K260635 · Nuvascular, Inc. · KRD · Jun 12, 2026 · Cardiovascular
Device Facts
Record ID
K260635
Device Name
HARBOR Occlusion Device
Applicant
Nuvascular, Inc.
Product Code
KRD · Cardiovascular
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HARBOR Occlusion Device is indicated for arterial embolization in the peripheral vasculature.
Device Story
Self-expanding braided nitinol mesh implant; used for arterial embolization in peripheral vasculature; delivered via microcatheter under fluoroscopic guidance; mechanical detachment system; radiopaque platinum marker band at proximal end; implant attached to 185 cm delivery system (stainless steel hypotube with nitinol wire); supplied sterile in hoop dispenser; single-use; physician-operated in clinical setting; provides vessel occlusion to manage blood flow; benefits include minimally invasive treatment of peripheral vascular conditions.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included compatibility evaluation with additional microcatheters and accelerated aging studies to support 1-year shelf-life extension.
Indicated for arterial embolization in the peripheral vasculature for patients requiring occlusion of peripheral vessels.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 12, 2026
Nuvascular, Inc.
Meadow Wang
Director, RA/QA
141 Innovation Dr., Suite 100
Irvine, California 92617
Re: K260635
Trade/Device Name: HARBOR Occlusion Device
Regulation Number: 21 CFR 870.3300
Regulation Name: Vascular embolization device
Regulatory Class: Class II
Product Code: KRD
Dated: February 26, 2026
Received: February 26, 2026
Dear Meadow Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260635 - Meadow Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260635 - Meadow Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FINN E.
DONALDSON
-S
Digitally signed by FINN
E. DONALDSON -S
Date: 2026.06.12
08:28:15 -04'00'
Finn Donaldson
Acting Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K260635
Device Name
HARBOR Occlusion Device
Indications for Use (Describe)
The HARBOR Occlusion Device is indicated for arterial embolization in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K260635 Page 1 of 9
# 510(k) SUMMARY
## SUBMITTER
Nuvascular Inc.
141 Innovation Dr. Suite 100, Irvine, CA 92617
Phone: 626-233-6758
Contact Person: Meadow S. Wang
Date Prepared: March 16, 2026
## DEVICE
Trade Name of the Device: Harbor Occlusion Device
Common Name: Vascular Embolization Device
Classification Name: Device, Vascular, for promoting embolization
Regulatory Class: Class II, 21 CFR 870.3300
Product Code: KRD
## PREDICATE DEVICE
Nuvascular Inc.: HARBOR Occlusion Device (KRD) (K250133)
## DEVICE DESCRIPTION
The HARBOR Occlusion Device is a self-expanding braided nitinol arterial embolization implant (Figure 1) supplied with components used for implantation (Figure 2). The Device has a radiopaque marker band attached to the proximal end of the implant. The implant is packaged collapsed within an Introducer sheath and attached to a Delivery System provided within a hoop dispenser.
Device implant sizing and dimensions are specified in Table 1. Based on its size, the device is designed to be used with commercially available microcatheters under fluoroscopy for delivery and implantation in the peripheral vasculature.

**Figure 1 Harbor Occlusion Device – Implant**
{5}
K260635 Page 2 of 9

Figure 2 Harbor Occlusion Device – Delivery System
The implant comes in various diameters, ranging from 4.0 to 13.5 mm for Unconstrained OD and 2.0 to 5.4 mm for Unconstrained Length.
