K260662 · Yofo Medical Technology Co., Ltd. · OAS · Jun 4, 2026 · Radiology
Device Facts
Record ID
K260662
Device Name
Dental Cone Beam Computed Tomography System
Applicant
Yofo Medical Technology Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Jun 4, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The Dental Cone Beam Computed Tomography System is intended to be used by medical institutions for X-ray image diagnosis on oral and maxillofacial areas through X-ray Cone Beam Computed Tomography (CBCT).
Device Story
Dental CBCT system; utilizes X-ray tube assemblies and flat panel detector to capture head/neck images; open-design gantry with electric lifting seat for upright patient positioning. System reconstructs 3D anatomical structures of oral/maxillofacial areas; provides multi-slice viewing, measurement, and contour plotting via workstation software. Operated by clinical staff in medical institutions; parameters adjusted via touch screen/control panel. Safety features include exposure switch and interlocks to prevent leakage. Modifications from predicate include inverter board and voice control board updates for stability and efficiency. Output supports clinical diagnosis and treatment planning.
Clinical Evidence
No clinical data included. Bench testing only: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 61223-3-7 (CBCT image quality, dose repeatability, and system accuracy) demonstrated performance equivalent to the predicate.
Technological Characteristics
Stationary anode tungsten target X-ray tube; amorphous silicon flat panel detector (153.6mm x 153.6mm, 100μm pixel pitch); 80-100kV tube voltage; 2-12mA tube current. Open gantry design with electric seat. Connectivity via workstation. Standards: IEC 60601-1, IEC 60601-1-2, IEC 61223-3-7.
Indications for Use
Indicated for X-ray image diagnosis of oral and maxillofacial areas in patients requiring CBCT imaging. Intended for use by medical institutions.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Dental Cone Beam Computed Tomography System (K243337)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 4, 2026
Yofo Medical Technology Co., Ltd.
% Kyra Kang
Landlink Healthcare Technology (Shanghai) Co., Ltd.
Rm. 1308, Baohua International Plz.,
W. Guangzhong Rd. 555, Jingan District
SHANGHAI, CHINA
Re: K260662
Trade/Device Name: Dental Cone Beam Computed Tomography System
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: OAS
Dated: February 27, 2026
Received: March 2, 2026
Dear Kyra Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260662 - Kyra Kang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260662 - Kyra Kang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260662 | ? |
| Please provide the device trade name(s). | | ? |
| Dental Cone Beam Computed Tomography System | | |
| Please provide your Indications for Use below. | | ? |
| The Dental Cone Beam Computed Tomography System is intended to be used by medical institutions for X-ray image diagnosis on oral and maxillofacial areas through X-ray Cone Beam Computed Tomography (CBCT). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
# K260662
## 510(k) Summary
### 1. Contact Details
Applicant Name: Yofo Medical Technology Co., Ltd.
Applicant Address: Building A1, National Health Industrial Park, Hi-tech
Industrial Development Zone, Hefei, Anhui, 230000,
P.R. China
Applicant Contact Telephone: 0551-65770683
Applicant Contact: Li Zhao
Applicant Contact Email: li.zhao@yofomedical.com
### 2. Device Name
| Device Trade Name: | Dental Cone Beam Computed Tomography System |
| --- | --- |
| Common Name: | X-ray, Tomography, Computed, Dental |
| Classification Name: | Computed tomography x-ray system |
| Regulation Number: | 892.1750 |
| Product Code: | OAS |
| Regulation Class: | II |
### 3. Legally Marketed Predicate Device
| Predicate # | Predicate Trade Name | Product Code |
| --- | --- | --- |
| K243337 | Dental Cone Beam Computed Tomography System | OAS |
### 4. Device Description Summary
The system consists of X-ray tube assemblies, collimators, flat panel detector, control devices (the touch screen, the control panel, the console panel and the exposure switch), auxiliary positioning devices (the lifting seat and the chin rest) and U-shape gantry (the X-ray tube assembly support, the base and the Detector Support), and workstation. The
1
{5}
system is an open design that allows patients to sit upright during a procedure. An electric powered seat is built into the scanner for proper patient positioning.
The system has 2 configurations: Configuration 1 has two X-ray tube assemblies (X-ray tube assembly A and X-ray tube assembly B) and two collimators (A and B), the Configuration 2 only has X-ray tube assembly A and collimator A.
The proposed device is a CBCT dental system utilizes X-ray to obtain cross-sectional images of the head or neck. The image software can reconstruct data and clearly display the 3D anatomical structure of the oral and maxillofacial areas, and achieve the multi-slice at any level, with accurate measurement, contour plots in the workstation software, supporting clinical applications in oral and maxillofacial diagnosis.
