Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button
K260353 · Arthrex, Inc. · HRS · Mar 3, 2026 · Orthopedic
Device Facts
Record ID
K260353
Device Name
Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button
Applicant
Arthrex, Inc.
Product Code
HRS · Orthopedic
Decision Date
Mar 3, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Humeral Plating System is indicated for fractures and fracture-dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. The Cerclage Button is intended for use with the humeral plating system and FiberTape® cerclage suture, to augment fracture stabilization with humeral plates in long bone fixation. The cerclage button is designed for use with the humeral plating system and may not be used alone.
Device Story
System consists of contoured titanium alloy humeral plates and 4.5 mm threaded cerclage buttons; used for internal fixation of proximal humerus fractures. Plates feature suture holes for soft tissue repair; compatible with various Arthrex 3.5 mm, 4.0 mm, and 4.5 mm screws. Devices provided sterile or non-sterile for single use. Surgeon implants device during orthopedic surgery to stabilize bone fragments; facilitates healing via rigid fixation. Output is physical stabilization of bone; aids clinical decision-making by providing structural support for fracture reduction. Benefits include improved stabilization in osteopenic bone.
Clinical Evidence
Bench testing only. 4-point bend testing (ASTM F382-17) performed to demonstrate statistical equivalence to primary predicate. MRI safety (force, torque, image artifact, RF heating) evaluated per FDA guidance and ASTM standards (F2052, F2119, F2182, F2213); device determined not to represent a new worst-case scenario.
Indicated for patients with proximal humerus fractures, fracture-dislocations, osteotomies, and non-unions, particularly those with osteopenic bone. Cerclage button indicated for use with humeral plating system and FiberTape® cerclage suture to augment fracture stabilization in long bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Arthrex Humeral Plating System and Cerclage Button (K243995)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 3, 2026
Arthrex, Inc.
Osiris Rios
Senior Regulatory Affairs Specialist
1370 Creekside Blvd.
Naples, Florida 34108
Re: K260353
Trade/Device Name: Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II
Product Code: HRS, JDQ
Dated: February 2, 2026
Received: February 3, 2026
Dear Osiris Rios:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260353 - Osiris Rios
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260353 - Osiris Rios
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260353 | ? |
| Please provide the device trade name(s). | | ? |
| Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button | | |
| Please provide your Indications for Use below. | | ? |
| The Humeral Plating System is indicated for fractures and fracture-dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
The Cerclage Button is intended for use with the humeral plating system and FiberTape® cerclage suture, to augment fracture stabilization with humeral plates in long bone fixation. The cerclage button is designed for use with the humeral plating system and may not be used alone. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260353
Page 1 of 4
# 510(k) Summary
| Date Prepared | 02/02/2026 |
| --- | --- |
| Submitter | Arthrex Inc.
1370 Creekside Boulevard
Naples, FL 34108-1945 |
| Contact Person | Name: Osiris Rios
Title: Senior Regulatory Affairs Specialist
Phone: 239-643-5553, ext. 71263
Email: Osiris.Rios@arthrex.com |
| Trade Name | Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button |
| Common Name | Plate, Fixation, Bone
Cerclage, Fixation |
| Product Code | HRS, JDQ |
| Classification Name | 21 CFR 888.3030: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories
21 CFR 888.3010: Bone fixation cerclage |
| Regulatory Class | II |
| Primary Predicate Device | K243995: Arthrex Humeral Plating System and Cerclage Button |
| Additional Predicate Devices | K180310: Depuy Synthes LCP Proximal Humerus Plate |
| Purpose of Submission | This Special 510(k) premarket notification is submitted to obtain clearance of the Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button. |
| Device Description | The Arthrex Humeral Plating (Anatomic Humeral Plate) and Cerclage Button consist of contoured humeral plates and a 4.5 threaded cerclage button for use in the repair of proximal humerus fractures. The devices are provided sterile (gamma) and non-sterile, for single use. The proposed plates range in length from 92 - 265 mm and include suture holes for soft tissue repair. The proposed devices are manufactured from titanium alloy conforming to ASTM F136. The proposed devices are compatible with Arthrex 3.5 mm screws (K241592 and K203294), Arthrex 4.0 mm screws (K150456, K143614, and K103705), and Arthrex 4.5 mm screws (K141735). |
{5}
K260353
Page 2 of 4
| Indications for Use | The Humeral Plating System is indicated for fractures and fracture-dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
The Cerclage Button is intended for use with the humeral plating system and FiberTape® cerclage suture, to augment fracture stabilization with humeral plates in long bone fixation. The cerclage button is designed for use with the humeral plating system and may not be used alone. |
| --- | --- |
| Performance Data | Arthrex conducted 4-point bend (ASTM F382-17) testing to demonstrate that the Arthrex Humeral Plating System perform statistically equivalent to the primary predicate device, Arthrex Humeral Plating System and Cerclage Button (K243995).
