← Product Code [QYT](/productcode/QYT) · K260754

# Fastep COVID-19 Antigen Pen Home Test; Fastep COVID-19 Antigen Pen Test (K260754)

_Assure Tech., LLC · QYT · Jun 5, 2026 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K260754

## Device Facts

- **Applicant:** Assure Tech., LLC
- **Product Code:** [QYT](/productcode/QYT.md)
- **Decision Date:** Jun 5, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3984
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Pediatric

## Indications for Use

The Fastep COVID-19 Antigen Pen Home Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19. This test is for use by individuals aged 14 years and older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment. Positive results do not rule out co-infection with other respiratory pathogens. This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek appropriate follow-up care from their healthcare provider. Performance characteristics for SARS-CoV-2 were established from September 2023 to February 2025 when SARS-CoV-2 Omicron variant was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.

## Device Story

The Fastep COVID-19 Antigen Pen is a single-use, lateral flow immunoassay for qualitative detection of SARS-CoV-2 nucleocapsid protein. The device features an integrated sample swab and test strip within a pen-like housing. The user collects an anterior nasal specimen, inserts the swab into the base containing a pre-filled buffer, and allows the mixture to migrate across a nitrocellulose membrane. If SARS-CoV-2 antigen is present, it binds to conjugate particles and is captured at the Test Line (T); a Control Line (C) confirms assay validity. Results are visually interpreted within 15 minutes. Designed for OTC home use or point-of-care settings, the device provides rapid results to inform clinical decision-making, though negative results require follow-up testing for symptomatic individuals. It benefits patients by enabling accessible, rapid screening for COVID-19.

## Clinical Evidence

Prospective clinical study (N=1077) of symptomatic subjects (age 2+) in simulated home-use environment. Compared to FDA-cleared RT-PCR. Results: PPA 84.5% (95% CI: 77.3%-89.7%); NPA 99.9% (95% CI: 99.4%-100.0%). Usability study (N=113) showed 99.1% critical task success. Readability study (N=565) showed 96.5% correct interpretation of mock panels.

## Technological Characteristics

Lateral flow immunochromatographic assay. Materials: nitrocellulose membrane, sample/conjugate/absorbent pads. Integrated swab/pen form factor. Buffer volume ~0.95 mL. Sterilization: Radiation (ISO 11137). Biocompatibility: ISO 10993-5, 10993-10, 10993-23. No electronics/software; purely visual interpretation.

## Regulatory Identification

For the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen in individuals 2 years and older.

## Predicate Devices

- Flowflex COVID-19 Antigen Home Test ([K230828](/device/K230828.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number
K260754

B Applicant
Assure Tech., LLC

C Proprietary and Established Names
Fastep COVID-19 Antigen Pen Home Test; Fastep COVID-19 Antigen Pen Test

D Regulatory Information
|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  QYT | Class II | 21 CFR 866.3984 - Over-The-Counter Test To Detect SARS-CoV-2 From Clinical Specimens | MI - Microbiology  |

## II Submission/Device Overview:

A Purpose for Submission:
To obtain 510(k) clearance for the Fastep COVID-19 Antigen Pen Home Test and Fastep COVID-19 Antigen Pen Test (hereafter referred to as the Fastep COVID-19 Pen Test).

B Measurand:
Nucleocapsid protein antigens from SARS-Coronavirus 2 (SARS-CoV-2)

C Type of Test:
Qualitative lateral flow immunoassay

## III Intended Use/Indications for Use:

A Intended Use(s):
See Indications for Use below.

B Indication(s) for Use:
Fastep COVID-19 Antigen Pen Home Test:
The Fastep COVID-19 Antigen Pen Home Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19.

This test is for use by individuals aged 14 years and older testing themselves, or adults testing individuals aged 2 years or older.

All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment.

Positive results do not rule out co-infection with other respiratory pathogens.

This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek appropriate follow-up care from their healthcare provider.

Performance characteristics for SARS-CoV-2 were established from September 2023 to February 2025 when SARS-CoV-2 Omicron variant was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.

## Fastep COVID-19 Antigen Pen Test:

The Fastep COVID-19 Antigen Pen Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19.

This test is for use by individuals aged 14 years and older testing themselves, or adults testing individuals aged 2 years or older.

All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment.

Positive results do not rule out co-infection with other respiratory pathogens.

This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek appropriate follow-up care from their healthcare provider.

Performance characteristics for SARS-CoV-2 were established from September 2023 to February 2025 when SARS-CoV-2 Omicron variant was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.

## C Special Conditions for Use Statement(s):

- OTC - Over The Counter
- IVD – For in vitro diagnostic use

K260754 - Page 2 of 20

{2}

# D Special Instrument Requirements:

Not applicable (N/A)

# IV Device/System Characteristics:

# A Device Description:

The Fastep COVID-19 Pen Test is a lateral flow immunochromatographic assay for the qualitative detection of SARS-CoV-2 nucleocapsid antigen from self-collected anterior nasal swab specimens from symptomatic individuals. The single-use device consists of a base with built-in buffer and a test "pen" containing an integrated swab and test strip (Figure 1). The test strip is composed of a sample pad, conjugate pad, nitrocellulose membrane, and absorbent pad, with designated control (C) and test (T) lines for result interpretation.

![img-0.jpeg](img-0.jpeg)
Figure 1: Fastep COVID-19 Pen Test

# Materials provided in test kit box:

- 1, 2, 4, 10, 20, and 25 foil packages each containing a test pen and base with extraction buffer ( $\sim 0.95 \mathrm{~mL}$ )
1 Quick Reference Instructions (QRI)

# Materials needed but not provided in kit box:

Timer

# B Principle of Operation:

The integrated swab is used to collect samples from both nostrils and then inserted into the base containing a built-in buffer, forming a sample mixture. This mixture migrates along the test strip across a nitrocellulose membrane containing two reagent lines: the test line and the control line.

