K260429 · Tandem Diabetes Care, Inc. · QJI · Apr 24, 2026 · Clinical Chemistry
Device Facts
Record ID
K260429
Device Name
Control-IQ+ technology
Applicant
Tandem Diabetes Care, Inc.
Product Code
QJI · Clinical Chemistry
Decision Date
Apr 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1356
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic, Pediatric
Indications for Use
Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold. Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater. Control-IQ+ technology is intended for use in pregnancy complicated by Type 1 diabetes mellitus, provided the linked CGM system is suitable for use in pregnancy. Control-IQ+ technology is intended for single patient use and requires a prescription.
Device Story
Software-only device (Control-IQ+) installed on compatible ACE insulin pumps; manages insulin delivery for Type 1 and Type 2 diabetes. Inputs: iCGM glucose readings, predicted glucose values, user-entered data (carbs, exercise, sleep). Operation: automatically adjusts basal insulin (increase/decrease/suspend) and delivers correction boluses based on predicted glucose thresholds. Used in clinic/home settings by patients; requires prescription. Healthcare providers use output to monitor glycemic control; device benefits include improved time-in-range and reduced glycemic variability. Pregnancy-specific use requires CGM compatible with pregnancy.
Clinical Evidence
Pivotal study 'CIRCUIT' (randomized controlled trial) evaluated safety/effectiveness in pregnancy complicated by Type 1 diabetes. 91 pregnant women (ages 18-45) enrolled before 14 weeks gestation at 14 sites. Primary endpoint: glycemic control (percent time-in-range 63-140 mg/dL) measured by CGM from 16 to 34 weeks gestation compared to control group.
Technological Characteristics
Software-only device; operates on compatible ACE insulin pumps. Uses iCGM data for closed-loop insulin delivery. Modes: Normal, Sleep, Exercise with fixed target ranges. Requires user input for weight and total daily insulin. Complies with ANSI AAMI ISO 14971:2019. No changes to software, firmware, or cybersecurity compared to predicate.
Indications for Use
Indicated for management of Type 1 diabetes (age 2+) and Type 2 diabetes (age 18+) including pregnancy complicated by Type 1 diabetes. Contraindicated for patients requiring <5 units/day insulin or weighing <20 lbs. Not for use in children <2 years.
Regulatory Classification
Identification
An interoperable automated glycemic controller is a device intended to automatically calculate drug doses based on inputs such as glucose and other relevant physiological parameters, and to command the delivery of such drug doses from a connected infusion pump. Interoperable automated glycemic controllers are designed to reliably and securely communicate with digitally connected devices to allow drug delivery commands to be sent, received, executed, and confirmed. Interoperable automated glycemic controllers are intended to be used in conjunction with digitally connected devices for the purpose of maintaining glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) An appropriate, as determined by FDA, clinical implementation strategy, including data demonstrating appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(A) The clinical data must be representative of the performance of the device in the intended use population and in clinically relevant use scenarios and sufficient to demonstrate appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(B) For devices indicated for use with multiple therapeutic agents for the same therapeutic effect (
*e.g.,* more than one type of insulin), data demonstrating performance with each product or, alternatively, an appropriate, as determined by FDA, clinical justification for why such data are not needed.(C) When determined to be necessary by FDA, the strategy must include postmarket data collection to confirm safe real-world use and monitor for rare adverse events.
(ii) Results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(iii) A detailed and appropriate, as determined by FDA, strategy to ensure secure and reliable means of data transmission with other intended connected devices.
(iv) Specifications that are appropriate, as determined by FDA, for connected devices that shall be eligible to provide input to (
*e.g.,* specification of glucose sensor performance) or accept commands from (*e.g.,* specifications for drug infusion pump performance) the controller, and a detailed strategy for ensuring that connected devices meet these specifications.(v) Specifications for devices responsible for hosting the controller, and a detailed and appropriate, as determined by FDA, strategy for ensuring that the specifications are met by the hosting devices.
(vi) Documentation demonstrating that appropriate, as determined by FDA, measures are in place (
*e.g.,* validated device design features) to ensure that safe therapy is maintained when communication with digitally connected devices is interrupted, lost, or re-established after an interruption. Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled and logged appropriately during and after the interruption to maintain patient safety.(vii) A detailed plan and procedure for assigning postmarket responsibilities including adverse event reporting, complaint handling, and investigations with the manufacturers of devices that are digitally connected to the controller.
(2) Design verification and validation documentation must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures to address device system hazards;
(ii) Design decisions related to how the risk control measures impact essential performance; and
(iii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include appropriate, as determined by FDA, and validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to connected devices;
(ii) Secure, accurate, and reliable means of data transmission between the controller and connected devices;
(iii) Sharing of necessary state information between the controller and any connected devices (
*e.g.,* battery level, reservoir level, sensor use life, pump status, error conditions);(iv) Ensuring that the controller continues to operate safely when data is received in a manner outside the bounds of the parameters specified;
(v) A detailed process and procedures for sharing the controller's interface specification with connected devices and for validating the correct implementation of that protocol; and
(vi) A mechanism for updating the controller software, including any software that is required for operation of the controller in a manner that ensures its safety and performance.