The Harbor Occlusion Device will be available in the following with delivery system configurations:
| Product Model Number | Harbor Occlusion Device OD, Unconstrained, DIM A (mm) | Harbor Occlusion Device Length, Unconstrained, DIM C (mm) | Treatable Vessel Diameter Range (mm) |
| --- | --- | --- | --- |
| 04-OCCLUDE-020 | 4.0 | 2.0 | 2.0 – 2.5 |
| 04-OCCLUDE-025 | 4.0 | 2.5 | 2.0 – 2.5 |
| 04-OCCLUDE-030 | 4.5 | 2.8 | 2.5 – 3.0 |
| 04-OCCLUDE-035 | 5.0 | 3.0 | 3.0 – 3.5 |
| 04-OCCLUDE-040 | 5.5 | 3.2 | 3.5 – 4.0 |
| 04-OCCLUDE-045 | 6.0 | 3.4 | 4.0 – 4.5 |
| 04-OCCLUDE-050 | 6.5 | 3.6 | 4.5 – 5.0 |
| 04-OCCLUDE-055 | 7.0 | 3.8 | 5.0 – 5.5 |
| 04-OCCLUDE-060 | 7.5 | 4.0 | 5.5 – 6.0 |
| 04-OCCLUDE-065 | 8.0 | 4.2 | 6.0 – 6.5 |
| 04-OCCLUDE-070 | 8.5 | 4.4 | 6.5 – 7.0 |
{6}
K260635 Page 3 of 9
| 04-OCCLUDE-080 | 9.5 | 4.6 | 7.0 – 8.0 |
| --- | --- | --- | --- |
| 04-OCCLUDE-090 | 10.5 | 4.8 | 8.0 – 9.0 |
| 04-OCCLUDE-100 | 11.5 | 5.0 | 9.0 – 10.0 |
| 04-OCCLUDE-110 | 12.5 | 5.2 | 10.0 – 11.0 |
| 04-OCCLUDE-120 | 13.5 | 5.4 | 11.0 – 12.0 |
The implant is delivered through a microcatheter on a delivery system attached to the implant.
The delivery system attached to the Harbor Occlusion Device implant is 185 cm in length and an outer diameter suitable for delivery through commercially available Microcatheters with ID range from 0.0165” to 0.033”.
| HARBOR Catalog Number | Microcatheter Description | Inner Diameter (in) | Usable Length (cm) |
| --- | --- | --- | --- |
| 04-OCCLUDE-020 | Terumo Headway Duo | 0.0165 | 156 |
| | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| 04-OCCLUDE-025 | Terumo Headway Duo | 0.0165 | 156 |
| | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| 04-OCCLUDE-030 | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| 04-OCCLUDE-035 | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| 04-OCCLUDE-040 | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| 04-OCCLUDE-045 | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| | Terumo Headway 21 | 0.021 | 156 |
| | Medtronic Phenom 21 | 0.021 | 160 |
| 04-OCCLUDE-050 | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| | Terumo Headway 21 | 0.021 | 156 |
| | Medtronic Phenom 21 | 0.021 | 160 |
| 04-OCCLUDE-055 | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| | Terumo Headway 21 | 0.021 | 156 |
| | Medtronic Phenom 21 | 0.021 | 160 |
| 04-OCCLUDE-060 | Terumo Headway 17 | 0.017 | 150 |
| | Terumo VIA 17 | 0.0175 | 154 |
| | Terumo Headway 21 | 0.021 | 156 |
| | Medtronic Phenom 21 | 0.021 | 160 |
| 04-OCCLUDE-065 | Terumo Headway 27 | 0.027 | 156 |
| | Terumo VIA 27 | 0.027 | 154 |
{7}
K260635 Page 4 of 9
| | Terumo PG Pro Stryker XT-27 Medtronic Phenom 27 | 0.027 0.027 0.027 | 165 150 160 |
| --- | --- | --- | --- |
| 04-OCCLUDE-070 | Terumo Headway 27 Terumo VIA 27 Terumo PG Pro Stryker XT-27 Medtronic Phenom 27 | 0.027 0.027 0.027 0.027 0.027 | 156 154 165 150 160 |
| 04-OCCLUDE-080 | Terumo Headway 27 Terumo VIA 27 Terumo PG Pro Stryker XT-27 Medtronic Phenom 27 | 0.027 0.027 0.027 0.027 0.027 | 156 154 165 150 160 |
| 04-OCCLUDE-090 | Terumo Headway 27 Terumo VIA 27 Terumo PG Pro Stryker XT-27 Medtronic Phenom 27 | 0.027 0.027 0.027 0.027 0.027 | 156 154 165 150 160 |
| 04-OCCLUDE-100 | Terumo Headway 27 Terumo VIA 27 Terumo PG Pro Stryker XT-27 Medtronic Phenom 27 Terumo VIA 33 | 0.027 0.027 0.027 0.027 0.027 0.033 | 156 154 165 150 160 133 |
| 04-OCCLUDE-110 | Terumo Headway 27 Terumo VIA 27 Terumo PG Pro Stryker XT-27 Medtronic Phenom 27 Terumo VIA 33 | 0.027 0.027 0.027 0.027 0.027 0.033 | 156 154 165 150 160 133 |
| 04-OCCLUDE-120 | Terumo Headway 27 Terumo VIA 27 Terumo PG Pro Stryker XT-27 Medtronic Phenom 27 Terumo VIA 33 | 0.027 0.027 0.027 0.027 0.027 0.033 | 156 154 165 150 160 133 |
{8}
K260635 Page 5 of 9
## INDICATIONS FOR USE
The Harbor Occlusion Device is indicated for arterial embolization in the peripheral vasculature.
