K260308 · Adaptiiv Medical Technologies, Inc. · MUJ · Feb 19, 2026 · Radiology
Device Facts
Record ID
K260308
Device Name
TrueFit Bolus
Applicant
Adaptiiv Medical Technologies, Inc.
Product Code
MUJ · Radiology
Decision Date
Feb 19, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TrueFit Bolus is indicated for and intended to be placed on the patient’s skin as an accessory to attenuate and/or compensate external beam (photon or electron) radiation during prescribed radiation therapy for the treatment of cancer or other non-malignant tissue conditions for which radiation therapy is indicated. The device is for a single patient’s use only and can be reused throughout the entirety of the treatment course. The device is designed by the radiation therapy professional using patient imaging data as input and must be verified and approved by the trained radiation therapy professional prior to use. The device is restricted to sale by or on the order of a physician and is by prescription only.
Device Story
TrueFit Bolus is a 3D-printed, patient-matched radiation therapy accessory. It uses patient imaging data from a treatment planning system as input to generate a digital design via the 3D Bolus Software Application (K213438). The output is an STL file used to 3D-print the bolus using MJF (polyamide/polyurethane) or SLA (methacrylate photopolymer resin). The device is used in clinical radiation therapy settings; operated by radiation therapy professionals. The bolus is placed on the patient's skin to act as a tissue-equivalent material, accounting for the build-up region of the treatment beam to ensure the prescribed dose is delivered to the skin surface or near-surface target. The healthcare provider verifies the design before use. The device benefits patients by providing a precise, patient-specific fit, improving dose distribution accuracy.
Clinical Evidence
Bench testing only. Evaluation included spatial fidelity (dimensional accuracy vs. STL), physical density (g/cc), and radiological density/uniformity (Hounsfield Units). Results: ≥95% of points within ±3 mm of STL; mean density 1.20 g/cc (SD 0.01); HU range 52–111. All samples met acceptance criteria.
Technological Characteristics
3D-printed patient-matched accessory. Materials: Polyamide, polyurethane (MJF), or methacrylate photopolymer resin (SLA). Sensing/Actuation: None (passive). Energy: None. Connectivity: Digital input from TPS via STL file. Sterilization: Not specified. Software: 3D Bolus Software Application (K213438) used for design.
Indications for Use
Indicated for patients prescribed external beam radiation therapy (photon or electron) for cancer or non-malignant tissue conditions requiring skin-surface dose attenuation or compensation.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 19, 2026
Adaptiiv Medical Technologies, Inc.
Olga Zhuk
Quality and Regulatory Manager
1559 Barrington St., Suite 200-D
Halifax, NS B3J 1Z8
Canada
Re: K260308
Trade/Device Name: TrueFit Bolus
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical charged-particle radiation therapy system
Regulatory Class: Class II
Product Code: MUJ
Dated: January 30, 2026
Received: January 30, 2026
Dear Olga Zhuk:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260308 - Olga Zhuk
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260308 - Olga Zhuk
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260308 | ? |
| Please provide the device trade name(s). | | ? |
| TrueFit Bolus | | |
| Please provide your Indications for Use below. | | ? |
| TrueFit Bolus is indicated for and intended to be placed on the patient’s skin as an accessory to attenuate
and/or compensate external beam (photon or electron) radiation during prescribed radiation therapy for the
treatment of cancer or other non-malignant tissue conditions for which radiation therapy is indicated.
The device is for a single patient’s use only and can be reused throughout the entirety of the treatment
course.
The device is designed by the radiation therapy professional using patient imaging data as input and must
be verified and approved by the trained radiation therapy professional prior to use.
The device is restricted to sale by or on the order of a physician and is by prescription only. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Special 510(k) Summary
K260308
## Contact Details
Applicant Name: Adaptiiv Medical Technologies, Inc.
Applicant Address: 1559 Barrington Street, Suite 200-D Halifax NS B3J 1Z8 Canada
Applicant Contact Telephone: 902-442-9091
Applicant Contact: Olga Zhuk
Applicant Contact Email: olga.zhuk@adaptiiv.com
## Device Name
Device Trade Name: TrueFit Bolus
Common Name: Medical charged-particle radiation therapy system
Classification Name: System, Planning, Radiation Therapy Treatment
Regulation Number: 892.5050
Product Code: MUJ
## Legally Marketed Predicate Devices
Predicate #: K243057
Predicate Trade Name: TrueFit Bolus
Product Code: MUJ
## Device Description Summary
TrueFit Bolus is a 3D printed patient-matched radiation therapy accessory that expands the application of external beam radiation therapy by providing a patient-specific fit.
