PDX nova is a portable dental X-ray system that captures radiographic images for dental diagnosis using intra-oral imaging receptors. Only trained and qualified dental practitioner or radiologist shall use PDX nova to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
Device Story
Handheld dental X-ray system; generates ionizing radiation to expose intra-oral image receptors. Used in dental operatory by dentists/radiologists for diagnostic imaging of teeth, jaws, and oral structures. Device features internal timer, 70 kVp/3 mA fixed output, and backscatter radiation shield. Operator selects exposure time via push buttons; device emits X-rays to capture images on external sensors. Output aids clinical diagnosis and treatment planning. Benefits include portability and diagnostic utility for dental pathologies.
Clinical Evidence
Bench testing performed per consensus standards (IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-65, 60601-1-6, 62366-1). Document claims results from both bench and clinical testing confirmed diagnostic image quality substantially equivalent to the predicate device; however, no specific clinical study data, sample sizes, or quantitative metrics (e.g., sensitivity/specificity) are provided in the summary.
Technological Characteristics
Handheld X-ray system; 70 kVp/3 mA fixed output; 0.4 mm focal spot; 1.5 mmAl total filtration; 200 mm source-to-skin distance. Powered by 14.8 VDC Li-Po battery. Includes backscatter shield. Complies with 21 CFR 1020.30/1020.31 and IEC 60601 series standards. Software validation per IEC 62304.
Indications for Use
Indicated for dental radiographic imaging in adults and children for diagnosis of diseases of teeth, jaws, and oral structures. For use by trained dental practitioners or radiologists.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 27, 2026
PDX Co., Ltd.
% Lee Strong
Quality Manager
Strong Systems, Inc.
803 John Anderson Dr.
ORMOND BEACH, FL
Re: K260367
Trade/Device Name: Portable Dental X-ray System Model PDX nova
Regulation Number: 21 CFR 872.1800
Regulation Name: Extraoral source x-ray system
Regulatory Class: Class II
Product Code: EHD
Dated: February 1, 2026
Received: April 24, 2026
Dear Lee Strong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260367 - Lee Strong
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260367 - Lee Strong
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
GABRIELA
M. RODAL -S
Digitally signed
by GABRIELA M. RODAL -S
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260367 | ? |
| Please provide the device trade name(s). | | ? |
| Portable Dental X-ray System Model PDX nova | | |
| Please provide your Indications for Use below. | | ? |
| PDX nova is a portable dental X-ray system that captures radiographic images for dental diagnosis using intra-oral imaging receptors. Only trained and qualified dental practitioner or radiologist shall use PDX nova to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260367
PDX nova
510(k) Summary
**Submitter**
PDX Co., Ltd.
Room 3209, 2F, 40 Simindae-ro 365beongil, Dongan-gu,
Anyang-si, Gyeonggi-do 14957, Republic of Korea
Tel: +82-(0)32-8068-7717
Fax: +82-(0)70-7585-7719
Contact: Hyuntaek Shim, Quality Manager
Date: January 26, 2025
**Device Classification**
Trade Names: Portable Dental X-ray System Model PDX nova
Common name: Portable X-ray System
Regulation Name: Extra-oral Source X-ray System
Regulation Number: 21 CFR 872.1800
Product code: EHD
Common Name: Unit, X-ray, Extraoral with Timer
Regulatory Class: II
510k Review Panel: Radiology
Submission Type: 510(k)
**Predicate Device**
The following predicate is a legally marketed, post-amendment device:
510(k) Number K221233
Submitters: ECOTRON Co., Ltd
Date Cleared: June 27, 2022
Device Name: DT-703
Regulation Name: Extra-oral Source X-ray
Regulation Number: 21 CFR 872.1800
Product Code: EHD
Common Name: Unit, X-ray, Extraoral with Timer
Regulatory Class: II
510k Review Panel: Radiology
Submission Type: 510(k)
{5}
2
# Device Description
PDX nova is a handheld x-ray device designed for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structure by exposing an x-ray image receptor to ionizing radiation. The x-ray source, a tube, is located inside the handheld device. The device is intended to be used with compatible intraoral image receptors.
This device is used in general dentistry and is supplied with an internal timer to control the duration of the x-ray source to the patient. The choice of an x-ray generator is a matter of functional utility in the dental operatory and personal preference by the medical professional.
