The device is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound tissue brightness parameters such as ultrasound beam attenuation and backscattering coefficient, and estimates of speed of sound, in internal structures of the body. The device is intended to be used by trained healthcare professionals, in a healthcare environment. The device is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen. In particular, the device is intended to provide: - Linear distance measurements of anatomical structures, - Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen, - Estimates of ultrasound tissue brightness parameters in the liver at selected ultrasound frequencies, - Measurement of brightness ratio between structures and in particular between the liver and the kidney, - Estimates of speed of sound in the liver. The shear wave speed measurements, ultrasound tissue brightness parameters, elastic properties estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of adult and pediatric patients with liver disease.
Device Story
Ultraportable ultrasound imaging system; consists of external transabdominal probe with mechanical vibrator and off-the-shelf host (Windows/macOS) running Hepatoscope App. Probe transmits/receives/digitizes ultrasound signals; transfers raw data via USB-C to host. Host software controls sequences, processes data, and displays images/quantifications. Used by trained healthcare professionals in clinical environments. Provides real-time 2D imaging guidance for elastography, tissue brightness, and speed of sound measurements. Output includes images, linear distance measurements, and quantitative parameters for liver/spleen assessment. Aids diagnosis, monitoring, and clinical management of liver disease.
Clinical Evidence
Bench testing only. Electrical safety testing per IEC 60601-2-37; bench testing on calibrated phantoms to validate signal processing/ultrasound sequence refinements for spleen stiffness; software verification and validation per FDA guidance.
Technological Characteristics
Ultraportable ultrasound system; external transabdominal probe with mechanical vibrator; USB-C connectivity; host-based software processing. Supports Windows and macOS. Connectivity via HL7 FHIR. Complies with IEC 60601-2-37. Software documentation level: Basic.
Indications for Use
Indicated for adult and pediatric patients with liver disease; used for abdominal imaging and physical property measurements (shear wave speed, tissue stiffness, brightness parameters, speed of sound) in the liver and spleen.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 28th, 2026
E-Scopics
Aurelie Gruener
QARA Director
235 Rue Léon Foucault
Aix-En-Provence, 13290
FRANCE
Re: K260589
Trade/Device Name: ES-Series
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX
Dated: February 20, 2026
Received: February 20, 2026
Dear Aurelie Gruener:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260589 - Aurelie Gruener
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260589 - Aurelie Gruener
Page 3
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.04.29 15:53:48 -04'00" For
Yanna Kang, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260589 | ? |
| Please provide the device trade name(s). | | ? |
| ES-Series | | |
| Please provide your Indications for Use below. | | ? |
| The device is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of
the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound tissue
brightness parameters such as ultrasound beam attenuation and backscattering coefficient, and estimates
of speed of sound, in internal structures of the body. The device is intended to be used by trained
healthcare professionals, in a healthcare environment.
The device is indicated for imaging of anatomical structures in the abdomen and measurements of physical
properties in the liver and the spleen.
In particular, the device is intended to provide:
- Linear distance measurements of anatomical structures,
- Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness
in the liver and the spleen,
- Estimates of ultrasound tissue brightness parameters in the liver at selected ultrasound frequencies,
- Measurement of brightness ratio between structures and in particular between the liver and the kidney,
- Estimates of speed of sound in the liver.
