K260428 · Boston Scientific Corporation · ODG · Jun 17, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K260428
Device Name
RIVOS™ EUS Access Device (M00553900)
Applicant
Boston Scientific Corporation
Product Code
ODG · Gastroenterology, Urology
Decision Date
Jun 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RIVOS EUS Access Device is used to access and electrosurgically puncture the transgastric or transduodenal wall into the following: the intra- or extra-hepatic bile ducts, or a pancreatic pseudocyst. The device is intended for use through the accessory channel of an ultrasound endoscope.
Device Story
Sterile, single-use ultrasound-guided access device; used in clinical settings by physicians. Device consists of sharp within access cannula for initial puncture; after sharp removal, guidewire is placed. Flexible catheter with electrosurgical tip is advanced over cannula/guidewire to create fistula using RF energy from compatible electrosurgical generators (e.g., ERBE VIO). Echogenic features allow real-time ultrasound visualization. Device attaches to curvilinear array echoendoscope accessory channel. Output is a created fistula allowing passage of accessory devices. Benefits include minimally invasive access to biliary/pancreatic structures for therapeutic intervention.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing and GLP ex-vivo thermal effects study comparing RIVOS to predicate devices across various generator power settings.
Technological Characteristics
Sterile, single-use catheter with electrosurgical cutting electrode. Features echogenic distal end for ultrasound visibility and radiopaque markers. Compatible with 0.025" guidewires and monopolar electrosurgical generators (3mm female plug). Complies with IEC 60601-1, IEC 60601-1-6, IEC 60601-2-2, IEC 60601-2-18, and IEC 60601-1-2. Adjustable sheath length and locking mechanism.
Indications for Use
Indicated for patients requiring transgastric or transduodenal access to intra- or extra-hepatic bile ducts or pancreatic pseudocysts via ultrasound endoscope.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
EchoTip Ultra High Definition Ultrasound Access Needle (K092359)
{0}
June 17, 2026
Boston Scientific Corporation
Yingying Gao
Fellow, Regulatory Affairs
100 Boston Scientific Way
Marlboro, Massachusetts 01752
Re: K260428
Trade/Device Name: RIVOS™ EUS Access Device
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: ODG, KNS
Dated: May 15, 2026
Received: May 15, 2026
Dear Yingying Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260428 - Yingying Gao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260428 - Yingying Gao
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee, Ph.D.,
MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260428 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| RIVOS™ EUS Access Device | | |
| Please provide your Indications for Use below. | | ? |
| The device is used to access and electrosurgically puncture the transgastric or transduodenal wall into the following: the intra- or extrahepatic bile ducts, or a pancreatic pseudocyst. The device is intended for use through the accessory channel of an ultrasound endoscope. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260428
Page 1 of 6
# 510(k) SUMMARY
1. Submitter:
Boston Scientific Corporation
100 Boston Scientific Way
Marlborough, MA 01752
Contact: Yingying Gao
Fellow, Regulatory Affairs
508-683-4356
yingying.gao@bsci.com
Secondary Contact: Ashley Santos
Regulatory Affairs Manager
508-683-4359
Ashley.santos@bsci.com
Date Prepared: February 9, 2026
2. Device:
| **Trade Name:** | RIVOS™ EUS Access Device |
| --- | --- |
| **Common Name:** | Endoscope and accessories |
| **Classification Name:** | Endoscopic Ultrasound System, Gastroenterology-Urology |
| **Regulation Number:** | 876.1500 |
| **Product Code:** | ODG, KNS |
| **Classification:** | Class II |
3. Predicate Devices:
Predicate device:
| **Trade Name:** | EchoTip Ultra High Definition Ultrasound Access Needle |
| --- | --- |
| **Common Name:** | Gastroenterology-urology biopsy instrument |
| **Classification Name:** | Biopsy Needle |
| **Manufacturer:** | Cook Medical LLC |
Traditional 510(k) Premarket Notification, RIVOS EUS Access Device
510(k) Summary Page 1 of 6
{5}
K260428
Page 2 of 6
| Clearance Number: | K092359 |
| --- | --- |
| Regulation Number: | 876.1075 |
| Product Code: | FCG |
| Classification: | Class II |
And
Predicate device:
| Trade Name: | Cystotome Cystoenterostomy Needle Knife |
| --- | --- |
| Common Name: | Endoscopic electrosurgical unit and accessories |
| Classification Name | Unit, electrosurgical, endoscopic (with or without accessories) |
| Manufacturer: | Cook Medical LLC |
| Clearance Number: | K022595 |
| Regulation Number: | 876.4300 |
| Product Code: | KNS |
| Classification: | Class II |
#### 4. Device Description:
Device Name: RIVOS™ EUS Access Device
The proposed RIVOS EUS Access Device is a sterile, single use ultrasound guided access device. The device comprises of a sharp contained within an access cannula, which facilitates puncture and access to a target structure. After sharp removal, a guidewire is placed through the access cannula and passed through to the desired location. The access cannula is contained within a flexible catheter with an electrosurgical tip that is advanced over the access cannula and guidewire for fistula creation.
