The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment. - The cemented primary humeral stem, long/revision stem, fracture stems, and all Equinoxe glenoids are intended for cemented fixation. - The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion of the surgeon. - The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon. - Humeral Heads are intended for use in cemented and press-fit applications. Clinical indications for the PRIMARY (P), LONG/REVISION (L), and FRACTURE (F) humeral components are as follows: [Table of indications omitted for brevity as per instructions to concatenate] The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head. The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
Device Story
Equinoxe Shoulder System; modular shoulder prosthesis for total or hemi-arthroplasty. Subject device introduces Activit-E (highly crosslinked UHMWPE with vitamin E) humeral liners for reverse total shoulder arthroplasty (rTSA). Device components include humeral stems (primary, long/revision, fracture), glenoids, and humeral heads; implanted by orthopedic surgeons in clinical settings. System restores joint mobility; addresses degenerative disease, fractures, and rotator cuff deficiency. Output is mechanical joint replacement; clinical decision-making based on surgeon assessment of joint pathology and bone quality. Benefits include restored function and pain relief.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included humeral liner-tray fatigue and disassembly, wear testing, particle analysis per ASTM F1877-24, biocompatibility per ISO 10993-1:2018, and bacterial endotoxin testing per USP <161>, USP <85>, and ANSI/AAMI ST72.
Technological Characteristics
Modular shoulder prosthesis. Subject components: Activit-E (highly crosslinked UHMWPE with vitamin E) humeral liners. Other components: metal/polymer. Sterile, single-use. Conforms to recognized performance standards for shoulder replacement. No geometric or compatibility changes from predicate.
Indications for Use
Indicated for skeletally mature patients with glenohumeral degenerative disease or fractures requiring total or hemi-arthroplasty. Specific indications include rheumatoid arthritis, osteoarthritis, osteonecrosis, post-traumatic degeneration, congenital abnormalities, humeral head necrosis, displaced tuberosity fractures, 3/4-part proximal humerus fractures, mid-humerus fractures, and revision of failed reconstructions. Reverse configuration indicated for degenerative disease with irreparable rotator cuff or failed arthroplasty with superior humeral head migration.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter Trays (K223833)
Exactech Equinoxe Reverse Shoulder System 36mm Glenosphere and Humeral Liner (K093275)
Exactech Equinoxe Small Reverse Shoulder System (K180632)
Exactech Equinoxe Small Reverse Shoulder Glenospheres (K182536)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 14, 2026
Exactech, Inc.
% Pedro Ravelo
Senior Regulatory Specialist
Advita Ortho
2320 NW 66th Court
Gainesville, Florida 32653
Re: K260583
Trade/Device Name: Equinoxe® Shoulder System
Regulation Number: 21 CFR 888.3660
Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: PHX, PAO, KWS, KWT
Dated: February 20, 2026
Received: February 20, 2026
Dear Pedro Ravelo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260583 - Pedro Ravelo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260583 - Pedro Ravelo
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FARZANA SHARMIN -S
Farzana Sharmin, Ph.D.
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement on last page.
510(k) Number (if known)
K260583
Device Name
Equinoxe® Shoulder System
Indications for Use (Describe)
The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment.
- The cemented primary humeral stem, long/revision stem, fracture stems, and all Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion of the surgeon.
- The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon.
- Humeral Heads are intended for use in cemented and press-fit applications.
Clinical indications for the PRIMARY (P), LONG/REVISION (L), and FRACTURE (F) humeral components are as follows:
| P | L | F | Indications |
| --- | --- | --- | --- |
| ✓ | ✓ | | Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems |
| ✓ | ✓ | | Congenital abnormalities in the skeletally mature |
| ✓ | | | Primary and secondary necrosis of the humeral head. |
| ✓ | | ✓ | Humeral head fracture with displacement of the tuberosities |
| ✓ | ✓ | | Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable |
| ✓ | ✓ | | Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved) |
| | | ✓ | Displaced three-part and four-part upper humeral fractures |
| | ✓ | | Spiral and other fractures of the mid-humerus (in combination with glenohumeral degenerative diseases) |
| | ✓ | | Revision of failed previous reconstructions when distal anchorage is required |
| ✓ | ✓ | | To restore mobility from previous procedures (e.g. previous fusion) |
The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (6/20)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{5}
K260583
Page 1 of 4
Equinoxe® Shoulder System
510(k) Summary
Applicant:
Exactech, Inc.
