K260804 · Momentis Surgical , Ltd. · SCV · Jul 22, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260804
Device Name
Anovo Surgical System (6Ne)
Applicant
Momentis Surgical , Ltd.
Product Code
SCV · General, Plastic Surgery
Decision Date
Jul 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4964
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Anovo Surgical System is an endoscopic instrument control system that is intended to assist in the accurate control of the Instrument ARMS during laparoscopic-assisted benign surgical procedures listed below. The Anovo Surgical System is indicated for use in adult patients. It is intended to be used by trained physicians in an operating room environment. The representative uses of the Anovo Surgical System are indicated for the following single-site, natural orifice, transvaginal benign procedures: - Total benign hysterectomy with salpingo-oophorectomy - Total benign hysterectomy with salpingectomy - Total benign hysterectomy - Salpingectomy - Oophorectomy - Adnexectomy - Ovarian cyst removal Single-site and multi-port transabdominal benign procedures: - Ventral hernia
Device Story
Robotic-assisted endoscopic instrument control system; assists surgeons in performing minimally invasive procedures. System comprises surgeon console, robotic control units (RCU), instrument arms, and support pedestals. Surgeon operates console to control instrument arms; system integrates with standard electrosurgical generators and visualization systems. Modification (6Ne) enables multi-port transabdominal access by separating control units into two floor-mounted pedestals; only one RCU active at a time using two instrument arms. Optional SOLOASSIST II endoscope arm allows console-based camera control. Used in operating rooms by trained physicians. Provides precise instrument manipulation; enhances surgical access and control; facilitates complex benign procedures via robotic assistance.
Clinical Evidence
No clinical data. Evidence includes bench testing (dimensional, mechanical, functional verification), cybersecurity testing, electrical safety/EMC testing (IEC 60601 series), human factors/usability validation in simulated OR environment with general surgeons, and design validation via female cadaver model to confirm workspace, access, and collision avoidance.
Technological Characteristics
Modular electromechanical surgical system. Components: surgeon console, robotic control units, instrument arms, multi-port pedestals. Connectivity: electrical wiring. Integration: standard electrosurgical generators and laparoscopic visualization systems. Standards: IEC 60601-1, IEC 60601-1-6, IEC 80601-2-77, IEC 60601-1-2, IEC 60601-4-2, IEC 60601-2-2. Software: robotic control and safety monitoring.
Indications for Use
Indicated for adult patients undergoing laparoscopic-assisted benign surgical procedures, including single-site natural orifice transvaginal procedures (hysterectomy, salpingo-oophorectomy, salpingectomy, oophorectomy, adnexectomy, ovarian cyst removal) and single-site or multi-port transabdominal procedures (ventral hernia).
Regulatory Classification
Identification
The Versius Surgical System is a robotically assisted surgical device that is intended to assist in the precise and accurate control of Versius Surgical endoscopic instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, forceps/pick-ups, needle holders, electrosurgery, and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrosurgery and suturing. It is indicated for adult patients 22 years of age and older, eligible for soft tissue minimal access surgery, for cholecystectomy.
Special Controls
In combination with the general controls of the FD&C Act, the modular electromechanical surgical system is subject to the following special controls:
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 22, 2026
Momentis Surgical, Ltd.
Maya Leib Shlomo
VP Qa/ra
6 Yoni Netanyahu St.
Or Yehuda, 6037604
Israel
Re: K260804
Trade/Device Name: Anovo Surgical System (6Ne)
Regulation Number: 21 CFR 878.4964
Regulation Name: Modular electromechanical surgical system
Regulatory Class: Class II
Product Code: SCV
Dated: March 11, 2026
Received: March 12, 2026
Dear Maya Leib Shlomo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260804 - Maya Leib Shlomo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260804 - Maya Leib Shlomo
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARK
TRUMBORE -S
Digitally signed by MARK TRUMBORE -S
Date: 2026.07.22 09:26:15 -04'00'
Mark Trumbore, Ph.D.
