The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults.
Device Story
Automatic digital blood pressure monitor; uses oscillometric method to measure systolic/diastolic blood pressure and pulse rate. Device consists of main unit (pump, valve, PCB, display) and inflatable arm cuff (22-32cm circumference). Pump inflates cuff; electronic valve manages deflation; semiconductor sensor converts cuff pressure changes into electrical signals. Software algorithm analyzes pulse waveform data to calculate blood pressure and pulse rate. Used in hospitals, clinics, or home environments by patients or clinicians. Output displayed on unit; aids in monitoring cardiovascular status. Powered by batteries or AC adapter via Type-C/Micro USB port (power only).
Clinical Evidence
No new clinical studies conducted. Clinical performance leveraged from predicate device (K170466) based on identical measurement principle, patient population, and accuracy specifications (±3 mmHg static pressure, ±5% pulse rate).
Technological Characteristics
Oscillometric measurement principle; semiconductor pressure sensor. Materials: cuff biocompatible per ISO 10993. Power: 4x AA/AAA batteries or AC adapter (Type-C/Micro USB). Standards: ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-30. Connectivity: None (USB port for power only).
Indications for Use
Indicated for non-invasive measurement of systolic and diastolic arterial blood pressure and pulse rate in adult patients.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Automatic Digital Blood Pressure Monitor LD-537 (K170466)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 28, 2026
Honsun (Nantong) Co., Ltd.
Jian Shen
Director of Regulatory Affairs
#8 Tongxing Rd., Economic&Technological, Development Area
Nantong, Jiangsu 226009
China
Re: K260260
Trade/Device Name: Automatic Digital Blood Pressure Monitor (LD-587, LD-528, LD-562, LD-581, LD-521, LD-527, LD-572, LD-569)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: April 2, 2026
Received: April 2, 2026
Dear Jian Shen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260260 - Jian Shen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260260 - Jian Shen
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN C. BROWNING -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260260 | |
| Device Name Automatic Digital Blood Pressure Monitor (LD-587, LD-528, LD-562, LD-581, LD-521, LD-527, LD-572, LD-569) | |
| Indications for Use (Describe) The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Sponsor: HONSUN (Nantong) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
# Summary of special 510(K)
510K number: K260260
Date prepared: 22-April-2026
## 1. Submitter
Company: HONSUN(Nantong)Co.,Ltd.
Establishment registration number: 3007740533
Address: No.8, Tongxing Road, Nantong Economic &Technological development Area, Jiangsu, 226009, China
Phone: +86-513-80580116
Fax: +86-513-80580080
Contact Person: Jian Shen
Email: Jason@lordmed.com
## 2. Modified Device information
Name of Device: Automatic Digital Blood Pressure Monitor
Model: LD-587, LD-528, LD-562, LD-581, LD-521, LD-527, LD-572, LD-569
Common Name: Non-invasive blood pressure measurement system
Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
Regulation number: 21 CFR 870.1130
Regulatory Class: II
Product Code: DXN
Review Panel: Cardiovascular
## 3. Cleared Device information
Name of Device: Automatic Digital Blood Pressure Monitor
Model: LD-537
510k number: K170466
Common Name: Non-invasive blood pressure measurement system
Regulation number: 21 CFR 870.1130
Regulatory Class: II
Product Code: DXN
Review Panel: Cardiovascular
## 4. Device description
The Automatic Digital Blood Pressure Monitor is an automatic, non-invasive, blood pressure measurement device that is intended to measure the systolic and diastolic arterial blood pressure and pulse rate. The systolic and diastolic pressure are determined using the oscillometric method, where the cuff is inflated with a pump and deflates via an automatic electronic valve. During the inflation measurements, a pump within the main unit slowly
K260260
Page 1 of 8
{5}
Sponsor: HONSUN (Nantong) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
inflates the cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the sensor to determine systolic pressure and diastolic pressure.
The Automatic Digital Blood Pressure Monitor consists of two parts: main unit and cuff. The main unit is mainly composed of pump, valve, PCB, enclosure and display. The cuff, which is applicable to arm circumference approximately between 22 and 32cm, includes the inflatable bladder and the shell.
# 5. Intended use and indications for use
The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults.
# 6. Operation principle
The Automatic Digital Blood Pressure Monitor is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic semiconductor sensor rather than stethoscope and aneroid or mercurial manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".
7. Comparisons of technological characteristic with the predicate device
| Features | Subject Device | Predicate Device | Remark |
| --- | --- | --- | --- |
| General Comparison | | | |
| Applicant | HONSUN (NANTONG) Co., Ltd. | HONSUN (NANTONG) Co., Ltd. | / |
| Device name | Non-invasive blood pressure measurement system | Non-invasive blood pressure | / |
| Model | LD-587, LD-528, LD-562, LD-581, LD-521, LD-527, LD-572, LD-569 | LD-537 | |
| Classification | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
K260260
Page 2 of 8
{6}
Sponsor: HONSUN (Nantong) Co.,Ltd.
