K260618 · Ulthera, Inc. · OHV · May 21, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260618
Device Name
Ulthera® System
Applicant
Ulthera, Inc.
Product Code
OHV · General, Plastic Surgery
Decision Date
May 21, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260618 · May 21, 2026
Ulthera® System
Ulthera, Inc.
Clinician-reported real-world data (RWD); Global post-market surveillance (PMS) data
RWD and PMS data were used to supplement clinical literature to demonstrate the safety and effectiveness of the Ulthera System for the expanded indication of treating skin laxity on the knees, specifically monitoring for adverse events such as joint, muscle, or tendon damage.
Real-world data collection; Retrospective collection of clinician-reported data
Patients treated for knee skin laxity in routine clinical practice
Not applicable for this study
Safety (adverse events, specifically joint, muscle, or tendon damage)
Global Post-Market Surveillance; Post-market surveillance
Global patient population treated with the Ulthera System
Not applicable for this study
Safety (orthopedic-based adverse events of the knee)
Indications for Use
The Ulthera® System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin. The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to: • lift the eyebrow • lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions • improve lines and wrinkles of the décolleté. • improve the appearance of skin laxity on the abdomen, anterior arms, posterior arms, and knees The Ulthera® System in conjunction with the Ulthera® DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to: • ensure proper coupling of the transducer to the skin • confirm appropriate depth of treatment such as to avoid bone
Device Story
System uses micro-focused ultrasound energy to induce thermal coagulation in dermal/subdermal tissue; results in skin contraction/tightening. Input: acoustic waves directed to treatment area; supplemental ultrasound imaging (DeepSEE transducer) for visualization up to 8mm depth. Output: visual display of tissue layers for operator guidance. Used in clinical settings by trained medical professionals. Imaging ensures proper transducer coupling and depth confirmation to avoid bone. Benefits: non-invasive aesthetic improvement of skin laxity and wrinkles; no hardware/software changes from predicate.
Clinical Evidence
Evidence includes a literature search identifying three clinical studies (n=44 total) evaluating knee skin laxity improvement. Endpoints included blinded clinician photographic assessment, Global Aesthetic Improvement Scores (GAIS), and patient satisfaction surveys. Results showed clinically significant improvement up to 180 days post-treatment. Safety data from studies and real-world evidence showed no unexpected adverse events, specifically no joint, muscle, or tendon damage.
Technological Characteristics
Focused ultrasound stimulator; low-energy acoustic deposition. System components: control console, handpiece, interchangeable transducers. Imaging: DeepSEE transducer (up to 8mm depth). Connectivity: USB Wi-Fi adapter. Power: 100-240 VAC. Software: version 2.1.6400, Windows 10 OS. Electrical safety: compliant with IEC standards.
Indications for Use
Indicated for adult patients seeking non-invasive dermatological aesthetic treatment to lift eyebrows, lift lax submental and neck tissue, improve décolleté lines/wrinkles, and improve skin laxity on the abdomen, anterior/posterior arms, and knees.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Ulthera® System (UC-1 Control Unit PRIME) (K250418)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 21, 2026
Ulthera, Inc.
Scott Jewett
Regulatory Affairs Manager
6501 Six Forks Rd.
Raleigh, North Carolina 27560
Re: K260618
Trade/Device Name: Ulthera® System
Regulation Number: 21 CFR 878.4590
Regulation Name: Focused Ultrasound Stimulator System For Aesthetic Use
Regulatory Class: Class II
Product Code: OHV, IYO
Dated: February 25, 2026
Received: February 25, 2026
Dear Scott Jewett:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260618 - Scott Jewett
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260618 - Scott Jewett
Page 3
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2026.05.21
21:49:12 -04'00"
Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260618 | |
| Device Name Ulthera® System | |
| Indications for Use (Describe) The Ulthera® System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin. The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to: • lift the eyebrow • lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions • improve lines and wrinkles of the décolleté. • improve the appearance of skin laxity on the abdomen, anterior arms, posterior arms, and knees The Ulthera® System in conjunction with the Ulthera® DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to: • ensure proper coupling of the transducer to the skin • confirm appropriate depth of treatment such as to avoid bone | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Ulthera, Inc.