## DEVICE Attributes
Table 3 list the attributes of HARBOR Occlusion Device
| Table 3 HARBOR Occlusion Device Attributes | |
| --- | --- |
| Device Attributes | Nuvascular Harbor Occlusion Device (Subject Device) |
| Device Classification | Class II, KRD, 21 CFR 870.3300 |
| Intended Use/Indications for Use Statement | Indicated for arterial embolization in the peripheral vasculature |
| Device Function | Embolization in the arterial peripheral vasculature |
| Anatomical Location | Peripheral Arterial Vasculature |
| Harbor (Implant) | Self-expanding, Nitinol mesh occlusion device with non-blood contact platinum core to improve radiopacity. The device has a radiopaque marker band at the proximal end and an uncoated Nitinol ball at one end for attaching to the delivery system. |
| Implant Material | Nitinol |
| Implant Shape | Single Layer Braid Device Single Lobe with Cylindrical shape |
| Delivery System Material | Stainless Steel Hypotube with Nitinol Wire |
| Introducer Sheath | HDPE, High Density Polyethylene |
| Radiopaque Marker | Platinum Marker Band at the proximal end of the Harbor Occlusion Device |
| Delivery Method | Delivered through the commercially available Microcatheters with ID range from 0.0165' to 0.033' |
| Delivery system Length | 185 cm |
| Detachment System | Mechanical Detachment |
| Materials of Construction for Implant | Nitinol Wire (Main Substances: Nickel: 55-57%, Titanium: 43-45%) |
| Materials of Construction for Delivery system | Stainless Steel Hypotube with inner Nitinol wire |
| Marker Bands | Platinum |
| Packaging | Harbor Occlusion Device includes implant attached to 185 cm long delivery system, loaded in a hoop dispenser |
| Sterilization Process | Ebeam |
| Method of Supply | Sterile, single use |
{9}
K260635 Page 6 of 9
## COMPARISION OF TECHNOLOGICAL CHARACTERISTICS OF THE HARBOR OCCLUSION DEVICE WITH THE PREDICATE DEVICE
The predicate device for this Special 510(k) submission is the HARBOR Occlusion Device (K250133), which is the submitter’s own legally marketed device previously cleared by U.S. Food and Drug Administration under K250133.
The subject device is a shelf-life extension from 6 months to 1 year based on the predicate device. The modification does not affect the intended use or the fundamental scientific technology of the predicate device. The modification was evaluated through the submitter’s design control process, including risk analysis and design verification and validation activities.
The subject device is to add additional compatible delivery microcatheters. Adding additional compatible delivery microcatheters does not change the intended use or the fundamental scientific technology of the predicate device. The modification was evaluated through the submitter’s design control process, including risk analysis and design verification and validation activities.