If SARS-CoV-2 nucleocapsid antigen is present in the sample, it binds to anti-SARS-CoV-2 conjugate particles and is captured at the test line, producing a visible, pink-colored line indicating a positive result. The sample continues to migrate to the control (C) region, which contains goat anti-mouse antibodies. To serve as a procedural control, a pink line will always appear in the control line region (C), which confirms proper sample flow and test performance and must be present before interpreting results.

K260754 - Page 3 of 20

{3}

Results must be interpreted visually between 15 and 30 minutes after inserting the test pen into the base containing the extraction buffer. A positive specimen will present two pink lines, one in the control region (C) and one in the test region (T) of the result window, indicating that SARS-CoV-2 antigen was detected. A negative specimen will present only one pink line next to the "C" region of the test window and indicates that SARS-CoV-2 antigen was not detected. The procedural control line must always appear in the "C" region. If a line in the "C" region is not visible after 15 minutes, then the result is invalid.

V Substantial Equivalence Information:

A Predicate Device Name(s):

Flowflex COVID-19 Antigen Home Test

B Predicate 510(k) Number(s):

K230828

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | Candidate Device K260754 | Predicate Device K230828  |
| --- | --- | --- |
|  Device Trade Name | Fastep COVID-19 Antigen Pen Home Test; Fastep COVID-19 Antigen Pen Test | Flowflex COVID-19 Antigen Home Test  |
|  Intended Use/Indications For Use | Fastep COVID-19 Antigen Pen Home Test: The Fastep COVID-19 Antigen Pen Home Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19. This test is for use by individuals aged 14 years and older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV- 2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment. | The Flowflex COVID-19 Antigen Home Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS- CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19 within the first 6 days of symptom onset. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. The Flowflex COVID-19 Antigen Home Test does not differentiate between SARS-CoV and SARS-CoV-2. All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re- tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV- 2. Negative results do not preclude SARS-CoV-2 infections or other  |

K260754 - Page 4 of 20

{4}

K260754 - Page 5 of 20
|  Device & Predicate Device(s): | Candidate Device K260754 | Predicate Device K230828  |
| --- | --- | --- |
|   | Positive results do not rule out co-infection with other respiratory pathogens. This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek appropriate follow-up care from their healthcare provider. Performance characteristics for SARS-CoV-2 were established from September 2023 to February 2025 when SARS-CoV-2 Omicron variant was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected. **Fastep COVID-19 Antigen Pen Test:** The Fastep COVID-19 Antigen Pen Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19. This test is for use by individuals aged 14 years and older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV- | pathogens and should not be used as the sole basis for treatment. Positive results do not rule out co-infection with other respiratory pathogens. Performance characteristics for SARS-CoV-2 were established from December 2022 to March 2023 of the SARS-CoV-2 pandemic when SARS-CoV-2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variant are emerging, performance characteristics may vary. This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.  |

{5}

K260754 - Page 6 of 20
|  Device & Predicate Device(s): | Candidate Device K260754 | Predicate Device K230828  |
| --- | --- | --- |
|   | 2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment. Positive results do not rule out co-infection with other respiratory pathogens. This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek appropriate follow-up care from their healthcare provider. Performance characteristics for SARS-CoV-2 were established from September 2023 to February 2025 when SARS-CoV-2 Omicron variant was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected. |   |
|  **General Device Characteristic Similarities**  |   |   |
|  Intended Use Setting | OTC | Same  |
|  Analyte | Nucleocapsid protein antigen from SARS-CoV-2 | Same  |
|  Technology | Lateral flow immunoassay | Same  |
|  Sample Type | Anterior nasal specimen | Same  |
|  Assay Type | Qualitative | Same  |
|  Interpretation | Visually read | Same  |
|  Reading Time | 15-30 min | Same  |
|  **General Device Characteristic Differences**  |   |   |
|  Swab Format | Integrated sterile swab | Individually packed sterile disposable nasal swab  |
|  Tube holder | No | Yes  |

{6}

VI Standards/Guidance Documents Referenced:

|  Document | Title | Publisher | Applicable study  |
| --- | --- | --- | --- |
|  21 CFR 866.3984 | Reclassification order for DEN220028 and special controls under 21 CFR 866.3984 | FDA/CDRH | All Studies  |
|  11137-1:2025 | Sterilization of health care products - Radiation - Part 1: Requirements for development validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)] | ISO | Sterility  |
|  11137-2:2013 | Sterilization of health care products — Radiation Part 2: Establishing the sterilization dose | ISO | Sterility  |
|  11137-3:2017 | Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development validation and routine control | ISO | Sterility  |
|  11737-1:2018/Amd 1:2021 | Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)] | ISO | Sterility  |
|  11737-2:2019 | Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition validation and maintenance of a sterilization process | ISO | Sterility  |
|  10993-5: 2009 | Biological Evaluation of Medical Devices - Tests for in vitro cytotoxicity | ISO | Biocompatibility  |
|  10993-10: 2021 | Biological Evaluation of Medical Devices –Tests for irritation and skin sensitization | ISO | Biocompatibility  |
|  10993-23:2021 | Biological Evaluation of Medical Devices – Tests for irritation | ISO | Biocompatibility  |

K260754 - Page 7 of 20

{7}

VII Performance Characteristics:

A Analytical Performance:

1. Precision Study:

The purpose of the study was to assess lot-to-lot variability of three different lots of the Fastep COVID-19 Pen Test. Gamma-irradiated SARS-CoV-2 USA_WA1/2020 was spiked in negative nasal fluid and aliquoted as follows.