(4) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices, and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the controller must, at a minimum, include:
(i) Commands issued by the controller, and associated confirmations the controller receives from digitally connected devices;
(ii) Malfunctions of the controller and malfunctions reported to the controller by digitally connected devices (
*e.g.,* infusion pump occlusion, glucose sensor shut down);(iii) Alarms and alerts and associated acknowledgements from the controller as well as those reported to the controller by digitally connected devices; and
(iv) Connectivity events (
*e.g.,* establishment or loss of communications).(5) The device must only receive glucose input from devices cleared under § 862.1355 (integrated continuous glucose monitoring system), unless FDA determines an alternate type of glucose input device is designed appropriately to allow the controller to meet the special controls contained within this section.
(6) The device must only command drug delivery from devices cleared under § 880.5730 of this chapter (alternate controller enabled infusion pump), unless FDA determines an alternate type of drug infusion pump device is designed appropriately to allow the controller to meet the special controls contained within this section.
(7) An appropriate, as determined by FDA, training plan must be established for users and healthcare providers to assure the safety and performance of the device when used. This may include, but not be limited to, training on device contraindications, situations in which the device should not be used, notable differences in device functionality or features compared to similar alternative therapies, and information to help prescribers identify suitable candidate patients, as applicable.
(8) The labeling required under § 809.10(b) of this chapter must include:
(i) A contraindication for use in pediatric populations except to the extent clinical performance data or other available information demonstrates that it can be safely used in pediatric populations in whole or in part.
(ii) A prominent statement identifying any populations for which use of this device has been determined to be unsafe.
(iii) A prominent statement identifying by name the therapeutic agents that are compatible with the controller, including their identity and concentration, as appropriate.
(iv) The identity of those digitally connected devices with which the controller can be used, including descriptions of the specific system configurations that can be used, per the detailed strategy submitted under paragraph (b)(1)(iii) of this section.
(v) A comprehensive description of representative clinical performance in the hands of the intended user, including information specific to use in the pediatric use population, as appropriate.
(vi) A comprehensive description of safety of the device, including, for example, the incidence of severe hypoglycemia, diabetic ketoacidosis, and other relevant adverse events observed in a study conducted to satisfy paragraph (b)(1)(i) of this section.
(vii) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(viii) For any controller with hardware components intended for multiple patient reuse, instructions for safely reprocessing the hardware components between uses.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K260429
B Applicant
Tandem Diabetes Care, Inc.
C Proprietary and Established Names
Control-IQ+ technology
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QJI | Class II | 21 CFR 862.1356 | CH – Clinical Chemistry |
## E Purpose for Submission:
The purpose of this submission is to expand the indications for use to include pregnancy complicated by type 1 diabetes mellitus.
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold.
Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
Control-IQ+ technology is intended for use in pregnancy complicated by Type 1 diabetes mellitus, provided the linked CGM system is suitable for use in pregnancy.
Control-IQ+ technology is intended for single patient use and requires a prescription.
## C Special Conditions for Use Statement(s):
Rx – For Prescription Use Only
Pregnant users should only utilize this System with a CGM that is suitable for use during pregnancy.
Only use U-100 insulin analogs that have been tested and found to be compatible for use in the pump. Use of insulin with lesser or greater concentration can result in under delivery or over delivery of insulin. This can cause hypoglycemia (low BG) or hyperglycemia (high BG) events.
When the CGM reading is automatically populated into the bolus calculator, only the current CGM reading is used to calculate the correction bolus. The trend arrow is not used in the dose calculation. Speak with your healthcare provider for recommendations on how best to utilize the arrows for your correction bolus dosing.
Control-IQ+ should not be used in anyone under the age of two years old. Control-IQ+ should also not be used in patients who require less than a total daily insulin dose of 5 units per day or who weigh less than 20 pounds, as those are the required minimum values needed in order for Control-IQ technology to operate safely.
The pump is magnetic resonance (MR) unsafe. You must take off your pump and leave it outside the procedure room.
DO NOT use Dexcom CGM readings to make diabetes treatment decisions or assess glucose control when taking hydroxyurea, a medication used in the treatment of diseases including cancer and sickle cell anemia. Hydroxyurea is known to interfere with glucose readings from the Dexcom sensor. Relying on sensor glucose values while taking hydroxyurea could result in missed hypoglycemia alerts or errors in diabetes management, such as giving a higher dose of insulin than necessary to correct falsely high sensor glucose values.
