← Product Code [FDA](/productcode/FDA) · K260314

# Ancora-SB (K260314)

_Aspero Medical, Inc. · FDA · May 1, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K260314

## Device Facts

- **Applicant:** Aspero Medical, Inc.
- **Product Code:** [FDA](/productcode/FDA.md)
- **Decision Date:** May 1, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Ancora-SB is an accessory to an endoscope. The Ancora-SB is intended for use with any standard endoscope that has an outer diameter of 9.0 – 9.4 mm and a working length of 1680 mm or greater. The Ancora-SB is intended for use with any standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 kPa (+2.6 kPa, -1.8 kPa). The device is indicated to ensure complete positioning of an endoscope in the small intestine by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment.

## Device Story

Ancora-SB is a single-use, dual-lumen silicone balloon overtube; slides over standard endoscopes. Major lumen features PEG-based hydrophilic coating to reduce friction; minor lumen connects to external balloon inflation unit. Distal end includes textured silicone balloon, soft end cap, and barium sulfate radiopaque marker. Operated by gastroenterologists during endoscopic procedures; balloon inflation stabilizes endoscope or facilitates advancement via plication in small intestine. Output is physical stabilization/advancement of endoscope; enables visualization and therapeutic access. Benefits include improved navigation and positioning within small intestine.

## Clinical Evidence

Bench testing only. Biocompatibility evaluated per ISO 10993-1:2018. Performance testing included friction testing of the new hydrophilic coating under simulated challenging conditions, bond strength testing for the modified end cap, and verification of radiopacity for the barium sulfate marker.

## Technological Characteristics

Dual-lumen silicone tubing; major lumen with PEG-based hydrophilic coating; minor lumen for air inflation. Distal end: textured silicone balloon, barium sulfate (25%) impregnated silicone radiopaque marker. Proximal end: silicone funnel handle with titanium dioxide. Dimensions: 38 mm balloon OD, 11 mm inner diameter, 132 cm working length. Non-sterile. Shelf life: 12 months. Energy source: external pneumatic balloon inflation unit (5.4 kPa).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Ancora-SB ([K231323](/device/K231323.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

May 1, 2026

Aspero Medical, Inc.
% Nancy Sauer
Owner and President
Nancy Sauer Regulatory Consulting, LLC
401 Spruce St.
Louisville, Colorado 80027

Re: K260314
Trade/Device Name: Ancora-SB
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FDA, FED
Dated: January 30, 2026
Received: January 30, 2026

Dear Nancy Sauer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K260314 - Nancy Sauer
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K260314 - Nancy Sauer
Page 3

Sincerely,

SHANIL P. HAUGEN -S

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K260314  |   |
|  Device Name Ancora-SB  |   |
|  Indications for Use (Describe) The Ancora-SB is an accessory to an endoscope. The Ancora-SB is intended for use with any standard endoscope that has an outer diameter of 9.0 – 9.4 mm and a working length of 1680 mm or greater. The Ancora-SB is intended for use with any standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 kPa (+2.6 kPa, -1.8 kPa). The device is indicated to ensure complete positioning of an endoscope in the small intestine by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K260314
Page 1 of 4
1

# 510(k) Summary

Date Prepared: April 21, 2026
Prepared by: Nancy Sauer

## Manufacturer and 510(k) Owner

Aspero Medical, Inc.
10835 Dover Street, Suite 800, Denver, CO 80021
Phone: (303) 834-7885
Official Contact: Dr. Mark Rentschler, CEO

## Submission Correspondent/Consultant

Nancy Sauer, MA
Nancy Sauer Regulatory Consulting llc
Louisville, CO 80027
E-mail: Nancy.sauer4@gmail.com

## Device Information

Trade Name: Ancora-SB
Common Name: Balloon overtube
Regulation Number: 21CFR 876.1500
Class: II
Product Codes: FDA, FED
Review Panel: Gastroenterology/Urology

## Predicate Device Information

Trade Name: Ancora-SB
Common Name: Balloon overtube
Regulation Number: 21CFR 876.1500
Class: II
Product Codes: FDA, FED
Review Panel: Gastroenterology/Urology

510(k) Number: K231323

## Device Description

The Ancora-SB is a single-use, close-fitting sleeve that slides freely over a currently marketed standard endoscope having an outer diameter of 9.0 – 9.4 mm and a working length of 1680 mm or greater. The Ancora-SB balloon overtube assembly is a dual lumen silicone tubing that is coaxially coupled to the endoscope by inserting the distal end of the endoscope into the open funnel lumen of the

510(k) – Modified Ancora-SB Overtube
Aspero Medical
Apr 21, 2026

{5}

K260314
Page 2 of 4
2

overtube’s proximal handle end assembly and by positioning the overtube as needed. The major lumen has a polyethylene glycol-based hydrophilic lubricious coating to facilitate endoscope insertion and advancement.

The minor lumen provides the air channel for balloon inflation. The Ancora-SB balloon overtube assembly is connected via an air supply tube to a compatible balloon inflation unit. An air supply tube adapter (luer connector) is provided in the package as an optional accessory to facilitate connection with inflation units where such an adapter is needed.

When the textured balloon near the distal end is inflated by the balloon inflation unit, it can either stabilize the position of endoscope in the intestine or it can facilitate advancement of the endoscope through a maneuver know as plication.