1) Negative sample (neat negative nasal fluid)
2) C5 sample (0.16x LoD)
3) C95 sample (1x LoD)
4) Moderate positive sample (3x LoD)

All sample aliquots tested in the study were randomized and blinded. For each concentration, 5 replicates were tested by 6 operators across 3 different testing sites (2 per site), between three (3) lots, and tested for 5 days (i.e., 3 lots x 6 operators x 5 replicates/run x 5 days). A total of 450 tests were tested for each panel member.

All negative samples and replicates prepared at 3x LoD demonstrated 100% agreement across the operators, lots, days, runs and sites tested. Results with concentrations at 0.16x LoD and 1x LoD yielded less than 100% positive agreement, however, results showed minor variability between lots and were deemed acceptable. The results are summarized below.

Table 1: Summary Results of Multi-lot Precision Study

|  Sample | Number of Positives / Number of Samples Tested (%) |   |   | Total number of positives / Total number of samples (%)  |
| --- | --- | --- | --- | --- |
|   |  Lot 1 | Lot 2 | Lot 3  |   |
|  Negative | 0/150 | 0/150 | 0/150 | 0/450  |
|   |  (0.0%) | (0.0%) | (0.0%) | (0.0%)  |
|  C5 sample (0.16x LoD) | 5/150 | 5/150 | 4/150 | 14/450  |
|   |  (3.3%) | (3.3%) | (2.6%) | (3.1%)  |
|  C95 sample (1x LoD) | 142/150 | 144/150 | 145/150 | 431/450  |
|   |  (94.6%) | (96.0%) | (96.6%) | (95.8%)  |
|  Moderate positive (3x LoD) | 150/150 | 150/150 | 150/150 | 450/450  |
|   |  (100.0%) | (100.0%) | (100.0%) | (100.0%)  |

2. Linearity:

N/A

3. Analytical Specificity/Interference:

a. Cross-Reactivity and Microbial Interference:

Cross-reactivity and microbial interference studies were conducted to determine if other respiratory pathogens/microbial flora that may be present in nasal swab samples could cause a false positive test result or interfere with the detection of a true positive result and cause a false negative result. A panel of viruses, high-prevalence disease agents, and normal or pathogenic flora were used for these studies. The microbial interference and the

K260754 - Page 8 of 20

{8}

cross-reactivity study were conducted simultaneously with samples tested in a randomized and blinded manner.

For the cross-reactivity study, dilutions of the panel organisms were prepared in negative nasal fluid and tested in triplicate in the absence of SARS-CoV-2. No cross-reactivity was observed with the organisms tested (Table 2).

For the microbial interference study, dilutions of the panel organisms were prepared in negative nasal fluid, in the presence of low levels (2x LoD;  $1.58 \times 10^{3} \mathrm{TCID}_{50} / \mathrm{mL}$ ) of Gamma- irradiated SARS-CoV-2 (USA_WA1/2020) and tested in triplicate. No microbial interference was observed (Table 2).

Table 2: Cross-Reactivity and Microbial Interference Study Results

|  Virus/Microorganism | Concentration | Units | Number of Positive Results/ Number of Replicates Tested  |   |
| --- | --- | --- | --- | --- |
|   |   |   |  Cross-Reactivity (without analyte) | Interference (2x LoD SARS-CoV-2)  |
|  Bordetella pertussis | 1.28 × 106 | CFU/mL | 0/3 | 3/3  |
|  Candida albicans | 2.28 × 106 | CFU/mL | 0/3 | 3/3  |
|  Chlamydia trachomatis | 6.25 × 106 | CFU/mL | 0/3 | 3/3  |
|  Chlamydia pneumoniae | 4.50 × 107 | IFU/mL | 0/3 | 3/3  |
|  Haemophilus influenzae | 1.28 × 108 | CFU/mL | 0/3 | 3/3  |
|  Legionella pneumophila | 3.50 × 107 | CFU/mL | 0/3 | 3/3  |
|  Mycoplasma pneumonia | 2.25 × 106 | CFU/mL | 0/3 | 3/3  |
|  Mycobacterium tuberculosis | 3.03 × 107 | CFU/mL | 0/3 | 3/3  |
|  Pseudomonas aeruginosa | 1.73 × 108 | CFU/mL | 0/3 | 3/3  |
|  Saccharomyces cerevisiae | 7.65 × 107 | CFU/mL | 0/3 | 3/3  |
|  S. aureus MSSA | 2.31 × 109 | CFU/mL | 0/3 | 3/3  |
|  S. aureus MRSA | 2.38 × 109 | CFU/mL | 0/3 | 3/3  |
|  Staphylococcus epidermidis | 3.05 × 109 | CFU/mL | 0/3 | 3/3  |
|  Streptococcus pneumoniae | 1.81 × 108 | CFU/mL | 0/3 | 3/3  |
|  Streptococcus pyogenes | 8.00 × 107 | CFU/mL | 0/3 | 3/3  |
|  Adenovirus | 2.88 × 106 | TCID50/mL | 0/3 | 3/3  |
|  Enterovirus 68 | 2.39 × 106 | TCID50/mL | 0/3 | 3/3  |
|  Human coronavirus 229E | 3.15 × 105 | TCID50/mL | 0/3 | 3/3  |
|  Human coronavirus NL63 | 1.18 × 105 | TCID50/mL | 0/3 | 3/3  |
|  Human coronavirus OC43 | 7.00 × 106 | TCID50/mL | 0/3 | 3/3  |
|  Human coronavirus HKU1 | 1.25 × 105 | TCID50/mL | 0/3 | 3/3  |
|  Human Metapneumovirus 9 | 1.00 × 105 | TCID50/mL | 0/3 | 3/3  |
|  Influenza A Virus | 2.63 × 105 | TCID50/mL | 0/3 | 3/3  |