## III Device Description
The Subject Device, Control-IQ+ technology (“Control-IQ+”) is a software-only device intended for the management of type 1 and type 2 diabetes mellitus. The device controls insulin delivery from a compatible alternate controller enabled insulin pump (ACE pump) based on inputs provided by a compatible integrated continuous glucose monitor (iCGM) and inputs provided by the user (e.g., carbohydrate intake, exercise, and sleep schedule). Control-IQ+ technology is meant to be installed on a compatible ACE pump.
Control-IQ+ technology has three different modes: Normal, Sleep, and Exercise. The glucose targets are not individually customizable in these modes but can change based on the mode selected. During Normal mode, Control-IQ+ technology aims to control glucose within a target
K260429 - Page 2 of 14
{2}
range of 112.5 – 160 mg/dL, during Sleep mode the target range is 112.5 – 120 mg/dL, and during Exercise mode the target range is 140 – 160 mg/dL.
Control-IQ+ technology includes an integrated feature whereby iCGM values are automatically populated into the glucose field of the integrated bolus calculator when Control-IQ+ technology is active (i.e., the device is operating in closed-loop mode). This feature is disabled when Control-IQ+ is turned off.
Control-IQ+ technology requires users to input their weight and their total daily insulin requirement, which should be established with the help of a health care provider before using the device.
## IV Substantial Equivalence Information:
A Predicate Device Name(s):
Control-IQ+ technology
B Predicate 510(k) Number(s):
K250798
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K260429 | K250798 |
| --- | --- | --- |
| Device Trade Name | Control-IQ+ technology | Same |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold. | Same |
K260429 - Page 3 of 14
{3}
| | Control-IQ+ technology is intended for single patient use and requires a prescription. | |
| --- | --- | --- |
| General Device Characteristic Differences | | |
| Intended Use Population | Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater.
Control-IQ+ technology is intended for use in pregnancy complicated by Type 1 diabetes mellitus, provided the linked CGM system is suitable for use in pregnancy. | Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater. |
V Standards/Guidance Documents Referenced:
- Special controls established under 21 CFR 862.1356
- ANSI AAMI ISO 14971:2019 Medical devices – Applications of risk management to medical devices
VI Performance Characteristics:
A. Non-Clinical Performance
Non-clinical performance was not applicable to the modifications made in this submission.
B. Clinical Studies
For the evaluation of Control-IQ+ technology in a clinical study, Control-IQ technology was installed on the t:slim X2 insulin pump with interoperable technology and paired with the Dexcom G6 iCGM. The clinical study was conducted in Canada and Australia, and the Dexcom G6 was approved for use during pregnancy in Canada for the majority of the time the study was active and in Australia for the duration of the clinical study. Most Dexcom G6 sensors were worn on the back of the upper arm. Some subjects chose to wear sensors in other locations, including the abdomen, lower back, and leg. The differences in performance between the Dexcom G6 (not cleared for use in the U.S. during pregnancy) and the Dexcom G7 are not expected to result in significantly different clinical outcomes.
Control-IQ technology does not have a pregnancy-specific design and in this study, it was adapted for use during pregnancy. Adaptations included using the device for the majority of time in sleep mode (target glucose 112.5 – 120 mg/dL) and adjusting basal rates, insulin to
K260429 - Page 4 of 14
{4}
carbohydrate ratios, and correction factors, generally making them more aggressive over time throughout pregnancy.
A total of 91 subjects were enrolled in this study, 46 in the intervention group and 45 in the standard care group. During the study, 6 subjects in the intervention group and 3 subjects in the standard care group used U-200 insulin. Since U-200 insulin products are not approved for use in insulin pumps in the U.S., the description of the demographics and results from the study, in the tables below, includes data only from participants who used U-100 insulin exclusively in the study. Two subjects began using U-200 insulin during week 34. For U-200 insulin users, transition back to U-100 insulin use occurred after delivery. Clinical outcomes were not significantly different between U-100 and U-200 users and conclusions from the study are unaffected by removal of the U-200 subjects from the analysis.
## Summary of Clinical Testing
The CIRCUIT (Closed-loop Insulin delivery by glucose Responsive Computer algorithms in Type 1 diabetes pregnancies) clinical study was conducted to evaluate the Control-IQ technology in pregnant women with type 1 diabetes.
| Study Feature | Description |
| --- | --- |
| Title | Closed-loop Insulin delivery by glucose Responsive Computer algorithms In Type 1 diabetes pregnancies (CIRCUIT) |
| Study Design | A parallel arm, open-label, multi-center, randomized controlled trial of pregnant women with pre-existing type 1 diabetes |
| Number of Sites | 15 total sites; 14 sites enrolled subjects
• 10 sites in Canada; 9 sites enrolled subjects
• 5 sites in Australia |
| Population | Pregnant women with type 1 diabetes were enrolled before 14 weeks gestation and followed until 6-weeks postpartum. Participants were randomized 1:1 to an intervention group or control group.
