K260765 · Azena Medical, LLC · NVK · Mar 10, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260765
Device Name
Gemini NOVA 810+980 Soft Tissue Laser
Applicant
Azena Medical, LLC
Product Code
NVK · General, Plastic Surgery
Decision Date
Mar 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Dental Soft Tissue Indications Incision, excision, vaporization, ablation and coagulation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications: - Excisional and incisional biopsies - Exposure of unerupted teeth - Fibroma removal - Frenectomy - Frenotomy - Gingival troughing for crown impressions - Gingivectomy - Gingivoplasty - Gingival incision and excision - Hemostasis and coagulation - Implant recovery - Incision and drainage of abscess - Leukoplakia - Operculectomy - Oral papillectomies - Pulpotomy - Pulpotomy as an adjunct to root canal therapy - Reduction of gingival hypertrophy - Soft tissue crown lengthening - Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa - Treatment of aphthous ulcers. - Vestibuloplasty - Tissue retraction for impression - Lesion (tumor) removal. Laser Periodontal Procedures. - Laser soft tissue curettage. - Laser removal of diseased, Infected, Inflamed and necrosed soft tissue within the periodontal pocket. - Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium. - Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility) - Reduction of bacterial level (decontamination) and inflammation Pain therapy - Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain, the temporary increase in local blood circulation; the temporary relaxation of muscle.
Device Story
Dual-wavelength (810 nm/980 nm) GaAlAs diode laser system; delivers laser energy via fiber-optic tips or photobiomodulation (PBM) adapters for oral soft-tissue procedures and pain therapy. System comprises base unit, cordless wand, wireless foot pedal, and rechargeable lithium-ion batteries. Operated by dental professionals in clinical environments. Laser energy transforms into thermal energy at tissue interface to perform incision, excision, ablation, coagulation, or topical heating. Output controlled via capacitive touchscreen; wireless communication (BLE) links wand, base, and foot pedal. Benefits include precise soft tissue management, periodontal decontamination, and temporary pain relief. Device replaces corded handpiece of predicate with cordless design; utilizes PBM adapters for beam expansion/spot size control.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser performance (IEC 60601-2-22, IEC 60825-1), battery safety (IEC 62133-2), and biocompatibility (ISO 10993).
Technological Characteristics
GaAlAs diode laser; 810/980 nm wavelengths; 0.1-2.0 W average power; 20 W peak power. Components: capacitive touchscreen, BLE wireless connectivity, lithium-ion batteries. Accessories: disposable fiber tips, PBM adapters (3mm, 7mm, 25mm). Reprocessing: cleaning/disinfection for wand; steam sterilization for PBM adapters. Software: Moderate Level of Concern, IEC 62304 compliant.
Indications for Use
Indicated for dental patients requiring oral soft tissue procedures (incision, excision, ablation, coagulation, biopsy, frenectomy, gingivectomy, etc.), periodontal pocket debridement/decontamination, and topical heating for temporary relief of minor muscle/joint pain, stiffness, arthritis, or muscle spasm.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 10, 2026
Azena Medical, LLC
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120a
Madison, New Jersey 07940
Re: K260765
Trade/Device Name: Gemini NOVA 810+980 Soft Tissue Laser
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II
Product Code: NVK, GEX, ILY
Dated: March 8, 2026
Received: March 9, 2026
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260765 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260765 - Dave Yungvirt
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K260765
Device Name
Gemini NOVA 810+980 Soft Tissue Laser
Indications for Use (Describe)
Dental Soft Tissue Indications
Incision, excision, vaporization, ablation and coagulation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications:
- Excisional and incisional biopsies
- Exposure of unerupted teeth
- Fibroma removal
- Frenectomy
- Frenotomy
- Gingival troughing for crown impressions
- Gingivectomy
- Gingivoplasty
- Gingival incision and excision
- Hemostasis and coagulation
- Implant recovery
- Incision and drainage of abscess
- Leukoplakia
- Operculectomy
- Oral papillectomies
- Pulpotomy
- Pulpotomy as an adjunct to root canal therapy
- Reduction of gingival hypertrophy
- Soft tissue crown lengthening
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
- Treatment of aphthous ulcers.
- Vestibuloplasty
- Tissue retraction for impression
- Lesion (tumor) removal.
Laser Periodontal Procedures.
- Laser soft tissue curettage.
- Laser removal of diseased, Infected, Inflamed and necrosed soft tissue within the periodontal pocket.
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.
- Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Reduction of bacterial level (decontamination) and inflammation
Pain therapy
FORM FDA 3881 (6/20)
PISC Publishing Services (301) 443-6740
{4}
- Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain, the temporary increase in local blood circulation; the temporary relaxation of muscle.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 2 of 2
{5}
K260765
# 510(k) Summary
## Submitter Information (21 CFR 807.92)
Azena Medical, LLC
1777 Botelho Drive, Suite 100, Walnut Creek, CA 94596 USA
Contact Person: Lindsay Tilton
Telephone: (925) 699-9091
Date Prepared: March 4, 2026
**Device Trade Name:** Gemini Nova 810 + 980 Soft Tissue Laser
**Common Name:** Dental Laser, Soft-Tissue, Powered
**Classification:** Class II
**Regulation:** 21 CFR 878.4810
**Product Codes:** NVK (primary); GEX; ILY
## Predicate Device
Gemini 2 (Gemini EVO) 810 + 980 Soft Tissue Laser (K210350)
## Indications for Use
Dental Soft Tissue Indications
Incision, excision, vaporization, ablation and coagulation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications:
- Excisional and incisional biopsies
- Exposure of unerupted teeth
- Fibroma removal
- Frenectomy
- Frenotomy
- Gingival troughing for crown impressions
- Gingivectomy
- Gingivoplasty
- Gingival incision and excision
- Hemostasis and coagulation
- Implant recovery
- Incision and drainage of abscess
- Leukoplakia
- Operculectomy
- Oral papillectomies
- Pulpotomy
- Pulpotomy as an adjunct to root canal therapy
- Reduction of gingival hypertrophy
Page 1
{6}
- Soft tissue crown lengthening
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
- Treatment of aphthous ulcers.
- Vestibuloplasty
- Tissue retraction for impression
- Lesion (tumor) removal.
## Laser Periodontal Procedures
- Laser soft tissue curettage.
- Laser removal of diseased, Infected, Inflamed and necrosed soft tissue within the periodontal pocket.
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.
- Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Reduction of bacterial level (decontamination) and inflammation
## Pain therapy
- Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain, the temporary increase in local blood circulation; the temporary relaxation of muscle.
## Device Description
The Gemini Nova 810 + 980 Soft Tissue Laser is a dual-wavelength diode laser system delivering 810 nm and 980 nm laser energy for oral soft-tissue procedures. Laser energy is delivered through fiber-optic disposable tips or photobiomodulation (PBM) adapters. The system includes a base unit, wand, wireless foot pedal, and rechargeable battery packs.
## Comparison of Technological Characteristics
| Feature | Gemini Nova | Predicate Device (Gemini 2 K210350) |
| --- | --- | --- |
| Laser Type | GaAlAs diode | GaAlAs diode |
| Wavelengths | 810 ±10 nm, 980 ±10 nm | 810 ±10 nm, 980 ±10 nm |
| Peak Power | 20 W | 150 W |
| Average Power | 0.1 – 2.0 W | 0.1 – 2.0 W |
| Energy Delivery | Fiber tips and PBM adapters | Fiber tips and PBM adapters |
{7}
Page 3
| Feature | Gemini Nova | Predicate Device (Gemini 2 K210350) |
| --- | --- | --- |
| Activation | Wireless Foot Switch | Wireless Foot Switch |
| Accessories | Disposable fiber tips.
Safety eyewear; Aluminum hand piece; 25mm PBM Adapter with spacers, 7mm and 3 mm intraoral PBM adapters, Wand battery (2), Barrier sleeves (10 ea. fiber tips, 5 ea. 3mm and 5 ea. 7mm) and the light adapters for the PBM tips (5) | Disposable fiber tips;
Safety eyewear; Aluminum hand piece; 25mm PBM Adapter with spacers, 7mm and 3 mm intraoral PBM adapters |
| Power Source | Internal lithium-ion battery packs, AC Adapter | Battery pack and AC adapter |
| User Interface | Capacitive touchscreen | Electroluminescent display |
| Connectivity | BLE wireless architecture connecting wand (cordless), base, and foot pedal | BLE wireless architecture connecting unit (corded handpiece attached to base) and foot pedal |
| Reprocessing | Cleaning and Low-Intermediate disinfection validated (Caviwipes) due to not being able to steam sterilize handpiece.
Steam sterilization validated for 3mm and 7mm PBM adapters | Cleaning and Steam sterilization validated for handpiece shell and 3mm/7mm PBM adapters |
| Intended Environment | Dental clinical environments | Dental clinical environments |
## Rationale for Technological Differences
### Peak Pulse Power
The Gemini NOVA has a lower maximum peak pulse power (20 W) than the predicate Gemini 2 (150 W); however, both devices operate within the same average power range (0.1–2.0 W), which primarily determines clinical performance. Bench testing confirmed accurate power output, and this difference does not raise new questions of safety or effectiveness.
### Accessories
The Gemini NOVA accessory set includes disposable fiber tips, safety eyewear, an aluminum handpiece, PBM light adapters (3 mm, 7 mm, and 25 mm with spacer), barrier sleeves, and lithium-ion battery packs for the cordless wand. In the predicate Gemini 2 system, photobiomodulation (PBM) energy is delivered through an external optical fiber assembly, while in the Gemini NOVA the external fiber cord has been eliminated as part
{8}
of the cordless wand design and PBM light adapters interface directly with the integrated fiber ferrule to provide equivalent beam expansion, spot size, and energy distribution at the tissue interface. The adapters are manufactured from the same optical fiber material used in the predicate device and do not alter wavelength, output power, pulse characteristics, or average power; therefore, these differences do not raise new questions of safety or effectiveness.