K260754 - Page 9 of 20

{9}

|  Virus/Microorganism | Concentration | Units | Number of Positive Results/ Number of Replicates Tested  |   |
| --- | --- | --- | --- | --- |
|   |   |   |  Cross-Reactivity (without analyte) | Interference (2x LoD SARS-CoV-2)  |
|  Influenza B Virus | 3.78×105 | TCID50/mL | 0/3 | 3/3  |
|  Parainfluenza virus 1 | 1.14×106 | TCID50/mL | 0/3 | 3/3  |
|  Parainfluenza virus 2 | 1.25×105 | TCID50/mL | 0/3 | 3/3  |
|  Parainfluenza virus 3 | 2.13×107 | TCID50/mL | 0/3 | 3/3  |
|  Parainfluenza virus 4b | 1.25×105 | TCID50/mL | 0/3 | 3/3  |
|  Respiratory syncytial virus | 3.78×105 | TCID50/mL | 0/3 | 3/3  |
|  Rhinovirus | 1.04×105 | TCID50/mL | 0/3 | 3/3  |
|  MERS-coronavirus | 4.00×105 | TCID50/mL | 0/3 | 3/3  |
|  Negative nasal fluid | N/A | N/A | 0/3 | 3/3  |

## b. Endogenous/Exogenous Substances Interference:

The Fastep COVID-19 Pen Test was evaluated for performance in the presence of a panel of common interfering endogenous and exogenous substances. Potentially interfering substances were prepared in negative nasal fluid to the recommended concentration. Virus-negative specimens were evaluated in triplicate on each of the 3 lots of the test device and by three different operators to confirm that the potentially interfering substances were not cross-reactive with the test.

Positive samples were prepared in negative nasal fluid containing Gamma-irradiated SARS-CoV-2 (USA-WA1/2020) at 2x LoD and were evaluated in the presence of interfering substances in triplicate on each of the 3 lots of the test device and three different operators to confirm that the substances did not interfere with the detection of SARS-CoV-2.

No cross-reactivity or interference was observed for any of the substances tested at the concentrations evaluated as shown in the table below.

Table 3: Interfering Substances Study Results

|  Interfering Substances | Concentration Tested | Number of Positive Results/ Number of Replicates Tested  |   |
| --- | --- | --- | --- |
|   |   |  Cross-Reactivity (without analyte) | Interference (2x LoD SARS-CoV-2)  |
|  Beclomethasone | 15% v/v | 0/3 | 3/3  |
|  Biotin | 3500 ng/mL | 0/3 | 3/3  |
|  Dexamethasone | 15% v/v | 0/3 | 3/3  |
|  Dyclonine Hydrochloride | 2 mg/mL | 0/3 | 3/3  |
|  Flunisolide | 15% v/v | 0/3 | 3/3  |
|  Hand sanitizer | 15% v/v | 0/3 | 3/3  |
|  Hand Soap | 15% v/v | 0/3 | 3/3  |

K260754 - Page 10 of 20

{10}

|  Interfering Substances | Concentration Tested | Number of Positive Results/ Number of Replicates Tested  |   |
| --- | --- | --- | --- |
|   |  | Cross-Reactivity (without analyte) | Interference (2x LoD SARS-CoV-2)  |
|  Homeopathic allergy relief (Histaminum hydrochloricum) | 15% w/v | 0/3 | 3/3  |
|  Homeopathic nasal wash (alkalol) | 5% v/v | 0/3 | 3/3  |
|  Leukocytes | 4.8 x 10^{6} cells/mL | 0/3 | 3/3  |
|  Molnupiravir | 10 mg/mL | 0/3 | 3/3  |
|  Mucin | 2.5 mg/mL | 0/3 | 3/3  |
|  Mupirocin | 10 mg/mL | 0/3 | 3/3  |
|  Nasal corticosteroids (Budesonide) | 15% v/v | 0/3 | 3/3  |
|  Nasal corticosteroids (fluticasone furate) | 5% v/v | 0/3 | 3/3  |
|  Nasal corticosteroids (fluticasone propionate) | 5% v/v | 0/3 | 3/3  |
|  Nasal corticosteroids (Mometasone furoate) | 15% v/v | 0/3 | 3/3  |
|  Nasal corticosteroids (Triamcinolone Acetonide) | 15% v/v | 0/3 | 3/3  |
|  Nasal decongestant (Galphimia glauca, Luffa opperculata, sabadilla) | 15% v/v | 0/3 | 3/3  |
|  Nasal gel | 5% v/v | 0/3 | 3/3  |
|  Nasal spray (Cromolyn sodium nasal solution) | 15% v/v | 0/3 | 3/3  |
|  Nasal spray (Oxymetazoline HCl) | 15% v/v | 0/3 | 3/3  |
|  Nasal spray (Phenylephrine HCl) | 15% v/v | 0/3 | 3/3  |
|  Nasal spray (Sodium Chloride & Preservatives) | 15% v/v | 0/3 | 3/3  |
|  Oral Anesthetic Cough Lozenge (Menthol) | 3 mg/mL | 0/3 | 3/3  |
|  Oseltamivir Phosphate (Tamiflu) | 15% w/v | 0/3 | 3/3  |
|  Remdesivir | 10 mg/mL | 0/3 | 3/3  |
|  Sore Throat & Cough Lozenges (Benzocaine, Dextromethorphan HBr) | 3 mg/mL | 0/3 | 3/3  |
|  Sore Throat Spray (Phenol) | 5% v/v | 0/3 | 3/3  |
|  Tobramycin | 50 ug/mL | 0/3 | 3/3  |
|  Whole Blood | 2.5% v/v | 0/3 | 3/3  |

