel diameter is smaller than that of the predicate device, it is the same diameter provided for other single channel gastroscopes indicated for use in the upper digestive tract (e.g., K112680). The subject device provides narrow-band imaging (NBI), red dichromatic imaging (RDI), and white light imaging (WLI) while the predicate device provides linked color imaging (LCI) and blue laser imaging (BLI). These technological differences do not raise different questions of safety and effectiveness when compared to the predicate device.

Validation of reprocessing instructions, a biological safety evaluation, software validation and cybersecurity testing, electrical safety and electromagnetic compatibility testing, and non-clinical testing, including comparative bench and animal studies were undertaken to demonstrate substantial equivalence to the predicate device.

### VII. Performance Data

The following performance data were provided to demonstrate substantial equivalence of the GIF-2T1500:

- Reprocessing method validation per the FDA's Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling"
- Biocompatibility testing per ISO 10993-1:2018 including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021), irritation (ISO 10993-23:2021)
- Software verification and validation and cybersecurity testing per the FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
- Electrical safety and EMC testing per IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, and IEC TR 60601-4-2

Page 3 of 4

{7}

OLYMPUS®

K260824

Page 4 of 4

Traditional 510(k)

Gastrointestinal Videoscope Olympus GIF-2T1500

• Mechanical testing and comparative testing to verify device performance

○ Durability
○ Thermal safety
○ Dimensional analysis
○ Photobiological safety
○ Transportation
○ Rates of air, water, and suction (Verification/Comparison)
○ Color performance
○ Field of View (FOV) and Direction of View (DOV)
○ Distortion (Verification/Comparison)
○ Resolution (Verification/Comparison)
○ Depth of Field (DOF)
○ Noise and dynamic range (Verification/Comparison)
○ Image Intensity Uniformity (Verification/Comparison)
○ Video Latency (Verification/Comparison)
○ Distal Attachment and Detachment

• An animal study was conducted to evaluate the performance of the Narrow Band Imaging (NBI) observation, TeXture and color enhancement Imaging (TXI), and Red Dichromatic Imaging (RDI) observation as well as the Brightness Adjustment Imaging with Maintenance of Contrast (BAI-MAC) function.

Clinical data is not required to demonstrate substantial equivalence.

### VIII. Conclusion

The non-clinical data demonstrate that the subject devices are as safe, as effective, and perform as well as or better than the identified predicate devices. Therefore, it is concluded that the Gastrointestinal Videoscope Olympus GIF-2T1500 is substantially equivalent to the identified predicate device.

Page 4 of 4

---

**Source:** [https://fda-staging.innolitics.com/device/K260824](https://fda-staging.innolitics.com/device/K260824)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).