The Gold Hue Custom Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Gold Hue Custom Abutments are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Gold Hue Custom Abutment are intended to be sent to Straumann for manufacturing at a validated milling center. Medentika Custom Abutments Gold Hue are intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. Medentika Custom Abutments Gold Hue are intended for use with the Straumann® CARES® System. All digitally designed Medentika Custom Abutments Gold Hue are intended to be manufactured at a Straumann® validated milling center. The final patient matched form is a Custom Abutment Gold Hue. Medentika abutments for the Nobel Biocare Nobel Active® 3.0 mm, Dentsply Sirona Astra Tech OsseoSpeed EV® 3.0 mm and TX® 3.0 mm, and Straumann Bone Level 2.9 implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Device Story
Customized prosthetic abutments; manufactured from titanium alloy (ASTM F136); feature titanium nitride (TiN) coating on emergence profile/coronal aspect for gold color to improve esthetics in thin gingival tissue. Inputs: digital design files from CAD/CAM software. Transformation: milling at Straumann-validated center to create patient-specific geometry matching implant-abutment connection. Output: one-piece custom abutment. Used in dental clinics; operated by dental professionals. Output used to support screw-retained single or cemented-retained single/multiple restorations. Benefits: natural-looking restoration; functional support for dental prostheses.
Clinical Evidence
Bench testing only. Evidence includes surface treatment evaluation (chemical composition, thickness, SEM, roughness, scratch, static/dynamic shear per ASTM F1044/F1160), dynamic fatigue testing (ISO 14801), dimensional analysis/reverse engineering of implant-to-abutment connections, and steam sterilization validation (ISO 17665-1/2). Coating safety supported by biocompatibility data (cytotoxicity, wear) referenced from K162848.
Technological Characteristics
Material: Ti6Al4V ELI (ASTM F136). Surface: Titanium nitride (TiN) coating. Connection: Implant-specific geometry. Manufacturing: CAD/CAM milling at validated center. Sterilization: Moist heat (supplied non-sterile). Design limits: 0.4-0.5mm minimum wall thickness; 30° max angulation; 4.0mm minimum post height.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple tooth prostheses supported by dental implants. Specific abutments for Nobel Biocare Nobel Active 3.0mm, Dentsply Sirona Astra Tech OsseoSpeed EV 3.0mm/TX 3.0mm, and Straumann Bone Level 2.9mm are restricted to maxillary lateral and mandibular central/lateral incisors.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Neodent Implant System – Custom Abutments (K250614)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 7, 2026
Institut Straumann AG
% Jennifer Jackson
Sr Dir, Regulatory & Quality NAM
Straumann USA, LLC
60 Minuteman Road
Andover, Massachusetts 01810
Re: K260460
Trade/Device Name: Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: February 11, 2026
Received: February 11, 2026
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260460 - Jennifer Jackson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260460 - Jennifer Jackson
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANDREW I. STEEN -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K260460
Device Name
Neodent Gold Hue Custom Abutments
Indications for Use (Describe)
The Gold Hue Custom Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Gold Hue Custom Abutments are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Gold Hue Custom Abutment are intended to be sent to Straumann for manufacturing at a validated milling center.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PBC Publishing Services (301) 443-6740
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K260460
Device Name
Medentika Custom Abutments Gold Hue
Indications for Use (Describe)
Medentika Custom Abutments Gold Hue are intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. Medentika Custom Abutments Gold Hue are intended for use with the Straumann® CARES® System. All digitally designed Medentika Custom Abutments Gold Hue are intended to be manufactured at a Straumann® validated milling center. The final patient matched form is a Custom Abutment Gold Hue.