4. Detection Limit and Assay Reportable Range:
a. Limit of Detection (LoD):

K260754 - Page 11 of 20

{11}

The Limit of Detection (LoD) for the Fastep COVID-19 Pen Test was established by testing various dilutions of Gamma-irradiated SARS-CoV-2 isolate (USA_WA1/2020) in negative nasal fluid. The LoD was determined as the lowest virus concentration that was detected  $\geq 95\%$  of the time (concentration at which at least 19 out of 20 replicates tested positive) for each of 3 lots. The LoD was established in two phases.

# Preliminary LoD Study

A series of 1/10, 1/100, 1/1000 and 1/10,000 dilutions were prepared from the Gamma-irradiated SARS-CoV-2 virus diluted into negative nasal fluid. Three replicates of each serial dilution were tested across each lot to determine the preliminary LoD concentration of the device. For each replicate,  $50~\mu \mathrm{l}$  of the sample aliquots were pipetted onto the swab for testing and processed according to the Instructions for Use (IFU). The lowest concentration with 9/9 positive results in the preliminary LoD study was further diluted by a series of 1:2, 1:4, and 1:8 dilutions where each dilution was tested in triplicate for each lot. The lowest concentration, with 9/9 positive results from each lot was considered the preliminary LoD for the virus strain. The results are summarized below.

Table 4: Preliminary LoD Study Summary

|  Isolate/Lineage | SARS-CoV-2 Concentration |   | Number of Positive Results/ Number of Replicates Tested  |
| --- | --- | --- | --- |
|   |  TCID50/mL | TCID50/Swab  |   |
|  SARS-CoV-2 (USA_WA1/2020) (Gamma-irradiated) | 7.9 x 104 | 3.9 x 103 | 9/9  |
|   |  7.9 x 103 | 3.9 x 102 | 9/9  |
|   |  7.9 x 102 | 3.9 x 101 | 9/9  |
|   |  7.9 x 101 | 3.9 x 100 | 0/9  |
|   |  3.9 x 102 | 1.9 x 101 | 6/9  |
|   |  1.9 x 102 | 9.5 x 100 | 0/9  |
|   |  9.9 x 101 | 4.9 x 100 | 0/9  |

# Confirmatory LoD Study

The preliminary LoD concentration, as well as concentrations above and below the LoD, were tested in a total of twenty (20) replicates using the same Gamma-irradiated SARS-CoV-2 virus for each of the same lots evaluated in the preliminary LoD study. To confirm the LoD for each lot, at least 19 of the 20 replicates should be positive. The final confirmation data set included the confirmed LoD concentration, and at least one concentration above and below the LoD to demonstrate  $100\%$  detection above the LoD and  $&lt; 95\%$  detection below the LoD. The results are summarized below.

Table 5: Confirmatory LoD Study Summary of SARS-CoV-2

|  Isolate/Lineage | SARS-CoV-2 Concentration |   | Number of Positive Results/ Number of Replicates Tested  |
| --- | --- | --- | --- |
|   |  TCID50/mL | TCID50/Swab  |   |
|  SARS-CoV-2 | 1.1 x 103(1.5x LoD) | 5.9 x 101 | 60/60  |
|   |  9.8 x 102(1.25x LoD) | 4.9 x 101 | 60/60  |

K260754 - Page 12 of 20

{12}

|  Isolate/Lineage | SARS-CoV-2 Concentration |   | Number of Positive Results/ Number of Replicates Tested  |
| --- | --- | --- | --- |
|   |  TCID_{50}/mL | TCID_{50}/Swab  |   |
|  (USA_WA1/2020) (Gamma-irradiated) | 7.9 x 10^{2} (1x LoD) | 3.9 x 10^{1} | 59/60  |
|   |  5.9 x 10^{2} (0.75x LoD) | 2.9 x 10^{1} | 50/60  |
|   |  3.9 x 10^{2} (0.5x LoD) | 1.9 x 10^{1} | 41/60  |

The LoD for the Fastep COVID-19 Pen Test was determined to be $7.9 \times 10^{2} \mathrm{TCID}_{50} / \mathrm{mL}$ which equates to $39.5 \mathrm{TCID}_{50} / \mathrm{swab}$ across all tested lots.

## b. International Standard LoD Study:

The LoD of the Fastep COVID-19 Pen Test was determined by evaluating different dilutions of SARS-CoV-2 antigen (NIBSC code: 21/368) in negative nasal fluid. SARS-CoV-2 Antigen (NIBSC 21/368) was used for testing 2 lots of the Fastep COVID-19 Pen Test. The LoD was determined as the lowest virus concentration that was detected $\geq 95\%$ of the time (i.e., concentration at which at least 19 out of 20 replicates tested positive). The LoD was established in two (2) phases.