The device is designed to be attached to the biopsy channel of a Curvilinear Array (CLA) Echoendoscope with a standard luer connection and delivered into the digestive tract. The device catheter length can be adjusted to accommodate different model echoendoscopes.
The RIVOS device has echogenic (visible under ultrasound) features at the distal end to facilitate real time visualization of the device under ultrasound.
The RIVOS device can be used with compatible BF Electrosurgical Generators: ERBE VIO® 3 or ERBE VIO® 300D and a Monopolar endoscopic cable with 3mm female plug.
Traditional 510(k) Premarket Notification, RIVOS EUS Access Device
510(k) Summary Page 2 of 6
{6}
K260428
Page 3 of 6
## 5. Intended use/ Indications for Use:
The RIVOS EUS Access Device is used to access and electrosurgically puncture the transgastric or transduodenal wall into the following: the intra- or extra-hepatic bile ducts, or a pancreatic pseudocyst. The device is intended for use through the accessory channel of an ultrasound endoscope.
The Indications for Use statement of the RIVOS EUS Access Device is not identical to the predicate devices; however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicates. Both the proposed and predicate devices have the same intended use for accessing and puncturing pancreaticobiliary anatomies or lesions from the gastrointestinal tract.
## 6. Technological Characteristics:
The RIVOS EUS Access Device performs initial puncture to access the target structure, allows for injection of contrast for visualization, facilitates guidewire passage and creation of a fistulous track using radio frequency (RF) energy. During puncture, the sharp and access cannula are advanced together as a single unit under ultrasound guidance. After access to the target structure is obtained, the sharp is removed from the device, leaving the access cannula within the target anatomy. A guidewire is inserted into the access cannula. Joined to the distal end of the catheter is an electrocautery tip, which contains a cutting electrode. The electrode can be activated by an electrosurgical generator and the electrocautery tip is advanced over the access cannula, creating a fistula for the passage of accessory devices.
It has the following technological characteristics:
- • Secure scope attachment
- • Adjustable sheath length and sheath locking mechanism
- • Device Scope Passability
- • Adjustable Device Extension Length
- • Puncture into target structure
- • Access a target structure
- • Ability to maintain access
- • Aspiration capability
- • Contrast/Saline Injection Capability
- • Guidewire Compatibility
Traditional 510(k) Premarket Notification, RIVOS EUS Access Device
510(k) Summary Page 3 of 6
{7}
K260428
Page 4 of 6
- Device visibility under ultrasound
- Device visibility under fluoroscopy
- Fistula Creation with Electrocautery
- Ergonomic handle design
The RIVOS EUS Access Device shares similar design characteristics with its predicate device Cook EchoTip Ultra High Definition Ultrasound Access Needle. Both devices incorporate adjustable lengths of sheath for Ultrasound endoscope attachment and possess similar puncture component configurations. For the RIVOS device, initial puncture is achieved by advancing the sharp and access cannula into the target structure, whereas the Cook EchoTip Ultra needle utilizes advancement of its stylet and needle for puncture. Both devices offer echogenic visibility under ultrasound imaging, facilitating accurate initial puncture.
There are differences between the RIVOS device and Cook EchoTip Ultra needle. The minimum endoscope working channel for the RIVOS device is 3.7 mm, which is greater than the 2.0 mm minimum endoscope working channel specified for the Cook EchoTip Ultra needle. The RIVOS device is compatible with a 0.025" guidewire, while the Cook EchoTip Ultra needle accommodates a larger 0.035" guidewire. Additionally, the RIVOS device features a shorter extension length. These distinctions do not raise concerns regarding safety or effectiveness.
