Electrosurgical Pencil and Accessories designed for general electrosurgical applications, including cutting and coagulation, illumination and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The electrosurgical pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. Electrosurgical Electrode Accessories designed for general electrosurgical applications, including cutting and coagulation. The electrosurgical electrode enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.
Device Story
Monopolar electrosurgical pencil and disposable electrode accessories; used to conduct high-frequency electrical current from electrosurgical unit to surgical site; enables cutting, coagulation, and smoke evacuation; includes illumination feature; operated by surgeon via push-button or foot switch; device components include electrode, handle, cable, and suction tube; used in clinical surgical settings; provides surgical effect via electrical energy; benefits include precise tissue management and smoke plume removal.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing (thermal effects on ex-vivo kidney, liver, and muscle tissue compared to predicate).
Technological Characteristics
Monopolar electrosurgical pencil; materials include stainless steel, Teflon, ABS, PC, POM, tungsten, silicone, ceramic, and epoxy resin; energy source: high-frequency electrical current; connectivity: 3-pin or 1-pin plug; sterilization: Ethylene Oxide (EO) to SAL 10^-6; dimensions: electrode length 100-330mm, diameter 2.36mm; cable length 2.5-3.2m; complies with IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2.
Indications for Use
Indicated for general electrosurgical applications including cutting, coagulation, illumination, and smoke removal (when used with evacuation system) in surgical procedures requiring remote conduction of electrosurgical current to target tissue.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
PlumePen (Elite, Pro and Ultra) Surgical Smoke Evacuation Pencil (K230547)
Reference Devices
Single Use Electrosurgical pencil with non-coated and non-stick electrode (K192542)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 7, 2026
Zhejiang shuyou Surgical Instrument Co., Ltd.
Jialong Wu
Quality Management Representative
Zhejiang shuyou Surgical Instrument Co., Ltd.
3689#, E. Yangguang Ave., Dipu St. Anji, Huzhou, Xhejiang 313300 CHN
Re: K260466
Trade/Device Name: Multifunctional Operational Dissectors (Electrosurgical Pencils); Multifunctional Operational Dissectors (Disposable Electrode)
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: February 24, 2026
Received: February 12, 2026
Dear Jialong Wu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260466 - Jialong Wu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260466 - Jialong Wu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JAMES H. JANG-S
Digitally signed by
JAMES H. JANG-S
Date: 2026.04.07
22:33:01 -04'00"
James Jang, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260466 | ? |
| Please provide the device trade name(s). | | ? |
| Multifunctional Operational Dissectors (Electrosurgical Pencils);
Multifunctional Operational Dissectors (Disposable Electrode) | | |
| Please provide your Indications for Use below. | | ? |
| Electrosurgical Pencil and Accessories designed for general electrosurgical applications, including cutting
and coagulation, illumination and for removing smoke generated by electrosurgery when used in
conjunction with an effective smoke evacuation system. The electrosurgical pencil enables the operator to
remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the
operative site for the desired surgical effect.
Electrosurgical Electrode Accessories designed for general electrosurgical applications, including cutting
and coagulation. The electrosurgical electrode enables the operator to remotely conduct an electrosurgical
current from the output connector of an electrosurgical unit to the operative site for the desired surgical
effect. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
K260466
# K260466 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K260466
1. Date of Preparation: 04/02/2026
2. Sponsor Identification
Zhejiang shuyou Surgical Instrument Co., Ltd.
3689#, East Yangguang Avenue, Dipu Street Anji Huzhou City, Zhejiang Province China
Establishment Registration Number: Not registered
Contact Person: Jialong Wu
Position: Quality Management Representative
Tel: +86-13735120013
Email: wujialong@shuyou-china.com
3. Identification of Proposed Device
Trade Name: Multifunctional Operational Dissectors (Electrosurgical Pencils)
Multifunctional Operational Dissectors (Disposable Electrode)
Common Name: Electrosurgical Smoke Evacuation Pencil
Regulatory Information
Classification Name: Electrosurgical, cutting & coagulation & accessories
Classification: II,
Product Code: GEI,
Regulation Number: 21 CFR 878.4400,
Review Panel: General & Plastic Surgery.
