The H100Plus Hemoglobin Testing System is intended for in vitro quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in venous whole blood and hemolysate from venous whole blood using ion-exchange high performance liquid chromatography (HPLC) on the H100Plus Hemoglobin Analyzer (HPLC). Hemoglobin A1c measurements are used as an aid in the diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk of developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. The H100Plus Hemoglobin Testing System is intended for Professional Use Only.
Device Story
H100Plus Hemoglobin Testing System is an automated, professional-use, in vitro diagnostic system for HbA1c measurement. It utilizes ion-exchange high performance liquid chromatography (HPLC) to separate hemoglobin species in venous whole blood or hemolysate samples. The system includes the H100Plus Hemoglobin Analyzer, chromatographic columns, specific eluents, calibrators, controls, and hemolytic agent. The analyzer automatically dilutes and lyses whole blood samples; pre-diluted samples are injected directly. Hemoglobin fractions are separated via ionic interactions within the column and detected at 415 nm. The system calculates HbA1c levels based on peak area ratios. Healthcare providers use these quantitative results to diagnose diabetes, identify at-risk patients, and monitor long-term glycemic control. The device provides standardized results traceable to DCCT/IFCC reference methods.
Clinical Evidence
Bench testing only. Performance evaluated via precision/reproducibility studies (n=720 measurements per sample), linearity (11 levels), and interference testing (endogenous/exogenous substances and hemoglobin variants HbS, HbC, HbD, HbE, HbA2, HbF). Method comparison study against the predicate device using 129 variant-free K2-EDTA venous whole blood samples showed high correlation (Deming slope 1.0060, Passing-Bablok slope 0.9974 for whole blood mode). Total error was calculated and found acceptable.
Technological Characteristics
Ion-exchange HPLC system. Components: H100Plus Hemoglobin Analyzer, hydrophilic polymer-filled HPLC columns, succinate buffer eluents (with <0.1% sodium azide), lyophilized hemoglobin calibrators/controls, and hemolytic agent. Detection at 415 nm. Connectivity via barcode/manual entry. Electrical safety/EMC compliant with IEC 61010-1, IEC 61326-1, and IEC 61326-2-6. Software/cybersecurity compliant with IEC 81001-5-1.
Indications for Use
Indicated for professional use in clinical settings for the quantitative determination of HbA1c in venous whole blood or hemolysate to aid in the diagnosis of diabetes mellitus, identification of individuals at risk for diabetes, and monitoring long-term glycemic control. Contraindicated for use during pregnancy, for gestational diabetes diagnosis, for newborn testing, or for patients with heterozygous sickle cell trait. Results must be interpreted with caution in patients with conditions altering red blood cell lifespan (e.g., anemia, hemoglobinopathies, chronic blood loss, hepatic/renal disease).
Regulatory Classification
Identification
A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.
Special Controls
(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA.
(2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following:
(i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable.
(ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method.
(iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent.
(iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
## DECISION SUMMARY
ASSAY AND INSTRUMENT
### I Background Information:
A 510(k) Number
K260468
B Applicant
Shenzhen Lifotronic Technology Co., Ltd.
C Proprietary and Established Names
H100Plus Hemoglobin Testing System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PDJ | Class II | 21 CFR 862.1373 - Hemoglobin A1c Test System | CH - Clinical Chemistry |
| LCP | Class II | 21 CFR 864.7470 - Glycosylated hemoglobin assay | HE - Hematology |
### II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Glycosylated Hemoglobin (HbA1c)
C Type of Test:
Quantitative Ion-exchange High Performance Liquid Chromatography (HPLC)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
### III Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use below.
#### B Indication(s) for Use:
The H100Plus Hemoglobin Testing System is intended for in vitro quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in venous whole blood and hemolysate from venous whole blood using ion-exchange high performance liquid chromatography (HPLC) on the H100Plus Hemoglobin Analyzer (HPLC).
