← Product Code [MQB](/productcode/MQB) · K260820

# VIVIX-S 4386W (FXRD-4386WA) (K260820)

_Vieworks Co., Ltd. · MQB · Jun 18, 2026 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K260820

## Device Facts

- **Applicant:** Vieworks Co., Ltd.
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** Jun 18, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.

## Device Story

VIVIX-S 4386W (FXRD-4386WA) is a flat panel X-ray detector; 43cm x 86cm imaging area. Device intercepts X-ray photons; scintillator emits visible light; amorphous silicon (a-SI) array converts light to electrical signals; signals digitized and processed by software. Integrated with operating PC and X-ray generator; used in clinical settings for diagnostic radiography. Output displayed on monitors for physician review; facilitates diagnostic decision-making by replacing film/screen systems. Benefits include digital image acquisition and transfer.

## Clinical Evidence

Bench testing only. Evaluated imaging performance parameters (MTF, DQE) comparing subject device to predicate. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. No clinical data provided.

## Technological Characteristics

Flat panel detector; amorphous silicon (a-SI) photodetector array; Cesium Iodide (CsI) scintillator. Dimensions: 43cm x 86cm. Connectivity: integrated with PC and X-ray generator. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC).

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- VIVIX-S 4386W ([K220239](/device/K220239.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** **U.S. FOOD & DRUG**  
ADMINISTRATION

June 18, 2026

Vieworks Co., Ltd  
% Juin Lee  
Regulatory Affairs Specialist  
Vieworks Co., Ltd.  
41-3, Burim-Ro 170beon-Gil, Dongan-Gu, Anyang-Si,  
GYEONGGI-DO, KOREA 14055

Re: K260820

Trade/Device Name: VIVIX-S 4386W(FXRD-4386WA)  
Regulation Number: 21 CFR 892.1680  
Regulation Name: Stationary X-Ray System  
Regulatory Class: Class II  
Product Code: MQB  
Dated: May 21, 2026  
Received: May 21, 2026

Dear Juin Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}

K260820 - Juin Lee

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K260820 - Juin Lee

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

GABRIELA M. RODAL -S Digitally signed by GABRIELA M. RODAL -S for

Lu Jiang, Ph.D.

Assistant Director

Diagnostic X-Ray Systems Team

DHT8B: Division of Radiological Imaging Devices and Electronic Products

OHT8: Office of Radiological Health

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

# Indications for Use

510(k) Number (if known)
K260820

Device Name
VIVIX-S 4386W (FXRD-4386WA)

Indications for Use (Describe)

VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}

510(k) #: K260820

# 510(k) Summary

Prepared on: 2026-05-21

## Contact Details

21 CFR 807.92(a)(1)

Applicant Name

Vieworks Co., Ltd.

Applicant Address

41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do 14055 Korea, Republic of

Applicant Contact Telephone

+827088764616

Applicant Contact

Ms. Juin Lee

Applicant Contact Email

jhlee@vieworks.com

## Device Name

21 CFR 807.92(a)(2)

Device Trade Name

VIVIX-S 4386W (FXRD-4386WA)

Common Name

Solid State X-Ray Imager (Flat Panel/Digital Imager)

Classification Name

Stationary x-ray system

Regulation Number

892.1680

Product Code(s)

MQB

## Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #

Predicate Trade Name (Primary Predicate is listed first)

Product Code

K220239

VIVIX-S 4386W

MQB

## Device Description Summary

21 CFR 807.92(a)(4)

VIVIX-S 4386W, a flat panel detector model named; FXRD-4386WA with imaging areas of 43cm x 86cm. The device intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create electrical signals. After the electrical signals are generated, it is converted to digital value, and the Software which acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors. These devices should be integrated with an operating PC and an X-Ray generator.

It can be utilized to digitalize x-ray images and transfer for radiography diagnostic.

## Intended Use/Indications for Use

21 CFR 807.92(a)(5)

VIVIX-S 4386W is indicated for digital imaging solution designed as a general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purposes of diagnostic procedures. It is not to be used for mammography. It is intended for both adult and pediatric populations.

## Indications for Use Comparison

21 CFR 807.92(a)(5)

The modification made on the subject device(Model Name: FXRD-4386WA) has no effect on the indications for use, therefore the indications for use remain same as the predicate device.

## Technological Comparison

21 CFR 807.92(a)(6)

The subject device incorporates a modified scintillator (GadOx to CsI) compared to the predicate device, representing a change in technological characteristics. As a result, differences are observed in key imaging performance parameters, including DQE and MTF.

{5}

However, the subject device employs the same fundamental imaging principles and detector architecture as the predicate device. Performance evaluation demonstrates that DQE and MTF remain comparable and within acceptable ranges for the intended use, and these differences do not raise new questions of safety or effectiveness. Therefore, the subject device is considered substantially equivalent to the predicate device.

## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Non-clinical performance testing was conducted to compare the subject device with the predicate device. Imaging performance parameters, including modulation transfer function (MTF) and detective quantum efficiency (DQE), were evaluated. The subject device demonstrated slightly higher values under the specified test conditions, indicating performance comparable to or better than the predicate device.

Electrical safety and electromagnetic compatibility (EMC) testing were conducted in accordance with IEC 60601-1 and IEC 60601-1-2, and all applicable requirements were met.

In summary, the subject device, FXRD-4386WA, has the same indications for use and similar technological characteristics as the predicate device and does not raise any new questions of safety or effectiveness. Therefore, the device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K260820](https://fda-staging.innolitics.com/device/K260820)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).