Medentika abutments for the Nobel Biocare Nobel Active® 3.0 mm, Dentsply Sirona Astra Tech OsseoSpeed EV® 3.0 mm and TX® 3.0 mm, and Straumann Bone Level 2.9 implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Implant System Compatibility Series (Series / Implant System / Implant diameter / Platform Diameters or Implant Connection):
| Medentika Series of the medical device | Manufacturer of the implant system | Compatible implant system | Implant Diameter (mm) | Platform Diameter (mm) |
| --- | --- | --- | --- | --- |
| E-Series | Nobel Biocare | Replace™ Select | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 |
| EV-Series | DENTSPLY Implants | ASTRA TECH OsseoSpeed® EV | 3.0, 3.6, 4.2, 4.8. 5.4 | 3.0, 3.6, 4.2, 4.8. 5.4 |
| F-Series | Nobel Biocare | NobelActive® CC | 3.0, 3.5, 4.3, 5.0, 5.5 | 3.0, 3.5, 4.3/5.0, 5.5 |
| H-Series | ZimVie | Biomet 3i Certain® Internal Connection | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| I-Series | ZimVie | Biomet 3i External Hex | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| K-Series | Nobel Biocare | Branemark System®, NobelSpeedy®, Groovy® | 3.3, 3.75,4.0,5.0 | 3.5, 4.1, 5.1 |
| L-Series | Straumann | Bone Level | 2.9, 3.3, 4.1, 4.8 | SC, NC, RC |
| N-Series | Straumann | Tissue Level | 3.3, 4.1, 4.8 | NNC, RN, WN |
| OT-Series | OSSTEM Implants | TS System | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | Mini, Regular |
| | HiOssen Implants® | ET System | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | |
| R-Series | ZimVie | Tapered Screw-Vent® | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5,5.7 |
| S-Series | Dentsply Implants | ASTRA TECH OsseoSpeed® TX | 3.0, 3.5, 4.0, 4.5, 5.0 | 3.0, 3.5/ 4.0, 4.5/ 5.0 |
| T-Series | Dentsply Implants | XiVE® S | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 |
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
{5}
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{6}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
# K260460 510(k) Summary
## Submitter's Contact Information
Submitter: Straumann USA, LLC
60 Minuteman Road
Andover, MA 01810
Registration No.: 1222315
Owner/Operator No.: 9005052
On behalf of:
Institut Straumann AG,
Peter Merian-Weg 12
Basal 4002 Switzerland
Laura Bleyendaal
Sr. Regulatory Affairs Specialist
Prepared By: Laura Bleyendaal
Contact Person: Jennifer M. Jackson, MS
Director of Regulatory Affairs
Phone Number: +1-978-747-2509
Fax Number: +1-978-747-0023
Date of Submission: May 7, 2026
## Name of the Device
Trade Names: Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
Common Name: Endosseous dental implant abutment
Classification Name: Endosseous dental implant abutment
Regulation Number: 21 CFR 872.3630
Device Classification: II
Product Code(s): NHA
Classification Panel: Dental
## Predicate Device(s)
Primary Predicate:
Institut Straumann AG
May 7, 2026
{7}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
- K250614 – Neodent Implant System – Custom Abutments
- Custom Abutments AS (angulated screw channel) for NGM and HS and ASC screws
## Reference Devices:
### Neodent
- K233857 – Neodent Implant System – Custom Abutments
- Custom Abutments AS for GM, Custom Abutments for NGM and HS
- K162848 – Straumann CARES Golden Ti/TiN Abutments
- Custom Abutments with same titanium nitride coating via same process
### Medentika
- K253341 – Medentika Custom Abutments Angulated Screw Channel (AS)
- Custom Abutments AS Series E, EV, F, H, I, K, L, N, OT, R, S and T
- K223113 – Medentika Abutment System, Medentika CAD/CAM Abutments
- Custom Abutments Series E, EV, F, and OT with a straight screw channel
- K150203 – Medentika CAD/CAM Abutments
- Custom Abutments Series E, F, H, I, K, L, N, R, S and T with a straight screw channel
- K242542 – Medentika CAD/CAM Abutments
- Custom Abutment Series L (SC) with a straight screw channel
- K162848 – Straumann CARES Golden Ti/TiN Abutments
- Custom Abutments with same titanium nitride coating via same process
## Device Description
The Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue can be used in combination with cemented prosthetics, e.g., crowns and superstructures, to reconstruct the function and esthetics of lost teeth. The Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue are one-piece abutments, which are customized abutments that are digitally designed and can only be milled and ordered from the Straumann validated milling center. The Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue have an implant-specific connection geometry for the respective compatible implant. The predicate devices are made from titanium
Institut Straumann AG
May 7, 2026
{8}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue K260460 510(k) Summary
alloy (titanium 6-aluminum 4-vanadium) conforming to ASTM F136. The purpose of this submission is to add Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue. The subject Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue feature a titanium nitride (TiN) coating on the emergence profile and coronal aspect of the abutment. The TiN coating results in the abutment having a gold color. The TiN coating serves to produce a more natural looking restoration; the gold color makes the presence of the abutment less noticeable, particularly in patients with thin gingival tissue.
