The HOTWIRE is indicated for creation of an atrial septal defect in the heart.
Device Story
The HOTWIRE™ is a sterile, single-use RF guidewire used to create an atrial septal defect. It consists of a stainless steel core wire jacketed in insulating polymer, featuring an uninsulated stainless-steel coil at the distal tip that acts as an electrode. The device is used in a clinical setting by a physician, typically in conjunction with a compatible third-party transseptal sheath/dilator and an RF electrosurgical generator. The physician advances the wire through the vasculature; once positioned, the generator delivers monopolar RF energy to the distal electrode to perform the septal puncture. The device provides fluoroscopic and echogenic visualization for guidance. The stiff body supports the advancement of subsequent wire-guided devices into the left atrium. By enabling controlled transseptal access, the device facilitates complex cardiac procedures.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench top performance testing, including visual inspection, simulated use, arc integrity, fracture resistance, flexing, and tensile strength tests. Biocompatibility was assessed and confirmed.
Technological Characteristics
Stainless steel core wire with insulating polymer jacket; uninsulated stainless-steel coil electrode at distal tip; tungsten marker coil for radiopacity. Monopolar RF energy delivery via compatible third-party electrosurgical generators. Sterilized by Gamma radiation (VDmax25, SAL 10^-6). Compliant with IEC 60601-2-2 for return electrodes.
Indications for Use
Indicated for creation of an atrial septal defect in the heart.
Regulatory Classification
Identification
A septostomy catheter is a special balloon catheter that is used to create or enlarge the atrial septal defect found in the heart of certain infants.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 27, 2026
Atraverse Medical
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Dr. Suite #510k
Saint Paul, Minnesota 55114
Re: K260292
Trade/Device Name: HOTWIRE™ RF Guidewire (901xxx); HOTWIRE™ RF Guidewire (902xxx)
Regulation Number: 21 CFR 870.5175
Regulation Name: Septostomy catheter
Regulatory Class: Class II
Product Code: DXF
Dated: January 29, 2026
Received: January 29, 2026
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260292 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260292 - Prithul Bom
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Katherine N. Trivedi -S
Digitally signed by
Katherine N. Trivedi -S
Date: 2026.02.27 21:00:48
-07'00'
Katherine Trivedi
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260292 | |
| Device Name HOTWIRE™ RF GUIDEWIRE (models 901XXX and 902XXX) | |
| Indications for Use (Describe) The HOTWIRE is indicated for creation of an atrial septal defect in the heart. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 5
# 510(K) SUMMARY
This 510(k) summary for HOTWIRE™ RADIOFREQUENCY (RF) GUIDEWIRE is submitted in accordance with the requirements of 21 CFR 807.87(h) and 807.92 and following the recommendation outlined in FDA Guidance document “The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notification [510(k)]” (issued on July 28, 2014).
Atraverse Medical, Inc. is submitting this Special 510(k) Premarket Notification for the HOTWIRE™ RF Guidewire. The HOTWIRE™ RF Guidewire has previously been cleared under the Traditional 510(k) Program (K252419), in which Atraverse Medical, Inc. is the legal manufacturer.
# SUBMITTER [807.92(a)(1)]
Atraverse Medical, Inc.
2611 S Coast Hwy 101 #204
Cardiff by the Sea, CA 92007
Contact Person: Charles Yang
SVP QA/RA
Telephone: +1 (760) 278-8098
Email: Charles.yang@atraversemedical.com
Date prepared: January 09, 2025
# DEVICE [807.92(a)(2)]
Table 1: Device Information
| Trade Name | HOTWIRE™ RF Guidewire |
| --- | --- |
| Common Name | Catheter, Septostomy |
| Classification Name | Catheter, Septostomy |
| Regulation | 21 CFR 870.5175 |
| Product Code | DXF |
| Regulatory Classification | Class II |
| Device Panel: | Cardiovascular |
# PREDICATE DEVICE [807.92(a)(3)]
HOTWIRE™ RF Guidewire (K252419)
{5}
# DEVICE DESCRIPTION [807.92(a)(4)]
# HOTWIRE™ Kit (901XXX):
The HOTWIRE™ is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and third-party RF electrosurgical generator(s), such as the Valleylab FT10 Electrosurgical Platform (K151649) which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572).
Table 2: Devices that have undergone Compatibility Testing
| Manufacturer | Device | 510(k) |
| --- | --- | --- |
| St Jude Medical / Abbott | Agilis NXT Steerable Introducer | K081645 |
| Medtronic | FlexCath Steerable Sheath | K183174 |
| Biosense Webster | CARTO VIZIGO Sheath | K170997 |
| Valleylab | FT10 Electrosurgical Platform | K151649 |
| Valleylab | E7507 Return Electrode | K822572 |
The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included Adapter Pin that connects to hand pieces used with compatible third-party RF electrosurgical generators.
