The BiWaze® Clear System is indicated for the loosening and mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and can provide supplemental oxygen when used with an oxygen supply. The system provides three therapies: Positive Expiratory Pressure (PEP), Oscillation (OSC), and delivery of saline via integrated nebulizer (NEB). The BiWaze Clear System is for use on adult or pediatric patients in acute care (aged 2 years and older) and home care environments (aged 5 years and older). The BiWaze Clear System is compatible with various patient interfaces including face mask, mouthpiece, a trach adapter for connection to an endotracheal or tracheostomy tube in acute and home care settings. The BiWaze Clear System may be used in-line with a ventilator in acute care environments only.
Device Story
BiWaze Clear System provides airway clearance and lung expansion therapy; utilizes PEP, high-frequency oscillation (up to 5 Hz), and integrated vibrating mesh nebulization (Aerogen Solo). Inputs include user-defined pressure/frequency settings via touchscreen; device delivers controlled air/oxygen flow through a dual-lumen breathing circuit with coaxial bacterial/viral filter. Used in acute and home care settings by clinicians or patients; provides real-time pressure monitoring and SpO2/heart rate display. Output facilitates mucus mobilization via mucociliary escalator; reduces airway obstruction and atelectasis. Bluetooth/Wi-Fi connectivity enables data transmission. Device assists in respiratory management, potentially reducing infection risk and improving gas exchange.
Clinical Evidence
Bench testing only. Performance testing evaluated device function with and without the added filter across various pressure/frequency settings and resistance/compliance profiles to simulate adult and pediatric patient conditions. Results demonstrated substantial equivalence to the predicate.
Technological Characteristics
Electro-mechanical device; provides PEP (≤30 cmH2O) and oscillation (≤70 cmH2O, 5 Hz). Materials comply with ISO 10993-1 and ISO 18562-1. Features touchscreen interface, Bluetooth/Wi-Fi connectivity, and dual-lumen breathing circuit with coaxial bacterial/viral filter. Powered by air/oxygen source. Integrated Aerogen Solo nebulizer.
Indications for Use
Indicated for loosening/mobilization of secretions, lung expansion, and treatment/prevention of pulmonary atelectasis in adult and pediatric patients (acute care ≥2 years; home care ≥5 years). Compatible with face mask, mouthpiece, or trach adapter; may be used in-line with ventilators in acute care.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 24, 2026
Abmrc, LLC
% Paul Dryden
Consultant
ProMedic Consulting, LLC
131 Bay Pt. Dr. NE
St. Petersburg, Florida 33704
Re: K260604
Trade/Device Name: BiWaze Clear System
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: NHJ
Dated: July 23, 2026
Received: July 24, 2026
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260604 - Paul Dryden
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260604 - Paul Dryden
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binoy J.
Mathews -S
Digitally signed by
Binoy J. Mathews -S
Date: 2026.08.24
13:43:05 -04'00'
For
James Lee, PhD
Division Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260604
Device Name
BiWaze Clear System
Indications for Use (Describe)
The BiWaze® Clear System is indicated for the loosening and mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and can provide supplemental oxygen when used with an oxygen supply. The system provides three therapies: Positive Expiratory Pressure (PEP), Oscillation (OSC), and delivery of saline via integrated nebulizer (NEB).
The BiWaze Clear System is for use on adult or pediatric patients in acute care (aged 2 years and older) and home care environments (aged 5 years and older).
The BiWaze Clear System is compatible with various patient interfaces including face mask, mouthpiece, a trach adapter for connection to an endotracheal or tracheostomy tube in acute and home care settings.
The BiWaze Clear System may be used in-line with a ventilator in acute care environments only.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
K260604
510(k) Summary
Page 1 of 6
Date Prepared: 24-Aug-26
Sponsor: ABMRC LLC
860 Blue Gentian Road
Suite 200
Eagan, MN 55121, USA
Sponsor Contact: Priyanka Paul - Director, Quality & Regulatory
Tel - +91-7299334392
Submission Correspondent: Paul Dryden
ProMedic, LLC
Proprietary or Trade Name: BiWaze Clear System.
Common/Usual Name: Noncontinuous ventilator (IPPB)
Classification: NHJ, 868.5905
Predicate Device: BiWaze Clear System, ABMRC LLC, K231728
Common/Usual Name: Noncontinuous ventilator (IPPB)
Classification: NHJ, 868.5905
# Device Description:
The proposed device is similar to the predicate cleared under K231728 except for the addition of Bluetooth connectivity and the addition of an FDA-cleared filter (K192713).
The BiWaze Clear System assists patients in loosening and mobilizing secretions as well as treating and preventing atelectasis by providing lung expansion and high frequency oscillation therapies. The oscillating lung expansion therapy of the BiWaze Clear System is intended to reduce airway obstructions caused by secretions occupying the lower airways, prevent respiratory tract infections, re-expand the collapsed areas of the lung, thereby enhancing gas exchanges and reducing inflammatory response.
BiWaze Clear delivers three therapies: Positive Expiratory Pressure (PEP), Oscillation (OSC), and nebulizer (NEB).