MRI force, torque, and image artifact testing was previously conducted in accordance with FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, ASTM F2052 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, ASTM F2119 Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants, ASTM F2182 Standard Test Method for Measurement of Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging and ASTM F2213 Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment. Arthrex determined that the proposed Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button device does not represent a new worst-case in terms of MR compatibility and will be adopted into the MRI labeling previously reviewed and accepted by FDA within primary predicate device, Arthrex Humeral Plating System and Cerclage Button (K243995).
Assessment of the physical product attributes including product, design, size, and materials has determined that the Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button does not introduce additional risks or concerns regarding sterilization and shelf-life. |
| Technological Comparison | The proposed Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button have the same intended use/indications, fundamental scientific technology, materials, packaging method, |
{6}
K260353
Page 3 of 4
sterility, shelf-life and MRI safety labeling as the primary predicate Arthrex Humeral Plating System and Cerclage Button (K243995) and additional predicate device Depuy Synthes LCP Proximal Humerus Plate (K180310).
The differences with the predicates are provided below:
- Design:
- The Anatomic Humeral Plate in the proposed Arthrex Humeral Plating System is contoured and curved whereas the lateral plate in the cleared primary predicate (K243995) and additional predicate (K180310) are straight.
- The Anatomic Humeral Plate in the proposed Arthrex Humeral Plating System is designed with 3 to 15 screw hole configurations and in multiple lengths with the overall length of the plates ranging from 92mm to 265 mm whereas the lateral plate in the cleared primary predicate (K243995) is designed with 3 to 17 screw hole configurations and in lengths 90.5 – 309 mm. The lengths for the proposed Anatomic Humeral Plates are within the lengths range of the primary predicate (K243995). The additional predicate (K180310) has less holes configurations than the proposed device. The lengths for the proposed Anatomic Humeral Plates are within the length range of the additional predicate (K180310).
- The Anatomic Humeral Plate in the proposed Arthrex Humeral Plating System accepts 4.5 mm cerclage buttons in the distal shaft whereas the lateral plate in the cleared primary predicate (K243995) accepts 3.5 mm cerclage buttons in the shaft.
- In addition to the Arthrex 3.5 mm Variable Angle Locking (VAL) Screws, Reinforced (K241592), 3.5mm VAL KreuLock™ Screws, Reinforced (K241592), Arthrex 3.5mm Low Profile Screws, Cortical (K203294), Arthrex 4.0 mm Low Profile Screws, and Cancellous (K150456, K143614, K103705); the proposed Anatomic Humeral Plate, in 7-hole and longer configurations, the most distal shaft
{7}
K260353
Page 4 of 4
| | screw holes accept 4.5 mm cortical or locking low profile screws (K141735). • Packaging: The proposed Cerclage Button (4.5mm) is packaged in a double Nylon/Nylon pouch configuration whereas the Cerclage Button (3.5 mm) cleared under K243995 was packaged in a double Poly/Tyvek pouch. However, the double Nylon/Nylon pouch configuration used for the proposed 4.5 mm Cerclage Button is identical to the packaging configuration (double Nylon/Nylon pouch) used for the lateral plates in the primary predicate device (K243995). Based on the basic design features, device configuration, intended use, fundamental scientific technology, technological characteristics and the performance data provided in this Special 510(k), Arthrex has determined that the proposed Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button are substantially equivalent to the primary predicate and additional predicate devices cleared via K243995 and K180310, respectively. Any differences between the proposed and predicate devices are considered minor and do not raise different questions concerning safety and effectiveness. |
| --- | --- |
| Conclusion | The Arthrex Humeral Plating System (Anatomic Humeral Plate) and Cerclage Button is substantially equivalent to the predicate devices cleared under K243995 and K180310 in which the basic design features and intended use are the same. Based on the indications for use, technological characteristics, and the summary of data submitted, Arthrex Inc. has determined that the proposed device is substantially equivalent to the currently marketed predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.