Indications for Use
Electrosurgical Pencil and Accessories designed for general electrosurgical applications, including cutting and coagulation, illumination and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The electrosurgical pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.
{5}
510(k) Summary
K260466
Electrosurgical Electrode Accessories designed for general electrosurgical applications, including cutting and coagulation. The electrosurgical electrode enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.
## Device Description
The electrosurgical pencil mainly consists of electrode, functional tube, insulating layer, plastic handle, manual button switch, cable, plug, Teflon material, suction tube joint, three-way valve. There are four main types of electrosurgical pencil, including SY-I, SY-II, SY-III and SY-IV. The design of the four types of electrosurgical pencil is different, and each type of electrosurgical pencil has differences in size and shape.
The disposable electrode is composed of electrode shaft, insulation layer, or plastic handle. There are six main types of disposable electrode, including ordinary electrode, tungsten needle electrode, tungsten wire electrode, silicone electrode, endoscopic electrode and orthopedic electrode. The size and shape of the six types of disposable electrode is different.
## 4. Identification of Predicate Devices and Reference Device
**Predicate Device**
510K Number: K230547
Trade Name: PlumePen (Elite, Pro and Ultra) Surgical Smoke Evacuation Pencil
**Reference Device**
510K Number: K192542
Trade Name: Single Use Electrosurgical pencil with non-coated and non-stick electrode
## 5. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ISO 10993-7:2008 Biological Evaluation Of Medical Devices-Part 7: Ethylene Oxide Sterilization Residuals
- ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10:2021 Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization
- ISO 10993-11:2017 Biological Evaluation of Medical Devices-Part 11: Tests for systemic toxicity
- ISO 10993-23:2021
- USP<151> Pyrogen Test
{6}
510(k) Summary
K260466
- ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ASTM F88/F88M-23 Standard test method for seal strength of flexible barrier materials
- ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
- IEC 60601-1:2005, AMD1: 2012, AMD2: 2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-2: 2017 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- IEC TR 60601-4-2: 2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- USP<85> Bacterial Endotoxins Test
The performance testing conducted on subject device and predicate device are listed below. All the test results demonstrate proposed device meet the requirements of its pre-defined acceptance criteria and intended uses.
- Appearance test
- Dimension test
- Conductivity test
- Detachable active electrode insert/pull force test
- Reliability testing of power cord
- High-frequency leakage current testing
- Dominant-frequency high-voltage
- High-frequency high-voltage test
- Corrosion resistance test
## Thermal Effects on Tissue
In accordance with the nonbinding FDA guidance on "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", part XI-E, the thermal effects on animal tissue of our device was compared with the predicate device K230547 Ex-vivo kidney, liver, and muscle tissues were used and the thermal damage zone sizes were compared for width, and depth. These tests showed that the thermal effects of proposed device are virtually identical to those shown in the predicate device.
{7}
510(k) Summary
K260466
The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 10⁻⁶. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 20 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 3-year shelf-life.
- Package Integrity Test – after environmental conditioning, simulated transportation testing in accordance to ASTM D4169-22 on final, packaged, and sterile device.
- Sterile Barrier Packaging performed on the proposed device:
- Visual Inspection per ASTM F1886/F1886M-16
- Seal Strength per ASTM F88/F88-21
- Dye Penetration per ASTM F1929-15
- Shelf-life of 5 years is validated using FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
The biocompatibility evaluations were conducted in accordance with the 2020 FDA Guidance document Use of International Standard ISO-10993-1, ‘Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process’. The cytotoxicity, sensitization, intracutaneous irritation, system toxicity, pyrogen and hemolysis tests were performed to demonstrate the biocompatibility of the device.
Tests on Electromagnetic Compatibility and Electrical Safety were performed in accordance with requirements per IEC 60601-1:2005, AMD1: 2012, AMD2: 2020, IEC 60601-1-2: 2014+A1:2020 and IEC TR 60601-4-2: 2024, and in particular we also conducted tests on high frequency surgical equipment accessories per IEC 60601-2-2: 2017.