Hemoglobin A1c measurements are used as an aid in the diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk of developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus.
The H100Plus Hemoglobin Testing System is intended for Professional Use Only.
#### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
- For in vitro diagnostic use only.
- HbA1c results must not be used during pregnancy or to diagnose gestational diabetes.
- The product is not intended for testing newborn specimens.
- In rapidly developing type 1 diabetes, an increase in HbA1c may be delayed relative to the acute increase in blood glucose. In such cases, diabetes must be diagnosed based on plasma glucose results and clinical findings rather than HbA1c alone.
- Conditions that alter red blood cell lifespan or turnover may cause HbA1c results that do not accurately reflect average blood glucose. These conditions may include recent significant or chronic blood loss, iron deficiency anemia, hemolytic anemia, hemoglobinopathies, thalassemia, hereditary spherocytosis, malignancy, and severe chronic hepatic or renal disease. HbA1c results from patients with these conditions must not be used alone for the diagnosis of diabetes, identification of individuals at risk of developing diabetes, or assessment of long-term glycemic control. Interpret the results with caution and consider glucose-based testing or other appropriate markers of glycemic control.
- The product is not intended for HbA1c testing in patients with heterozygous sickle cell trait.
#### D Special Instrument Requirements:
H100Plus Hemoglobin Analyzer(HPLC)
K260468 - Page 2 of 14
{2}
# IV Device/System Characteristics:
# A Device Description:
The H100Plus Hemoglobin Testing System consists of the following components:
1. H100Plus Hemoglobin Analyzer(HPLC): automated instrument for running the H100Plus HbA1c HPLC assay to measure the content of HbA1c in human whole blood.
2. H100Plus HbA1c Eluents A, B, and C: contain succinate buffer with <0.1% sodium azide as a preservative, used for elution of different substances in HPLC. Eluents A is distributed in 800 mL and 1600 mL packages; Eluent B and C are distributed in 800 mL package only.
3. H100Plus HbA1c Calibrator (2 levels): lyophilized powders containing hemoglobin should be reconstituted with labeled volume of the H100Plus Hemolytic Agent L.
4. H100Plus HbA1c Control Material (2 levels): lyophilized powders containing hemoglobin should be reconstituted with labeled volume of the H100Plus Hemolytic Agent L.
5. H100Plus Hemoglobin Chromatographic Column (HPLC): assembled HPLC columns filled with hydrophilic polymer for use with the H100Plus Hemoglobin Analyzer (HPLC), with 5,000 or 10,000 test capacities.
6. H100Plus Hemolytic Agent L: intended for pre-treatment of human whole blood to lyse red blood cells for releasing glycated and other hemoglobin molecules, can also be used to reconstitute H100Plus HbA1c Calibrator and H100Plus HbA1c Control Material, distributed in 250 mL, 2,500 mL, and 5,000 mL packages.
# B Principle of Operation:
The device is based on the principle of ion-exchange high performance liquid chromatography (HPLC) to separate hemoglobin species in a HPLC column. The device uses venous whole blood samples in blood collection tubes (K2 EDTA or K3 EDTA) or manually pre-diluted and hemolyzed venous blood samples in Eppendorf tubes. Whole blood samples are automatically diluted and lysed by the instrument prior to being injected into the column. Pre-diluted (hemolysate) samples are directly injected into the column without the dilution step in the instrument.
A programmed buffer with stepwise eluents containing different salt concentrations delivers the sample into the column. Hemoglobin species are separated through ionic interactions with the column material and are detected at the detection window at wavelengths of 415 nm. The device automatically collects chromatographic data for each hemoglobin fraction and calculates the HbA1c levels based on the peak area ratio.
# C Instrument Description Information:
1. Instrument Name:
H100Plus Hemoglobin Analyzer(HPLC)
2. Specimen Identification:
K260468 - Page 3 of 14
{3}
Sample identification is via barcoding or manual entry.
