K260511 · Shenzhen Ulike Smart Electronics Co., Ltd. · OHS · Mar 16, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260511
Device Name
Ulike Reglow Light Therapy Device (UM10)
Applicant
Shenzhen Ulike Smart Electronics Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Mar 16, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ulike Reglow Light Therapy Device is an over the counter device that is intended for use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne.
Device Story
Light therapy mask for home use; operates via external controller; utilizes 272 LEDs to emit multi-wavelength light (465nm, 590nm, 630nm, 830nm) to facial skin; four selectable modes (Glow, Anti-Aging, Restoring, Clear) combine specific wavelengths to target wrinkles and acne; includes integrated protective eye-shield to prevent ocular exposure; powered by 3.7V lithium battery; user-operated; intended to improve skin appearance through photobiological effects.
Clinical Evidence
Bench testing only. Evidence includes photobiological safety (IEC 62471), electrical safety (IEC 60601-1, 60601-2-57, 60601-2-83), battery safety (IEC 62133-2), and wavelength/irradiance performance testing. Biocompatibility verified per ISO 10993-5, 10993-10, and 10993-23.
Technological Characteristics
Mask-form factor with 272 LEDs; multi-wavelength output (465, 590, 630, 830 nm); intensity 1-40mW/cm2; powered by 3.7V 2600mAh lithium battery; controller-based mode selection; biocompatible materials per ISO 10993; electrical safety per IEC 60601 series; photobiological safety per IEC 62471.
Indications for Use
Indicated for adults (22 years and older) for the treatment of full face wrinkles and mild to moderate inflammatory acne.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 16, 2026
Shenzhen Ulike Smart Electronics Co., Ltd.
Blue Yang
Registration Director
810, Bldg. 1, Xunmei Science And Technology Plaza, # 8 Keyuan Rd., Science Park Community,
Yuehai Sub-District, Nanshan District
Shenzhen, Guangdong 518000
China
Re: K260511
Trade/Device Name: Ulike Reglow Light Therapy Device (UM10)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP
Dated: February 13, 2026
Received: February 17, 2026
Dear Blue Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260511 - Blue Yang
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K260511 - Blue Yang
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.03.16
18:40:24 -04'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260511 | ? |
| Please provide the device trade name(s). | | ? |
| Ulike Reglow Light Therapy Device (UM10) | | |
| Please provide your Indications for Use below. | | ? |
| Ulike Reglow Light Therapy Device is an over the counter device that is intended for use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☐ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater) | ? |
{4}
Shenzhen Ulike Smart Electronics Co., Ltd.
# 510(k) Summary of K260511
## I. Submitter
Shenzhen Ulike Smart Electronics Co., Ltd.
Address: 810, Building 1, Xunmei Science and Technology Plaza, No. 8 Keyuan Road, Science Park Community, Yuehai Sub-District, Nanshan District, Shenzhen 518000, Guangdong, P.R. China
Contact person: Blue Yang
Email: blue@ulike.com
The date the summary was prepared: 03/06/2026
## II. Device
Name of Device: Ulike Reglow Light Therapy Device
Model(s): UM10
Common or Usual Name: Light based over the counter wrinkle reduction, Over-The-Counter Powered Light Based Laser For Acne
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHS, OLP
Regulation Number: 21 CFR 878.4810
## III. Device Description
Ulike Reglow Light Therapy Device adopts 272 light emitting diodes (LED) with the multi-wavelength of $630\mathrm{nm} \pm 10\mathrm{nm}, 830\mathrm{nm} \pm 10\mathrm{nm}, 590\mathrm{nm} \pm 10\mathrm{nm}$ and $465\mathrm{nm} \pm 10\mathrm{nm}$ spectrum to irradiate the facial area to realize its therapeutic effect, including reducing wrinkles and treating mild to moderate inflammatory acne, which improves overall skin appearance.
The Ulike Reglow Light Therapy Device adopts the form of a mask that contains LEDs on the inner surface of the main unit. The device is with user-friendly operation through a controller which is connected to the main unit to control the device, such as turn on/off the device, switch mode.
The device offers four selectable operating modes via the external controller:
- Mode 1 (Glow mode): $630\mathrm{nm} + 830\mathrm{nm} + 590\mathrm{nm}$
- Mode 2 (Anti-Aging mode): $630\mathrm{nm} + 830\mathrm{nm}$
- Mode 3 (Restoring mode): $590\mathrm{nm} + 830\mathrm{nm}$
- Mode 4 (Clear mode): $465\mathrm{nm} + 630\mathrm{nm} + 830\mathrm{nm}$
To use the device, user should place the mask over the face and use the controller to operate.