K260618
510(k) Summary
Traditional 510(k)
Ulthera® System Expanded Indications (Knees)
# 510(k) Summary
## 1 APPLICANT
Company's Name: Ulthera, Inc.
Company's Address: 6501 Six Forks Road
Raleigh, NC 27615
Telephone: 919.215.4879
Contact Person: Scott Jewett, Regulatory Affairs Manager
Date Prepared: 12 May 2026
## 2 DEVICE
Device Name: Ulthera® System
Classification Name: Focused Ultrasound Stimulator System for Aesthetic Use
21 C.F.R § 878.4590, Focused Ultrasound Stimulator Use
Regulatory Class: Class II
Product Codes: OHV, IYO
Applicable Guidances: Focused Ultrasound Stimulator System for Aesthetic Use
Marketing Clearance of Diagnostic Ultrasound Systems and Transducers
## 3 PREDICATE DEVICE
Ulthera® System (UC-1 Control Unit PRIME), Ulthera, Inc., K250418
Page 1 of 5
{5}
Ulthera, Inc. Ulthera® System Expanded Indications (Knees)
K260618
510(k) Summary Traditional 510(k)
# 4 DEVICE DESCRIPTION
The Ulthera® System consists of the Ulthera® Control Unit (with system software), a handpiece with cable, and interchangeable transducers. The device produces controlled tissue coagulation below the skin surface (epidermis) within the first few millimeters of tissue (dermis) using highly focused, low-energy ultrasound deposition. The Ulthera® System directs micro-focused acoustic waves to the treatment area at desired depths without affecting or requiring a secondary action to protect the skin surface. The operator may also use the device's supplemental imaging capability to visualize the treatment area and aid in assuring full/proper skin contact of the Ulthera® System transducer to the skin in the target area.
# 5 INTENDED USE / INDICATIONS FOR USE
The Ulthera® System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin.
The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to:
- Lift the eyebrow
- Lift lax submental (beneath the chin) and neck tissue, which can also affect the appearance of lax tissue in the submental and neck regions
- Improve lines and wrinkles of the décolleté
- Improve the appearance of skin laxity on the abdomen, anterior arms, posterior arms, and knees
The Ulthera® System, in conjunction with the Ulthera® DeepSEE® transducer, allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:
- Ensure proper coupling of the transducer to the skin
- Confirm appropriate depth of treatment such as to avoid bone
# 6 SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The purpose of this 510(k) notification relates to expanded indications for the Ulthera® System to improve the appearance of skin laxity on the knees. Low-intensity, highly focused ultrasound is the main technological principle of both the subject and the predicate device. In both devices, focused ultrasound energy is delivered below the skin and produces discrete points of thermal coagulation that results in contraction of the skin, which produces a lifting or tightening effect and improves appearance of lax tissue, lines, and wrinkles. There are no changes in design, hardware, or software between subject and predicate device. Comparison between the subject and predicate device are summarized in Table 1 below:
{6}
Ulthera, Inc.
Ulthera® System Expanded Indications (Knees)
K260618
510(k) Summary
Traditional 510(k)
Table 1: Substantial Equivalence Table Comparing Subject & Predicate Device
| | Ulthera® System (Subject) | Ulthera® System K250418 (Predicate) | Similarities and significant differences to the predicate |
| --- | --- | --- | --- |
| Intended Use | The Ulthera® System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin. | The Ulthera® System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin. | Identical |
| Indications for Use | The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to: • lift the eyebrow • lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions • improve lines and wrinkles of the décolleté. • improve the appearance of skin laxity on the abdomen, anterior arms, posterior arms, and knees The Ulthera® System in conjunction with the Ulthera® DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to: • ensure proper coupling of the transducer to the skin • confirm appropriate depth of treatment such as to avoid bone | The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to: • lift the eyebrow • lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions • improve lines and wrinkles of the décolleté. • improve the appearance of skin laxity on the abdomen, anterior arms, and posterior arms The Ulthera® System in conjunction with the Ulthera® DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to: • ensure proper coupling of the transducer to the skin • confirm appropriate depth of treatment such as to avoid bone | Different Additional indications added for the knees. Clinical data, real-world evidence (RWE), and post market surveillance (PMS) data have been provided to demonstrate the safety and effectiveness of the device for the expanded indications and support the expansion of the subject device’s indications for use. |
| User Population | Treatment of adult patient population by trained medical professional | Treatment of adult patient population by trained medical professional | Identical |
| Main System Components | • Control console • Handpiece • Transducers | • Control console • Handpiece • Transducers | Identical |
| Additional Components Required for Operation | • ACLF Power Cord with pigtail adapter • Wibu USB Access Key | • ACLF Power Cord with pigtail adapter • Wibu USB Access Key | Identical |
| Console Dimensions | Height: <16.69" (424 mm) Width: 19.4" (493.6 mm) Depth: 13.1" (333 mm) | Height: <16.69" (424 mm) Width: 19.4" (493.6 mm) Depth: 13.1" (333 mm) | Identical |
Page 3 of 5
{7}
Ulthera, Inc.
