K260432 · Venocare, Inc. · FOZ · Sep 2, 2026 · General Hospital
Device Facts
Record ID
K260432
Device Name
Ruby Intravascular Catheter
Applicant
Venocare, Inc.
Product Code
FOZ · General Hospital
Decision Date
Sep 2, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The RUBY Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The RUBY Intravascular Catheter is suitable for use with power injectors. The 20 Gauge RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 10 kg) and adults. The 22 Gauge and 24 Gauge RUBY Intravascular Catheter are indicated for use in pediatric patients (with body mass higher or equal to 3.5 kg) and adults.
Device Story
Radiopaque, single-use, sterile intravascular catheter; 1.30-inch length. Clinician inserts into peripheral vascular system; device features blood control valve to prevent flow upon needle retraction; includes catheter guiding element with atraumatic tip deployed via slider. Active safety mechanism prevents sharp injuries. Used for blood sampling, pressure monitoring, or fluid administration; compatible with power injectors. Output is vascular access; healthcare provider uses device to facilitate therapy or monitoring. Benefits include reduced sharp injury risk and controlled blood flow during insertion.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including performance testing per ISO 10555-1/5, biocompatibility per ISO 10993 series, and sterilization/packaging validation per ISO 11135, ISO 11607, and ASTM standards.
Indicated for pediatric patients (≥10 kg for 20G; ≥3.5 kg for 22G/24G) and adults requiring vascular access for blood sampling, blood pressure monitoring, or intravenous fluid administration.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
9/2/2026
Venocare, Inc.
Ivenette Guzman
Vice-President of Quality Assurance & Regulatory Affairs
3750 NW 87th Ave.
Suite 260
Doral, Florida 33178
Re: K260432
Trade/Device Name: Ruby Intravascular Catheter
Regulation Number: 21 CFR 880.5200
Regulation Name: Intravascular Catheter
Regulatory Class: Class II
Product Code: FOZ
Dated: July 31, 2026
Received: August 3, 2026
Dear Ivenette Guzman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260432 - Ivenette Guzman
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260432 - Ivenette Guzman
Page 3
Sincerely,
GANG PENG -S
Gang Peng for
David Wolloscheck
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260432
Device Name
RUBY Intravascular Catheter
Indications for Use (Describe)
The RUBY Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The RUBY Intravascular Catheter is suitable for use with power injectors.
The 20 Gauge RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 10 kg) and adults.
The 22 Gauge and 24 Gauge RUBY Intravascular Catheter are indicated for use in pediatric patients (with body mass higher or equal to 3.5 kg) and adults.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
VENOCARE
# K260432 - 510(K) SUMMARY
| Device Name | RUBY Intravascular Catheter | |
| --- | --- | --- |
| Indications for Use | The RUBY Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The RUBY Intravascular Catheter is suitable for use with power injectors. The 20 Gauge RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 10 kg) and adults. The 22 Gauge and 24 Gauge RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 3.5 kg) and adults. | |
| Submitter Information | Submitter Name | Venocare, Inc. |
| | Contact Person | Raul Leyte-Vidal |
| | Telephone Number | 786.870.4020 |
| | Address | 3750 NW 87^{th} Avenue, Suite 260 Doral, FL 33178 |
| | Date of Preparation | September 1, 2026 |
| Subject Device | Trade Name | RUBY Intravascular Catheter |
| | Common Name | Catheter, Intravascular, therapeutic, short-term less than 30 days |
| | Classification Name | Intravascular catheter |
| | Class | Class II |
| | Regulation Number | 21 CFR § 880.5200 |
| | Product Code | FOZ |
| | 510(k) Number | K260432 |
| | Classification Panel | General Hospital |
| Predicate Device | Trade Name | RUBY Intravascular Catheter (20 Gauge) |
| | Common Name | Catheter, Intravascular, therapeutic, short-term less than 30 days |
| | Classification Name | Intravascular catheter |
| | Class | Class II |
Page 1 of 7
{5}
VENOCARE®
| | Regulation Number | 21 CFR § 880.5200 | |
| --- | --- | --- | --- |
| | Product Code | FOZ | |
| | 510(k) Number | K243105 | |