To prevent irradiation of LED lights to eyes during the treatment, Ulike Reglow Light Therapy Device has incorporated protective eye-shield which blocks light energy from LEDs.
{5}
Shenzhen Ulike Smart Electronics Co., Ltd.
# IV. Indications for Use
Ulike Reglow Light Therapy Device is an Over the Counter device that is intended for the use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne.
# V. Technological comparison with the Predicate Device
The Ulike Reglow Light Therapy Device has the same intended use and operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the Ulike Reglow Light Therapy Device may be found substantially equivalent to its predicate device.
Ulike Reglow Light Therapy Device is compared with the following Predicate Device in terms of intended use, design, specifications and performance
| Comparison Items | Subject Device | Predicate Device 1 | Remar k |
| --- | --- | --- | --- |
| 510(k) number | / | K243492 | / |
| Trade Name | Ulike Reglow Light Therapy Device | Ulike Reglow Light Therapy Device | / |
| Manufacturer | Shenzhen Ulike Smart Electronics Co., Ltd. | Shenzhen Ulike Smart Electronics Co., Ltd. | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Product code | OHS,OLP | OHS,OLP | Same |
| Device classification | Class II | Class II | Same |
| Indications for use/ Intended use | Ulike Reglow Light Therapy Device is an Over the Counter device that is intended for the use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne. | Ulike Reglow Light Therapy Device is an Over the Counter device that is intended for the use in the treatment of full face wrinkles and treatment of mild to moderate inflammatory acne. | Same |
| Prescription or OTC | OTC | OTC | Same |
| Location for use | Face | Face | Same |
| Power supply | Controller:3.7V, 2600mAh lithium battery, 9.62Wh | Controller:3.7V, 2600mAh lithium battery, 9.62Wh | Same |
| Light source | LED | LED | Same |
| Wavelength | Mode 1: 630nm+830nm+590nmMode 2: 630nm+830nmMode 3: 590nm+ 830nmMode 4: 465nm+ | Mode 1: 630nm+830nm+590nmMode 2: 630nm+830nmMode 3: 590nm+ 830nmMode 4: 465nm+ | Same |
{6}
Shenzhen Ulike Smart Electronics Co., Ltd.
| | 630nm+830nm | 630nm+830nm | |
| --- | --- | --- | --- |
| LED Intensity | 1-40mW/cm2 | 1-40mW/cm2 | Same |
| Treatment Time | Mode 1:5 minutes each timeMode 2:8 minutes each timeMode 3:8 minutes each timeMode 4:5 minutes each time | Mode 1:5 minutes each timeMode 2:8 minutes each timeMode 3:8 minutes each timeMode 4:5 minutes each time | Same |
| Electrical safety | IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-57IEC 60601-2-83 | IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-57IEC 60601-2-83 | Same |
| Eye safety | IEC 62471 | IEC 62471 | Same |
| Biocompatibility | ISO 10993-5ISO 10993-10ISO 10993-23 | ISO 10993-5ISO 10993-10ISO 10993-23 | Same |
| Modifications | 1.Add component suppliersAdd a battery supplier with the same specificationsAdd LEDs supplier with the same specifications2. Add a kind of package contents: not providing adaper, only providing Type C chargingOriginal version:Reglow Light TherapyDevice*1+Rechargeable Remote*1+UserManual*1+Adapter*1Added version:Reglow Light TherapyDevice*1+Rechargeable Remote*1+UserManual*1+Type-C charging cable*1 | / | Substainially equivalent |
{7}
Shenzhen Ulike Smart Electronics Co., Ltd.
# Summary of performance testing
1. IEC 62133-2:2017+2021, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portablesealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
2. IEC 62471:2006 Photobiological safety of lamps and lamp systems
3. Wavelength and Irradiance Performance Test Report
4. IEC 60601-1:2005+A1:2012+A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, relevant clauses
5. IEC 60601-2-57:2011 Medical electrical equipment–Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use, relevant clauses
6. IEC 60601-2-83:2019 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment, relevant clauses
Conclusion: Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device Ulike Reglow Light Therapy Device is as safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.