# 3. Specimen Sampling and Handling:
Venous whole blood samples collected in K2-EDTA or in K3-EDTA can be stored at 10-30°C for up to 24 hours, at 2-8°C for up to 7 days, or at -80°C for up to 6 month.
Manually diluted and hemolyzed samples should be tested immediately.
# 4. Calibration:
Calibration using the H100Plus HbA1c Calibrator (2 levels) should be performed:
- After installation of the analyzer for the first time;
- After performing maintenance or troubleshooting;
- After replacing column, filter, or different lots of Eluent;
- Analyzer is reused after it is shelved for more than 7 days;
- At least once every 30 days during normal use;
- At any time when QC is out of range or obvious inaccuracy of analysis results are detected.
# 5. Quality Control:
To ensure the reliability of test results, perform quality control testing, using two levels of controls, after powering on the analyzer and prior to daily testing. Additionally, QC should be repeated after any re-calibration. The quality control results shall fall within the established quality control ranges. If results are out of range, instrument status, reagent or calibrator shall be checked for reasons and perform re-calibration if necessary.
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
D-100 HbA1c, D-100 HbA1c Calibrator Pack
# B Predicate 510(k) Number(s):
K151321
# C Comparison with Predicate(s):
D
| Device & Predicate Device(s): | K260468 | K151321 |
| --- | --- | --- |
| Device Trade Name | H100Plus Hemoglobin Testing System | D-100 HbA1c, D-100 HbA1c Calibrator Pack |
K260468 - Page 4 of 14
{4}
| General Device Characteristic Similarities | | |
| --- | --- | --- |
| Intended Use/Indications For Use | Intended for in vitro quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP%) in human whole blood, as an aid in the diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk of developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. For Professional Use Only. | Same |
| Test principle | Ion exchange high performance liquid chromatography (HPLC) | Same |
| Standardization/ Traceability | Traceable to the Diabetes Control and Complications Trial (DCCT) reference method and IFCC. Certified via the National Glycohemoglobin Standardization Program (NGSP) | Same |
| General Device Characteristic Differences | | |
| Matrices | K2-EDTA, K3-EDTA | K2-EDTA, K3-EDTA Potassium Oxalate/ Sodium Fluoride, Sodium Citrate, Sodium Heparin, Lithium Heparin |
| Measuring range | 3.0% to 20.0% (NGSP); 9.3 – 195.1 mmol/mol HbA1c (IFCC) | 3.5 to 20% (NSGP); 15 – 195 mmol/mol HbA1c (IFCC) |
# VI Standards/Guidance Documents Referenced:
- IEC 61010-1 Edition 3.1 2017-01 CONSOLIDATED VERSION, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)]
- IEC 61326-1 Edition 3.0 2020-10, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- IEC 61326-2-6 Edition 4.0 2025-06, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment
- AAMI TIR69:2017/(R2020), Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems
- IEEE ANSI USEMCSC C63.27-2021, American National Standard for Evaluation of Wireless Coexistence
K260468 - Page 5 of 14
{5}
- ISO 17511 Second edition 2020-04, In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples
- CLSI EP06 2nd Edition, Evaluation of the Linearity of Quantitative Measurement Procedures
- CLSI EP05-A3 (Reaffirmed: September 2019), Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
- CLSI EP07 3rd Edition, Interference Testing in Clinical Chemistry