Assistant Director, THT4A1: Robotically-Assisted Surgical Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K260804
Device Name
Anovo Surgical System (6Ne)
Indications for Use (Describe)
The Anovo Surgical System is an endoscopic instrument control system that is intended to assist in the accurate control of the Instrument ARMS during laparoscopic-assisted benign surgical procedures listed below.
The Anovo Surgical System is indicated for use in adult patients. It is intended to be used by trained physicians in an operating room environment.
The representative uses of the Anovo Surgical System are indicated for the following single-site, natural orifice, transvaginal benign procedures:
- Total benign hysterectomy with salpingo-oophorectomy
- Total benign hysterectomy with salpingectomy
- Total benign hysterectomy
- Salpingectomy
- Oophorectomy
- Adnexectomy
- Ovarian cyst removal
Single-site and multi-port transabdominal benign procedures:
- Ventral hernia
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K260804
| Submitter | Momentis Surgical Ltd. 6 Yoni Netanyahu St. Or Yehuda, Israel 6037604 |
| --- | --- |
| Contact Person: | Maya Leib Shlomo, VP of QA/RA Maya.leib@momentissurgical.com Tel.: 972-5-088-52822 |
| Date: | March 11^{th}, 2026 |
| Device Name: | Anovo Surgical System model 6Ne |
| Classification: | 21 CFR 878.4964 Modular electromechanical surgical system, Product Code SCV, Class 2 |
| Predicate: | Versius Surgical System (DEN230078) |
| Reference Device: | Anovo Surgical System model 6Ne (K251761) |
| Description: | The Anovo Surgical System is a robotic-assisted device intended to assist in the accurate control of endoscopic instruments in minimally invasive surgery. The subject Anovo Surgical System 6Ne is a modification of the reference device (K251761) to include a multi-port approach in the device labeling for transabdominal Ventral Hernia procedures. To support multiport access, the Anovo Surgical System 6Ne modifications include providing an additional Robotic Control Unit (RCU) in which the two Control Units of the reference are now separated (Control Unit I and Control Unit II). Each of the Control Units is mounted on Multi-port Pedestals as a floor-mounted support system, while maintaining the established system architecture, intended use, and fundamental operating principles of the cleared system. Only one (1) RCU can be used at a time, and thus the surgery is performed with 2 Instruments ARMs. The system includes the Endoscope Arm (FDA-cleared SOLOASSIST II by AKTORmed), which is an off-the-shelf optional accessory that can be connected to the Surgeon Console and allows the surgeon to control the endoscope's movements via the Surgeon Console. The Endoscope Arm is attached to the rail of the surgical table or mounted on the Endoscope Arm Cart and covered by Sterile Drape. The Endoscope Arm is transported and stored using a Trolley. The Anovo Surgical System's components are all connected using electrical wiring. The system is designed to be used in conjunction with standard commercially available Electrosurgical Generators and standard commercially available laparoscopic surgery visualization systems. |
| Indications for Use: | The Anovo Surgical System is an endoscopic instrument control system that is intended to assist in the accurate control of the Instrument ARMS during laparoscopic-assisted benign surgical procedures listed below. The Anovo Surgical System is indicated for use in adult patients. It is intended to be used by trained physicians in an operating room environment. The representative uses of the Anovo Surgical System are indicated for the following single-site, natural-orifice, transvaginal benign procedures: |
Page 1 of 4
{5}
K260804
- Total benign hysterectomy with salpingo-oophorectomy
- Total benign hysterectomy with salpingectomy
- Total benign hysterectomy
- Salpingectomy
- Oophorectomy
- Adnexectomy
- Ovarian cyst removal
Single-site and multi-port transabdominal benign procedures:
- Ventral Hernia
# Comparison of Technological Characteristics:
The subject Anovo Surgical System 6Ne is substantially equivalent to the predicate Versius Surgical System. It also shares many technological features with the reference Anovo Surgical System (K251761), sharing the same intended use and similar technological characteristics, and principles of operation. All systems include a surgeon console, robotic units, support systems, instruments, and accessories, enabling independent control of surgical tools through abdominal ports under endoscopic visualization. The subject device supports a two-instrument, multiport configuration using two separate control units without changes to electrical functionality. It is therefore similar to the predicate in that it is a multimodular system. Technologically, the subject device also remains similar to the cleared Anovo Surgical System 6Ne (K251761), sharing the same fundamental design, principles of operation, and core technological characteristics. Additionally, there are no differences in the procedure for use of the system; the surgeon and surgical teams perform system set-up in the same manner, where the only difference is the multi-ports (specifically two ports) Vs. Single site of the instrument arms. Verification and validation testing, including bench, electrical safety and EMC, human factor, and cadaver studies, confirmed that the Anovo 6Ne is capable of performing its intended use. Therefore, the modifications have been evaluated for maintaining substantial equivalence.
