K260751 · Piraeus Medical, Inc. · MOF · Aug 27, 2026 · Cardiovascular
Device Facts
Record ID
K260751
Device Name
35 Helix Guidewire
Applicant
Piraeus Medical, Inc.
Product Code
MOF · Cardiovascular
Decision Date
Aug 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The 35 Helix Guidewire is intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature.
Device Story
0.035" nitinol guidewire; facilitates access to neuro and peripheral vasculatures; features distal radiopaque polymer jacket and hydrophilic coating to reduce friction. Used by physicians in clinical settings to introduce/position catheters and interventional devices. Operated manually; includes introducer and torque device for steering. Fluoroscopic visualization used for guidance. Benefits patient by enabling precise navigation of tortuous vasculature for interventional procedures.
Clinical Evidence
Bench testing only. Performance verified via dimensional verification, visual inspection, radiopacity (ASTM F640), delivery/compatibility/retraction, pushability, tip flexibility, torqueability, particulate testing, coating lubricity/durability, tensile strength, kink resistance, corrosion resistance, and coating integrity. Biocompatibility testing (ISO 10993) confirmed non-cytotoxic, non-sensitizing, non-irritating, non-pyrogenic, and non-hemolytic properties. Usability and design validation performed per IEC 62366-1.
Technological Characteristics
Nitinol core wire; distal radiopaque loaded polymer section (13.75 cm); hydrophilic coating on distal segment. 0.035" diameter. Sterilized via Ethylene Oxide (ISO 11135). Shelf life 6 months (ASTM F1980). Accessories: introducer, torque device. Radiopacity per ASTM F640.
Indications for Use
Indicated for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. Not intended for use in the coronary vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 27, 2026
Piraeus Medical, Inc.
Sarah Kabanuk
COO
7351 Kirkwood Lane N, Suite 116
Maple Grove, Minnesota 55369
Re: K260751
Trade/Device Name: 35 Helix Guidewire
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: MOF, DQX
Dated: July 23, 2026
Received: July 24, 2026
Dear Sarah Kabanuk:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260751 - Sarah Kabanuk
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260751 - Sarah Kabanuk
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260751 | ? |
| Please provide the device trade name(s). | | ? |
| 35 Helix Guidewire | | |
| Please provide your Indications for Use below. | | ? |
| The 35 Helix Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260751 Page 1
# 510(k) Summary, K260751
# A. Submitter Information
Submitter's Name:
Address:
Contact Person:
Telephone:
Email:
Date of Preparation:
Piraeus Medical, Inc.
7351 Kirkwood Lane N, Suite 116
Maple Grove, MN 55369, USA
Sarah Kabanuk
507-782-0184
skabanuk@piraeusmedical.com
March 6th, 2026
# B. Correspondent Information
Contact Name:
Telephone:
Sarah Kabanuk
507-782-0184
# C. Subject Device
Proprietary Name:
Common/Usual Name:
Classification Name:
Product Codes:
Regulation:
35 Helix™ Guidewire
Guidewire
Guide, Wire, Catheter, Neurovasculature
MOF, DQX
21 CFR 870.1330
# D. Predicate Device
Proprietary Name:
Common/Usual Name:
Classification Name:
Product Codes:
Regulation:
Manufacturer:
510(k) #:
Aristotle® Colossus® Guidewire
Guidewire
Guide, Wire, Catheter, Neurovasculature
MOF, DQX
21 CFR 870.1330
Scientia Vascular, Inc.
K222437
# E. Device Description
The 35 Helix Guidewire is a 0.035" diameter guidewire designed to facilitate access to neuro and peripheral vasculatures. It is supplied sterile for single use only and is available in a standard stiffness profile with a length of 180 cm. The guidewire is offered pre-shaped with either a straight tip or a round curve distal tip. The guidewire is constructed of nitinol and features a distal radiopaque polymer jacket to facilitate fluoroscopic visualization. A hydrophilic coating on the distal segment is to reduce friction during manipulation in vessels. The guidewire is provided with an introducer, to aid with guidewire insertion into a catheter hub and/or rotating hemostasis valve (RHV), and a torque device to attach to the proximal portion of the guidewire to facilitate steering.