- CLSI EP09c 3rd Edition, Measurement Procedure Comparison and Bias Estimation Using Patient Samples
- CLSI EP25 2nd Edition, Evaluation of Stability of In Vitro Medical Laboratory Test Reagents
- IEC 81001-5-1 Edition 1.0 2021-12, Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle
### VII Performance Characteristics (if/when applicable):
#### A Analytical Performance:
##### 1. Precision/Reproducibility:
Precision studies were conducted with the H100Plus Hemoglobin Testing System using four K2-EDTA venous whole blood samples at concentrations near 5%, 6.5%, 8%, and 12% HbA1c (S1, S2, S3, and S4 respectively) and two quality control samples at concentrations near 6% and 11% HbA1c (QC1 and QC2 respectively). Samples were either analyzed as whole blood samples or as manually prepared hemolysate samples. Precision was evaluated using three H100Plus Hemoglobin Analyzer (HPLC) instruments with three lots of reagents. Samples were analyzed in duplicate per run, two runs per day, for 20 days. For each sample there were 720 measurements. Results in NGSP (%) and IFCC units (mmol/mol) are shown in the tables below for samples diluted and hemolyzed automatically by the analyzer (whole blood) or manually (hemolysate):
All Analyzers Combined, Whole Blood, % HbA1c (NGSP) units
| Sample | Mean Values (%) | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 4.99 | 0.80 | 0.00 | 0.19 | 0.25 | 0.82 | 1.18 |
| S2 | 6.50 | 0.59 | 0.00 | 0.14 | 0.29 | 0.90 | 1.13 |
| S3 | 8.02 | 0.52 | 0.00 | 0.15 | 0.23 | 0.88 | 1.06 |
| S4 | 12.07 | 0.37 | 0.00 | 0.08 | 0.13 | 0.90 | 0.98 |
| QC1 | 5.12 | 0.76 | 0.00 | 0.25 | 0.27 | 0.91 | 1.24 |
| QC2 | 10.57 | 0.39 | 0.00 | 0.08 | 0.14 | 0.96 | 1.05 |
Note: Negative variance component values have been set to zero
K260468 - Page 6 of 14
{6}
All Analyzers Combined, Whole Blood, mmol/mol (IFCC) units
| Sample | Mean Values mmol/mol | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 31.08 | 1.41 | 0.00 | 0.33 | 0.43 | 1.43 | 2.08 |
| S2 | 47.50 | 0.88 | 0.00 | 0.21 | 0.43 | 1.35 | 1.68 |
| S3 | 64.19 | 0.71 | 0.00 | 0.20 | 0.31 | 1.20 | 1.44 |
| S4 | 108.38 | 0.45 | 0.00 | 0.10 | 0.16 | 1.10 | 1.20 |
| QC1 | 32.45 | 1.32 | 0.00 | 0.43 | 0.46 | 1.56 | 2.14 |
| QC2 | 91.98 | 0.49 | 0.00 | 0.10 | 0.18 | 1.20 | 1.32 |
Note: Negative variance component values have been set to zero
All Analyzers Combined, Hemolysate, % HbA1c (NGSP) units
| Sample | Mean Values (%) | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 4.99 | 0.77 | 0.00 | 0.16 | 0.37 | 0.83 | 1.21 |
| S2 | 6.48 | 0.58 | 0.00 | 0.22 | 0.26 | 0.93 | 1.15 |
| S3 | 8.02 | 0.50 | 0.00 | 0.11 | 0.18 | 0.90 | 1.05 |
| S4 | 12.03 | 0.37 | 0.00 | 0.13 | 0.11 | 0.86 | 0.95 |
Note: Negative variance component values have been set to zero
All Analyzers Combined, Hemolysate, mmol/mol (IFCC) units
| Sample | Mean Values mmol/mol | %CV, Percent CV for All Analyzers Combined | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Between-Lot | Between-Analyzer | Total |
| S1 | 31.00 | 1.36 | 0.00 | 0.28 | 0.66 | 1.46 | 2.12 |
| S2 | 47.33 | 0.87 | 0.00 | 0.33 | 0.39 | 1.38 | 1.71 |
| S3 | 64.12 | 0.69 | 0.00 | 0.14 | 0.25 | 1.23 | 1.44 |
| S4 | 107.97 | 0.46 | 0.00 | 0.16 | 0.14 | 1.04 | 1.16 |
Note: Negative variance component values have been set to zero