Table 1 compares the subject device, HARBOR Occlusion Device, to the predicate device, HARBOR Occlusion Device (K250133).
| Device Attributes | Nuvascular HARBOR Occlusion Device (Subject Device) | Nuvascular HARBOR Occlusion Device (Predicate Device) (K250133) | Justification for Difference (If Present) |
| --- | --- | --- | --- |
| Device Classification | Class II, KRD, 21 CFR 870.3300 | Class II, KRD, 21 CFR 870.3300 | Same classification |
| Intended Use/ Indications for Use Statement | Indicated for arterial embolization in the peripheral vasculature | Indicated for arterial embolization in the peripheral vasculature | Same Intended Use/Indication for Use Statement |
| Device Function | Arterial embolization in the peripheral vasculature | Arterial embolization in the peripheral vasculature | Same device function |
| Anatomical Location | Peripheral Arterial Vasculature | Peripheral Arterial Vasculature | Use for the same anatomical location |
| HARBOR (Implant) | Self-expanding, Nitinol mesh occlusion device with non-blood contact platinum core to improve radiopacity. The device has a radiopaque marker band at the proximal end and an uncoated Nitinol ball at one end for attaching to the delivery system. | Self-expanding, Nitinol mesh occlusion device with non-blood contact platinum core to improve radiopacity. The device has a radiopaque marker band at the proximal end and an uncoated Nitinol ball at one end for attaching to the delivery system. | Same Implant |
| Product Model Number | 04-OCCLUDE-020 04-OCCLUDE-025 04-OCCLUDE-030 04-OCCLUDE-035 04-OCCLUDE-040 04-OCCLUDE-045 04-OCCLUDE-050 04-OCCLUDE-055 | 04-OCCLUDE-020 04-OCCLUDE-025 04-OCCLUDE-030 04-OCCLUDE-035 04-OCCLUDE-040 04-OCCLUDE-045 04-OCCLUDE-050 04-OCCLUDE-055 | Same Product Model Number |
{10}
K260635 Page 7 of 9
| | 04-OCCLUDE-060 04-OCCLUDE-065 04-OCCLUDE-070 04-OCCLUDE-080 04-OCCLUDE-090 04-OCCLUDE-100 04-OCCLUDE-110 04-OCCLUDE-120 | 04-OCCLUDE-060 04-OCCLUDE-065 04-OCCLUDE-070 04-OCCLUDE-080 04-OCCLUDE-090 04-OCCLUDE-100 04-OCCLUDE-110 04-OCCLUDE-120 | |
| --- | --- | --- | --- |
| HARBOR (Implant) OD, Unconstrained | 4.0 mm 4.0 mm 4.5 mm 5.0 mm 5.5 mm 6.0 mm 6.5 mm 7.0 mm 7.5 mm 8.0 mm 8.5 mm 9.5 mm 10.5 mm 11.5 mm 12.5 mm 13.5 mm | 4.0 mm 4.0 mm 4.5 mm 5.0 mm 5.5 mm 6.0 mm 6.5 mm 7.0 mm 7.5 mm 8.0 mm 8.5 mm 9.5 mm 10.5 mm 11.5 mm 12.5 mm 13.5 mm | The same unconstrained OD |
| HARBOR (Implant) Unconstrained Length | 2.0 mm 2.5 mm 2.8 mm 3.0 mm 3.2 mm 3.4 mm 3.6 mm 3.8 mm 4.0 mm 4.2 mm 4.4 mm 4.6 mm 4.8 mm 5.0 mm 5.2 mm 5.4 mm | 2.0 mm 2.5 mm 2.8 mm 3.0 mm 3.2 mm 3.4 mm 3.6 mm 3.8 mm 4.0 mm 4.2 mm 4.4 mm 4.6 mm 4.8 mm 5.0 mm 5.2 mm 5.4 mm | Same, Implant Unconstrained Length |
| Target Vessel Diameter | 2.0 mm-2.5 mm 2.5 mm-3.0 mm 3.0 mm-3.5 mm 3.5 mm-4.0 mm 4.0 mm-4.5 mm 4.5 mm-5.0 mm 5.0 mm-5.5 mm 5.5 mm-6.0 mm 6.0 mm-6.5 mm 6.5 mm-7.0 mm 7.0 mm-8.0 mm 8.0 mm-9.0 mm 9.0 mm-10.0 mm 10.0 mm-11.0 mm 11.0 mm-12.0 mm | 2.0 mm-2.5 mm 2.5 mm-3.0 mm 3.0 mm-3.5 mm 3.5 mm-4.0 mm 4.0 mm-4.5 mm 4.5 mm-5.0 mm 5.0 mm-5.5 mm 5.5 mm-6.0 mm 6.0 mm-6.5 mm 6.5 mm-7.0 mm 7.0 mm-8.0 mm 8.0 mm-9.0 mm 9.0 mm-10.0 mm 10.0 mm-11.0 mm 11.0 mm-12.0 mm | Same Target Vessel Diameter Scope. |
{11}
K260635 Page 8 of 9
| Implant Shape | Single Layer Braid Device Single Lobe with Cylindrical shape | Single Layer Braid Device Single Lobe with Cylindrical shape | Same Implant Shape |
| --- | --- | --- | --- |
| Loading Device Method | Introducer Sheath, HDPE | Introducer Sheath, HDPE | Same Loading Method |
| Radiopaque Marker Band | Platinum Marker Band at the proximal end of the HARBOR Occlusion Device | Platinum Marker Band at the proximal end of the HARBOR Occlusion Device | Same Radiopaque Marker Band |