# Indications for Use
PDX nova is a portable dental X-ray system that captures radiographic images for dental diagnosis using intra-oral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use PDX nova to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
# Comparison of Characteristics with Predicate
The following table compares technological and other characteristics of the subject and predicate devices.
| Characteristic | PDX nova (Subject Device) | ECOTRON DT-703 (K221233) | Comparison |
| --- | --- | --- | --- |
| Indications for Use | PDX nova is a portable dental X-ray system that captures radiographic images for dental diagnosis using intra-oral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use PDX nova to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children. | The DT-703 is a portable dental X-ray system that captures radiographic images for dental diagnosis using intra-oral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use DT-703 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children. | Same |
| Classification & Product Code | 872.1800; EHD | 872.1800; EHD | Same |
| Size: Body | 370 x 322 x 176 mm | 259(L) x 277(H) x Ø165 mm (10”L x 11”H x 6.5”W) | Difference of design |
| Weight | 2.6 kg (5.72 lbs) | 1.6 kg (3.5 lbs) | Higher weight |
| Energy Source | Rechargeable 14.8 VDC, 1,500 mAh Li-Po battery pack Battery Charger Input: 100-240 VAC 50/60 Hz, 1.5A Output: 16.8 VDC 3.0 A. | Rechargeable 22.2 VDC, 1,000 mAh Li-Po battery pack Battery Charger Input: 100-240 VAC 50/60 Hz, 0.4A Output: 25.5 VDC 0.9 A. | Similar |
{6}
| User Interface | Up and down push buttons for exposure time LCD display | Left and right push keys for exposure time LCD display | Difference of design |
| --- | --- | --- | --- |
3
{7}
| Characteristic | PDX nova (Subject Device) | ECOTRON DT-703 (K221233) | Comparison |
| --- | --- | --- | --- |
| Exposure times | 0.1 – 0.5 sec (0.01 sec. steps) | 0.02 – 0.5 sec (0.01 sec. steps) | Higher minimum, same steps |
| kVp | 70 kVp fixed | 70 kVp fixed | Same |
| mA | 3 mA fixed | 3 mA fixed | Same |
| Focal Spot Size | 0.4 mm | 0.3 mm | Lower precision |
| Tube | D-041(CANON) | CEI OX/70-3 | Different manufacturer |
| Total Filtration | 1.5 mmAl | 1.5 mmAl | Same |
| Source to Skin Distance (SSD) | 200 mm | 200 mm | Same |
| Electrical Safety Standards | IEC 60601-1:2005/ A2:2020 | IEC 60601-1:2005 / A1:2012 | Similar standards, different editions |
| EMI Standards | IEC 60601-1-2:2014/ A1:2020 | IEC60601-1-2: 2014 | Similar standards, different editions |
| Performance Standards | 21 CFR 1020.30, 1020.31 IEC 60601-1-3:2008/ A2:2021 IEC 60601-2-65:2012/ A2:2021 | 21 CFR 1020.30, 1020.31; IEC60601-1-3;2013 IEC60601-2-65:2012 | Similar standards, different editions |
The above comparison shows the subject and predicate devices have substantially similar technological characteristics. The exposure time is similar when comparing the subject device to the predicted device. The differences of the device are minor and do not raise new issues of safety and effectiveness.
## Substantial Equivalence Discussion
The subject and predicate are both Handheld/Portable X-ray Systems intended for dentist or dental hygienists as an extra-oral x-ray source for producing diagnostic x-ray images.
As shown in the comparison table above...
- Both devices have the same classification and product code.
- Both devices have similar indications for use and intended use.
- Both devices have equivalent power source of 70kVp and 3mA.
- Both devices have the same backscatter radiation protection shield and cone that are permanently attached.
Safety and effectiveness of the PDX nova device was further documented by the following:
- Cybersecurity... in accordance with guidance 'Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions' (21 CFR Part 820).
- Software...the software validation report is included in the submission conforming to IEC 62304.
- Risk Management...the risk management report was completed conforming to ISO 14971.
- EMC and Electrical...these test reports conform to IEC 60601-1 and 60601-1-2 standards.
- Performance Test...the performance test reports are for specifications and usability, conforming to IEC 60601-2-65, 60601-1-3, 60601-1-6 and 62366-1 standards.
{8}
5
# Specific Guidance Document
There are three FDA Specific Guidance Documents associated with the device: Radiation Safety Consideration for X-ray Equipment Designed for Hand-Held Use. This submission utilized this guidance to develop this device to ensure the safety of this device for both the operators and the patients. The Guidance document: "Content of Premarket Submissions for Device Software Functions" and "Pediatric Information for X-ray Imaging Device Premarket Notifications". Details of this guidance are provided within the Software Validation Report.
# Labels
The labels on the device show that this device conforms to the following:
21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products,
21 CFR 1020.30: Diagnostic x-ray systems and their major components,
21 CFR 1020.31: Radiographic Equipment
# Performance Testing
Performance testing was conducted to support the device's indications for use and demonstrate substantial equivalence including:
1. Tests based on consensus industry standards confirmed the performance capabilities of the subject device.
2. Results from both bench and clinical testing confirmed diagnostic image quality substantially equivalent to the predicate device
# Conclusion
The comparison analysis and performance test results demonstrate the subject device is at least as safe and effective as the legally marketed predicate. The technological differences between the subject device and the predicate device do not raise new questions of safety and effectiveness. The PDX nova subject device is therefore substantially equivalent to the ECOTRON DT-703 predicate device and thus warrants clearance from FDA for marketing activities in the United States.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.