| Device Description | The RUBY Intravascular Catheter consists of a radiopaque catheter having a usable length of 1.30 inches. The device is a single use, sterile intravascular catheter designed to be inserted into a patient's vascular system to sample blood, monitor pressure, or administer fluids intravenously. The RUBY Intravascular Catheter's hub has a blood control valve mechanism that automatically activates to stop the blood flow in the catheter hub when the needle is removed from the catheter during initial insertion by the clinician. Blood flow from the catheter is restricted immediately after needle retraction until a secure luer connection has been made. The RUBY Intravascular Catheter is provided with a catheter guiding element with an atraumatic tip design positioned within the formed lumen between the catheter and the needle and connected to a slider used to deploy and retract the catheter guiding element. The RUBY Peripheral Intravascular Catheter also has an active safety mechanism and container that prevents sharp injuries. | | |
| Intended Use | The RUBY Intravascular Catheter is intended to be inserted in the patient's vascular system for short-term use to sample blood, monitor blood pressure, or administer fluids intravenously. | | |
| Technological Characteristics | The key modifications made to the subject device when compared to the predicate device are as follows: | | |
| | 1. Change to the catheter guiding element design and composition | | |
| | 2. Changes to the catheter wing and strain relief | | |
| Attribute | Predicate Device RUBY Intravascular Catheter (K243105) | Subject Device RUBY Intravascular Catheter (K260432) | Comparison |
| Owner | Venocare, Inc. | Venocare, Inc. | Same |
| Classification | Class II | Class II | Same |
| 510(k) Status | K243105 | K260432 | N/A |
| Indications for Use | The RUBY Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be used with | The RUBY Intravascular Catheter is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be used with | Different. The 22-gauge and 24-gauge devices expand the indicated pediatric population to patients with body mass ≥3.5 kg. Simulated-use testing, performance testing, and |
Page 2 of 7
{6}
VENOCARE®
| | consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The RUBY Intravascular Catheter is suitable for use with power injectors. The RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 10 kg) and adults. | consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The RUBY Intravascular Catheter is suitable for use with power injectors. The 20 Gauge RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 10 kg) and adults. The 22 Gauge and 24 Gauge RUBY Intravascular Catheter is indicated for use in pediatric patients (with body mass higher or equal to 3.5 kg) and adults. | biocompatibility assessments support the difference and do not raise new questions of safety or effectiveness. |
| --- | --- | --- | --- |
| **Commercial Name** | RUBY Intravascular Catheter | RUBY Intravascular Catheter | N/A |
| **Duration of Use** | Short term (< 30 days) | Short term (< 30 days) | Same |
| **Device Components** | Catheter Length: 1.30 inches Catheter Diameter: 20 GA | Catheter Length: 1.30 inches Catheter Diameter: 22, 24 GA | Same. Different. Performance testing done on the device shows that this change does not introduce new questions of safety and effectiveness. |
| | Needle Length: 1.30 inches | Needle Length: 1.30 inches | Same |
| | Needle Diameter: 20GA Catheter– 22GA Needle | Needle Diameter: 22GA Catheter– 24GA Needle | Different. Performance testing done on the device shows that this change does not introduce new |
Page 3 of 7
{7}
VENOCARE
| | | 24GA Catheter– 26GA Needle | questions of safety and effectiveness. |
| --- | --- | --- | --- |
| | Catheter Guiding Element Length: > 1.30 inches | Catheter Guiding Element Length: > 1.30 inches | Same |
| | Catheter Guiding Element Diameter: .007 inches (20GA) | Catheter Guiding Element Diameter: .006 inches (22GA) .006 inches (24GA) | Different. Performance testing was done on the device to show that this difference does not raise new questions of safety and effectiveness. |
| | Catheter Luer | Catheter Luer | Same |
| | Needle Safety Housing | Needle Safety Housing | Same |
| | Needle Cover | Needle Cover | Same |
| | Support tubes | Support tubes | Same |
| | Spring | Spring | Same |
| | Porous Plug | Porous Plug | Same |
| | Button | Button | Same |
| | Slider | Slider | Same. |
| Means of Insertion | Percutaneous, over a catheter guiding element | Percutaneous, over a catheter guiding element | Same |
| Insertion Site | Peripheral | Peripheral | Same |
| Device Materials | Catheter Shaft: TPU | Catheter Shaft: TPU | Same |
| | Needle: Stainless Steel | Needle: Stainless Steel | Same |
| | Catheter Guiding Element: Nitinol and Pebax | Catheter Guiding Element: Nitinol, Pebax, Polyethylene Terephthalate (PET) Heat Shrink | Different. Simulated-use testing, performance testing, and biocompatibility assessments support the difference and do not raise new questions of safety or effectiveness. |