## Indications for Use
## Neodent Gold Hue Custom Abutments
The Gold Hue Custom Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Gold Hue Custom Abutments are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Gold Hue Custom Abutment are intended to be sent to Straumann for manufacturing at a validated milling center.
## Medentika Custom Abutments Gold Hue
Medentika Custom Abutments Gold Hue are intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. Medentika Custom Abutments Gold Hue are intended for use with the Straumann® CARES® System. All digitally designed Medentika Custom Abutments Gold Hue are intended to be manufactured at a Straumann® validated milling center. The final patient matched form is a Custom Abutment Gold Hue.
Medentika abutments for the Nobel Biocare Nobel Active® 3.0 mm, Dentsply Sirona Astra Tech OsseoSpeed EV® 3.0 mm and TX® 3.0 mm, and Straumann Bone Level 2.9 implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Institut Straumann AG
May 7, 2026
{9}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue K260460 510(k) Summary
Implant System Compatibility Series (Series / Implant System / Implant diameter / Platform Diameters or Implant Connection):
| Medentika Series of the medical device | Manufacturer of the implant system | Compatible implant system | Implant Diameter (mm) | Platform Diameter (mm) |
| --- | --- | --- | --- | --- |
| E-Series | Nobel Biocare | Replace™ Select | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 |
| EV-Series | DENTSPLY Implants | ASTRA TECH OsseoSpeed® EV | 3.0, 3.6, 4.2, 4.8, 5.4 | 3.0, 3.6, 4.2, 4.8, 5.4 |
| F-Series | Nobel Biocare | NobelActive® CC | 3.0, 3.5, 4.3, 5.0, 5.5 | 3.0, 3.5, 4.3/5.0, 5.5 |
| H-Series | ZimVie | Biomet 3i Certain® Internal Connection | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| I-Series | ZimVie | Biomet 3i External Hex | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| K-Series | Nobel Biocare | Branemark System®, NobelSpeedy®, Groovy® | 3.3, 3.75,4.0,5.0 | 3.5, 4.1, 5.1 |
| L-Series | Straumann | Bone Level | 2.9, 3.3, 4.1, 4.8 | SC, NC, RC |
| N-Series | Straumann | Tissue Level | 3.3, 4.1, 4.8 | NNC, RN, WN |
| OT-Series | OSSTEM Implants | TS System | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | Mini, Regular |
| | HiOssen Implants® | ET System | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | |
| R-Series | ZimVie | Tapered Screw-Vent® | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5,5.7 |
| S-Series | Dentsply Implants | ASTRA TECH OsseoSpeed® TX | 3.0, 3.5, 4.0, 4.5, 5.0 | 3.0, 3.5/4.0, 4.5/5.0 |
| T-Series | Dentsply Implants | XiVE® S | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 |
Institut Straumann AG
May 7, 2026
{10}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
# Technological Characteristics
## Neodent Comparison
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables.
Table 1. Table of Substantial Equivalence – Neodent Indications for Use Statement.
| | Indications for Use Statement |
| --- | --- |
| Subject Device Neodent Custom Abutments Gold Hue | The Gold Hue Custom Abutment is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Gold Hue Custom Abutments are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Gold Hue Custom Abutment are intended to be sent to Straumann for manufacturing at a validated milling center. |
| Primary Predicate device K250614 Neodent Implant System – Custom Abutments JJGC Indústria e Comércio de Materiais Dentários S.A. | Custom Abutment AS: The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center. |
| Reference device K233857 Neodent Implant System – | The Custom Abutment Ti with Screw is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments Ti with Screw are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom |
Institut Straumann AG
May 7, 2026
{11}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
| Custom Abutments | Indications for Use Statement |
| --- | --- |
| JJGC Indústria e Comércio de Materiais Dentários S.A. | Abutment Ti with Screw are intended to be sent to Straumann for manufacturing at a validated milling center. The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center. |
| Reference device K162848 Straumann® CARES® Golden Ti/TiN Abutments Institut Straumann AG | Straumann® CARES® Golden Ti/TiN Abutments Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures. The Straumann CARES Golden Ti/TiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained. |
Institut Straumann AG
May 7, 2026
{12}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
Table.2 Substantial Equivalence – Neodent Comparative Summary of Technological Characteristics.