Device: Automatic Digital Blood Pressure Monitor
| Regulation | | | |
| --- | --- | --- | --- |
| Classification and Code | Class II, DXN | Class II, DXN | Same |
| Intended use | The device is a non-invasive blood pressure monitoring instrument intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate in adults. | The Scian Automatic Digital Blood Pressure Monitor LD-518&LD-537 are devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Besides, they can detect the appearance of irregular heartbeat during measurement, and give a warning signal once the irregular heartbeat is detected. | Discussion 1 |
| Patient Populations | Adults | Adults | Same |
| Performance Comparison | | | |
| Principle | Oscillometric method | Oscillometric method | Same |
| Anatomical sites | Upper Arm | Upper Arm | Same |
| Where used (hospital, home, ambulance, etc.) | Hospital, clinics, home | Hospital, clinics ,home | Same |
| Blood Pressure Measurement | 40 to 180 mmHg (DIA diastolic pressure)60 to 260 mmHg (SYS, systolic | 40 to 260 mmHg (blood pressure)40 to 160 beats/minute | Discussion 2 |
K260260
Page 3 of 8
{7}
Sponsor: HONSUN (Nantong) Co.,Ltd.
Device: Automatic Digital Blood Pressure Monitor
| nt | pressure)40 to 160 beats/minute (PUL, pulse rate) | | (pulse rate) | |
| --- | --- | --- | --- | --- |
| Pulse rate measurement | ±3 mmHg for static pressure± 5% of the reading for the pulse rate | | ±3 mmHg for static pressure±5% of the reading for the pulse rate | Same |
| Cuff Deflation | Automatic deflation | | Automatic deflation | Same |
| Energy used and/or delivered | 4*AA/AAA batteries; Type-C/Micro USB AC adapter | | 4xAA batteries: AC adapter | Discussion 3 |
| Human factors | Blood pressure | | Blood pressure | Same |
| Performance | Measuring systolic and diastolic blood pressure and pulse rate of adult individual | | Measuring systolic and diastolic blood pressure and pulse rate of adult individual | Same |
| Compatibility with environment and other devices | Temperature: 5°C~40°C, Relative humidity :15%~90%RH, Atmospheric: 70KPa~106KPa. | | Temperature: 10°C~40°C, Relative humidity :15%~85%RH, Atmospheric: 700hPa~1060hPa. | Discussion 4 |
| Cuff range | 22cm~32cm | | 22cm~42cm (Standard)22cm~32cm(optional) | Same |
| Function | Memory recall:2*100 sets | LD-587/LD-528 | ● Memory recall:2*90 sets;● Voice broadcast● WHO Blood Pressure Classification Indication● Setting the Date and Time● Irregular Heartbeat Detector | Discussion 5 |
| | Memory recall:2*90 sets | LD-562/LD-581/LD-527/LD-569 | | Same |
| | Memory recall:90 sets | LD-521/LD-572 | | |
K260260
Page 4 of 8
{8}
Sponsor: HONSUN (Nantong) Co.,Ltd.
Device: Automatic Digital Blood Pressure Monitor
| | WHO Blood Pressure Classification Indication | LD-587/LD-528/LD-562/LD-581/LD-521/LD-527/LD-569 | | |
| --- | --- | --- | --- | --- |
| | Voice broadcast | LD-587/LD-528 | | |
| | Irregular Heartbeat Detector | LD-587/LD-528/LD-562/LD-581/LD-521/LD-572/LD-527/LD-569 | | |
| | Setting the Date and Time | LD-587/LD-528 | | |
| EMC, Electrical and biological Safety Comparison | | | | |
| Biocompatibility | Cuff, according to ISO-10993 | | Cuff, according to ISO-10993 | Same |
| Electrical safety | According to IEC 60601-1-2
According to IEC 60601-1
According to IEC 60601-1-11 | | According to IEC 60601-1-2
According to IEC 60601-1
According to IEC 60601-1-11 | Same |
## Justification of difference:
## Discussion 1: Intended use
The indication for use is same as the predicate devices. They are intended for the measurement of systolic and diastolic arterial blood pressure and pulse rate. The difference description of the intended use occurred is considered not raise any issues in safety and effectiveness and they can be considered substantially equivalent in safety and effectiveness.
## Discussion 2: Blood Pressure Measurement
The blood pressure measurement range of the predicated device is 40 to 260 mmHg, while the measurement range of the proposed device under review has been further subdivided into 40 to 180 mmHg (DIA diastolic pressure) and 60 to 260 mmHg (SYS, systolic pressure). The actual measurement ranges for both are identical. Therefore, the difference is considered not raise any issues in safety and effectiveness.
