HipGuide is a medical device software application designed to assist surgeons in evaluating implant positioning, alignment, and anatomical relationships during hip replacement procedures. HipGuide supports intra-operative clinical decision making by providing visual overlays or measurements taken manually on fluoroscopic images. It is not intended to replace the surgeon’s judgment or serve as a diagnostic device. Rx only.
Device Story
HipGuide is a software application used intra-operatively by surgeons during hip replacement procedures. It takes live or captured fluoroscopic images from a mobile C-arm as input. The device functions as a manual measurement tool, allowing the surgeon to place and adjust visual overlays, points, and lines directly on the displayed image to assess distances and angles. It does not modify, store, or transmit patient data. The output is a visual overlay on the host imaging system, which the surgeon uses to support clinical decision-making regarding implant positioning and alignment. The device does not contact the patient or interact with implants; it serves as an adjunct to the surgeon's judgment.
Clinical Evidence
Bench testing only. No clinical data. Accuracy evaluated using ~100 linear and angular measurements on calibrated test objects. Results showed linear accuracy of ±5mm and angular accuracy of ±5°, consistent with predicate performance and literature.
Technological Characteristics
Software-only medical device. Operates on mobile C-arm systems with flat-panel detectors. Provides manual measurement tools (points, lines, visual overlays). No patient contact. No automated algorithmic processing; relies on manual user-controlled placement. Standalone software application.
Indications for Use
Indicated for surgeons performing hip replacement procedures to assist in evaluating implant positioning, alignment, and anatomical relationships via manual measurements on intra-operative fluoroscopic images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Orthopedic Driven Imaging, LLC
Michael Mccaig
General Manager
4424 NW 13th St., Suite B-6
Gainesville, FL 32609
May 8, 2026
Re: K260321
Trade/Device Name: HipGuide (V 1.0.0.0)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: April 2, 2026
Received: April 2, 2026
Dear Michael Mccaig:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260321 - Michael Mccaig
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260321 - Michael Mccaig
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260321 | ? |
| Please provide the device trade name(s). | | ? |
| HipGuide (V 1.0.0.0) | | |
| Please provide your Indications for Use below. | | ? |
| HipGuide is a medical device software application designed to assist surgeons in evaluating implant positioning, alignment, and anatomical relationships during hip replacement procedures. HipGuide supports intra-operative clinical decision making by providing visual overlays or measurements taken manually on fluoroscopic images. It is not intended to replace the surgeon’s judgment or serve as a diagnostic device. Rx only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: | 510(k) Summary | Prepared on: 2026-01-30 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Orthopedic Driven Imaging, LLC | |
| Applicant Address | 4424 NW 13th Street, Suite B-6 Gainesville FL 32609 United States | |
| Applicant Contact Telephone | (901) 603-1309 | |
| Applicant Contact | Dr. Michael McCaig | |
| Applicant Contact Email | mac@orthodriven.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | HipGuide (V 1.0.0.0) | |
| Common Name | Medical image management and processing system | |
| Classification Name | System, Image Processing, Radiological | |
| Regulation Number | 892.2050 | |
| Product Code(s) | LLZ | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K210136 | PhantomMSK Hip | LLZ |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| HipGuide is a software application that provides calibrated measurement overlays, alignment guides, and visual reference markers on fluoroscopic images to support intraoperative decision-making during orthopedic procedures, including hip replacement surgery. HipGuide's principle of operation is manual, user-controlled placement and adjustment of points, lines, and visual guides directly on the displayed fluoroscopic image, enabling assessment of distances and angles without altering the underlying image data. HipGuide is used during procedures involving the hip and surrounding anatomy and does not contact the patient or interact with implants. The user interface consists of on-screen tools that allow surgeons to create, move, and remove measurement elements using standard system controls. The device interacts only with the host imaging system by displaying overlays on top of the live or captured fluoroscopic image and does not modify, store, or transmit patient data. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| HipGuide is a medical device software application designed to assist surgeons in evaluating implant positioning, alignment, and anatomical relationships during hip replacement procedures. HipGuide supports intra-operative clinical decision making by providing visual overlays or measurements taken manually on fluoroscopic images. It is not intended to replace the surgeon's judgment or serve as a diagnostic device. Rx only. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
| The subject device utilizes the same principle of operation as the predicate device as they are both designed to make measurements using intra-operative radiographic images acquired by a mobile c arm. HipGuide is substantially equivalent to the predicate device PhantomMSK Hip (K210136) with respect to intended use, technological characteristics, and performance. | | |
{5}
Technological Comparison
21 CFR 807.92(a)(6)
The predicate and subject devices are similar in the types of measurements they make and in the environments in which they are used, namely, to landmark and make measurements in connection with total hip arthroplasty (THA) procedures using images acquired by a mobile C-arm in the operating room. The predicate device is mounted on a monitor cart connected to the C-arm video output, whereas the subject device is operated directly on the C-arm. The predicate device can use a grid to correct image distortion caused by image intensifiers and can work with flat-panel detectors. The subject device works with images produced by flat-panel detectors, which have twice the image area of a 12-inch image intensifier, providing images wide enough to capture both hips at once without requiring grid calibration. The predicate device offers algorithmic automated landmarking with manual adjustments and AI-powered software to simplify landmarking and measurements. The subject device relies on manual landmarking and the placement of measurement tools.
Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The bench testing was performed in connection with the Software Verification Testing. The Bench Testing Report provides a summary of those activities and compares those results to several scientific publications on this subject to provide evidence that the measurement accuracy of the subject device is well within the limits suggested by the literature. No guidance documents or FDA recognized consensus standards were used.
Measurement Accuracy Testing: Nearly 100 linear and angular measurements were evaluated using calibrated test objects with known dimensions and angular positions. HipGuide demonstrated a measurement accuracy of $\pm 5\mathrm{mm}$ for linear measurements and $\pm 5^{\circ}$ for angular measurements. These results support that HipGuide's measurement performance is consistent with the predicate device and suitable for its intended intraoperative use.
Not Applicable
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.