• N = 46 in the intervention group: t:slim X2 insulin pump with Control-IQ technology and Dexcom G6 CGM (closed-loop group)
• N=45 in the standard care group: continuation of pre-study insulin delivery method, plus use of the Dexcom G6 CGM (standard care group)
88 women completed the study through 6 weeks after delivery. 9 of the 88 participants who completed the study (6 in the intervention arm and 3 in the standard care arm) used U-200 insulin at some point during the study. Those participants have been excluded from the summary tables below. |
K260429 - Page 5 of 14
{5}
K260429 - Page 6 of 14
| | Key Inclusion Criteria
• Aged 18-45 years (inclusive)
• A diagnosis of type 1 diabetes, as defined by Diabetes Canada, for at least 12 months
• Currently on insulin therapy (≥ 3 injections or CSII)
• HbA1c ≥ 6.2% and < 10%, measured any time during pregnancy prior to enrollment
Key Exclusion Criteria
• Current treatment with drugs known to interfere with glucose metabolism, such as high dose synthetic corticosteroids
• Total daily dose of insulin < 8 or > 250 units/day at screening
• Current use of Tandem Control-IQ, DIY looping system, 670G in Auto Mode, or alternate closed-loop system |
| --- | --- |
| Protocol Overview/Synopsis | After consent was signed and participants were enrolled, participants continued to use their personal insulin delivery method (pump or MDI) during a 1-week Dexcom CGM run-in.
Participants who completed the CGM run-in by 15 weeks gestation were then randomized 1:1 to either their current insulin modality + CGM (standard care group) or the Tandem t:slim X2 insulin pump with Control-IQ technology (intervention group), and started closed-loop use (if required) by 15 weeks + 6-7 days.
Follow-up visits occurred at 16 weeks ± 7 days, 20 weeks ± 7 days, 24 weeks ± 7 days, 28 weeks ± 7 days, 32 weeks ± 7 days, 34 weeks ± 7 days, 36 weeks ± 7 days, and 6 weeks postpartum.
HbA1c was recorded at enrollment, 24 weeks, and 34 weeks. It was also recorded at 6 weeks postpartum, if the subject had it collected during a standard of care visit in the timeframe of ± 2 weeks of study visit.
Questionnaires were completed at Visit 1 (enrollment, 7-13 weeks 6 days gestation), Visit 5 (24 weeks gestation), Visit 8 (34 weeks gestation), and Visit 11 (6 weeks postpartum). The partner questionnaire was also completed at the same visits if the partner consented to participate. |
| Primary Endpoint | The primary efficacy endpoint defined by the sponsor for the study was the glycemic control as reflected by percent glucose time in range (63-140 mg/dL) per day assessed by CGM glucose measurement (16 weeks until 34 weeks 6 days |
{6}
| | gestation) compared between intervention and standard care groups.
FDA’s assessment of clinical study results was primarily based on results through the end of pregnancy and through 6 weeks postpartum. |
| --- | --- |
| Safety Results | No unexpected device-related safety issues occurred during pregnancy. Considering only subjects who used U-100 insulin during the study, there were 2 episodes of diabetic ketoacidosis in the intervention group and 1 in the standard care group. There were no severe hypoglycemia events in the intervention group and 1 in the standard care group. There were 2 miscarriages in the intervention group that were not attributed to device use. |
Device Use During the Study (N = 79 subjects, excluding those who used U-200 insulin at any point during the study)
During the CIRCUIT study, 32 of 38 subjects who were randomized into the Control-IQ arm used Control-IQ for more than 90% of the time and median was 94.7%.
| Frequency of Closed-Loop System Use in the Control-IQ Group (n = 38) | |
| --- | --- |
| Median [IQR] | 94.7 [92.3, 95.7] |
| > 90% | 32 |
| 80 – 90% | 5 |
| 70 – 80% | 0 |
| 60 – 70% | 0 |
| 50 – 60% | 0 |
| < 50% | 1 used closed-loop for 1 week |
| 0% | 0 |
Subjects were advised to use the device in sleep mode (target glucose 112.5 – 120 mg/dL) throughout the day and night. From the start of pump use through delivery, subjects used sleep mode 84.4% of time, compared to 11.7% of the time in normal mode and 3.9% of the time in exercise mode. The time spent in sleep mode decreased in the 7 days after delivery to 52.6%. The table below shows the fraction of time spent in each mode from starting device use through 6 weeks postpartum.
| Percent of Time Spent in Different Activity Modes in Control-IQ Arm | | | |
| --- | --- | --- | --- |
| | Pump Start to Delivery | Delivery to 7 days Postpartum | Delivery to 6 Weeks Postpartum |
| Normal Mode (SD) | 11.7 (18.3) | 42.8 (44.4) | 38.2 (40.1) |
| Exercise Mode (SD) | 3.9 (16.4) | 4.6 (19.1) | 9.4 (23.5) |
| Sleep Mode (SD) | 84.4 (23.0) | 52.6 (44.1) | 52.5 (40.0) |
Healthcare providers in the study also advised subjects to adjust their basal rate settings through pregnancy so that the programmed total basal settings were at least equal to the average total basal insulin use for the previous 1-2 weeks. After 20 weeks gestation, subjects
K260429 - Page 7 of 14
{7}
were advised by their providers to program their total daily basal rate settings to be 20% more than the average total basal used in the previous 1-2 weeks.