## 2. Linearity:
Linearity was evaluated for the H100Plus Hemoglobin Testing System with three lots of reagents using a dilution series consisting of 11 levels of HbA1c prepared by mixing high and low K2-EDTA whole blood pools. The expected sample concentrations were 3.04%, 4.72%, 6.40%, 8.08%, 9.76%, 11.44%, 13.12%, 14.80%, 16.48%, 18.16%, and 19.84% HbA1c. The samples were either analyzed as whole blood samples or were manually hemolyzed before analyzing. Five replicates were tested at each concentration. The measured mean values were compared to the expected values. The linear regressions for the three lots of reagents are the following:
### Whole Blood
NGSP unit:
1st lot of reagent: Y=0.9976X - 0.0003, R2=1.0000
K260468 - Page 7 of 14
{7}
2nd lot of reagent: Y=0.9993X - 0.0023, R2=0.9999
3rd lot of reagent: Y=0.9994X - 0.0229, R2=1.0000
# **IFCC unit:**
1st lot of reagent: Y=0.9977X - 0.0854, R2=1.0000
2nd lot of reagent: Y=0.9992X - 0.0235, R2=0.9999
3rd lot of reagent: Y=0.9996X - 0.2595, R2=1.0000
# **Hemolysate**
# **NGSP unit:**
1st lot of reagent: Y=0.9973X - 0.0183, R2=0.9999
2nd lot of reagent: Y=1.0014X - 0.0296, R2=1.0000
3rd lot of reagent: Y=0.9999X - 0.0244, R2=1.0000
# **IFCC unit:**
1st lot of reagent: Y=0.9974X - 0.2502, R2=0.9999
2nd lot of reagent: Y=1.0017X - 0.2954, R2=1.0000
3rd lot of reagent: Y=1.0000X - 0.2639, R2=1.0000
Results of the linearity study support the claimed measuring range of the device of 3.0% to 20.0% HbA1c (9.3-195.1 mmol/mol).
# 3. Analytical Specificity/Interference:
Interference studies were performed to assess the impact of endogenous and exogenous substances on the performance of the H100Plus Hemoglobin Testing System. Pooled K2-EDTA venous whole blood at two HbA1c levels (~6.5 and ~8.0% HbA1c) spiked with potential interferent was analyzed in replicates of two and compared to the same whole blood pool without interferent (control). The highest concentrations at which no significant interference (a difference of no more than ±6% from the control sample as defined by the sponsor) was observed are summarized below:
Endogenous Substances
| Interferent | Highest Concentration Tested with No Significant Interference (mg/dL) |
| --- | --- |
| Unconjugated Bilirubin | 100 |
| Conjugated Bilirubin | 100 |
| Glucose | 2,000 |
| Rheumatoid Factors | 750 IU/mL |
| Lipemia | 6,000 |
| Total Protein | 21,000 |
| Triglycerides | 2,000 |
| Albumin (human) | 20,000 |
Exogenous Substances
| Interferent | Highest Concentration Tested with No Significant Interference (mg/dL) |
| --- | --- |
| Acetylcysteine | 330 |
K260468 - Page 8 of 14
{8}
| Interferent | Highest Concentration Tested with No Significant Interference (mg/dL) |
| --- | --- |
| Ampicillin-Na | 1,000 |
| Ascorbic acid | 300 |
| Cefoxitin | 2,500 |
| Heparin | 5,000 U/L |
| Levodopa | 20 |
| Methyldopa | 30 |
| Metronidazole | 200 |
| Doxycyclin | 50 |
| Rifampicin | 64 |
| Cyclosporine | 5 |
| Acetaminophen | 200 |
| Ibuprofen | 500 |
| Theophylline | 100 |
| Phenylbutazone | 400 |
| Salicylic acid | 60 |
### Cross Reactivity with Hemoglobin Derivatives:
Potential interference from Labile HbA1c, Carbamylated Hb, Acetylated Hb were evaluated. Two HbA1c concentrations of K2-EDTA pooled venous whole blood (~6.5 and ~8.0% HbA1c) were incubated at 37°C for 3 hours to prepare hemoglobin derivatives. The %HbA1c values were compared to the same control sample with no potential interferent present. Significant interference was defined by the sponsor as ≥6% change in HbA1c value of the mean of the test samples relative to the mean of the control samples. The test results and conclusions are as follows:
- Labile A1c - (up to 2000 mg/dL glucose ) does not interfere with this assay.