## Indications for Use

The Ancora-SB is an accessory to an endoscope. The Ancora-SB is intended for use with any standard endoscope that has an outer diameter of 9.0 – 9.4 mm and a working length of 1680 mm or greater. The Ancora-SB is intended for use with any standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 kPa (+2.6 kPa, -1.8 kPa). The device is indicated to ensure complete positioning of an endoscope in the small intestine by either oral or anal insertion, and assist with optical visualization, diagnosis, and endoscopic treatment.

## Comparison of Modified Ancora-SB to Predicate Device

Aspero Medical, Inc. submitted this traditional 510(k) because to gain clearance for changes in materials and specification changes that met the criteria for a new 510(k) when analyzed according to the criteria in the 2017 FDA guidance document “Deciding When to Submit a 510(k) for a Change to an Existing Device.”

There are no changes to the indications for use or the intended use of the Ancora-SB.

The table below compares the technological characteristics of the modified Ancora-SB to those of the originally cleared Ancora-SB.

|  Characteristic | Modified Ancora-SB | Original Ancora-SB  |
| --- | --- | --- |
|  Target Areas of Use | Small intestine | Small intestine  |
|  Device Introduction Site | Oral or anal | Oral or anal  |
|  Device components | Single use splinting tube Air supply assembly Luer adapter | Single use splinting tube Air supply assembly Luer adapter  |

510(k) – Modified Ancora-SB Overtube
Aspero Medical
Apr 21, 2026

{6}

K260314
Page 3 of 4
3

|  Characteristic | Modified Ancora-SB | Original Ancora-SB  |
| --- | --- | --- |
|  Operating Principle and Design | Dual lumen silicone tubing with major lumen having hydrophilic lubricious coating to facilitate endoscope insertion and minor lumen as air channel for balloon inflation. Proximal end is a funnel handle with endoscope access, air tubing attachment, and irrigation syringe attachment to activate lubricious coating. Distal end features a soft rounded end cap, a radiopaque marker, and a silicone balloon with a textured surface that contacts the mucosa during use. | Dual lumen silicone tubing with major lumen having hydrophilic lubricious coating to facilitate endoscope insertion and minor lumen as air channel for balloon inflation. Proximal end is a funnel handle with endoscope access, air tubing attachment, and irrigation syringe attachment to activate lubricious coating. Distal end features a soft rounded end cap, a radiopaque marker, and a silicone balloon with a textured surface that contacts the mucosa during use.  |
|  Balloon Outer Diameter | 38 mm | 38 mm  |
|  Balloon Injection Volume | 43 mL | 43 mL  |
|  Insertion Tube Maximum Insertion Width | 15.9 mm | 15.9 mm  |
|  Insertion Tube Inner Diameter | 11 mm | 11 mm  |
|  Insertion Tube Working Length | 132 cm | 132 cm  |
|  Insertion Tube Total Length | 140 cm | 140 cm  |
|  Air Flow Tube Length | 370 cm | 370 cm  |
|  Materials | Funnel handle: Silicone with titanium dioxide, external laser marking Tube: Silicone with PEG-based hydrophilic coating of the lumen interior Balloon: Silicone Radiopaque feature on distal end: Barium sulfate (25%) impregnated silicone | Funnel handle: Silicone, external printing – polyurethane-based ink Tube: Silicone with PVP-based hydrophilic coating of the lumen interior Balloon: Silicone Radiopaque feature on distal end: platinum  |
|  Balloon | Single textured balloon on distal end of splinting tube | Single textured balloon on distal end of splinting tube  |
|  Balloon Inflation Method | Standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 (+2.6/-1.8) kPa. | Standard flexible endoscopy balloon inflation unit with a set pressure of 5.4 (+2.6/-1.8) kPa.  |
|  Proximal End Connections | Endoscope insertion port Fluid/flush port Air/inflation port | Endoscope insertion port Fluid/flush port Air/inflation port  |
|  Sterility | Non-sterile | Non-sterile  |

510(k) – Modified Ancora-SB Overtube
Aspero Medical
Apr 21, 2026

{7}

K260314
Page 4 of 4
4

|  Characteristic | Modified Ancora-SB | Original Ancora-SB  |
| --- | --- | --- |
|  Shelf Life | 12 months | 12 months  |

The primary differences are in the formulation of the hydrophilic coating of the overtube lumen and a revised proprietary test method. These interrelated changes are intended to ensure acceptable levels of friction even in challenging use cases.

This 510(k) also reports a change in the radiopaque feature at the distal end of the overtube.

## Summary of Non-Clinical Testing

Biocompatibility was evaluated in accordance with ISO 10993-1: 2018 and the associated FDA guidance document. This evaluation demonstrated acceptable biocompatibility for the device’s intended use.

Bench testing was limited to evaluation of characteristics affected by the material and/or specification changes.

- The new hydrophilic coating demonstrated acceptably low friction under test conditions simulating challenging use cases.
- The modified end cap continued to meet specifications for a secure bond between the cap and the rest of the overtube.
- The modified end cap was found to have acceptable radiopacity.

## Conclusion

The modified Ancora-SB is substantially equivalent to the originally cleared Ancora-SB. This conclusion is based on the following information:

- The indications for use and the operating principle are the same.
- There are a limited number of technological changes that do not raise new issues of safety or effectiveness when compared to the predicate device.
- Appropriate methods exist to evaluate the technological changes, and the results of testing support a conclusion of substantial equivalence.

510(k) – Modified Ancora-SB Overtube
Aspero Medical
Apr 21, 2026

---

**Source:** [https://fda-staging.innolitics.com/device/K260314](https://fda-staging.innolitics.com/device/K260314)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