| Comparison | Neodent Gold Hue Custom Abutments | K250614 Neodent Implant System – Custom Abutments | K233857 Neodent Implant System – Custom Abutments | K162848 Straumann CARES Golden Ti/TiN Abutments | Equivalence |
| --- | --- | --- | --- | --- | --- |
| FDA product code | NHA | NHA | NHA | NHA | Identical |
| Implant-Abutment Connection | Grand Morse (GM) Narrow Grand Morse (NGM) Helix Short (HS) | NGM HS | GM NGM HS | CrossFit (NC, RC) SynOcta (NN, RN, WN) | Equivalent. No new compatible implant system is introduced with this submission. |
| Channel Solution | Straight and angulated | Angulated (NGM, HS) | Straight (GM, NGM, HS) Angulated (GM) | Straight | Equivalent. No new channel solution for a new implant abutment connection is introduced. |
| Abutment Designs | Titanium blank with pre-manufactured implant-interface Customized at validated milling center | Titanium blank with pre-manufactured implant-interface Customized at validated milling center | Titanium blank with pre-manufactured implant-interface Customized at validated milling center | Titanium blank with pre-manufactured implant-interface Customized at validated milling center | Equivalent |
| Prosthesis Attachment^{1} | Cement retained | Cement retained | Cement retained | Cement retained | Equivalent |
| Abutment & Abutment Screw Materials | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 Titanium nitride coating on abutments | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 | Commercially pure Titanium (grade 4) conforming to ISO 5832-2 Titanium nitride coating on abutments | Equivalent; the addition of a titanium nitride coating is supported by the predicate Straumann CARES Golden Ti/TiN Abutments as well as coating testing. |
| Sterilization | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile Moist heat sterilized by end user | Identical |
| Usage- All components | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | Identical |
| Design Limits | | | | | |
| Minimum wall thickness to screw channel (mm) | 0.4 | 0.4 | 0.4 | Not specified in 510(k) summary | Equivalent |
| Gingival Height (mm) | Straight screw channel: GM: 0.6 - 8.6 mm NGM: 0.6 - 8.6 mm | Angulated screw channel NGM: 0.6 - 8.6mm | The maximum gingiva height that the blank allows milling to considering the 4mm | Not specified in 510(k) summary | Equivalent |
Institut Straumann AG
May 7, 2026
{13}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
| Comparison | Neodent Gold Hue Custom Abutments | K250614 Neodent Implant System – Custom Abutments | K233857 Neodent Implant System – Custom Abutments | K162848 Straumann CARES Golden Ti/TiN Abutments | Equivalence |
| --- | --- | --- | --- | --- | --- |
| | HS for use with tissue level implant system, gingival height is given by the implant design Angulated screw channel GM: 0.6 – 8.6 mm NGM: 0.6 - 8.6 mm HS: for use with tissue level implant system, gingival height is given by the implant design | HS: for use with tissue level implant system, gingival height is given by the implant design | minimum abutment post height is the following: Straight screw channel: GM: 0.6 - 8.6 mm NGM: 0.6 - 8.6 mm HS for use with tissue level implant system, gingival height is given by the implant design Angulated screw channel GM: 0.6 – 8.6 mm | | |
| Max angulation of abutment body to implant axis (degrees) | 30° | 30° | 30° | 30° | Equivalent |
| Minimum abutment post height (mm) (height above the abutment collar/gingival height) (for single unit restorations) | 4.0 | 4.0 | 4.0 | Not specified in 510(k) summary | Equivalent |
| Max screw channel angulation | Straight Angulated up to 25° from the implant axis | Angulated up to 25° from the implant axis | Straight Angulated up to 25° from the implant axis | Straight | Equivalent |
| Design workflow | Computer aided design (CAD) | Computer aided design (CAD) | Computer aided design (CAD) | Computer aided design (CAD) | Equivalent |
| Manufacturing workflow | Milled at Straumann validated milling center | Milled at Straumann validated milling center | Milled at Straumann validated milling center | Milled at Straumann validated milling center | Equivalent |
# Medentika Comparison
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables.