- Positive Expiratory Pressure (PEP): During PEP, the system delivers a programmed positive pressure which the patient exhales against to open and expand the patient's airways. The nebulizer can be configured to run during PEP therapy to help move saline throughout the airways.
- Oscillation (OSC): During OSC, the system oscillates the airways with pulses of positive pressure to thin secretions and mobilizes them from the lower airways to the upper airways so they can be coughed or suctioned out. The nebulizer can be configured to run during OSC therapy to help move saline throughout the airways.
- Nebulize (NEB): During NEB, the system powers only the Aerogen Solo vibrating mesh nebulizer. This therapy gives the patient a break from PEP or OSC while the patient receives their nebulized saline.
The BiWaze Clear System can be used in conjunction with the various patient interfaces such as facemask, mouthpiece or a trach adapter which connects to a patient's endotracheal or tracheostomy tube, or in-line with a ventilator. It is intended to deliver therapy to pediatric and adult patients in acute, post-acute and home care
{5}
510(k) Summary
Page 2 of 6
K260604
settings.
The BiWaze Clear System provides a closed-circuit therapy with the Dual Lumen Breathing Circuit that prevents aerosolized exhalation of air from escaping the handset or breathing tube before being filtered by a coaxial bacterial/viral filter.
The BiWaze Clear System consists of a main controller and single patient, multi-use breathing circuit which includes various patient interface options.
- Mouthpiece
- Face Mask
- Trach Adapter
- Two Step Adaptor
### Principle of Operation
The BiWaze Clear System delivers positive expiratory pressure for lung expansion therapy and produces high-frequency oscillations up to 300 beats per minute (5 Hz) and peak positive pressures ≤ 70 cmH₂O to break up mucus and move it up the mucociliary escalator.
The device can deliver nebulized saline during oscillation and positive expiratory pressure phases. The nebulizer can also be used independently to deliver aerosol saline.
### Indications for Use:
The BiWaze Clear System is indicated for the loosening and mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and can provide supplemental oxygen when used with an oxygen supply. The system provides three therapies: Positive Expiratory Pressure (PEP), Oscillation (OSC), and delivery of saline via integrated nebulizer (NEB).
The BiWaze Clear System is for use on adult or pediatric patients in acute care (aged 2 years and older) and home care environments (aged 5 years and older).
The BiWaze Clear System is compatible with various patient interfaces including face mask, mouthpiece, a trach adapter for connection to an endotracheal or tracheostomy tube in acute and home care settings.
The BiWaze Clear System may be used in-line with a ventilator in acute care environments only.
### Patient Population:
The BiWaze Clear System is for use on adult or pediatric patients in acute care (aged 2 years and older) and home care environments (aged 5 years and older).
### Environments of use:
The BiWaze Clear System is used in acute care (aged 2 years and older) and home care environments (aged 5 years and older).
### Substantial Equivalence Discussion
The BiWaze Clear System is viewed as substantially equivalent to the predicate devices because:
{6}
K260604
510(k) Summary
Page 3 of 6
Indications – The proposed indications for use for mobilization of secretions, lung expansion therapy, treatment and prevention of pulmonary atelectasis, and ability to provide supplemental oxygen when used with compressed oxygen are identical to the predicate.
Patient Population – The patient population is identical to the predicate.
Environment of Use – The environment of use is identical to the predicate. However, BiWaze Clear use in-line with ventilator is intended only in the acute care environment.
Technology – Functionally, the performance and therapy mode functions are similar to the predicate device.
Performance Data: Performance bench testing was conducted related to the addition of a filter to the breathing circuit. The comparative testing to the predicate was found to be substantially equivalent.
Biocompatibility: The design and materials in the gas and fluid pathways are identical to the predicate, BiWaze Clear System (K231728).
### Discussion of Differences
The difference between the proposed device and the predicate are:
- Enabling Bluetooth connectivity
- Addition of an FDA-cleared filter to the breathing circuit
These changes do not raise different questions of safety and effectiveness compared to the predicate and can be considered substantially equivalent.
### Nonclinical Testing Summary
Based upon the modification the following nonclinical testing performed as part of demonstrating substantial equivalence to the predicate.
- Software verification and validation was performed
- Cybersecurity
- Wireless co-existence testing
- IEC 60601-1
- IEC 60601-1-2
- Testing was conducted to assess device performance with the addition of a filter to the breathing circuit. The comparative testing was performed with and without the filter. Testing with the filter was found to be substantially equivalent to the predicate. Testing was performed with PAP and OSC therapies and included various, clinically relevant, pressures and frequencies (low, medium, and high) with a range of resistance and compliance settings to simulate patient condition and age (adult vs. pediatric).
In all cases the modified device met it performance specifications and can be considered substantially equivalent to the predicate.