The shear wave speed measurements, ultrasound tissue brightness parameters, elastic properties
estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of
adult and pediatric patients with liver disease. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K260589 | 510(k) Summary | Prepared on: 2026-03-31 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | E-Scopics | |
| Applicant Address | 235 Rue Léon Foucault Aix-en-Provence 13290 France | |
| Applicant Contact Telephone | 0033664366992 | |
| Applicant Contact | Mrs. AURELIE GRUENER | |
| Applicant Contact Email | aurelie.gruener@e-scopics.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | ES-Series | |
| Common Name | Ultrasonic pulsed doppler imaging system | |
| Classification Name | System, Imaging, Pulsed Doppler, Ultrasonic | |
| Regulation Number | 892.1550 | |
| Product Code(s) | IYN, IYO, ITX | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K232336 | Proprietary name: Hepatoscope; filed under ES Series V2 in K232336 | IYN |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| ES Series is an ultraportable ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging and quantitative imaging studies. ES Series consists of a Software App (running on a consumer off-the-shelf Selected Host) and an accessory probe. The system produces images and quantifications, which are displayed on the monitor of the Selected Host. ES Series is operated from the Selected Host and allows the user to perform measurements, to capture images, and to generate printable reports. The product is an ultraportable software-based ultrasound imaging device that is composed of an external transabdominal ultrasound probe equipped with a mechanical vibrator to produce shear waves in tissue, and connected to an off-the-shelf software host via a standard USB-C cable. The system software is operated by the host and is responsible for controlling the probe, for processing ultrasound data, and for exam management. Operating functionalities performed by the digital probe are limited to the transmission, the reception and the digitization of ultrasound signals, before transferring them to the processing host via the USB-C cable. This processing host runs the Hepatoscope App, which ensures the following functions: - Control of the ultrasound transmit/receive sequences by the probe; - Processing of the raw ultrasound data that are received from the probe; - Display of the reconstructed imaging information (images, quantitative parameters...) This ultrasound imaging device allows the operator to acquire information coming from quantitative imaging modalities such as elastography, tissue brightness parameters, speed of sound. This can be done under real-time 2D imaging guidance. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| The device is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and | | |
{5}
provides measurements of shear wave speed and tissue stiffness, ultrasound tissue brightness parameters such as ultrasound beam attenuation and backscattering coefficient, and estimates of speed of sound, in internal structures of the body. The device is intended to be used by trained healthcare professionals, in a healthcare environment.
The device is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.
In particular, the device is intended to provide:
- Linear distance measurements of anatomical structures,
- Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen,
- Estimates of ultrasound tissue brightness parameters in the liver at selected ultrasound frequencies,
- Measurement of brightness ratio between structures and in particular between the liver and the kidney,
- Estimates of speed of sound in the liver.
The shear wave speed measurements, ultrasound tissue brightness parameters, elastic properties estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of adult and pediatric patients with liver disease.
## Indications for Use Comparison
**21 CFR 807.92(a)(5)**
The Intended Use and Indications for Use remain identical to the predicate.
## Technological Comparison
**21 CFR 807.92(a)(6)**
ES Series has the same technological characteristics as the Predicate Device: same operating modalities, principles, features (B-mode, 2D-Transient Elastography, Shear Wave Speed estimation, attenuation coefficient, backscattering coefficient, and Speed of Sound), and associated combinations of modes.
- Spleen stiffness workflow: The signal processing and ultrasound sequence have been refined to target higher stiffness values as in the spleen. Performance data demonstrate that these updates are substantially equivalent to the Predicate Device and do not raise new questions of safety or effectiveness.
- Support of macOS platforms: The addition of the macOS operating system as a compatible host, alongside the existing Windows operating system, has been validated by performance data. This expansion is considered substantially equivalent to the Predicate Device and does not raise different questions of safety and effectiveness.
- Addition of connectivity: The implementation of connectivity to EHR using HL7 FHIR standard does not raise different questions of safety and effectiveness to the Predicate Device as demonstrated by performance data.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
**21 CFR 807.92(b)**
A series of non-clinical tests were performed to ensure that improvement of technological characteristics of ES Series do not affect the safety or effectiveness of the device. The following performance data were collected and analyzed:
- Electrical safety testing were conducted on the ES Series device. The device complies with the IEC 60601-2-37.
- Bench testing were conducted on calibrated phantoms to demonstrate the accuracy of the signal processing and ultrasound sequence refinements for spleen stiffness.
- Software Verification and Validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions.” The software for this device was considered as Basic Documentatio Level.
Clinical testing was not deemed necessary to demonstrate substantial equivalence.
## Conclusion:
ES Series demonstrated to be substantially equivalent to its Predicate Device as they have the same intended use and indications for use, and the technological differences do not raise issues of safety and effectiveness.
This has been demonstrated via safety & performance testing between the ES Series and its Predicate Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.