## Preliminary LoD Study

Serial dilutions of SARS-CoV-2 antigen (1:10, 1:100, and 1:1000) were prepared in negative nasal fluid. Three replicates were tested on 2 lots of the test device. For each replicate $50~\mu \mathrm{L}$ of virus dilution was applied to the swab and it was processed according to the IFU. The lowest concentration with 6/6 positive results was further diluted in a series of 1:2, 1:4, and 1:8 dilutions. The lowest concentration with 6/6 positive results following this dilution series was considered the preliminary LoD. The results are summarized below.

Table 6: Preliminary LoD for SARS-CoV-2 Antigen (NIBSC code: 21/368)

|  Concentration (IU/mL) | Concentration on swab (IU/swab) | Number of Positive Results/ Number of Replicates Tested  |
| --- | --- | --- |
|  2.0 x 10^{3} | 1.0 x 10^{2} | 6/6  |
|  2.0 x 10^{2} | 1.0 x 10^{1} | 0/6  |
|  2.0 x 10^{1} | 1.0 x 10^{0} | 0/6  |
|  1.0 x 10^{3} | 5.0 x 10^{1} | 6/6  |
|  5.0 x 10^{2} | 2.5 x 10^{1} | 0/6  |
|  2.5 x 102 | 1.2 x 101 | 0/6  |

The Fastep COVID-19 Pen Test can detect the SARS-CoV-2 Antigen (NIBSC 21/368) at the preliminary LoD of $1.0 \times 10^{3} \mathrm{IU} / \mathrm{mL}$ in negative nasal fluid.

## Confirmatory LoD Study

The preliminary LoD concentration, a concentration above (1.2x LoD), and a concentration below (0.75x LoD) were tested on the same 2 lots. A total of 20 replicates were tested for the confirmatory dilutions for SARS-CoV-2 Antigen (NIBSC 21/368) for each lot. To confirm the LoD for each lot, at least 19 of the 20 replicates should be positive. The results are summarized below.

K260754 - Page 13 of 20

{13}

Table 7: Confirmatory LoD for SARS-CoV-2 Antigen (NIBSC code: 21/368)

|  Concentration (IU/mL) | Concentration on swab (IU/swab) | Number of Positive Results/ Number of Replicates Tested  |
| --- | --- | --- |
|  1.2 x 10^{3} | 6.0 x 10^{1} | 40/40  |
|  1.0 x 10^{3} | 5.0 x 10^{1} | 40/40  |
|  7.5 x 10^{2} | 3.7 x 10^{1} | 39/40  |

The confirmatory LoD for the International Standard for SARS-CoV-2 antigen (NIBSC code: 21/368) in negative nasal fluid was confirmed to be 750 IU/mL which equates to 37.5 IU/swab.

c. **Assay Reportable Range:**
N/A

5. **High Dose Hook Effect:**
A high-dose hook effect study was conducted to evaluate whether high concentrations of SARS-CoV-2 produced false negative results. Gamma-irradiated SARS-CoV-2 was serially diluted in negative nasal fluid. Following preparation, 50 μL of each sample was applied to swabs, which were then processed according to the IFU. Each concentration was tested in five replicates across 3 lots of the test device and by three operators. The results demonstrated no evidence of a high-dose hook effect across the concentrations evaluated in this study as shown below.

Table 8: High-Dose Hook Effect Study Results

|  Concentration (TCID_{50}/mL) | Dilution | Positive/ Total | % Positive  |
| --- | --- | --- | --- |
|  7.9 x 10^{5} | 1000x LoD | 15/15 | 100.0%  |
|  7.9 x 10^{4} | 100x LoD | 15/15 | 100.0%  |
|  7.9 x 10^{3} | 10x LoD | 15/15 | 100.0%  |
|  7.9 x 10^{2} | 1x LoD | 15/15 | 100.0%  |

6. **Inclusivity:**
Analytical reactivity testing for the Fastep COVID-19 Pen Test was conducted to ensure that the device can adequately detect a variety of SARS-CoV-2 strains. A selection of temporally, geographically, and genetically diverse SARS-CoV-2 strains were diluted in negative nasal fluid and tested at multiple concentrations. Each concentration was tested in 5 replicates per lot across 3 lots (15 replicates). Results are summarized below and demonstrate that the test can detect SARS-CoV-2 across the variants evaluated.

K260754 - Page 14 of 20

{14}

Table 9: Inclusivity Results

|  SARS-CoV-2 Variants | Lowest Concentration (TCID_{50}/mL) with 100% detection  |
| --- | --- |
|  SARS-CoV-2 virus (USA_WA1/2020) (Wild-Type) | 7.90×10^{2}  |
|  B.1.1.7 (Alpha) | 8.88×10^{2}  |
|  B.1.351 (Beta) | 3.78×10^{3}  |
|  P.1 (Gamma) | 3.15×10^{3}  |
|  B.1.617.2 (Delta) | 1.04×10^{3}  |
|  B.1.1.529 (Omicron) | 6.26×10^{2}  |
|  BA.2.3 (Omicron) | 5.85×10^{1}  |
|  KP.2 (Omicron) | 3.30×10^{3}  |
|  KP.3 (Omicron) | 5.13×10^{3}  |
|  BA.2.86 (Omicron) | 4.80×10^{3}  |

7. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

a. **Internal controls:**

The Fastep COVID-19 Pen Test contains a built-in internal procedural control. The appearance of the pink line in the control line region (C) indicates that sufficient flow of the sample occurred and that the test reagents are functioning properly. If no visible signal appears on the control line, the test result is invalid.

b. **Sample Stability Study:**

Samples collected in the OTC setting will not undergo storage as the IFU instructs the users to immediately proceed from sample collection to the testing steps.

c. **Shelf-Life Stability Study:**

A real-time stability study was conducted to evaluate stability and determine the shelf-life of the unopened kit. To validate shelf-life, 3 unopened Fastep COVID-19 Pen Test kit lots were stored at 2-8°C and 30°C. At predefined intervals, three (3) replicates/per lot were evaluated using the following test panel: negative nasal fluid, and two (2) positive samples contrived in negative nasal fluid [2x (1.58×10³ TCID₅₀/mL), 5x (3.95×10³ TCID₅₀/mL)] spiked with Gamma-irradiated SARS-CoV-2. At the time of clearance, all study data met the protocol defined acceptance criteria and support storage of the test kits from 2-30°C (36-86°F) for up to 8 months.