## Discussion 3: Energy used and/or delivered
The charging method of the proposed device is same as predicate device. While the
K260260
Page 5 of 8
{9}
Sponsor: HONSUN (Nantong) Co.,Ltd.
Device: Automatic Digital Blood Pressure Monitor
specification of batteries and adapters are different. The charging interface of the proposed device has been changed to Type-C/Micro USB port. Note: the micro-USB/Type-C port is limited to power delivery only and has no actual communication functionality. The battery of the target device has been changed to 4 * AA or 4AAA batteries. The test results in IEC 60601-1 demonstrated that the proposed device is safety and the performance satisfied the specification. Therefore, the difference is considered would not raise any issues in safety and effectiveness and they can be considered substantially equivalent.
## Discussion 4: Compatibility with environment and other devices
The operation condition of the proposed device is different from predicate device. The IEC 60601-1 and IEC60601-1-2 can prove equivalent safety and performance. Therefore, the difference is considered would not raise any issues in safety and effectiveness and they can be considered substantially equivalent.
## Discussion 5: Function
The memory recall times of the proposed device is more than predicate device. The ISO 81060-2-30 can prove equivalent safety and performance. Therefore, the difference is considered would not raise any issues in safety and effectiveness and they can be considered substantially equivalent.
## Conclusion
Based on above comparative analysis, the subject device performs comparably to the predicate device LD-537 (K170466). Thus, the subject device is substantially equivalent to the predicate device.
## 8. Assessment of Non-clinical testing
## 8.1 Electrical safety and Electromagnetic Compatibility Testing
The electrical safety and EMC testing were conducted on the subject device. The related standards are shown as follows:
| Standard | Descriptions |
| --- | --- |
| ANSI AAMI ES 60601-1 | Medical electrical equipment-Part 1: General requirements for basic safety and essential performance |
| IEC 60601-1-2 | Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance- Collateral Standard: Electromagnetic disturbances -Requirements and tests |
| IEC 60601-1-11 | Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment |
| ISO 80601-2-30 | Medical electrical equipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers |
K260260
Page 6 of 8
{10}
Sponsor: HONSUN (Nantong) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
## 8.2 Biocompatibility Evaluation
The biocompatibility evaluations of the subject device were conducted in accordance with the International Standard ISO 10993-1 “Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing within a Risk Management Process” and FDA biocompatibility guidance, the Cytotoxicity test, Sensitization test, Irritation test are needed. The subject devices use the same patient-contact materials(cuff) with the predicate device, so the biocompatibility is not affected and it will not affect the effectiveness or raise any safety issues.
## 8.3 Software Verification and Validation Evaluation
Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained In Medical Devices-Guidance for Industry and FDA Staff, May 2005.
## 9. Clinical Testing
### Clinical Data Leveraging Statement:
The subject device leverages the clinical testing data from its predicate device, model LD-537 (K170466). As demonstrated in the comparison table (Section 7), the subject device shares the same intended patient population (adults), identical blood pressure and pulse rate measurement ranges, and identical accuracy specifications ( $\pm 3$ mmHg for static pressure, $\pm 5\%$ of reading for pulse rate) as the predicate device. Both devices utilize the same fundamental oscillometric measurement principle and are intended for use in the same environments (hospital, clinic, home).
Given these core equivalencies in design, intended use, and performance specifications, the clinical data supporting the safety and effectiveness of the predicate device LD-537 is directly applicable and sufficient to support the substantial equivalence of the subject device models LD-587, LD-528, LD-562, LD-581, LD-521, LD-527, LD-572, LD-569. The minor modifications identified in Discussions 1 through 5 are not related to clinical performance and do not affect the validity of the leveraged clinical data. Therefore, no new clinical studies are required for the subject device.
## 10. Conclusion
The subject device shares the same intended patient population (adults), identical blood pressure and pulse rate measurement ranges, and identical accuracy specifications ( $\pm 3$ mmHg for static pressure, $\pm 5\%$ of reading for pulse rate) as the predicate device. Both devices utilize the same fundamental oscillometric measurement principle and are intended for use in the same environments (hospital, clinic, home).
Base on the performance tests and associate with the non-clinical testing's assessment, the subject device performs comparably to the predicate device.
Furthermore, the clinical data from the predicate device (K170466) is leveraged and remains
K260260
{11}
Sponsor: HONSUN (Nantong) Co., Ltd.
Device: Automatic Digital Blood Pressure Monitor
valid for the subject device due to the identical patient population, measurement ranges, and accuracy specifications. Thus, the subject device is substantially equivalent to the predicate device.
K260260
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.