2320 NW 66th Court
Gainesville FL, 32653
Phone: (352) 377-1140
Fax: (352) 378-2617
Applicant Contact:
Pedro Ravelo
Senior Regulatory Specialist
Telephone: (352) 377-1140
Fax: (352) 378-2617
Date:
April 10, 2026
Device Trade Name:
Equinoxe® Shoulder System
Common Name:
Shoulder Prosthesis, Reverse Configuration
Classification Name:
Shoulder joint metal/polymer semi-constrained
cemented prosthesis
Shoulder joint metal/polymer non-constrained
cemented prosthesis
Regulation Number:
888.3660, 888.3650
Product Code:
PHX, KWS, KWT, PAO
Legally Marketed Predicate Devices:
| Predicate Number | Predicate Trade Name | Product Code |
| --- | --- | --- |
| K223833 | Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter Trays | PHX, KWS, KWT |
Reference Devices:
| Reference Number | Reference Trade Name | Product Code |
| --- | --- | --- |
| K063569 | Equinoxe Reverse Shoulder System | PHX, KWT |
| K093275 | Exactech Equinoxe Reverse Shoulder System 36mm Glenosphere and Humeral Liner | PHX, KWS, KWT |
| K180632 | Exactech Equinoxe Small Reverse Shoulder System | PHX, KWS, KWT |
| K182536 | Exactech Equinoxe Small Reverse Shoulder Glenospheres , Exactech Equinoxe Reverse | PHX, KWS, KWT |
{6}
K260583
Page 2 of 4
| Shoulder Humeral Liners, Exactech Equinoxe | Small Reverse Shoulder Glenoid Plates, Exactech Equinoxe Reverse Shoulder Locking Cap |
| --- | --- |
## Device Description Summary
The proposed Equinoxe Activit-E Humeral Liners are humeral liners intended to be used in reverse total shoulder arthroplasty (rTSA) as a part of the Equinoxe® Shoulder System. The proposed devices are constructed of highly crosslinked UHMWPE containing vitamin E referred to as Activit-E. There are no proposed geometric or compatibility changes as compared to the predicate humeral liner devices cleared in K223833.
## Intended Use/Indications for Use
The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi- arthroplasty is determined by the surgeon to be the preferred method of treatment.
- The cemented primary humeral stem, long/revision stem, fracture stems, and all Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion of the surgeon.
- The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon.
- Humeral Heads are intended for use in cemented and press-fit applications.
Clinical indications for the PRIMARY (P), LONG/REVISION (L), and FRACTURE (F) humeral components are as follows:
| P | L | F | Indications |
| --- | --- | --- | --- |
| ✓ | ✓ | | Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems |
| ✓ | ✓ | | Congenital abnormalities in the skeletally mature |
| ✓ | | | Primary and secondary necrosis of the humeral head. |
| ✓ | | ✓ | Humeral head fracture with displacement of the tuberosities |
| ✓ | ✓ | | Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable |
| ✓ | ✓ | | Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved) |
| | | ✓ | Displaced three-part and four-part upper humeral fractures |
{7}
K260583
Page 3 of 4
| | ✓ | | Spiral and other fractures of the mid-humerus (in combination with glenohumeral degenerative diseases) |
| --- | --- | --- | --- |
| | ✓ | | Revision of failed previous reconstructions when distal anchorage is required |
| ✓ | ✓ | | To restore mobility from previous procedures (e.g. previous fusion) |
The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemiarthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
## Indications for Use Comparison
The proposed and the predicate devices have the same indications for use.
## Technological Comparison
The proposed and predicate devices have the same intended use and basic fundamental scientific technology. There are no proposed geometric or compatibility changes as compared to the predicate humeral liner devices. The only technological difference between the proposed and predicate devices is the material. The proposed devices are constructed of highly crosslinked UHMWPE containing vitamin E referred to as Activit-E, whereas the predicate devices are constructed of conventional UHMWPE. The rationale for substantial equivalence of the proposed to the predicate cleared devices is based on consideration of the following aspects of the devices:
- The proposed and predicate devices are composed of similar biocompatible materials.
- The proposed and predicate devices have the same design features.
- The proposed and predicate devices are provided sterile for single use only.
- The proposed and predicate devices conform to recognized performance standards for shoulder replacement devices.
{8}
K260583
Page 4 of 4
# Non-Clinical and/or Clinical Tests Summary & Conclusions
The following non-clinical testing and engineering analyses were performed to demonstrate that the Equinoxe® Shoulder System performs as intended and is substantially equivalent to the identified predicate devices:
- Humeral Liner-Tray fatigue testing
- Humeral Liner-Tray disassembly testing
- Humeral Liner Mode I wear testing
- Post-wear testing particle analysis per ASTM F1877-24
- Biocompatibility per ISO 10993-1:2018
- Bacterial endotoxins per USP <161>, USP <85> and ANSI/AAMI ST72
The differences in material do not change the intended use, safety, or performance requirements of the proposed devices, nor do they adversely affect their safety or effectiveness. This conclusion is based on consideration of the preclinical testing and analysis including humeral liner-tray fatigue and disassembly testing, humeral liner wear testing and particle analysis, and biocompatibility assessment completed to establish substantial equivalence of the proposed devices to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.