Providers also recommended that study subjects strengthen the correction factor to be at least as strong as 1620 mg/dL divided by the average total daily insulin dose in the previous 1-2 weeks. They recommended that carb ratios should be individualized but ratios at least as strong as 400 divided by the total daily insulin dose should be considered. Carb ratios/mealtime insulin doses were also strengthened as needed as pregnancy progressed beyond 20 weeks gestation. Subjects in both study arms were instructed by providers in the study to bolus 10-15 minutes before meals in the first trimester, 20-30 minutes before meals in the second trimester, and 30-45 minutes before meals in the third trimester. Both study arms also received guidance to make diet adjustments that may include meals consisting of higher fiber carbohydrates with protein to reduce the rise in blood sugar after meals, in addition to smaller snacks throughout the day to limit the amount of carbohydrate eaten during meals. Weight and total daily dose settings were not routinely updated in Control-IQ during pregnancy.
Study subjects did not enter carbs above what they were eating (i.e., "ghost carbs" were not used). If subjects wanted to manually request additional insulin, they used the built in functions of the Control-IQ system to request additional insulin correction, and if necessary to manually override the recommended insulin dose to request additional insulin. The use of manual overrides to meal boluses was rare (e.g., <1% of meal boluses), and there were no pregnancy-specific instructions provided to subjects on the use of this feature.
The study protocol included 11 scheduled visits with study staff, including visits for enrollment, randomization, delivery, 6 weeks postpartum, and 7 visits from week 16 gestation to week 36 gestation (weeks 16, 20, 24, 28, 32, 34, and 36). Nearly every study visit was completed in both arms of the study, except for missed visits due to miscarriage or early delivery before 36 weeks.
Study subjects had frequent contact with their medical clinic, with a mean of 19 contacts per participant in both arms of the study. There were 60 unscheduled research contacts in the Control-IQ group and 32 unscheduled research contacts in the Standard Care group. The reasons for regular clinic contacts and unscheduled research contacts in the Control-IQ arm were due to questions or issues regarding the devices or needing supplies and in the Standard Care arm were due to questions related to the CGM or supplies.
Participant Demographics (N = 79 subjects, excluding those who used U-200 insulin at any point during the study)
| Characteristic | Closed-Loop
n = 38 | Standard Care
n = 41 |
| --- | --- | --- |
| Age (yrs) | | |
| Mean ± SD | 31.6 ± 5 | 31.6 ± 5 |
| Race and Ethnicity – no. (%)^{a} | | |
| European or Mediterranean | 32 (87) | 32 (78) |
| Indigenous | 2 (5) | 2 (5) |
| South Asian | 1 (3) | 2 (5) |
| Other | 2 (5) | 5 (12) |
K260429 - Page 8 of 14
{8}
K260429 - Page 9 of 14
| Duration of diabetes – years | 20.4 [13.9, 25.9] | 19.5 [15.2, 24.1] |
| --- | --- | --- |
| Median [IQR] | | |
| BMI (kg/m²) | | |
| Mean ± SD | 27.5 ± 4.7 | 28.4 ± 4.3 |
| Education | | |
| Primary education (age 5-11) | 1 (2.6) | 0 (0) |
| Secondary education (age 11-17) | 5 (13.2) | 9 (22.0) |
| Trade/technical school degree | 12 (31.6) | 8 (19.5) |
| University undergraduate degree | 11 (28.9) | 14 (34.1) |
| University higher degree | 9 (23.7) | 10 (24.4) |
| Week of gestation | | |
| At enrollment, mean ± SD | 11.0 ± 2.0 | 10.9 ± 1.7 |
| At randomization, mean ± SD | 12.5 ± 2.0 | 12.2 ± 1.7 |
| At closed-loop initiation, mean ± SD | 13.3 ± 1.6 | NA |
| Pregnancy History | | |
| Primiparous – no. (%) | 13 (34.2) | 12 (29.3) |
| Previous pregnancy loss – no. (%)^{b} | 15 (39.5) | 13 (31.7) |
| Pre-pregnancy factors – no. (%) | | |
| • Folic acid supplementation | 23 (60.5) | 26 (63.4) |
| • Alcohol consumption (current preg.) | 4 (10.5) | 5 (12.2) |
| • Cigarette smoking (current preg.) | 3 (7.9) | 2 (4.9) |
| • Preconception care^{c} | 19 (50.0) | 18 (43.9) |
| Glycated hemoglobin level (%) during early pregnancy | | |
| 6.2 to < 6.5 – no. (%) | 2 (5.3) | 3 (7.3) |
| 6.5 to < 7.0 – no. (%) | 16 (42.1) | 14 (34.1) |