# F. Intended Use
The 35 Helix Guidewire is intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature.
# G. Indications for Use
The 35 Helix Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
# H. Comparison of Technological Characteristics
Shown in the table below is the comparison of technological characteristics for the 35 Helix Guidewire to those of the predicate device, Aristotle Colossus Guidewire (K222437).
{5}
K260751 Page 2
Table 1: Comparison of Predicate and Subject Devices
| Device Attribute | Predicate Device (K222437) | Subject Device (K260751) |
| --- | --- | --- |
| Regulation / Product Code | Class II, MOF, DQX, 21 CFR 870.1330 | Same |
| Indications for Use | The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | The 35 Helix Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. |
| Intended Use | The Aristotle Colossus Guidewire is intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature. | The 35 Helix Guidewire is intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature. |
| Wire Diameter Maximum | 0.035" (0.88 mm) | Same |
| Wire Diameter Profile | 0.018"- 0.035" | 0.024" - 0.035" |
| Device Length | 150 cm to 300 cm | 180 cm |
| Flex (Tip) Length | 35 cm | 25 cm |
| Tip Type and Shape | Straight, Shapeable | Pre-Shaped Straight Pre-Shaped Round Curve |
| Flexibility (stiffness) | Standard | Same |
| Wire Material | Core Wire: Stainless Steel | Core Wire: Nitinol |
| Coatings | Distal End: Hydrophilic | Same |
| | Distal Coated Length: 46 cm | Distal Coated Length: 80 cm |
| | Proximal End: Polytetrafluoroethylene (PTFE) | Proximal End: None |
| Distal Portion Radiopacity | Tip includes a radiopaque platinum wire marker coil that is 10 cm long | Tip includes a radiopaque loaded polymer section that is 13.75 cm long |
| Accessories | Guidewire Introducer, Torque Device, Shaping Mandrel | Guidewire Introducer, Torque Device |
| Packaging Configuration | High-Density Polyethylene (HDPE) Tube, Tyvek Pouch, Carton, Shipper Box | Same |
| Sterilization Method | 100% Ethylene Oxide (EO) | Same |
| Shelf Life | 1 year | 6 months |
### I. Performance Data Supporting Substantial Equivalence
A summary of the supportive testing is presented in Table 2. These tests were performed per company approved protocols, test methods, and performance standards.
{6}
K260751 Page 3
Table 2: Test and Performance Specifications
| Test | Test Method Summary [standard and/or guidance] | Result |
| --- | --- | --- |
| Dimensional Verification | The guidewire diameter, usable length and coating length were measured [ISO 11070, FDA guidance “Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling” (FDA-2018-D-1775)]. | Pass |
| Visual Inspection | The guidewire was examined for surface defects and extraneous matter which could cause trauma to vessels during use. The distal tip was inspected to confirm the presence of smooth, rounded edges and correct tip configuration [ISO 11070, FDA-2018-D-1775]. | Pass |
| Radiopacity | Radiopacity testing per ASTM F640. | Pass |
| Delivery, Compatibility, and Retraction (Trackability and Simulated Use) | The guidewire can be inserted, tracked and removed per the instructions for use (IFU) within a simulated use model without damage to the guidewire [ FDA-2018-D-1775]. | Pass |
| Pushability | The proximal shaft of the guidewire has sufficient stiffness that the user can push the guidewire to the target anatomy without buckling [FDA-2018-D-1775]. | Pass |
| Tip Flexibility | Measure force to deflect guidewire tips at 5 mm, 10 mm, and 20 mm test lengths [FDA-2018-D-1775]. | Pass |
| Torqueability/Torque Strength | Torque response (average input-to-output lag) and torque turns to failure were measured in an anatomical model [FDA-2018-D-1775]. | Pass |
| Particulate | Particulates of various size ranges counted after simulated use in a tortuous path [FDA-2018-D-1775]. | Pass |
| Coating Lubricity and Durability | Frictional force of coated guidewires was determined after simulated use in a tortuous path [FDA-2018-D-1775]. | Pass |