- Labile A1c - (up to 60 mg/dL Acetaldehyde) does not interfere with this assay.
- Carbamylated Hb - (up to 10 mg/dL Potassium cyanate) does not interfere with this assay.
- Acetylated Hb - (up to 500 mg/dL acetylsalicylic acid) does not interfere with this assay.
Results demonstrated that there was no cross reactivity with these hemoglobin derivatives at tested levels and the labeling states that at physiologically occurring concentrations there is no interference from labile A1c, carbamylated hemoglobin or acetylated hemoglobin.
### Hemoglobin Variant Interference:
Hemoglobin variant studies were performed using K2-EDTA venous whole blood samples known to contain hemoglobin variants HbS, HbC, HbD, HbE, HbA2, and HbF. The samples containing the hemoglobin variants were tested using the Lifotronic Hemoglobin Test System (HPLC) and the results compared to those obtained using comparator methods that have been demonstrated to be free from interference by these hemoglobin variants. The sponsor defined significant interference as ≥6% change in HbA1c values in the presence of the hemoglobin variants relative to the comparator method. The following tables summarize the tested samples and results:
K260468 - Page 9 of 14
{9}
Hemoglobin Variant Samples
| Hemoglobin Variant | Number of Samples | Variant Concentration Range (%) | Range of HbA1c Concentration (%) |
| --- | --- | --- | --- |
| HbS | 25 | 38 – 42 | 5.8 – 10.4 |
| HbC | 25 | 30 – 38 | 5.7 – 10.0 |
| HbD | 25 | 40 – 42 | 5.7 – 11.7 |
| HbE | 27 | 23 – 32 | 5.8 – 11.8 |
| HbA2 | 25 | 3.5 – 6.7 | 5.0 – 9.0 |
| HbF | 25 | 3.0 – 30 | 5.5 – 10.4 |
Summary of Hemoglobin Variant Test Results
| Hemoglobin Variant | %Bias, Percent Relative Bias Observed Relative to the Comparator Method | | | |
| --- | --- | --- | --- | --- |
| | ~ 6.5 % HbA1c | | ~ 8.0% HbA1c | |
| | %Bias | Range of %Bias | %Bias | Range of %Bias |
| HbS | 0.69 | -1.61 to 0.81 | 0.63 | -0.87 to 1.93 |
| HbC | 0.65 | -1.25 to 0.88 | 0.60 | -0.80 to 1.73 |
| HbD | 0.81 | -1.88 to 0.92 | 0.84 | -1.37 to -1.54 |
| HbE | 0.78 | -1.72 to 0.98 | 0.68 | -0.85 to 1.87 |
| HbA2 | 0.78 | -1.89 to 0.93 | 0.69 | -1.52 to 1.08 |
| HbF | 0.55 | -1.61 to 0.76 | 0.81 | -1.46 to 0.98 |
The results demonstrate no significant interference for HbS, HbC, HbE, HbD, HbA2, and HbF at the concentrations stated in the table above.
# 4. Detection Limit and Assay Reportable Range:
The reportable range for this device is 3.0% – 20.0% HbA1c (NGSP) or 9.3 – 195.1 mmol/mol HbA1c (IFCC).
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The H100Plus Hemoglobin Testing System is traceable to the International Federation of Clinical Chemistry (IFCC) reference calibrators. The assay is certified by the National Glycohemoglobin Standardization Program (NGSP) and IFCC. The NGSP and IFCC certification expires in one year. See the NGSP website for current certification at https://www.NGSP.org.