Table 3. Table of Substantial Equivalence – Medentika Indications for Use Statement.
| | Indications for Use Statement |
| --- | --- |
| Subject Device | Medentika Custom Abutments Gold Hue are intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. Medentika Custom Abutments Gold Hue are intended for use with the Straumann® CARES® System. All digitally designed Medentika Custom Abutments Gold Hue are intended to be manufactured at a Straumann® validated milling center. The final patient matched form is a Custom Abutment Gold Hue. |
| Medentika Custom Abutments Gold Hue | Medentika abutments for the Nobel Biocare Nobel Active® 3.0 mm, Dentsply Sirona Astra Tech OsseoSpeed EV® 3.0 mm and TX® 3.0 mm, and Straumann Bone Level 2.9 implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only. Implant System Compatibility Series (Series / Implant System / Implant diameter / Platform Diameters or Implant Connection): |
Institut Straumann AG
May 7, 2026
{14}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
| | Indications for Use Statement | | | | |
| --- | --- | --- | --- | --- | --- |
| | Medentika Series of the medical device | Manufacturer of the implant system | Compatible implant system | Implant Diameter (mm) | Platform Diameter (mm) |
| | E-Series | Nobel Biocare | Replace™ Select | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 |
| | EV-Series | DENTSPLY Implants | ASTRA TECH OsseoSpeed® EV | 3.0, 3.6, 4.2, 4.8, 5.4 | 3.0, 3.6, 4.2, 4.8, 5.4 |
| | F-Series | Nobel Biocare | NobelActive® CC | 3.0, 3.5, 4.3, 5.0, 5.5 | 3.0, 3.5, 4.3/5.0, 5.5 |
| | H-Series | ZimVie | Biomet 3i Certain® Internal Connection | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | I-Series | ZimVie | Biomet 3i External Hex | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | K-Series | Nobel Biocare | Branemark System®, NobelSpeedy®, Groovy® | 3.3, 3.75,4.0,5.0 | 3.5, 4.1, 5.1 |
| | L-Series | Straumann | Bone Level | 2.9, 3.3, 4.1, 4.8 | SC, NC, RC |
| | N-Series | Straumann | Tissue Level | 3.3, 4.1, 4.8 | NNC, RN, WN |
| | OT-Series | OSSTEM Implants | TS System | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | Mini, Regular |
| | | HiOssen Implants® | ET System | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | |
| | R-Series | ZimVie | Tapered Screw-Vent® | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5,5.7 |
| | S-Series | Dentsply Implants | ASTRA TECH OsseoSpeed® TX | 3.0, 3.5, 4.0, 4.5, 5.0 | 3.0, 3.5/ 4.0, 4.5/ 5.0 |
| | T-Series | Dentsply Implants | XiVE® S | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 |
| | Indications for Use Statement | | | | |
| --- | --- | --- | --- | --- | --- |
| Predicate Device | Medentika CAD/CAM Abutments Medentika Preface CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. | | | | |
| K223113 | Medentika Preface is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center. The final patient matched form is a MedentiCAD abutment. | | | | |
| Medentika CAD/CAM Abutments | Medentika abutments for the Dentsply Sirona Astra Tech OsseoSpeed EV 3.0mm and TX 3.0mm implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only. Implant System Compatibility Series (Series / Implant System / Implant diameter / Platform Diameters or Implant Connection): | | | | |
| (Portions relevant to current submission) | Medentika Series of the medical device | Manufacturer of the implant system | Compatible implant system | Implant Diameter (mm) | Platform Diameter (mm) |
| | E-Series | Nobel Biocare | Replace™ Select | 6.0 | 6.0 |
| | EV-Series | DENTSPLY Implants | ASTRA TECH OsseoSpeed® EV | 3.0, 3.6, 4.2, 4.8, 5.4 | 3.0, 3.6, 4.2, 4.8, 5.4 |
| | F-Series | Nobel Biocare | NobelActive® CC | 5.5 | WP 5.5 |
| | OT-Series | OSSTEM Implants HiOssen Implants | TS System ET System | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | Mini, Regular |
Institut Straumann AG
May 7, 2026
{15}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
| | Indications for Use Statement |
| --- | --- |
| Reference device | Straumann® CARES® Golden Ti/TiN Abutments |
| K162848 | Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures. |
| Straumann® CARES® Golden Ti/TiN Abutments | The Straumann CARES Golden Ti/TiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained. |
| Institut Straumann AG | |
Table 4. Medentika Comparative Summary of Technological Characteristics.