The Lifting Seat moves up and down via buttons on the system Control Panel allowing positioning and adjustment of patients at different heights; the Touch Screen facilitates the adjustment of projection parameters and the control of actions of the device; the Gating and the shielding door realize the safety interlock of the X-ray beam exposure to prevent the dose leakage caused by abnormal operation; the Exposure Switch controls the exposure output of the whole machine in real time during the scanning process.
There are four FOV sizes, 20cm x 18cm high resolution for diagnostic imaging of maxillary region, mandibular region, TMJ region and other craniofacial region. The 20cm x 18cm high resolution FOV is only available in Configuration 1. 17cm x 10.5cm high resolution, 8cm x8cm Super Resolution and 4cm x4cm Super Resolution for diagnostic imaging of maxillary region and mandibular region. Under High Resolution mode, the patients will receive lower dose radiation compared to Super Resolution. The radiation dose for each scanning mode differs between pediatric and adult patients.
The following modifications have been made to the Dental Cone Beam Computed Tomography System (K243337), which primarily includes:
- Inverter Board Updates: Enhancements to component layout, power supply efficiency, and PCB routing to improve thermal performance, production quality, and system stability.
- Voice Control Board Optimization: Circuit refinements including the addition of a filter and removal of unused components to ensure signal integrity and manufacturing efficiency.
- Labeling & Component Updates: Corrections to packaging symbols and updates to power cords to comply with U.S. national difference requirements.
All indications for use and the core imaging functions of the device remain unchanged.
2
{6}
## 5. Indications for Use
The Dental Cone Beam Computed Tomography System is intended to be used by medical institutions for X-ray image diagnosis on oral and maxillofacial areas through X-ray Cone Beam Computed Tomography (CBCT).
## 6. Technological Comparison
Table 1 Comparison of the Subject and Predicate Device
| Characteristics | Proposed device | Predicate device (K243337) | Discussion |
| --- | --- | --- | --- |
| **Indication for use** | The Dental Cone Beam Computed Tomography System is intended to be used by medical institutions for X-ray image diagnosis on oral and maxillofacial areas through X-ray Cone Beam Computed Tomography (CBCT). | The Dental Cone Beam Computed Tomography System is intended to be used by medical institutions for X-ray image diagnosis on oral and maxillofacial areas through X-ray Cone Beam Computed Tomography (CBCT). | Same |
| **Principle of Operation** | Cone Beam X-ray | Cone Beam X-ray | Same |
| **Generator** | High Frequency, Constant Potential | High Frequency, Constant Potential | Same |
| **X-ray tube** | Stationary anode | Stationary anode | Same |
| **Tube Voltage** | 80kV~100kV | 80kV~100kV | Same |
| **Tube Current** | 2-12 mA, 2-10mA | 2-12 mA, 2-10mA | Same |
| **Target material** | Tungsten | Tungsten | Same |
| **Focal spot** | 0.5 mm | 0.5 mm | Same |
| **Anode capacity** | 49 kHU | 49 kHU | Same |
| **Detector** | | | |
| **Type** | Amorphous Silicon | Amorphous Silicon | Same |
| **Total Pixel Area** | 153.6 mm x 153.6 mm | 153.6 mm x 153.6 mm | Same |
3
{7}
| Pixel Matrix | 1536 x1536 | 1536 x1536 | Same |
| --- | --- | --- | --- |
| Pixel Pitch | 100μm² | 100μm² | Same |
| Energy Range | 40-160 kVp | 40-160 kVp | Same |
| **Gantry** | | | |
| Resolution angle / Orbital angle | 360° | 360° | Same |
| Orientation | Vertical | Vertical | Same |
### 7. Non-Clinical Tests Summary
Based on the risk analysis of the modifications above, the following verification and validation activities were performed to establish substantial equivalence:
#### EMC & Electrical Safety:
- IEC 60601-1 (General safety): Testing was conducted on the modified inverter board and voice control board to verify electrical safety, dielectric strength, and mechanical integrity. All tests passed with no deviations.
- IEC 60601-1-2 (Electromagnetic compatibility): EMC testing was performed to confirm that the modifications do not introduce electromagnetic interference or affect immunity. The device met all requirements.
- IEC TR 60601-4-2 (EMC/EMI guidance): Applied as guidance for test setup and risk management; no non-conformities were identified.
#### Performance testing:
- IEC 61223-3-7 (Constancy and acceptance testing for CBCT): Performance verification was conducted including image quality, dose repeatability, and system accuracy. Results demonstrated that the modified device performs equivalently to the predicate device.
#### Clinical Performance Data:
No clinical performance data is included in this submission.
4
{8}
# 8. Conclusion
The proposed device has the same intended use as the cleared predicate device (K243337). Notwithstanding the modifications described in this Summary, comprehensive risk analysis and subsequent verification and validation testing have demonstrated that these modifications do not adversely affect the safety and effectiveness of the device. Therefore, the proposed device is Substantially Equivalent to the predicate device.
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.