## Power Source
The Gemini NOVA utilizes rechargeable lithium-ion battery packs to support cordless operation. Battery safety testing confirmed compliance with applicable standards, and this difference does not raise new questions of safety or effectiveness.
## User Interface / Display
The Gemini NOVA uses a 5”×5” capacitive touchscreen, while the predicate device uses an electroluminescent display with capacitive touch controls. Because both devices use the same touch-control method and usability testing confirmed effective operation, this difference does not raise new questions of safety or effectiveness.
## Connectivity
The Gemini NOVA incorporates Bluetooth Low Energy communication between system components, whereas the predicate uses a different connectivity configuration. This functionality only supports system control and does not affect laser energy delivery, and therefore does not raise new questions of safety or effectiveness.
## Reprocessing Method
The Gemini NOVA uses validated cleaning and intermediate-level disinfection procedures for reusable components except for the 3mm and 7mm PBM adapters, which still can be steam sterilized. The predicate device uses steam sterilization. Reprocessing validation confirmed effective cleaning and microbial reduction, and this difference does not raise new questions of safety or effectiveness.
## Performance Data
The following non-clinical testing was conducted to support substantial equivalence of the Gemini Nova system.
## Electrical Safety
Electrical safety testing was performed to demonstrate compliance with IEC 60601-1 requirements for medical electrical equipment. Testing confirmed the device met requirements for basic safety and essential performance.
Page 4
{9}
Page 5
# Electromagnetic Compatibility (EMC)
Electromagnetic compatibility testing was conducted according to IEC 60601-1-2:2014 + A1:2020. Testing evaluated both emissions and immunity performance and confirmed that the device meets applicable EMC requirements.
# Radio Frequency Emissions
Testing was conducted in accordance with FCC Part 15 Subpart B and ICES-003 for unintentional radiators. The Gemini Nova system met Class A limits for conducted and radiated emissions.
# Wireless Coexistence
Wireless communication functions were evaluated for compliance with applicable RF emission and immunity standards including ETSI EN 301 489-1 and ETSI EN 301 489-17. Testing demonstrated compliant wireless performance.
# Battery Safety Testing
Rechargeable lithium-ion battery packs used in the system were evaluated in accordance with IEC 62133-2:2017 safety requirements.
The battery packs met all applicable safety requirements.
# Laser Performance
Laser output characteristics were evaluated in accordance with IEC 60601-2-22 requirements for medical laser equipment and IEC 60825-1 for laser product safety, including classification, labeling, and user protection requirements. Testing verified that the laser output power remained within $\pm 20\%$ of the selected setting and confirmed compliance with applicable safety requirements for Class IV medical laser systems.
# Biocompatibility
Biocompatibility testing of patient-contacting components was conducted in accordance with ISO 10993 standards. Testing included:
- Cytotoxicity (ISO 10993-5)
- Sensitization (ISO 10993-10)
- Irritation / intracutaneous reactivity (ISO 10993-10)
Test results demonstrated that the materials were non-cytotoxic, non-sensitizing, and non-irritating.
# Reprocessing Validation
{10}
Cleaning and intermediate-level disinfection procedures were validated to confirm effective removal of biological soil and reduction of microbial contamination on reusable components.
## Software Verification and Validation/Cybersecurity
Software development and verification activities were conducted in accordance with IEC 62304 lifecycle processes, and the software was classified as Moderate Level of Concern. Verification and validation testing confirmed correct system operation, and cybersecurity controls—including user authentication, encrypted wireless communication, and secure software update mechanisms—were implemented to support the confidentiality, integrity, and availability of the system.
## Bench Testing
Nonclinical bench performance testing was conducted to verify that the Gemini NOVA 810+980 Soft Tissue Laser meets its design specifications and performs safely and effectively under intended use conditions. Testing evaluated laser output accuracy and stability, PBM optical delivery performance, mechanical durability and battery cycling performance, under worst-case and extended-use conditions. Results demonstrated that laser output remained within ±20% of stated values, optical performance and spot size were consistent with specifications, and the device maintained mechanical and electrical integrity following simulated lifetime use. All testing met predefined acceptance criteria and supports substantial equivalence to the predicate device.
## Clinical Data
No clinical studies were conducted or submitted as part of this 510(k) submission.
Clinical data were not required to demonstrate substantial equivalence because the subject device has the same intended use and similar technological characteristics as the predicate device.
Non-clinical testing and bench testing were sufficient to support substantial equivalence.
## Conclusion
The Gemini Nova 810 + 980 Soft Tissue Laser is substantially equivalent to the predicate device and does not raise new questions of safety or effectiveness. To the best of the manufacturer's knowledge, the predicate device has not been subject to any safety-related recalls.
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.