| Comparison | Medentika Custom Abutments Gold Hue | K253341 Medentika Custom Abutments AS | K223113 Medentika Preface CAD/CAM abutments | K242542 Medentika CAD/CAM Abutments (PreFace, TI-Forms) | K150203 Medentika Preface Abutments | K162848 Straumann CARES Golden Ti/TiN Abutments | Equivalence |
| --- | --- | --- | --- | --- | --- | --- | --- |
| FDA product code | NHA | NHA | NHA | NHA, PNP | NHA | NHA | Identical |
| Series | E, EV, F, H, I, K, L, N, OT, R, S, T | E, EV, F, H, I, K, L, N, OT, R, S, T | E, EV, F, OT | E, EV, F, GM, H, I, K, L, LX, MG, N, OT, R, S, T, Y | E, F, H, I, K, L, N, R, S, T | CrossFit (NC, RC) SynOcta (NN, RN, WN) | Equivalent. No new series is added. No new compatible implant system is introduced, in terms of series, implant diameter and platform diameter with this submission. |
| Abutment Designs | Straight screw channel Titanium blank Ø11.5 and 16mm with pre-manufactured implant-interface Customized at validated milling center Angulated screw channel Titanium blank Ø13 with pre-manufactured implant-interface | Titanium blank Ø13 with pre-manufactured implant-interface Customized at validated milling center | Titanium blank Ø11.5 and 16mm with pre-manufactured implant-interface Customized at validated milling center | Titanium Blank, Ø 11.5 and 11.8 mm with pre-manufactured implant-interface Customized at validated milling center or to be manufactured according to the digital dentistry workflow | Titanium blank Ø11.5 and 16mm with pre-manufactured implant-interface Customized at validated milling center | Titanium blank with pre-manufactured implant-interface Customized at validated milling center | Equivalent |
Institut Straumann AG
May 7, 2026
{16}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
| Comparison | Medentika Custom Abutments Gold Hue | K253341 Medentika Custom Abutments AS | K223113 Medentika Preface CAD/CAM abutments | K242542 Medentika CAD/CAM Abutments (PreFace, TI-Forms) | K150203 Medentika Preface Abutments | K162848 Straumann CARES Golden Ti/TiN Abutments | Equivalence |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Customized at validated milling center | | | | | | |
| Prosthesis Attachment | Cement retained | Cement retained | Cement retained | Cement retained | Cement retained | Cement retained | Equivalent |
| Compatible Implant Body Diameter (mm) | 2.9-7.0 | 2.9-7.0 | 3.0-7.0 | 2.9 – 8.0 | 3.0-6.5 | 3.3-4.8 | Equivalent |
| Abutment & Abutment Screw Materials | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 Titanium nitride coating on abutments | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 | Ti6Al4V ELI, medical grade 5, conforming to ASTM F136 | Commercially pure Titanium (grade 4) conforming to ISO 5832-2 Titanium nitride coating on abutments | Equivalent; the addition of a titanium nitride coating is supported by the predicate Straumann CARES Golden Ti/TiN Abutments as well as coating testing. |
| Sterilization | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile Moist heat sterilized by end user | Supplied non-sterile | Identical |
| Usage- All components | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | Identical |
| Design Limits | | | | | | | |
| Minimum wall thickness to screw channel (mm) | 0.4 Straight screw channel 0.5 Angulated Screw channel | 0.5 | 0.4 | 0.4 | 0.4 | Not specified in 510(k) summary | Equivalent |
| Minimum gingival height in stock component (mm) | 0.1 (for N series abutments for use with tissue level implant systems, gingival height is given by the implant design) | 0.1 (for N series abutments for use with tissue level implant systems, gingival height is given by the implant design) | 0.1-0.25 | Not supplied in 510(k) summary | 0.00-0.23 | Not specified in 510(k) summary | Equivalent Labeling includes specific warnings for gingival heights less than 0.5mm |
| Maximum gingival height (mm) | 8.3 | 8.3 | 5 | 5 | 6 | Not specified in 510(k) summary | Equivalent |
Institut Straumann AG
May 7, 2026
{17}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
| Comparison | Medentika Custom Abutments Gold Hue | K253341 Medentika Custom Abutments AS | K223113 Medentika Preface CAD/CAM abutments | K242542 Medentika CAD/CAM Abutments (PreFace, TI-Forms) | K150203 Medentika Preface Abutments | K162848 Straumann CARES Golden Ti/TiN Abutments | Equivalence |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Max angulation of abutment body to implant axis (degrees) | 30° | 30° | 30° | 30° | 30° | 30° | Equivalent |