# HOTWIRE™ + Handpiece Kit (902XXX):
The HOTWIRE™ + Handpiece Kit is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ + Handpiece Kit is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and the HOTWIRE™ System RF Generator, which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572).
Table 3: Devices that have undergone Compatibility Testing
| Manufacturer | Device | 510(k) |
| --- | --- | --- |
{6}
| St Jude Medical / Abbott | Agilis NXT Steerable Introducer | K081645 |
| --- | --- | --- |
| Medtronic | FlexCath Steerable Sheath | K183174 |
| Biosense Webster | CARTO VIZIGO Sheath | K170997 |
| Atraverse Medical | HOTWIRE™ System RF Generator | |
| Valleylab | E7507 Return Electrode | K822572 |
| McKesson Argent | 22-ESRSC Return Electrode | K092761 |
The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included handpiece that connects with the HOTWIRE™ System RF Generator.
## INDICATIONS FOR USE [807.92(a)(5)]
The HOTWIRE™ is indicated for creation of an atrial septal defect in the heart.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS [807.92(a)(6)]
The technological characteristics of the HOTWIRE™ is highly analogous to the technological characteristics of the Predicate device (K252419). Substantial equivalence is determined based on the following similarities:
- Same intended use/indications for use
- Same principles of operation
- Same fundamental scientific technology
- Same basic guidewire design
- Same guidewire construction material
Atraverse Medical, Inc. shall demonstrate that the device modifications made to the Predicate device remains substantially equivalent to the Predicate device. Specifics of this substantial equivalence determination is included in the discussion below.
The HOTWIRE™ RF Guidewire with device modifications (“Subject Device”) has identical or similar intended use, technological characteristics, and principles of operation as the Predicate
Page 3 of 5
{7}
device. A device comparison table is provided in this section in Table 3. This table provides a concise comparison between the relevant features of the Subject and the Predicate device.
Table 3: HOTWIRE™ Comparison Table
| Characteristic | Subject Device Compared to Predicate Device (K252419) | Comment |
| --- | --- | --- |
| Indications for Use | Identical | The subject device and predicate device are indicated for creation of an atrial septal defect in the heart. |
| Fundamental scientific technology | Identical | The subject device and predicate device a device rely on use of controlled RF energy by the user for transseptal puncture. |
| Principal of Operation | Identical | The subject device and predicate device are operator controlled. RF energy is delivered via a compatible RF Generator / electrosurgical handpiece to the RF wire distal tip by the operator. |
| Method of Supply | Identical | Both the subject and predicate device are single use; Sterilized by Gamma radiation – Method VDmax25; Sterility Assurance Level of 10-6 |
| Technological characteristics (Dimensions, design, materials) | Similar | Both the predicate and subject device share the same fundamental design, in terms of: • Insulation • Dimensions (e.g., overall outer diameter, overall length) • Configuration (i.e., tip shape) The Subject Device differs from the predicate as follows: • Material (i.e., Dymax 1184-M-T) The minor design differences (i.e., adding of material, core wire grind profile) between the Subject and Predicate device do not raise new questions of safety or effectiveness. |
| Energy Source | Identical | The subject device and predicate device can be used with 510(k) cleared compatible RF electrosurgical generator. |
| Packaging configuration | Identical | The Subject and Predicate device packaging configuration are identical. The HOTWIRE™ RF Guidewire (K252419) packaging was validated after environmental conditioning per ASTM D4332, transportation simulation per ASTM 4169 and 1 year accelerated aged equivalent. |
{8}
Page 5 of 5
# PERFORMANCE DATA [807.92(b)]
Non-clinical bench top performance testing was completed to demonstrate safety and effectiveness and substantial equivalence of the subject device to the predicate device. All test requirements were met as specified by applicable standards and test protocols. The following verification and validation activities were completed to demonstrate the safety and effectiveness of the subject device:
- RF Guidewire Testing
- Visual inspection
- Simulated use test
- Arc integrity test
- Fracture resistance test
- Flexing test
- Tensile Strength
**Biocompatibility**: Biocompatibility assessed for subject device. No impact to biocompatibility requirements. Refer to BEP-001 attached to the eSTAR.
# CONCLUSION
The HOTWIRE™ RF Guidewire w/ device modifications (Subject Device) met all specified criteria and did not raise new safety or performance questions. Based on the 510(k) summary and information provided herein, we conclude that the subject device, HOTWIRE™ RF Guidewire is substantially equivalent in its intended use, design, material, performance, and the underlying fundamental scientific technology used, to the Predicate device (K252419).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.