| 7.0 to < 7.5 – no. (%) | 9 (23.7) | 9 (22.0) |
| 7.5 to < 8.0 – no. (%) | 4 (10.5) | 4 (9.8) |
| ≥ 8.0 to 10 – no. (%) | 7 (18.4) | 11 (26.8) |
| Mean HbA1c ± SD | 7.29 ± 0.96 | 7.47 ± 1.01 |
| Range | 6.2 – 9.6 | 6.3 – 9.6 |
| Using CGM prior to study – no. (%) | | |
| All brands | 32 (84.2) | 37 (90.2) |
| Dexcom | 19 (50.0) | 13 (31.7) |
| Medtronic | 1 (2.6) | 5 (12.2) |
| Abbott | 12 (31.6) | 19 (46.3) |
| Insulin delivery – no. (%) | | |
| Multiple daily injections | 19 (50.0) | 21 (51.2) |
| Insulin pump (open-loop) | 19 (50.0) | 20 (48.8) |
| Total daily insulin – units/day | | |
| Mean ± SD | 47.4 ± 17.5 | 51.5 ± 22.3 |
| Total daily insulin – units/kg/day | | |
| Mean ± SD | 0.6 ± 0.2 | 0.7 0.6 ± 0.2 0.2 |
$^{a}$ Participants self-reported race on a fixed category questionnaire with an optional write-in answer. For the Closed-Loop group, 37 participants reported results, where other race or ethnicity groups included 1 participant who self-identified as East Asian and 1 who self-identified as Jewish. For the Standard Care group, 41 participants reported results, where 2 self-identified as Black, 2 as Middle Eastern, and 1 as multiethnic.
$^{b}$ Pregnancy losses include previous miscarriages and previous terminations.
$^{c}$ Preconception care means that participants were seen in a diabetes in pregnancy clinic prior to pregnancy.
{9}
Overall Glycemic Outcomes (N = 79 subjects, excluding those who used U-200 insulin at any point during the study)
Glycemic control as reflected by percent glucose time in range (63-140 mg/dL) per day assessed by CGM glucose measurement (16 weeks until end of pregnancy) was compared between Control-IQ and Standard Care groups. For subjects who only used U-100 insulin during the study, the mean time in pregnancy-specific glucose range (63-140 mg/dL) increased from 55.5% at baseline to 67.2% in the Control-IQ arm and from 48.4% at baseline to 51.8% in the Standard Care arm from 16 weeks to end of pregnancy. The adjusted group difference [Control-IQ minus Standard Care] was 12.6 percentage points (95% CI 9.4, 15.9); p <0.001) higher in the Control-IQ group. The results presented in the table below show the glycemic control from baseline to end of pregnancy.
| Outcome | Baseline | | Intervention (16 weeks to end of pregnancy) | | Adjusted Treatment Difference (95% CI) |
| --- | --- | --- | --- | --- | --- |
| | Control-IQ (n = 38) | Standard Care (n = 41) | Control-IQ (n = 38) | Standard Care (n = 41) | |
| % time in glucose range 63-140 mg/dL | 55 ± 11.1 | 48.4 ± 15.9 | 67.2 ± 8.9 | 51.8 ± 13.3 | 12.6 (9.4, 15.9) |
| % time in glucose range < 63 mg/dL | 3.4 ± 3.5 | 2.9 ± 3.2 | 1.2 ± 1.0 | 2.1 ± 1.8 | -1.1 (-1.6, -0.6) |
| % time in glucose range > 140 mg/dL | 41.1 ± 12.6 | 48.7 ± 16.3 | 31.6 ± 8.9 | 46.0 ± 14.0 | -11.3 (-14.6, -7.9) |
| Mean glucose level (mg/dL) | 139.4 ± 18.8 | 152.1 ± 30.1 | 130.0 ± 9.2 | 146.3 ± 22.1 | -9.1 (-13.6, -4.6) |
| Glucose SD (mg/dL) | 51.5 ± 11.6 | 56.2 ± 13.6 | 38.9 ± 6.7 | 51.7 ± 12.2 | -9.9 (-12.7, -7.1) |
| Glucose coefficient of variation % | 37.0 ± 6.2 | 36.9 ± 5.6 | 29.8 ± 3.7 | 35.1 ± 4.2 | -5.3 (-6.6, -4.0) |
Intrapartum and Postpartum Glycemic Outcomes (excluding those who used U-200 insulin at any point during the study)
During intrapartum and postpartum periods, subjects dealt with the challenges of laboring and breastfeeding. 34 of the 38 subjects continued to use their pump through delivery. Two subjects were not allowed to continue closed-loop therapy during their Caesarean deliveries due to hospital policy, one subject stopped closed-loop therapy 11 days prior to delivery, and one subject stopped closed-loop therapy 1 week after initiating it in pregnancy because of infusion site pain. Subjects who used the device in the 24-hour period leading up to delivery had 79.3% time in pregnancy-specific glucose range (63-140 mg/dL) compared to 63.6% of
K260429 - Page 10 of 14
{10}
subjects in the standard care group. At 1 week postpartum, the Control-IQ arm had 86.2% time in range (70-180 mg/dL) compared to 74.3% in the Standard Care arm, and from delivery to 6 weeks postpartum, the Control-IQ arm had 83.3% time in range compared to 65.1% in the Standard Care arm.