{7}
K260604
510(k) Summary
Page 4 of 6
Table - Comparison of Subject vs. Predicate
| | Subject Device BiWaze Clear | Predicate BiWaze Clear – K231728 | Comparison |
| --- | --- | --- | --- |
| Product Code | NJH - Noncontinuous ventilator (IPPB) | NJH - Noncontinuous ventilator (IPPB) | Identical |
| CFR | 868.5905 | 868.5905 | Identical |
| Indications for Use | The BiWaze® Clear System is indicated for the loosening and mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and can provide supplemental oxygen when used with an oxygen supply. The system provides three therapies: Positive Expiratory Pressure (PEP), Oscillation (OSC), and delivery of saline via integrated nebulizer (NEB). The BiWaze Clear System is for use on adult or pediatric patients in acute care (aged 2 years and older) and home care environments (aged 5 years and older). The BiWaze Clear System is compatible with various patient interfaces including face mask, mouthpiece, a trach adapter for connection to an endotracheal or tracheostomy tube in acute and home care settings. The BiWaze Clear System may be used in-line with a ventilator in acute care environments only. | The BiWaze Clear System is indicated for the loosening and mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and can provide supplemental oxygen when used with an oxygen supply. The BiWaze Clear System is for use on adult or pediatric patients in acute care (aged 2 years and older) and home care environments (aged 5 years and older). The BiWaze Clear System may be used with patient interfaces including face mask, mouthpiece, a trach adapter to endotracheal or tracheostomy tube in acute and home care environments. The BiWaze Clear System may be used in-line with a ventilator in acute care environment only. | Similar Language updated for clarity and added details of nebulizer use with saline previously cleared. |
| Environment of Use | Hospital, Sub-acute facilities, Nursing care Homecare | Hospital, Sub-acute facilities, Nursing care Homecare | Identical |
| Patient Population | Adult, Child > 2 years old (Acute care) Adult, Child > 5 years (Home care) | Adult, Child > 2 years old (Acute care) Adult, Child > 5 years (Home care) | Identical |
| Fundamental technology | Electro-Mechanical device Air or Oxygen | Electro-Mechanical device Air or Oxygen | Identical |
| Therapy Modes | Positive Expiratory Pressure (PEP), Oscillation (OSC), Nebulizer (NEB) | Positive Expiratory Pressure (PEP), Oscillation (OSC), Nebulizer (NEB) | Identical |
{8}
K260604
# 510(k) Summary
Page 5 of 6
| | Subject Device BiWaze Clear | Predicate BiWaze Clear – K231728 | Comparison |
| --- | --- | --- | --- |
| Positive Expiratory Pressure (PEP) / CPEP | Controlled static flow with positive pressure ≤ 30 cmH2O | Controlled static flow with positive pressure ≤ 30 cmH2O | Identical |
| Oscillations (OSC) / CHFO | Controlled continuous flow with frequencies up to 300 beats per minute (5 Hz) and peak positive pressures ≤ 70 cmH2O | Controlled continuous flow with frequencies up to 300 beats per minute (5 Hz) and peak positive pressures ≤ 70 cmH2O | Identical |
| NEB / Aerosol | Controlled continuous constant pressure with in- line nebulizer delivering saline | Controlled continuous constant pressure with in- line nebulizer delivering saline | Identical |
| Aerosol delivery | Mouthpiece, Face Mask Trach Adapter, Two Step Adaptor | Mouthpiece, Face Mask Trach Adapter, Two Step Adaptor | Identical |
| Nebulizer | Aerogen® Solo | Aerogen® Solo | Identical |
| Patient Circuit | Disposable circuit including handset with connection for in-line nebulizer. | Disposable circuit including handset with connection for in-line nebulizer. | Identical |
| Patient Interface | Acute care: Mouthpiece, Facemask, Insert into ventilator, Adapter to a patient's endotracheal tube or tracheostomy tube. Home care: Mouthpiece, Facemask, Adapter to a patient's tracheostomy tube. | Acute care: Mouthpiece, Facemask, Insert into ventilator, Adapter to a patient's endotracheal tube or tracheostomy tube. Home care: Mouthpiece, Facemask, Adapter to a patient's tracheostomy tube. | Identical |
| SpO2 Sensor | Supports connection and displays SpO2 values and heart rate from a SpO2 sensor | Supports connection and displays SpO2 values and heart rate from a SpO2 sensor | Identical |
| Settings | On/Off Touch Screen Frequency selection Pressure adjustment for OSC Pressure adjustment for PEP mode Pressure manometer | On/Off Touch Screen Frequency selection Pressure adjustment for OSC Pressure adjustment for PEP mode Pressure manometer | Identical |
| Wireless Connectivity | Bluetooth and Wi-Fi | Wi-Fi | Similar Testing demonstrated subject device was as safe and effective with enabled Bluetooth. |
| Biocompatibility | ISO 10993-1 ISO 18562-1 | ISO 10993-1 ISO 18562-1 | Identical |
{9}
K260604
510(k) Summary
Page 6 of 6---
# **Substantial Equivalence Conclusion**
The BiWaze Clear device has been determined to be substantially equivalent to BiWaze Clear, K231728, based upon the proposed modifications and that the subject device’s performance is unchanged. The intended use, indication for use, technological characteristics and principles of operation are similar to the predicate device. It can be concluded that the proposed device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.