# Performance Evaluation:
The following performance testing was conducted to demonstrate substantial equivalence to the predicate device:
# Bench Testing
Bench testing demonstrates that the subject device's design output meets the design input requirements. The testing conducted consisted of dimensional measurements and mechanical and functional verification and demonstrated that the Anovo System met its defined design requirements and specifications, including workspace, access, reachability, safety alerts, collision, and performance.
# Cybersecurity
Anovo Surgical System 6Ne implements robust security controls to safeguard the integrity and security of the system's operation. Cybersecurity testing was performed, and the information provided in this submission demonstrates compliance with section 524B of the FD&C Act, consistent with current "FDA Guidance: Cybersecurity in Medical Devices - Quality System Considerations and Content of Premarket Submissions".
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical Safety and EMC testing were conducted using a third-party
Page 2 of 4
{6}
K260804
Accredited laboratory in accordance with the current versions of IEC 60601-1 (basic safety and essential performance), IEC 60601-1-6 (usability), IEC 80601-2-77 (basic safety and essential performance of robotically assisted surgical equipment), IEC 60601-1-2 (Electromagnetic disturbances), IEC 60601-4-2 (Electromagnetic Immunity), and IEC 60601-2-2 (high-frequency surgical equipment).
# Human Factors
Human Factors and Usability Engineering Process was performed according to the requirements of:
- ISO/IEC 62366-1:2015, Medical devices – Application of usability engineering to medical devices.
- Guidance for Industry and Food and Drug Administration Staff - Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016)
- Guidance for Industry and Food and Drug Administration Staff - Content of Human Factors Information in Medical Device Marketing Submissions (May 29, 2026)
Representative US General Surgeons and OR Staff evaluated the Anovo Surgical System 6Ne, configured for multi-port, usage in a simulated OR environment by performing predefined critical tasks after a training session.
The Human Factor Usability Validation demonstrated that the Anovo Surgical System 6Ne, configured for multi-port access can be used by representative users during the performance of multi-port transabdominal laparoscopic-assisted Ventral Hernia surgical procedures, while maintaining an adequate range of motion, appropriate spatial separation of components, and absence of collisions, unintended contact, or hazardous interference. The analysis of the results demonstrated that all relevant use-related risks were found to be acceptable, and there is no residual use-related risk.
# Cadaver Study
Momentis has performed design validation using a female cadaver model to ensure that the Anovo Surgical System 6Ne, configured for multi-port access, meets its safety and performance requirements. The validation was performed by conducting procedures according to the intended use in an operating room environment.
A female cadaver model was selected for this study as it closely replicates human anatomy and provides a realistic environment for verifying proper setup and positioning of the multi-port carts, Control Units, and Endoscope Cart, as well as confirming appropriate camera orientation. The Anovo Surgical System 6Ne, configured for multi-port access used for Ventral Hernia procedures, met all the predefined specific requirements related to clinical compatibility, workspace, access, reach, collision avoidance, performance, and safety alerts.
# Conclusion:
The subject and the predicate device have the same intended use, similar indications for use, and technological characteristics. The minor differences in indications for use and technological characteristics do not raise different questions of safety or effectiveness. The modification to the Anovo Surgical System 6Ne is addressed with the same types of data as were used to support the predicate device, and the outcome of the performance evaluation (design verification and validation) does not present any new risks.
Page 3 of 4
{7}
# **K260804**
In addition, the modified device appropriately addresses all the special controls in the existing classification regulation. Therefore, the subject Anovo Surgical System Model 6Ne is substantially equivalent to its predicate device.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.