| Compatible with Microcatheter | Delivered and compatible with the commercialized microcatheters from ID 0.0165' to 0.033' See all compatible microcatheters in the proposed IFU. | Delivered and compatible with Terumo Headway 17, 0.017' or Headway 27, 0.027' | To support adding additional compatible microcatheters, supporting data is in below test reports: TR-04-0018 Revision B Design Verification and Validation Test Report TR-04-0027 Revision C, Particulate Test for HARBOR Occlusion Device |
| Delivery System Length | 185 cm | 185 cm | Same Delivery System Length |
| Detachment System | Mechanical Detachment | Mechanical Detachment | Same Mechanical Detachment Method |
| **Device Material** | | | |
| Materials of Construction for Implant | Nitinol Wire (Main Substances: Nickel: 55-57%, Titanium: 43-45%) | Nitinol Wire (Main Substances: Nickel: 55-57%, Titanium: 43-45%) | Same material of Implant |
| Materials of Construction for Delivery System | Stainless Steel Hypotube with inner Nitinol Delivery Wire | Stainless Steel Hypotube with inner Nitinol Delivery Wire | Same material of Delivery System |
| Marker Bands | Platinum | Platinum | Same material of Marker Bands |
| **Other Attributes** | | | |
| Packaging Configuration | Harbor Occlusion Device includes implant attached to 185 cm long delivery system, loaded in a hoop dispenser | Harbor Occlusion Device includes implant attached to 185 cm long delivery system, loaded in a hoop dispenser | Same Packaging Configuration |
| Sterilization Process | Ebeam | Ebeam | Same Sterilization Method |
| Method of Supply | Sterile, single use | Sterile, single use | Same Supply Method |
| Shelf-Life | 1 Year | 6 Months | TR-04-0023 Revision B is for 1 year Shelf-Life Study Report |
{12}
K260635 Page 9 of 9
# SUMMARY OF HARBOR OCCLUSION DEVICE PERFORMANCE TEST
The data presented in this submission demonstrate that the subjective device, HARBOR Occlusion Device has successfully completed the following performance testing activities and performs comparably to the predicate device, HARBOR Occlusion Device(K250133):
- Bench performance testing was conducted to evaluate compatibility with additional commercially available microcatheters listed in the proposed IFU. The test results demonstrated that the device maintains acceptable functional performance with the proposed compatible microcatheters.
- Shelf-life testing was conducted to support an extension of the device shelf life to 1 year. The results demonstrated that the device continues to meet applicable performance specifications following the requirements of accelerated aging and design specifications.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE:
The data presented in this submission demonstrates the technological similarity and equivalency of the subject HARBOR Occlusion Device when compared with the predicate device, HARBOR Occlusion Device (K250133).
The Subject device,
- has the same intended use as the predicate device,
- has the same indication for use as the predicate device,
- uses the same operating principle as the predicate device,
- uses the same materials as the predicate device,
- incorporates the same design and construction as the predicate device,
- has the same unconstrained OD range as the predicate device,
- has the same unconstrained Length range as the predicate device,
- has the same packaging configuration and method as the predicate device,
- has the same sterilization method as the predicate device.
In summary, the subject HARBOR Occlusion Device described in this submission is substantially equivalent to the predicate device, HARBOR Occlusion Device (K250133).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.