Ulthera® System Expanded Indications (Knees)
K260618
510(k) Summary
Traditional 510(k)
| | Ulthera® System (Subject) | Ulthera® System K250418 (Predicate) | Similarities and significant differences to the predicate |
| --- | --- | --- | --- |
| Console Connection Ports | 6 | 6 | Identical |
| Weight | Weight: ≤27 lbs (12.2 kg) | Weight: ≤27 lbs (12.2 kg) | Identical |
| Display | 18.5” screen; aspect ratio of 16:9 with 1920 x 1080 resolution | 18.5” screen; aspect ratio of 16:9 with 1920 x 1080 resolution | Identical |
| Power Source | 100-240 VAC, 50/60 Hz, 3A max Fuse: (2) 5x20mm, 6.3A fast acting, 250V | 100-240 VAC, 50/60 Hz, 3A max Fuse: (2) 5x20mm, 6.3A fast acting, 250V | Identical |
| Software | version 2.1.6400 | version 2.1.6400 | Identical |
| Treatment Modes | Amplify See.Plan.Treat | Amplify See.Plan.Treat | Identical |
| Ultrasound Color Maps | Black & White (monochromatic) Bone (blue scale) Twilight (multicolor) | Black & White (monochromatic) Bone (blue scale) Twilight (multicolor) | Identical |
| Wireless Connectivity | Connection via USB Wi-Fi Adapter | Connection via USB Wi-Fi Adapter | Identical |
| Operating System | Windows 10 | Windows 10 | Identical |
| Electrical Safety & EMC | Compliant with relevant IEC standards for console and ACLF power cord | Compliant with relevant IEC standards for console and ACLF power cord | Identical |
# 7 PERFORMANCE DATA
## 7.1 Clinical Testing
To support the device’s indications for use, a comprehensive literature search was performed, and three (3) unique clinical studies were identified that used the Ulthera® System to improve the appearance of skin laxity on the knees. These studies included 44 total subjects, which is comparable to and/or exceeds the number of subjects evaluated in the previous 510(k) clearances for the Ulthera® System and other focused ultrasound stimulator systems for aesthetic use.
These clinical studies demonstrated clinically significant skin improvement up to 180 days post-treatment via a variety of endpoints, including blinded clinician photographic assessment, physician and subject Global Aesthetic Improvement Scores (GAIS), and patient satisfaction surveys. From a safety perspective, there were no unexpected or severe adverse events (AEs), and all observed AEs resolved without long-term sequelae.
Page 4 of 5
{8}
Ulthera, Inc.
Ulthera® System Expanded Indications (Knees)
K260618
510(k) Summary
Traditional 510(k)
Additionally, the Company also obtained real-world data (RWD) gathered from clinicians who regularly treat the knee region with the Ulthera® System. No unexpected adverse events were reported, including no reports of joint, muscle, or tendon damage.
Lastly, there has not been any orthopedic-based adverse events of the knee reported from global post marketing surveillance of the device. Collectively, the totality of evidence supports the safety of the Ulthera® System for treating the knees.
# 8 CONCLUSIONS
The subject device has the same general intended use and principle of operation as the predicate device, and there have been no changes in device hardware, software, or firmware compared to the predicate device. An analysis of clinical, real-world, and post marketing data demonstrates that the device is as safe and effective as the predicate device for the requested indications for use.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.