K260456 · Traferox Technologies, Inc. · KDN · Jun 1, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K260456
Device Name
VI°Port™ Liver Preservation System
Applicant
Traferox Technologies, Inc.
Product Code
KDN · Gastroenterology, Urology
Decision Date
Jun 1, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Indications for Use
The VI°Port™ Liver Preservation System is intended to be used for the static hypothermic preservation of livers during transportation and eventual transplantation into a recipient, using cold storage solutions indicated for use with the livers. The intended organ storage time for the VI°Port™ Liver Preservation System is up to 15 hours. When clinically accepted static hypothermic preservation times are exceeded, the livers should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient.
Device Story
VI°Port™ Liver Preservation System provides static hypothermic storage (4°C to 8°C) for human livers during transport; utilizes insulated chamber with phase-change material (5°C transition) pre-conditioned in freezer; includes internal organ support, temperature probe, and logger; non-sterile, single-use, non-contact device; mobile unit with wheels and telescopic handle; temperature data displayed on external screen and via optional mobile application using Bluetooth; used by transplant personnel; assists surgeons in maintaining organ viability during transit; benefits patient by preserving organ integrity for transplantation.
Clinical Evidence
No clinical data. Bench testing only, including software/cybersecurity, electrical/mechanical/thermal safety, electromagnetic compatibility, transportation, thermal qualification, human factors, packaging/shipping, and lifetime testing.
Technological Characteristics
Insulated chamber; phase-change material (approx. 5°C transition); temperature probe and logger; Bluetooth connectivity; mobile form factor with wheels/handle; non-sterile; single-use; non-contact with organ. Software for temperature logging and display.
Indications for Use
Indicated for static hypothermic preservation of livers in adult patients during transport for transplantation, using cold storage solutions, for up to 15 hours.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}

June 1, 2026
Traferox Technologies, Inc.
Anuradha Iyer
Regulatory Affairs Manager
3505 Laird Rd. Unit 16
Mississauga, ON L5L 5Y7 CAN
Re: K260456
Trade/Device Name: VIPort™ Liver Preservation System
Regulation Number: 21 CFR 876.5880
Regulation Name: Isolated Kidney Perfusion And Transport System And Accessories
Regulatory Class: Class II
Product Code: KDN
Dated: April 30, 2026
Received: April 30, 2026
Dear Anuradha Iyer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260456 - Anuradha Iyer
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260456 - Anuradha Iyer
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee, Ph.D., M.B.A.
Anthony C. Lee, Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K260456 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| VI°Port™ Liver Preservation System | | | |
| Please provide your Indications for Use below. | | | ? |
| The VI°Port™ Liver Preservation System is intended to be used for the static hypothermic preservation of livers during transportation and eventual transplantation into a recipient, using cold storage solutions indicated for use with the livers. The intended organ storage time for the VI°Port™ Liver Preservation System is up to 15 hours. When clinically accepted static hypothermic preservation times are exceeded, the livers should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient. | | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
page 1/2 - K260456
510(k) #: K260456
# 510(k) Summary
Prepared on: 2026-04-30
## Contact Details
21 CFR 807.92(a)(1)
Applicant Name
Traferox Technologies Inc.
Applicant Address
3505 Laird Rd. Unit 16 Mississauga ON L5L 5Y7 Canada
Applicant Contact Telephone
+1 437-983-8971
Applicant Contact
Ms. Anuradha Iyer
Applicant Contact Email
aiyer@traferox.com
## Device Name
21 CFR 807.92(a)(2)
Device Trade Name
VI°Port™ Liver Preservation System
Common Name
System, Perfusion, Kidney
Classification Name
Isolated kidney perfusion and transport system and accessories
Regulation Number
876.5880
Product Code(s)
KDN
## Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K192869
SherpaPak™ Liver Transport System
KDN
## Device Description Summary
21 CFR 807.92(a)(4)
The VI°Port Liver Preservation System is designed to store, preserve, and transport human livers donated for transplantation. The target temperature range maintained during storage is 4°C to 8°C. The VI°Port Liver Preservation System is designed to protect the livers during transport and to display and log the internal device temperature on an external screen. Additionally, the internal temperature may be monitored via an optional mobile application. The system is designed to be mobile, with wheels and a telescopic handle to maneuver the device throughout its travel.
The system consists of an insulated chamber, specialized phase-change material with a transition temperature of approximately 5°C, an internal support for the organ, and a temperature probe and logger. The temperature logger has Bluetooth connectivity. The system is a non-sterile single use device and does not contact the organ. The VI°Port Liver Preservation System is intended to be used with cold storage solutions indicated for use with the liver.
VI°Port Liver Preservation System principle of operation is in the pre-conditioning and use of specialized phase change material cooling packs. The cooling packs are preconditioned in a freezer (at least -20°C) for a minimum of 48 hours before setting up the VI°Port Liver Preservation System.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The VI°Port™ Liver Preservation System is intended to be used for the static hypothermic preservation of livers during transportation and eventual transplantation into a recipient, using cold storage solutions indicated for use with the livers.
The intended organ storage time for the VI°Port™ Liver Preservation System is up to 15 hours.
{5}
page 2/2 - K260456
When clinically accepted static hypothermic preservation times are exceeded, the livers should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device has the same indications for use as its predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
The proposed VI*Port™ Liver Preservation System is substantially equivalent to its predicate SherpaPak™ Liver Transport System (K192869) with regards to:
- Intended use/Indications for use
- Technological characteristics (Fundamental scientific technology); and
- Safety and effectiveness.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The following non-clinical bench testing has been performed for the VI*Port Liver Preservation System to demonstrate the device's safety and effectiveness.
- Software and Cybersecurity Testing
- Basic Safety (electrical, mechanical, thermal)
- Electromagnetic Compatibility
- Transportation Testing
- Thermal Qualification Testing
- Human Factors Testing
- Packaging and Shipping Testing
- Lifetime Testing
Not Applicable
The non-clinical testing conducted demonstrates that the proposed device does not raise any new issues of safety and effectiveness and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.