| Tip Pull and Tensile Strength | Tensile strength / Tip Pull / Peak tensile force of the entire length of the guidewire [ISO 11070, FDA-2018-D-1775]. | Pass |
| Kink Resistance | There shall be no kinking of shaft (permanent deformation) after simulated use [FDA-2018-D-1775]. | Pass |
| Corrosion Resistance | Corrosion resistance evaluated the metallic components related to corrosion contacting the intended fluid path [ISO 11070]. | Pass |
| Fracture and Flexing | The guidewire does not fracture, loosen, or fail during repeated, tortuous use [ISO 11070]. | Pass |
| Compatibility with Other Devices | The guidewire is compatible with catheters and interventional devices up to the minimum size indicated in the device labeling [FDA-2018-D-1775]. | Pass |
| Usability and Design Validation | Physicians evaluated subject and predicate guidewires for various performance characteristics in an anatomically relevant model [IEC 62366-1, FDA guidance “Applying Human Factors and Usability Engineering to Medical Devices” (FDA-2011-D-0469)]. | Pass |
| Accessory Compatibility | The guidewires shall be compatible with their packaged accessories (introducer and torque device). | Pass |
{7}
K260751 Page 4
| Coating Integrity | Coating coverage was assessed after simulated use in a tortuous path anatomical model [ISO 11070, FDA-2018-D-1775]. | Pass |
| --- | --- | --- |
# J. Biocompatibility Testing
The following biocompatibility testing was conducted for the 35 Helix Guidewire:
Table 3: Summary of Subject Device Biocompatibility Testing Performed
| Test | Test Method Summary | Conclusion of Testing |
| --- | --- | --- |
| Cytotoxicity (MEM Elution) | Test in accordance with ISO 10993-5 | Non-cytotoxic |
| Sensitization | Test in accordance with ISO 10993-10 | Non-sensitizing |
| Irritation | Tested in accordance with ISO 10993-23 | Non-irritating |
| Acute Systemic Toxicity | Tested in accordance with ISO 10993-11 | Non-toxic |
| Material Mediated Rabbit Pyrogen | Tested in accordance with ISO 10993-11 and USP <151> | Non-pyrogenic |
| Direct Contact and Extract Method Hemolysis | Tested in accordance with ISO 10993-4 and ASTM F756 | Non-hemolytic |
| Platelet and Leukocyte Count Assay | Tested in accordance with ISO 10993-4 | The subject device had similar or lesser impact on platelet and leukocyte counts when compared to the predicate. |
| Sc5b-9 Complement Activation | Tested in accordance with ISO 10993-4 | The subject device had similar potential to activate the complement system when compared to the predicate. |
| In vitro blood flow loop | Tested in accordance with ISO 10993-4 | The thrombus deposition on the subject device was similar to or less than that of the comparator device. |
| Partial Thromboplastin Time (PTT) Test | Tested in accordance with ISO 10993-4 | No effect on the PTT. The subject device and the predicate are considered similar. |
# K. Sterilization
The 35 Helix Guidewire is sterilized using a validated EO process with a sterility assurance level of 1x10⁻⁶ validated per the overkill method in accordance with ISO 11135, “Sterilization of Health-Care Products - Ethylene Oxide - Requirements for The Development, Validation and Routine Control of A Sterilization Process For Medical Devices”, and tested for EO and ethylene chlorohydrin (ECH) residuals and bacterial endotoxin levels.
# L. Shelf Life and Packaging
Accelerated aging testing based on ASTM F1980 was conducted to verify packaged device performance to support a 6-month shelf-life claim. Device and package performance was verified by functional and performance testing.
# M. Conclusions
The subject device, 35 Helix Guidewire, has the same intended use, same indications for use, same fundamental design and similar device features as the predicate device. The differences in technological characteristics have been evaluated through testing and risk evaluation, and do not raise new or different questions of safety and effectiveness, supporting that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.