← Product Code [FYA](/productcode/FYA) · K260407

# Principle & Will ANSI/AAMI PB 70 (Sterile/Non-sterile) Level 3 Surgical Gowns (Reinforced) and Principle & Will ANSI/AMMI PB70 (Sterile/Non-sterile) Level 4 Surgical Gowns/Isolation Gowns (K260407)

_Principle & Will Co., Ltd. · FYA · Aug 26, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K260407

## Device Facts

- **Applicant:** Principle & Will Co., Ltd.
- **Product Code:** [FYA](/productcode/FYA.md)
- **Decision Date:** Aug 26, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** PCCP

## Indications for Use

Level 3: "The Principle & Will ANSI/AAMI PB 70 Sterile Level 3 Surgical Gown (Reinforced) is a sterile, single-use surgical apparel intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate matter. The non-sterile Level 3 Surgical Gown (Reinforced) is intended for distribution to re-packager/re-labeler or sterilizer establishments for sterilization using a validated ethylene oxide sterilization method in accordance with ISO 11135 prior to marketing to end users." The Principle & Will ANSI/AAMI PB 70 Sterile/Non-sterile Level 3 Surgical Gowns (Reinforced) meet the ANSI/AAMI PB70 Level 3 liquid barrier performance requirements, as well as the other performance requirements of ASTM F2407_F2407M-23a Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities. Level 4: "The Principle & Will ANSI/AAMI PB 70 Sterile Level 4 Surgical Gown/Isolation Gown is a sterile, single-use surgical apparel intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate matter. The non-sterile Level 4 Surgical Gown/Isolation Gown is intended for distribution to re-packager/re-labeler or sterilizer establishments for sterilization using a validated ethylene oxide sterilization method in accordance with ISO 11135 prior to marketing to end users." The Principle & Will ANSI/AAMI PB 70 Sterile/Non-sterile Level 4 Surgical Gown/Isolation Gown meets the ANSI/AAMI PB 70 Level 4 liquid barrier and ASTM F1671 viral barrier performance requirements as well as the other performance requirements of ASTM F2407_F2407M-23a Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities and ASTM F3352 F3352M-23 Standard Specification for Isolation Gowns Intended for Use in Healthcare Facilities.

## Device Story

Single-use surgical/isolation gowns; provide barrier protection against microorganisms, body fluids, and particulate matter; worn by operating room personnel during surgical procedures. Level 3 gowns meet ANSI/AAMI PB70 Level 3 requirements; Level 4 gowns meet ANSI/AAMI PB70 Level 4 and ASTM F1671 viral barrier requirements. Non-sterile versions intended for further processing/sterilization via ethylene oxide (ISO 11135) before end-user distribution.

## Clinical Evidence

No clinical data; bench testing only per ASTM F2407/F2407M-23a, ASTM F3352/F3352M-23, ANSI/AAMI PB70, and ASTM F1671 standards.

## Technological Characteristics

Single-use surgical/isolation gowns; Level 3 and Level 4 barrier performance per ANSI/AAMI PB70; compliant with ASTM F2407/F2407M-23a and ASTM F3352/F3352M-23; non-sterile versions require ethylene oxide sterilization per ISO 11135.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD & DRUG
ADMINISTRATION

August 26, 2026

Principle & Will Co., Ltd.

% Jing Ke

Regulatory Affairs Specialist

Principle & Will Biotech (PingHu) Co., Ltd.

Contact Address

Re: K260407

Trade/Device Name: Principle & Will ANSI/AAMI PB 70 (Sterile/Non-sterile) Level 3 Surgical Gowns (Reinforced) and Principle & Will ANSI/AMMI PB70 (Sterile/Non-sterile) Level 4 Surgical Gowns/Isolation Gowns

Regulation Number: 21 CFR 878.4040

Regulation Name: Surgical Apparel

Regulatory Class: Class II

Product Code: FYA, FYC

Dated: July 31, 2026

Received: July 31, 2026

Dear Jing Ke:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993

www.fda.gov

{1}

K260407 - Jing Ke

Page 2

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP), titled "Predetermined Change Control Plans (PCCP) – Shelf Life Claim Addition" as version "A/0." Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device

{2}

K260407 - Jing Ke

Page 3

Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Bifeng Qian, Ph.D., M.D.

Assistant Director

DHT4C: Division of Infection

Control Devices

OHT4: Office of Surgical and

Infection Control Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# **Indications for Use**

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K260407

Device Name

Principle & Will ANSI/AAMI PB 70 (Sterile/Non-sterile) Level 3 Surgical Gowns (Reinforced) and Principle & Will ANSI/AMMI PB70 (Sterile/Non-sterile) Level 4 Surgical Gowns/Isolation Gowns

Indications for Use (Describe)

Level 3: "The Principle & Will ANSI/AAMI PB 70 Sterile Level 3 Surgical Gown (Reinforced) is a sterile, single-use surgical apparel intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate matter. The non-sterile Level 3 Surgical Gown (Reinforced) is intended for distribution to re-packager/re-labeler or sterilizer establishments for sterilization using a validated ethylene oxide sterilization method in accordance with ISO 11135 prior to marketing to end users." The Principle & Will ANSI/AAMI PB 70 Sterile/Non-sterile Level 3 Surgical Gowns (Reinforced) meet the ANSI/AAMI PB70 Level 3 liquid barrier performance requirements, as well as the other performance requirements of ASTM F2407_F2407M-23a Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities.

Level 4: "The Principle & Will ANSI/AAMI PB 70 Sterile Level 4 Surgical Gown/Isolation Gown is a sterile, single-use surgical apparel intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate matter. The non-sterile Level 4 Surgical Gown/Isolation Gown is intended for distribution to re-packager/re-labeler or sterilizer establishments for sterilization using a validated ethylene oxide sterilization method in accordance with ISO 11135 prior to marketing to end users." The Principle & Will ANSI/AAMI PB 70 Sterile/Non-sterile Level 4 Surgical Gown/Isolation Gown meets the ANSI/AAMI PB 70 Level 4 liquid barrier and ASTM F1671 viral barrier performance requirements as well as the other performance requirements of ASTM F2407_F2407M-23a Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities and ASTM F3352 F3352M-23 Standard Specification for Isolation Gowns Intended for Use in Healthcare Facilities.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)

☑ Over-The-Counter Use (21 CFR 801 Subpart C)

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

---

**Source:** [https://fda-staging.innolitics.com/device/K260407](https://fda-staging.innolitics.com/device/K260407)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