HbA1c results are provided to users in two different units: NGSP equivalent units (%) and IFCC equivalent units (mmol/mol). These two units are converted using the Master Equation: NGSP (%) = 0.09148 × IFCC (mmol/mol) + 2.152.
# 6. Assay Cut-Off:
Not applicable.
K260468 - Page 10 of 14
{10}
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
Method comparison study was performed using 129 variant-free K2-EDTA venous whole blood samples (3.0 to 20.0% HbA1c concentration range) comparing the candidate device results with those from the predicate device (i.e., Bio-Rad D-100 HbA1c Test) performed at a NGSP secondary reference laboratory. The study was conducted using both the whole blood and hemolysate modes of the H100Plus Hemoglobin Testing System. The distribution of samples spanning the measuring interval is as follows:
| HbA1c levels, % | # of samples tested | % of sample tested |
| --- | --- | --- |
| ≤ 5 | 6 | 4.7 |
| 5.0 – 6.0 | 17 | 13.2 |
| 6.0 – 6.5 | 31 | 24.0 |
| 6.5 – 7.0 | 33 | 25.6 |
| 7.0 – 8.0 | 20 | 15.5 |
| 8.0 – 9.0 | 11 | 8.5 |
| > 9.0 | 11 | 8.5 |
| Total | 129 | 100 |
Regression analyses were performed for the candidate device versus the comparator method. Summary of results are as follows:
Whole blood mode, NGSP unit:
| Method | Slope | 95% CI | y-intercept | 95% CI |
| --- | --- | --- | --- | --- |
| Deming | 1.0060 | 1.0002 to 1.0117 | -0.0237 | -0.0721 to 0.0247 |
| Passing-Bablok | 0.9974 | 0.9815 to 1.0068 | 0.0513 | -0.0248 to 0.1515 |
Hemolysate mode, NGSP unit:
| Method | Slope | 95% CI | y-intercept | 95% CI |
| --- | --- | --- | --- | --- |
| Deming | 1.0070 | 0.9983 to 1.0156 | -0.0197 | -0.0839 to 0.0449 |
| Passing-Bablok | 0.9958 | 0.9800 to 1.0071 | 0.0649 | -0.0186 to 0.1740 |
Bias Estimation for the whole blood mode:
| HbA1c levels (%) | Deming | | Passing-Bablok | |
| --- | --- | --- | --- | --- |
| | Bias (%) | %Bias | Bias (%) | %Bias |
| 5.0 | 0.0063 | 0.1260 | 0.0383 | 0.7660 |
| 6.5 | 0.0153 | 0.2354 | 0.0344 | 0.5292 |
| 8.0 | 0.0243 | 0.3038 | 0.0305 | 0.3813 |
| 12.0 | 0.0483 | 0.4025 | 0.0201 | 0.1675 |
K260468 - Page 11 of 14
{11}
Bias Estimation for the hemolysate mode:
| HbA1c levels (%) | Deming | | Passing-Bablok | |
| --- | --- | --- | --- | --- |
| | Bias (%) | %Bias | Bias (%) | %Bias |
| 5.0 | 0.0153 | 0.3060 | 0.0439 | 0.8780 |
| 6.5 | 0.0258 | 0.3969 | 0.0376 | 0.5785 |
| 8.0 | 0.0363 | 0.4538 | 0.0313 | 0.3912 |
| 12.0 | 0.0643 | 0.5358 | 0.0145 | 0.1208 |
#### Total Error calculations and estimation:
The bias estimates from the method comparison study results and the precision estimates determined in the precision study were used to determine the total error of the device at each of the HbA1c levels listed in the tables below. Total error was calculated by the following equation: \(\% \mathrm{TE} = |\% \mathrm{Bias}| + 1.96 \times \% \mathrm{CV} \times (1 + \% \mathrm{Bias} / 100)\)
Total Error for the whole blood mode:
| Regression | HbA1c levels, % | %Bias | %CV | %TE |
| --- | --- | --- | --- | --- |
| Deming | 5.0 | 0.1260 | 1.18 | 2.44 |
| | 6.5 | 0.2354 | 1.13 | 2.46 |
| | 8.0 | 0.3038 | 1.06 | 2.39 |