| | Catheter Luer: | Catheter Luer: | Different. Material colorants changed to |
Page 4 of 7
{8}
VENOCARE
| | TPU + Colorant 20GA (Pink) | TPU + Colorant 22GA (Blue) 24GA (Yellow) | represent gauge size. Simulated-use testing, performance testing, and biocompatibility assessments support the difference and do not raise new questions of safety or effectiveness. |
| --- | --- | --- | --- |
| | Needle Safety Housing: Polycarbonate | Needle Safety Housing: Polycarbonate | Same |
| | Needle Cover: Polypropylene | Needle Cover: Polypropylene | Same |
| | Support Tubes: Stainless Steel | Support Tubes: Stainless Steel | Same |
| | Spring: Stainless Steel | Spring: Stainless Steel | Same |
| | Porous Plug: PTFE | Porous Plug: PTFE | Same |
| | Button: Polycarbonate | Button: Polycarbonate | Same |
| | Slider: Polycarbonate + Colorant 20GA (Pink) | Slider: Polycarbonate + Colorant 22GA (Blue) 24GA (Yellow) | Different. Material colorants changed to represent gauge size. Simulated-use testing, performance testing, and biocompatibility assessments support the difference and do not raise new questions of safety or effectiveness. |
| Catheter Proximal Configuration | Luer Connection | Luer Connection | Same |
| Catheter Distal Configuration | Open Ended | Open Ended | Same |
| Number of Lumens | Single Lumen | Single lumen | Same |
| Power injection maximum flow rate | 6mL/sec | 6mL/sec (22GA) 3mL/sec (24GA) | Different. Performance testing done on the device shows that this change does not introduce new questions of safety and effectiveness. |
Page 5 of 7
{9}
VENOCARE
| Sterility | Provided Sterile | Provided Sterile | Same |
| --- | --- | --- | --- |
| Available Packaging Configurations | Standalone Device | Standalone Device | Same |
| Safety and Performance Tests | Verification and validation tests were designed and performed in accordance with Design Controls as per 21 CFR § 820.30. The following tests were conducted per guidance documents and standards in determining the substantial equivalence of the RUBY Intravascular Catheter to the predicate RUBY Intravascular Catheter (K243105). | | |
| Performance | Applicable Standards: - ISO 11135:2014 – Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)] - ISO 10555-1: 2013 - Intravascular catheters - Sterile and single-use intravascular catheters - Part 1: General requirements [Including AMENDMENT 1 (2017)] - ISO 10555-5: 2023 - Intravascular catheters -- Sterile and single-use catheters -- Part 5: Over-needle peripheral catheters. - ISO 7864: 2016 - Sterile hypodermic needles for single use - Requirements and test methods. - ISO 23908: 2011 - Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles introducers for catheters and needles used for blood sampling. - ISO 9626: 2016 - Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods | | |
| Biocompatibility | Applicable Standards: - ISO 10993-1: 2018 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process - ISO 10993-12: 2021 - Biological evaluation of medical devices - Part 12: Sample preparation and reference materials. - ISO 10993-18: 2020 - Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process. | | |
| Sterilization and Packaging | Applicable Standards: - ISO 10993-7:2008 - Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009) AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)] - ASTM F2250-13: 2018 - Standard Practice for Evaluation of Chemical Resistance of Printed Inks and Coatings on Flexible Packaging Materials. - ASTM F1980:2021 – Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. - ISO 11607-1: 2019 – Packaging for terminally sterilized medical devices – Part 1: Requirements for materials sterile barrier systems and packaging systems. | | |
Page 6 of 7
{10}
VENOCARE
| | - ISO 11607-2: 2019 - Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming sealing and assembly processes. - ASTM D4169-22 – Standard Practice for Performance Testing of Shipping Containers and systems - ASTM D4332-22 – Standard Practice for Conditioning Containers Packages or Packaging Components for Testing |
| --- | --- |
| **Shelf Life** | Referenced assessment of baseline and real-time aged samples supports a product shelf-life of 6 months. |
| **Conclusion** | The selective non-clinical tests referenced support the substantial equivalence of the RUBY 22 Gauge and 24 Gauge Intravascular Catheters to the predicate RUBY 20 Gauge Intravascular Catheter (K243105) for the proposed indications for use. The comparison of the technological characteristics and the intended use do not raise new questions of safety and effectiveness. |
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.