| | Control-IQ | Standard Care | Adjusted Treatment Difference (95% CI) |
| --- | --- | --- | --- |
| 24 hours up to delivery | | | |
| n | 34 | 39 | |
| % time in glucose range 63-140 mg/dL | 79.3 ± 14.9 | 63.6 ± 19.8 | 13.4 (5.1, 21.7) |
| % time in glucose range < 63 mg/dL | 1.0 ± 1.8 | 1.1 ± 2.1 | -0.1 (-1.0, 0.7) |
| % time in glucose range > 140 mg/dL | 19.7 ± 14.4 | 35.3 ± 19.7 | -13.1 (-21.3, -5) |
| Mean glucose level (mg/dL) | 118.2 ± 11.9 | 130.5 ± 20.4 | -9.2 (-17.1, -1.3) |
| Glucose SD (mg/dL) | 29.5 ± 12.5 | 34.7 ± 10.9 | -3.9 (-9.4, 1.6) |
| Glucose coefficient of variation % | 24.6 ± 9.0 | 26.3 ± 5.9 | -1.1 (-4.4, 2.2) |
| First Week Postpartum | | | |
| n | 34 | 40 | |
| % time in glucose range 70-180 mg/dL | 86.2 ± 7.1 | 74.3 ± 14.0 | 10.5 (5.3, 15.6) |
| % time in glucose range < 70 mg/dL | 1.5 ± 1.8 | 3.3 ± 2.3 | -2.2 (-3.1, -1.3) |
| % time in glucose range > 180 mg/dL | 12.3 ± 7.5 | 22.4 ± 14.4 | -8.2 (-13.6, -2.8) |
| Mean glucose level (mg/dL) | 134.6 ± 13.7 | 146.3 ± 21.9 | -7.9 (-16.1, 0.4) |
| Glucose SD (mg/dL) | 38.2 ± 7.9 | 50.6 ± 14.2 | -10.6 (-15.8, -5.4) |
| Glucose coefficient of variation % | 28.4 ± 4.9 | 34.3 ± 6.6 | -5.9 (-8.6, -3.2) |
| From Delivery to 6 Weeks Postpartum | | | |
| n | 34 | 40 | |
| % time in glucose range 70-180 mg/dL | 83.3 ± 5.1 | 65.1 ± 13.9 | 12 (7.9, 16.2) |
| % time in glucose range < 70 mg/dL | 1.4 ± 1.1 | 2.9 ± 2.0 | -1.6 (-2.2, -1.0) |
| % time in glucose range > 180 mg/dL | 15.3 ± 5.5 | 32.0 ± 14.2 | -10.2 (-14.4, -5.9) |
| Mean glucose level (mg/dL) | 139.5 ± 9.1 | 162.6 ± 24.8 | -16.6 (-21.3, -12.0) |
| Glucose SD (mg/dL) | 41.2 ± 6.0 | 61.1 ± 14.5 | -17.1 (-21.4, -12.8) |
| Glucose coefficient of variation % | 29.5 ± 3.8 | 37.2 ± 5.2 | -7.6 (-9.6, -5.7) |
K260429 - Page 11 of 14
{11}
HbA1c Outcomes (N = 79 subjects, excluding those who used U-200 insulin at any point during the study)
For subjects who only used U-100 insulin during the study, the HbA1c decreased from 7.3% at baseline to 5.9% at 24 ± 4 weeks gestation and to 6.2% at 34 weeks ± 4 weeks gestation in the Control-IQ arm and from 7.5% at baseline to 6.5% at 24 weeks ± 4 weeks gestation and to 6.6% at 34 weeks ± 4 weeks gestation in the Standard Care arm.