K260754 - Page 15 of 20

{15}

K260754 - Page 16 of 20

d. Transport Simulation Study

The transport simulation study assessed the performance of the Fastep COVID-19 Pen Test under simulated worst-case shipping conditions for both summer and winter conditions within the continental United States. The temperature profiles were designed to mimic typical transport scenarios, with temperatures fluctuating between extreme highs and lows. The packages were tested using both a summer and winter thermal challenge profile. Tests were placed into an incubator set to different temperatures, following the order and cycle periods outlined below.

Table 10: Summer Profile
|  Cycle No. | Temperature | Cycle Period Hours  |
| --- | --- | --- |
|  1 | 40°C | 8  |
|  2 | 22°C | 4  |
|  3 | 40°C | 6  |
|  4 | 30°C | 56  |
|  5 | 40°C | 6  |

Table 11: Winter Profile
|  Cycle No. | Temperature | Cycle Period Hours  |
| --- | --- | --- |
|  1 | -10°C | 8  |
|  2 | 18°C | 4  |
|  3 | -10°C | 6  |
|  4 | 10°C | 56  |
|  5 | -10°C | 6  |

Baseline performance was assessed before subjecting the test kits to the temperature conditions (T0), followed by post-distribution testing [Td (Td-s= summer, Td-w= winter)] to evaluate any changes. Negative control samples, low positive samples at 2x LoD (1.58×10³ TCID₅₀/ mL), and high positive samples at 5x LoD (3.95×10³ TCID₅₀/ mL) were used for testing. The study involved 3 lots of the test device, with one operator testing each lot using three (3) replicates before and after the final temperature cycle challenge. Test results were recorded as positive, negative, or invalid.

All samples produced expected results following exposure to both simulated summer and winter conditions. The results support the stability of the device under anticipated shipping conditions.

8. Assay Cut-Off:
N/A

B Comparison Studies:

1. Method Comparison with Predicate Device:
See section C (clinical studies) below.

2. Matrix Comparison:
The Fastep COVID-19 Pen Test is only intended for use with direct anterior nasal swab specimens. As no other specimen or sample type is claimed for this device, a matrix comparison study is not applicable.

{16}

# C Clinical Studies:

# 1. Clinical Performance Study:

A prospective lay-person clinical study was conducted between September 2023 and February 2025 to assess the performance of the Fastep COVID-19 Pen Test when compared to an FDA cleared RT-PCR assay. The study prospectively enrolled symptomatic subjects at 7 geographically distinct study sites located in the United States. Enrolled subjects were 2 years of age or older and exhibited symptoms of infection consistent with COVID-19.

Testing was performed in a simulated home-use environment. Two anterior nasal swab (AN) specimens were collected from each participant. One swab was collected by study personnel or healthcare professionals, placed into a transport tube containing viral transport media and transported to a central laboratory for testing with a highly sensitive RT-PCR test. The comparator test was performed according to the comparator test IFU. The other swab was collected according to the Fastep COVID-19 Pen Test's QRI: either self-collected by a lay user aged  $\geq 14$  years or collected by an adult (parent/guardian) from individuals aged 2 to  $&lt; 14$  years and was tested immediately on-site using the Fastep COVID-19 Pen Test. The order of collection for the investigational and comparator AN swab was randomized.

One thousand seventy-seven (1077) samples were deemed evaluable. The demographics of the evaluable subjects are shown below:

Table 12: Patient Demographics

|  Characteristics of the study population |   | N=1077 | Percent (%)  |
| --- | --- | --- | --- |
|  Sex | Male | 447 | 41.5%  |
|   |  Female | 630 | 58.5%  |
|  Age | ≥2 and <14 years of age | 219 | 20.3%  |
|   |  14-21 years of age | 107 | 9.9%  |
|   |  22-64 years of age | 686 | 63.7%  |
|   |  ≥65 years of age | 65 | 6.0%  |
|  Race | Asian | 32 | 3.0%  |
|   |  Black or African American | 138 | 12.8%  |
|   |  White | 860 | 79.9%  |
|   |  Native Hawaiian or Pacific Islander | 1 | 0.1%  |
|   |  Mixed race/ Biracial/Other | 24 | 2.2%  |
|   |  American Indian or Alaska Native | 2 | 0.2%  |
|   | Prefer not to say | 20 | 1.9%  |

The performance of the Fastep COVID-19 Pen Test when compared to FDA-cleared highly sensitive RT-PCR molecular assays are presented in the table below.