The RIVOS EUS Access Device exhibits design characteristics similar to its predicate device, the Cook Cystotome Cystoenterostomy Needle Knife. Both devices employ electrocautery for fistula creation and are compatible with monopolar electrosurgical generator(s) and 3 mm female plug monopolar endoscopic cables. Fistula formation is achieved via the diathermic ring on the outer catheter.
There are differences between the RIVOS device and Cook Cystotome. The RIVOS device features an 8Fr outer sheath with electrosurgical capabilities, whereas the Cook Cystotome utilizes a 10Fr outer sheath for cautery. In addition, the RIVOS device claims compatibility with multiple monopolar generators. These variations in outer sheath size and generator(s) compatibility do not raise questions of safety and effectiveness. A GLP Ex-Vivo Thermal Effects study of RIVOS device and Cook Cystotome device comparison was completed with all compatible generators on both high and low power settings and concluded a substantial equivalence between these two devices.
Traditional 510(k) Premarket Notification, RIVOS EUS Access Device
510(k) Summary Page 4 of 6
{8}
K260428
Page 5 of 6
# 7. Performance Data:
# Biocompatibility Testing:
The biocompatibility evaluation for the RIVOS EUS Access Device conducted in accordance with the FDA Guidance “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process” Guidance for Industry and Food and Drug Administration Staff” (Issued on: September 8, 2023.), and ISO 10993-1 “ Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process” as recognized by FDA.
The RIVOS EUS Access Device is considered Surface Device: Breached or Compromised Surfaces Contact for a duration of less than 24 hours. The following tests were performed:
- Cytotoxicity
- Sensitization
- Irritation/intracutaneous reactivity
- Material mediated pyrogenicity
- Acute systemic toxicity
# Electrical safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the RIVOS EUS Access Device. The RIVOS device, tested together with compatible generator(s) and Active Cord, complies with the IEC 60601-1, IEC 60601-1-6, IEC 60601-2-2 and IEC 60601-2-18 standards for safety. Additional Confirmatory Radiated Emissions testing was conducted to confirm electromagnetic compliance with the established test thresholds per IEC 60601-1-2 for EMC.
# Performance Testing Summary:
Non-clinical performance bench testing was successfully completed to establish substantial equivalence between the proposed RIVOS EUS Access Device and its predicate devices Cook EchoTip Ultra Access Needle and Cook Cystotome. The following tests were conducted on the RIVOS EUS Access Device:
Traditional 510(k) Premarket Notification, RIVOS EUS Access Device
510(k) Summary Page 5 of 6
{9}
K260428
Page 6 of 6
- 1 Luer Interface
- 2 Scope Attachment Torsional Strength
- 3 Handle Tensile
- 4 Device Passability
- 5 Device Outer Diameter
- 6 Device Tip Flexibility
- 7 Device Extension Length
- 8 Sharp to Access Cannula Protrusion Length
- 9 Access Cannula Stiffness
- 10 Injection Force
- 11 Device Echogenicity
- 12 Device Radiopacity
- 13 Cautery Tip Dilation
- 14 Puncture Trajectory
- 15 Access Cannula Axial Compression
- 16 Image Quality
- 17 Access Cannula Kink Resistance
- 18 Access Cannula Lubricity
- 19 Device Durability
- 20 Sharp Removal Force
- 21 Puncture Force
- 22 Dilation Actuation Force
- 23 Handle Rotation
- 24 Adjustable Working Length
- 25 Guidewire Passability Force
A GLP Ex-Vivo Thermal Effects study of RIVOS device and Cook Cystotome device comparison was completed at all compatible generators on both high and low power settings and concluded a substantial equivalence between these two devices.
## 8. Conclusion:
Boston Scientific has demonstrated that proposed RIVOS EUS Access Device is substantially equivalent to the currently marketed Cook EchoTip Ultra High Definition Ultrasound Access Needle (K092359) and Cook Cystotome Cystoenterostomy Needle Knife (K022595).
Traditional 510(k) Premarket Notification, RIVOS EUS Access Device
510(k) Summary Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.