6. Clinical Test Conclusion
No clinical study is included in this submission.
{8}
510(k) Summary
K260466
# 7. Substantially Equivalent (SE) Comparison
Table 1. General Comparison
| Item | Proposed Device | Predicate Device K230547 | Reference Device K192542 | Remark |
| --- | --- | --- | --- | --- |
| Device Name | Multifunctional Operational Dissectors (Electrosurgical Pencils) Multifunctional Operational Dissectors (Disposable Electrode) | PlumePen (Elite, Pro and Ultra) Surgical Smoke Evacuation Pencil | Single Use Electrosurgical pencil with non-coated and non-stick electrode; | / |
| Product Code | GEI | GEI | GEI | Same |
| Regulation No. | 21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 | Same |
| Class | II | II | II | Same |
| Indications for Use | Electrosurgical Pencil and Accessories designed for general electrosurgical applications, including cutting and coagulation, illumination and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The electrosurgical pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. Electrosurgical Electrode Accessories designed for general electrosurgical applications, including cutting and coagulation. The electrosurgical electrode enables the operator to remotely conduct an electrosurgical current from the output connector of an | The PlumePen® Elite, PlumePen® Ultra, and PlumePen® Pro is designed for general electrosurgical applications, including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. Indicated for use to remove smoke plume from the surgical site and to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the target tissue for the desired surgical effect. | The devices are used to cut and coagulate soft tissue by means of high frequency electrical current during an electrosurgical procedure. | Similar |
{9}
510(k) Summary
K260466
| | electrosurgical unit to the operative site for the desired surgical effect. | | | |
| --- | --- | --- | --- | --- |
| Prescription or OTC | Prescription | Prescription | Prescription | Same |
| Energy delivery | High frequency electrical current/energy | High frequency electrical current/energy | High frequency electrical current/energy | Same |
| Polarity | Monopolar | Monopolar | Monopolar | Same |
| Rated Voltage | 3kV | 5kVpk | 4kVp, 5kVp | Different |
| Shelf-life | 5 years | 3 years | 3 years | Different |
| Single use/reusable | Single patients use device | Single patients use device | Single patients use device | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
| Design | Monopolar electrosurgical pencil for cutting, coagulation, illumination and removing smoke, and with different electrode. The electrode tips as blade, needle, ball, straight, curved, circular, square, B-shape, triangular, shovel, hook and serrated shape. | Monopolar electrosurgical pencil for cutting, coagulation and removing smoke, and with different blade electrode tips | Monopolar electrosurgical pencil for cutting and coagulation, and with different electrode tips as blade, needle and ball | Similar |
| Switching method of electrosurgical pencil | Push button or foot switch | Unknown | Push button, rocker switch or foot switch | Same |
| Cable of the electrosurgical pencil | 2.5~3.2m with 3-pins plug or 1-pin plug | Unknown | 3~5m with 3-pins plug or 1-pin plug | Similar |
| Electrode length | 100~330mm | Unknown | 69mm~152mm | Similar |
| Electrode diameter | 2.36mm | Unknown | 2.36mm | Same |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Particular requirements | IEC 60601-2-2 | IEC 60601-2-2 | IEC 60601-2-2 | Same |
| Biocompatibility | | | | |
| Cytotoxicity | No cytotoxicity | Conform with ISO 10993 standards | Conform with ISO 10993 standards | Same |
| Intracutaneous | No intracutaneous reactivity | | | |
6 / 9
{10}
510(k) Summary
K260466
| Reactivity | | | | |
| --- | --- | --- | --- | --- |
| Sensitization | No sensitization | | | |
| Acute System Toxicity | No acute system Toxicity | | | |
| Pyrogen | No pyrogen | | | |
| Hemolysis | No hemolysis | | | |
| Patient-contact component and material | Electrosurgical Pencil: stainless steel, Teflon coating, PE heat shrink tubing, white powder, ABS, PC, POM, grey powder, tungsten, PTFE heat shrink tubing, PVDF heat shrink tubing, blue powder, silicone coating
Disposable Electrode: stainless steel, Teflon coating PTFE heat shrink tubing, PE heat shrink tubing, blue power, ABS, PVDF heat shrink tubing, tungsten, yellow power, silicone coating, ceramic, epoxy resin spray powder | Unknown | Stainless steel, Teflon coating, ABS, PVC, polyolefin shrink, wrap and/or PTFE shrink wrap or ABS/HIPS overmold | Different |
| Sterilization | | | | |
| Sterile/non-sterile | Sterile | Sterile | Sterile | Same |
| Method | EO sterilized | EO sterilized | EO sterilized | Same |
| SAL | 10^{-6} | 10^{-6} | 10^{-6} | Same |
## Similar - Indications for Use
The proposed electrosurgical pencils and predicate device share the same intended use and share the same functions, including cutting, coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The proposed disposable electrode, as a component of the electrosurgical pencils, has the functions of cutting and coagulation. These two functions are covered by the predicate device.