Table 13: SARS-CoV-2 Performance

|   | Comparator Positive | Comparator Negative | Total  |
| --- | --- | --- | --- |
|  Candidate Positive | 109 | 1 | 110  |

K260754 - Page 17 of 20

{17}

|   | Comparator Positive | Comparator Negative | Total  |
| --- | --- | --- | --- |
|  Candidate Negative | 20 | 947 | 967  |
|  Total | 129 | 948 | 1077  |
|  Positive Percent Agreement (PPA) | 84.5% (109/129) (95% C.I.: 77.3%-89.7%)  |   |   |
|  Negative Percent Agreement (NPA) | 99.9% (947/948) (95% C.I.: 99.4%-100.0%)  |   |   |

2. Clinical Sensitivity:

Please refer to Section C (Clinical Studies) above for clinical validation. The PPA for the test is as follows:

SARS-CoV-2: 84.5% (109/129): 95% C.I.: 77.3%-89.7%

3. Clinical Specificity:

Please refer to Section C (Clinical Studies) above for clinical validation. The NPA for the test is as follows:

SARS-CoV-2: 99.9% (947/948): 95% C.I.: 99.4%-100.0%

4. Clinical Cut-Off:

This test is a qualitative test with a binary positive/negative signal and there is no clinical cut-off for the test.

5. Usability and Readability Study:

a. Usability Study:

A usability study was conducted to assess the lay users' ability to understand the IFU and to adequately execute the Fastep COVID-19 Pen Test workflow accordingly.

A total of 113 subjects were enrolled in the human factors assessment studies concurrently with the clinical evaluation at three sites. Out of these subjects, 63 participants self-collected specimens and performed testing using the investigational test, and 50 had their collection performed by another lay user.

Participants were evaluated for their ability to correctly perform critical and non-critical tasks associated with the testing process. Overall, 99.1% of all critical and 81.9% of non-critical tasks associated with the Fastep COVID-19 Pen Test were performed correctly. Results are summarized below.

Table 14: Critical vs. Non-Critical Tasks Correctly Performed

|  Steps | Steps performed correctly | Total number of steps | Percentage of steps performed correctly  |
| --- | --- | --- | --- |
|  Critical | 1008 | 1017 | 99.1%  |
|  Non-Critical | 185 | 226 | 81.9%  |
|  Sum | 1193 | 1243 | 96.0%  |

Overall, the majority of subjects who participated in the usability evaluation found the instructions clear and easy to follow, and none reported difficulty collecting a sample or

K260754 - Page 18 of 20

{18}

performing the test. Five participants reported difficulty understanding the instructions but were still able to successfully perform the test. Following completion of the usability evaluation, all participants completed a comprehension questionnaire. Results demonstrated a high level of understanding of the test purpose, result interpretation, and appropriate follow-up actions for positive and invalid results.

b. **Readability Study:**

A readability study was conducted to evaluate whether lay users could correctly interpret test results for the Fastep COVID-19 Pen Test in a simulated home-use environment.

A total of 565 lay users interpreted blinded mock test panels representing positive (including low-positive), negative, and invalid results.

Mock test panels included at least one result from each of the following: a negative result, a low positive result at no more than 1.5 - 2x LoD, a positive result at 5x LoD, and an invalid result. Each investigational test with the mock result was blinded and coded with a sample identification number. 96.5% (545/565 tests) of the mock tests were interpreted correctly. Overall agreement for positive, negative, and invalid mock results was 98.9%, 95.3%, and 92.0%, respectively, indicating that most users were able to correctly interpret test results and therefore, the performance was acceptable.

D Other Supportive Study/Device Information:

1. **Flex Studies:**

To assess the robustness of the Fastep COVID-19 Pen Test, flex studies were conducted that assessed all major aspects of the test procedure (e.g., sample volume, reading time, extraction buffer volume, swab elution time and procedure) and variability of environmental test conditions that the test may be subjected to when in use (e.g., disturbance during run, device orientation, lighting, various temperature and humidity stress conditions). Testing was performed by spiking the swab with contrived positive sample generated by diluting Gamma-irradiated SARS-CoV-2 virus into negative nasal fluid at 2x LoD (1.58×10³ TCID₅₀/mL). Flex studies support the robustness of the test under the intended use conditions and demonstrate a low risk of erroneous results.

2. **Serial Testing:**

As a mitigation for the low performance of antigen tests very early and at the tail end of infection, the Intended Use for this test device (and associated Instructions for Use) states that negative results are presumptive, and it includes the need for repeat testing (i.e., test at least twice over three days with at least 48 hours between tests). Although the data, when stratified by symptom onset have performance estimates with insufficient statistical confidence, the clinical study data set of this and similar studies for test devices of a similar principle and design, indicate that such mitigation is needed.

This mitigation is supported by data generated by the National Institutes for Health (NIH) and the University of Massachusetts Chan Medical School (in collaboration with the FDA) demonstrating that repeat testing over multiple days improves test performance and increases the likelihood that a COVID-19 antigen test will accurately detect an infection. These results have informed the FDA's general understanding that repeat testing after a negative result

K260754 - Page 19 of 20

{19}

from a COVID-19 antigen test reduces the risk of a false negative result. Please refer to the following studies for additional details:

- Finding a Needle in the Haystack: Design and Implementation of a Digital Site-less Clinical Study of Serial Rapid Antigen Testing to Identify Asymptomatic SARS-CoV-2 Infection –
https://www.medrxiv.org/content/10.1101/2022.08.04.22278274v1

- Performance of Screening for SARS-CoV-2 using Rapid Antigen Tests to Detect Incidence of Symptomatic and Asymptomatic SARS-CoV-2 Infection: findings from the Test Us at Home prospective cohort study –
https://www.medrxiv.org/content/10.1101/2022.08.05.22278466v1

E Expected Values/Reference Range:
Not applicable. An individual sample is expected to be negative for SARS-CoV-2.

VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K260754 - Page 20 of 20

---

**Source:** [https://fda-staging.innolitics.com/device/K260754](https://fda-staging.innolitics.com/device/K260754)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
