K260317 · International Life Sciences · QWJ · Apr 9, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260317
Device Name
Artelon Convenience Kits
Applicant
International Life Sciences
Product Code
QWJ · General, Plastic Surgery
Decision Date
Apr 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FlexBand, FlexPatch, and FlexBand Plus: FlexBand, FlexPatch, and FlexBand Plus are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists. FlexBand, FlexPatch, and FlexBand Plus are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. FlexBand, FlexPatch, and FlexBand Plus is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provide a degradable scaffold that is incorporated into the patient's own tissue. Twist and FlexBand Anchors: The Twist and FlexBand Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications: • Shoulder: Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair. • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Illiotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair. • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair. • Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment. • Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC. • Hip: Acetabular labral repair.
Device Story
Convenience kits containing surgical implants and instrumentation for soft tissue reinforcement and fixation; includes FlexBand/FlexPatch/FlexBand Plus implants, Twist/FlexBand anchors, and surgical tools (drill bits, guides, drivers, K-wires). Implants provide degradable scaffold for tissue incorporation; anchors secure soft tissue to bone. Used in OR by surgeons for orthopedic procedures (rotator cuff, ligament repairs, etc.). Kits facilitate surgical workflow by bundling necessary components. Implants do not provide primary mechanical strength; sutures/fixation devices provide strength. Benefits include reinforced soft tissue healing via scaffold integration.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including mesh molar mass characterization, suture retention, tensile testing, anchor pull-out strength, insertion/removal torque, and torque to failure.
Technological Characteristics
Implants consist of degradable scaffold materials (FlexBand/FlexPatch/FlexBand Plus) and metallic bone fixation fasteners (Twist/FlexBand anchors). Instrumentation includes stainless steel/standard surgical materials. Sterilization via EtO. No software or electronic components.
Indications for Use
Indicated for soft tissue reinforcement and soft tissue-to-bone attachment in surgical procedures. Patient population includes individuals requiring repair of tendons and ligaments in shoulder, knee, foot/ankle, elbow, hand/wrist, and hip. Contraindicated for use as a replacement for normal body structure or as the sole provider of mechanical strength for tendon/ligament repairs.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 9, 2026
International Life Sciences
Rachel Roberts
Sr. Staff Specialist, Regulatory Affairs
8601 Dunwoody Pl #250
Sandy Springs, Georgia 30350
Re: K260317
Trade/Device Name: Artelon Convenience Kits
Regulation Number: 21 CFR 878.3300
Regulation Name: Surgical Mesh
Regulatory Class: Class II
Product Code: QWJ, OWW, MBI
Dated: January 30, 2026
Received: January 30, 2026
Dear Rachel Roberts:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260317 - Rachel Roberts
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260317 - Rachel Roberts
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thomas Mcnamara -S
For: Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260317 | ? |
| Please provide the device trade name(s). | | ? |
| Artelon Convenience Kits | | |
| Please provide your Indications for Use below. | | ? |
| FlexBand, FlexPatch, and FlexBand Plus:
FlexBand, FlexPatch, and FlexBand Plus are intended for use in surgical procedures for reinforcement of
soft tissue where weakness exists.
FlexBand, FlexPatch, and FlexBand Plus are also intended for reinforcement of soft tissues that are
repaired by suture or other fixation devices during tendon and ligament repair surgery including
reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament,
lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or
extra-articular ligaments.
FlexBand, FlexPatch, and FlexBand Plus is not intended to replace normal body structure or provide the full
mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial
collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or
other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation
devices, used to attach the tissue to the bone, provide mechanical strength for the tendon repair. The
products reinforce soft tissue and provide a degradable scaffold that is incorporated into the patient's own
tissue.
Twist and FlexBand Anchors:
The Twist and FlexBand Anchors are indicated for attachment of soft tissue to bone. This product is
intended for the following indications:
• Shoulder: Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Repair, Bankart Lesion
Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion
Repair.
• Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament
Repair, Extra Capsular Reconstruction, Illiotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion
Repair.
• Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair,
Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
• Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment.
• Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction,
TFCC.
• Hip: Acetabular labral repair. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
stryker
510(K) SUMMARY: ARTELON CONVENIENCE KITS
(a)(1). Submitted by:
International Life Science
8601 Dunwoody Pl #250
Sandy Springs, GA 30350
Date: January 30, 2026
Contact Person: Rachel Roberts
Senior Regulatory Staff Specialist
(901) 302-5801
Secondary Contact: Leslie Fitch
Director of Regulatory Affairs
(a)(2). Proprietary Name: Artelon Convenience Kits
Common Name: Surgical Mesh & Smooth or threaded metallic bone fixation fastener
Classification Name and Reference: 21 CFR 878.3300 & 21 CFR 888.3040 – Class II
Device Product Code, Device Panel: QWJ, OWW, MBI
(a)(3). Primary Predicate Device(s): FlexBand, FlexPatch, and FlexBand Plus (K230316) and ATL Anchors (K200503)
(a)(4). Device Description
The subject device, Artelon convenience kits, contains implants and instrumentation for surgical procedures for soft tissue reinforcement and soft tissue fixation to bone. The kits include suture, FlexBand or Twist anchors, FlexBand Dynamic Matrix implants, and instrumentation. The names of the Artelon convenience kit options are FlexBand Fix Kit, FlexBand Solo Kit, FlexBand Multi Kits, FlexBand Twist Kits, FlexBand Drill Kits, and FlexBand Twist Driver Kits. The implants in the kits are the FlexBand or FlexBand Plus, Twist, or FlexBand anchors, which are the same or similar products as cleared under K230316 and K200503. The implants may be sold individually without being included in a convenience kit. The subject device instrumentation includes drill bits, tissue protectors, drivers, drill guides, inserter, K-wires, and anchor caddy.
(a)(5). Indications for Use
FlexBand, FlexPatch, and FlexBand Plus are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists.
FlexBand, FlexPatch, and FlexBand Plus are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral
Artelon Convenience Kits
K260317
{5}
stryker
ligament or other tendons or extra-articular ligaments.
FlexBand, FlexPatch, and FlexBand Plus is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provide a degradable scaffold that is incorporated into the patient's own tissue.
The Twist and FlexBand Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:
- Shoulder: Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Illiotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
- Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment.
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.
- Hip: Acetabular labral repair.
(a)(6). Technological Characteristics Comparison
For the subject device (FlexBand and FlexBand Plus), there are no changes to the indications for use, product codes, principle of operation, or implant materials. The subject devices have the same design and range of sizes as the predicate device (K230316). The subject devices (FlexBand and Twist anchors) have the same range of sizes and are a similar design as the predicate device (K200503). An alternative sterilization method, EtO, is included in this submission for the subject device kits.
(b)(1). Substantial Equivalence – Non-Clinical Evidence
Non-clinical performance bench testing (mesh molar mass characterization, mesh suture retention, mesh tensile testing, anchor pull-out strength, mesh + anchor pull-out strength, anchor insertion and removal torque, and anchor torque to failure) was performed to demonstrate equivalence to the predicate devices. Bacterial endotoxin testing, sterilization testing, and packaging testing information was also provided.
(b)(2). Substantial Equivalence - Clinical Evidence
N/A – Clinical testing was not necessary for the determination of substantial equivalence.
Artelon Convenience Kits
K260317
{6}
stryker
(b)(3). Substantial Equivalence – Conclusions
The evaluation performed and supporting documentation within this submission demonstrates that the subject device, Artelon convenience kits, are comparable to the predicate devices. In conclusion, based on similarities between the indications for use, intended use, principles of operation, materials, design, size offerings, shelf life, product codes, the subject device, Artelon convenience kits, are substantially equivalent to and is as safe and effective as, the legally marketed predicate devices, FlexBand, FlexPatch, and FlexBand Plus (K230316) and ATL Anchors (K200503).
Artelon Convenience Kits
K260317
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.