| Minimum abutment post height (mm) (height above the abutment collar/gingival height) (for single unit restorations) | 4.0 | 4.0 | 4.0 | 4.0 | 4.0 | Not specified in 510(k) summary | Equivalent |
| Max screw channel angulation | Straight Angulated up to 25° from the implant axis | up to 25° from the implant axis | Straight | Straight | Straight | Straight | Equivalent |
| Design workflow | Computer aided design (CAD) | Computer aided design (CAD) | Computer aided design (CAD) | Computer aided design (CAD) | Computer aided design (CAD) | Computer aided design (CAD) | Equivalent |
| Manufacturing workflow | Milled at Straumann validated milling center | Milled at Straumann validated milling center | Milled at Straumann CARES validated milling center | Milled at an FDA registered Medentika validated milling center or at point-of-care | Milled at Straumann CARES validated milling center | Milled at Straumann validated milling center | Equivalent |
Institut Straumann AG
May 7, 2026
{18}
Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue K260460 510(k) Summary
# Materials
The Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue are made of Titanium alloy (titanium 6-aluminum 4-vanadium) conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). The emergence profile and coronal aspect of the Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue are titanium nitride coated.
# Performance data
Nonclinical testing data submitted, referenced or relied upon to demonstrate substantial equivalence in this 510(k) includes:
- The surface treatment evaluation has been performed in accordance with 'Section 11 of Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutment'. The Titanium Nitride coating information such as chemical composition analysis, coating thickness, SEM imaging, surface roughness, scratch test data, static shear and tension testing was provided. Shear testing of the TiN coating was performed per ASTM F1044 Standard Test Method for Shear Testing of Calcium Phosphate and Metallic Coatings. Dynamic shear testing of the TiN coating was performed per ASTM F1160 Standard Test Method for Shear and Bending Fatigue of Calcium Phosphate and Metallic Medical and Composite Calcium Phosphate/ Metallic Coatings.
- Referenced from K253341, K223113, K242542, K150203 (Medentika GmbH) and K233857, K163194, K223638, K220251 (JJGC) Dynamic fatigue testing according to Root-form endosseous dental implants and endosseous dental abutments- class II special controls guidance document for industry and FDA staff and Guidance for industry and ISO 14801: 2016 Dentistry — implants — dynamic loading test for endosseous dental implant.
- Referenced from K223113, K242542, K150203 and K170838 (Medentika GmbH) was dimensional analysis and reverse engineering of the implant-to-abutment connection platform, including an assessment of maximum and minimum dimensions of critical design aspects, tolerances, and cross-sectional images of the submission device and compatible OEM implant body, OEM abutment, and OEM fixation screw.
- Referenced from K223113 (Medentika GmbH) and K250614 (JJGC) was steam sterilization validations according to ISO 17665-1: Sterilization of health care products- Moist heat- Part
Institut Straumann AG
May 7, 2026
{19}
Institut Straumann AG
May 7, 2026
14
# Traditional 510(k) Submission
Neodent Gold Hue Custom Abutments, Medentika Custom Abutments Gold Hue
K260460 510(k) Summary
1: Development, validation and routine control of a sterilization process for medical devices and ISO/TS 17665-2: Sterilization of health care products- Moist heat- Part 2: Guidance on the application of ISO 17665-1.
- Referenced from K162848 (Institut Straumann AG) was titanium nitride coating process and supporting data (cytotoxicity, chemical characterization, coating composition, coating thickness, coating porosity, coating mean volume percent voids, coating surface roughness and coating wear against dental tips)
# Conclusion
The data included in this submission demonstrates substantial equivalence to the predicate device listed above. Performance testing and comparison to previous clearances show that the subject devices are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.