And both of these devices enable the operator remotely to conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. The principles and functions of the proposed devices and predicate devices are the same, except for slight differences in the textual descriptions. Furthermore, a comparison test was conducted between the proposed device and predicate device. The test results showed that there was no significant difference between the performance of the proposed device and that of the predicate device. Therefore, this difference will not
{11}
510(k) Summary
K260466
affect the safety and effectiveness of the proposed device.
## Different - Rated Voltage
The rated voltage of the proposed electrosurgical pencil is different from the predicate devices 1. However, the rated voltage of proposed electrosurgical pencil complies with IEC 60601-1 standard, the difference in rated voltage is just the difference in device design. Therefore, this difference on rated voltage is considered not to affect the safety and effectiveness of the proposed device.
## Different- Shelf-life
The shelf-life of the proposed electrosurgical pencil is different from the predicate devices 1. However, the proposed device underwent a shelf-life test, and all the test results of the aged devices met the acceptance criteria. Therefore, this difference on rated voltage is considered not to affect the safety and effectiveness of the proposed device.
## Similar - Design
The proposed device is the same as the predicate device, both monopolar electrosurgical pencil different electrodes that can perform cutting, coagulation and smoke removal functions. However, the electrodes in the proposed device come in various shapes to be used in different medical situations. Electrodes of different shapes were also subjected to performance tests to verify the performance of the electrodes. Both the proposed device and the predicate device underwent comparative tests, and the test results indicated that there was no significant difference in performance between the two devices. Therefore, the difference will not affect the safety and effectiveness of the proposed device.
## Similar - Cable of the electrosurgical pencil
The lengths of the cables for the proposed device and predicate device are slightly different, but the cable lengths do not affect the cutting and coagulation functions of the electrodes. Moreover, both the proposed device and the reference product are cables with 3-pins plug or 1-pin plug. Additionally, both the proposed device and the predicate device underwent comparative tests, and the test results indicated that there was no significant difference in performance between the two devices. Therefore, the difference will not affect the safety and effectiveness of the proposed device.
## Similar - Electrode length
The electrode length of the predicate device is unknown, but the electrode length of the proposed device is similar to that of the reference device. Both proposed device and reference device have multiple different electrode lengths. The proposed device offers more electrode lengths for doctors to make appropriate choices based on the patient's condition. Additionally, both the proposed device and the predicate device underwent comparative tests, and the test results indicated that there was no significant difference in performance between the two devices. Therefore, the difference will not affect the safety and effectiveness of the proposed device.
8 / 9
{12}
510(k) Summary
K260466
## Different - Material
The patient-conduct component and material of the predicated device 1 are unknown. However, the biocompatibility tests were conducted on the proposed device and the test result showed that the proposed device does not raise the adverse effect on the material. Therefore, the difference will not affect the safety and effectiveness of the proposed device.
## 8. Substantially Equivalent (SE) Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and perform as well as or better than the legally marketed predicate device K230547 and reference device K192542.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.