| | 12.0 | 0.4025 | 0.98 | 2.33 |
| Passing-Bablok | 5.0 | 0.7660 | 1.18 | 3.10 |
| | 6.5 | 0.5292 | 1.13 | 2.76 |
| | 8.0 | 0.3813 | 1.06 | 2.47 |
| | 12.0 | 0.1675 | 0.98 | 2.09 |
Total Error for the hemolysate mode:
| Regression | HbA1c levels, % | %Bias | %CV | %TE |
| --- | --- | --- | --- | --- |
| Deming | 5.0 | 0.3060 | 1.21 | 2.68 |
| | 6.5 | 0.3969 | 1.15 | 2.66 |
| | 8.0 | 0.4538 | 1.05 | 2.52 |
| | 12.0 | 0.5358 | 0.95 | 2.41 |
| Passing-Bablok | 5.0 | 0.8780 | 1.21 | 3.27 |
| | 6.5 | 0.5785 | 1.15 | 2.85 |
| | 8.0 | 0.3912 | 1.05 | 2.46 |
| | 12.0 | 0.1208 | 0.95 | 1.99 |
#### 2. Matrix Comparison:
The effect of anticoagulants on Hemoglobin A1c determination by the H100Plus Hemoglobin Testing System was determined by comparing HbA1c results obtained from venous K2-EDTA and venous K3-EDTA samples. 50 paired samples in each anticoagulant with %HbA1c concentrations ranging from 3.1% to 20.0% were collected from 50 donors and evaluated once on the device. The following are results from ordinary least squares regression analysis:
| Sample | Unit | Slope | Intercept | \( R^2 \) |
| --- | --- | --- | --- | --- |
| | NGSP, % | 0.9995 | 0.0244 | 0.9997 |
K260468 - Page 12 of 14
{12}
| Whole blood | IFCC (mmol/mol) | 0.9995 | 0.2576 | 0.9997 |
| --- | --- | --- | --- | --- |
The results support the use of the HbA1c assay with samples collected in K2-EDTA and K3-EDTA.
### C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Clinical Cut-Off
Not applicable.
4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
### D Expected Values/Reference Range:
The sponsor provided the following expected values in their labeling:
| Hemoglobin A1c | | Suggested Diagnosis |
| --- | --- | --- |
| NGSP % | IFCC mmol/mol | |
| ≥6.5 | ≥48 | Diabetic[1-3] |
| 5.7–6.4 | 39–47 | Pre-Diabetic[2] |
| <5.7 | <39 | Non-Diabetic |
[1]World Health Organization. Use of glycated Haemoglobin (HbA1c) in the diagnosis of diabetes mellitus: abbreviated report of a WHO consultation. Geneva, 2011:1-25.
[2]American Diabetes Association Professional Practice Committee. Diagnosis and Classification of Diabetes: Standards of Care in Diabetes—2026. Diabetes Care. 2026;49 (Suppl 1):S27-S49.
[3]International Expert Committee. Report on the Role of the A1c Assay in the Diagnosis of Diabetes. Diabetes Care 2009, 32 (7), 1327-1334.
### E Other Supportive Instrument Performance Characteristics Data:
• EMC and Electrical Safety
K260468 - Page 13 of 14
{13}
The sponsor provided documentation certifying that acceptable electrical safety and (EMC) testing had been performed and the H100Plus Hemoglobin Testing System was found to be compliant.
- Software and Cybersecurity
The sponsor provided software and cybersecurity documentation that was reviewed and found to be acceptable.
- Additional studies were performed and found to be acceptable to support claims of 10,000 run of the HPLC column, the operation conditions of the system, and no carryover from sample to sample.
### VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
### IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K260468 - Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.