| | Baseline | | Intervention (24 weeks ± 4 gestational weeks gestation) | | Adjusted Treatment Difference (95% CI) |
| --- | --- | --- | --- | --- | --- |
| | Control-IQ (n = 38) | Standard Care (n = 41) | Control-IQ (n = 38) | Standard Care (n = 41) | |
| HbA1c (%) | 7.3 ± 0.9 | 7.5 ± 1.0 | 5.9 ± 0.5 | 6.5 ± 0.9 | -0.5 (-0.7, -0.2) |
| | Baseline | | Intervention (34 weeks ± 4 gestational weeks gestation) | | Adjusted Treatment Difference (95% CI) |
| | Control-IQ (n = 38) | Control-IQ (n = 38) | Control-IQ (n = 38) | Standard Care (n = 41) | |
| HbA1c (%) | 7.3 ± 0.9 | 7.5 ± 1.0 | 6.2 ± 0.5 | 6.6 ± 0.9 | -0.3 (-0.6, -0.1) |
Safety Results (N = 82 subjects, excluding those who used U-200 insulin at any point during the study)
Adverse events were recorded after randomization through 6 weeks postpartum for all subjects. The table below provides the types of adverse events by treatment arm (n = 82) post-randomization through 6 weeks postpartum, excluding the 9 participants who used U-200 insulin during the study.
| | Number of Events | |
| --- | --- | --- |
| | Control-IQ (n = 40 randomized participants) | Standard Care (n = 42 randomized participants) |
| Total number of adverse events | 47 | 56 |
| Severe Hypoglycemia (SH) Eventsa | 0 | 1 |
| Diabetic Ketoacidosis (DKA) Eventsb | 2 | 1 |
| Other Serious Adverse Events (SAES) | 1 | 7 |
| Anaphylaxis | 0 | 1 |
| Endometritis from retained placenta | 0 | 1 |
| Headache | 1 | 0 |
| Nausea and Vomiting | 0 | 1 |
| Perianal abscess | 0 | 1 |
| Persistent episiotomy wound infection | 0 | 1 |
| Pneumonia/Influenza A | 0 | 1 |
K260429 - Page 12 of 14
{12}
| Viral gastroenteritis | 0 | 1 |
| --- | --- | --- |
| Non-serious Adverse Events | | |
| Hyperglycemia with or without Ketosis | 1 | 0 |
| Related to Study Device | | |
| Hyperglycemia with or without Ketosis | 3 | 0 |
| Not Related to Study Device | | |
| Hypoglycemia (non-severe) | 1 | 0 |
| Other reported adverse events | 39 | 47 |
$^a$A severe hypoglycemic event is defined as a hypoglycemic event that a) required assistance of another person due to altered consciousness, and b) required another person to actively administer carbohydrate, glucagon, or other resuscitative actions. One episode occurred in the Standard Care group 12 days postpartum.
$^b$DKA events meeting DCCT criteria.
## Maternal and Neonatal Outcomes
Maternal outcomes, including rates of Cesarean delivery, maternal weight gain, and new or worsening hypertension were not significantly different between the intervention and standard care arms. Neonatal outcomes, including small or large for gestational age, shoulder dystocia, hypoglycemia treated with dextrose hyperbilirubinemia, or neonatal intensive care $\geq 1$ day were also not significantly different between the intervention and standard care arms. Overall, the maternal and neonatal outcomes were similar in both groups but favored the intervention group using Control-IQ.
## C. Other Supportive Device Performance Characteristics Data
### Human Factors
Human factors validation studies were provided in K2423823, K232382, K200467, and DEN190034. No differences in critical tasks were identified between adults with type 1 diabetes and pregnant women with type 1 diabetes. People with type 1 diabetes do not face new critical tasks or perform new critical tasks with this device as a result of their pregnancy status. Device use during the clinical study was generally the same as how it is currently used on the market, and adaptations that were recommended during the clinical study are part of normal device use.
### Software
The Control-IQ+ technology in this submission does not include any software changes compared to the predicate device, Control-IQ+ technology (K250798). The ACE pump software versions 7.4.3 and 7.6 (cleared in K232382) and Control-IQ (cleared in K200467) were used in the CIRCUIT study. However, when this device is marketed, it will be for ACE pump software version 7.10.3 and Control-IQ+ (first cleared K232382). The main difference in the iAGC from the prior versions of the pump to the latest ACE pump software version 7.10.3 is the addition of the Control-IQ+ feature set. The differences between Control-IQ and Control IQ+ have not changed since it was integrated into the ACE pump software beginning with version pump software version 7.9. Differences between Control-IQ and Control-IQ+ include allowed minimum and maximum weight and total daily insulin, correction factor range, maximum extended bolus duration, and basal rate clipping. These differences did not affect the use of this device during the CIRCUIT clinical study.
K260429 - Page 13 of 14
{13}
K260429 - Page 14 of 14
## Cybersecurity
The Control-IQ+ technology in this submission does not include any changes to the Software, Firmware, Cybersecurity & Interoperability compared to the predicate device Control-IQ+ technology (K250798). Additionally, there are no significant alterations to the design of security controls or the operational functionality of Control-IQ+ technology in comparison to the predicate device Control-IQ+ technology (K250798).
## VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.