Patient imaging data from the treatment planning system (TPS) are used as inputs to generate digital design of the radiation therapy bolus (TrueFit) by 3D Bolus Software Application (K213438), previously developed by Adaptiv. The resulting output Stereolithography (STL) file is compatible with the third-party 3D printers. A TrueFit Bolus can be 3D-printed using MJF with polyamide or polyurethane, or SLA with methacrylate photopolymer resin, based on the user's preference.
The bolus is used in radiation therapy when a patient requires the total prescription dose to be delivered on or near the skin surface. The bolus acts as a tissue-equivalent material placed on the patient skin to account for the build-up region of the treatment beam.
## Intended Use/Indications for Use
TrueFit Bolus is indicated for and intended to be placed on the patient's skin as an accessory to attenuate and/or compensate external beam (photon or electron) radiation during prescribed radiation therapy for the treatment of cancer or other non-malignant tissue conditions for which radiation therapy is indicated.
The device is for a single patient's use only and can be reused throughout the entirety of the treatment course.
The device is designed by the radiation therapy professional using patient imaging data as input and must be verified and approved by the trained radiation therapy professional prior to use.
1 of 3
{5}
The device is restricted to sale by or on the order of a physician and is by prescription only.
# Indications for Use Comparison
The updated Indications for Use for the TrueFit Bolus clarify and more explicitly describe the tissue conditions, without changing the fundamental intended use of the device. The device continues to be placed on the patient's skin as an accessory to attenuate and/or compensate external beam (photon or electron) radiation during prescribed radiation therapy. The added language, including "other non-malignant tissue conditions for which radiation therapy is indicated," clarifies the scope of tissue conditions treated within the established context of radiation therapy and reflects current clinical practice. This clarification does not introduce a new disease, condition, or distinguishable patient population. The intended patient population remains patients who are prescribed radiation therapy and require an accessory device to achieve the prescribed radiation dose distribution. The updated Indications for Use do not alter the device's mechanism of action, clinical application, or conditions of use. No new risks are introduced, and the safety and effectiveness of the TrueFit Bolus remain unchanged. Therefore, the updated Indications for Use do not constitute a new intended use.
# Technological Comparison
The matter of the design change is addition of new material (methacrylate resin) using the SLA printing technology. The addition of rigid clear resin as a material for TrueFit Bolus expands the potential clinical use of the device due to the material transparency. It increases usability of the device during the fit of the device on the patient's anatomy.
# Non-Clinical Tests Summary & Conclusions
Verification and Validation activities have been conducted using worst-case geometrical test samples and real-patient final devices.
The scope of the testing included evaluation of spatial fidelity and relevant radiological and physical properties directly affecting the performance and safety of the device. A summary of the testing is provided in Table 1.
Table 1. Summary of Non-Clinical Testing
| Test | Objective | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Look & Feel | Evaluate surface quality, visual conformity to design, and usability. | High-quality surface finish; no visible defects; printed device visually matches STL; label and anatomical markers readable. | All samples met acceptance criteria and were rated high quality. |
| Spatial Fidelity | Confirm dimensional accuracy of printed device compared to STL design. | ≥95% of measured points within ±3 mm of STL. | All samples had ≥95% of measured points within ±3 mm. |
| Physical Density | Confirm material density consistent with tissue-equivalent properties. | Mean density between 0.90–1.30 g/cc; SD ≤ ±0.10 g/cc. | Mean density 1.20 g/cc; SD 0.01 g/cc; all criteria met. |
| Radiological Density & Uniformity | Confirm radiological properties suitable for radiation therapy use. | Mean HU between –100 and 500; uniformity within ±100 HU. | HU range 52–111; uniformity within tolerance; all criteria met. |
The results of the testing approaches demonstrated:
- Acceptable spatial fidelity, ensuring a precise fit of the device on the patient's anatomy and accurate delivery of radiation to the target treatment tissue.
- Physical density consistent with tissue-equivalent material.
- Radiological properties appropriate for treatment planning and dose delivery.
{6}
Verification and validation results support that the TrueFit Bolus is safe and effective for use in clinical conditions and remains substantially equivalent to the predicate device.
3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.