K260246 · Bolt Medical, Inc. · PPN · Aug 24, 2026 · Cardiovascular
Device Facts
Record ID
K260246
Device Name
SEISMIQ™ Intravascular Lithotripsy System
Applicant
Bolt Medical, Inc.
Product Code
PPN · Cardiovascular
Decision Date
Aug 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SEISMIQ™ IVL System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, iliofemoral, popliteal, and infra-popliteal arteries. Not for use in the coronary, carotid or cerebral vasculature.
Device Story
System comprises reusable console and single-use balloon catheter; used in peripheral vasculature to treat stenotic, calcified lesions. Catheter delivered to lesion site via peripheral arterial system; balloon partially inflated. Console delivers energy through cabling to emitters inside balloon; emitters generate pulsatile mechanical energy (lithotripsy) to disrupt calcium deposits. Disruption allows subsequent low-pressure balloon dilatation to restore blood flow. Operated by clinicians in clinical settings. Output is mechanical energy for vessel wall modification; clinical benefit is improved vessel patency via facilitated dilatation.
Clinical Evidence
No new clinical data provided. Leverages clinical evidence (RESTORE ATK and RESTORE BTK studies) and porcine in vivo chronic GLP study from predicate K250225, as modifications do not impact delivered therapy.
Technological Characteristics
System includes balloon catheter and console. Materials and sterilization methods identical to predicate. Energy source: pulsatile mechanical energy (lithotripsy). Features: variable balloon diameters/lengths, fluoroscopic visibility. Biocompatibility per ISO 10993-1:2020. Bench testing covers tensile strength, kink resistance, RBP, compliance, fatigue, and sonic output.
Indications for Use
Indicated for lithotripsy-enhanced balloon dilatation of calcified and non-calcified lesions in peripheral arteries (iliac, femoral, iliofemoral, popliteal, infra-popliteal). Contraindicated for coronary, carotid, or cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
SEISMIQ Intravascular Lithotripsy System (K250225)
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 24, 2026
Bolt Medical, Inc.
Stephanie Onstot
Regulatory Affairs Manager
2131 Faraday Ave.
Carlsbad, California 92008
Re: K260246
Trade/Device Name: SEISMIQ™ Intravascular Lithotripsy System
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: PPN
Dated: July 24, 2026
Received: July 24, 2026
Dear Stephanie Onstot:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260246 - Stephanie Onstot
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260246 - Stephanie Onstot
Page 3
Sincerely,
GREGORY W. O'CONNELL -S Digitally signed by GREGORY W. O'CONNELL -S Date: 2026.08.24 14:26:55 -04'00'
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260246 | ? |
| Please provide the device trade name(s). | | ? |
| SEISMIQ^{™}Intravascular Lithotripsy System | | |
| Please provide your Indications for Use below. | | ? |
| The SEISMIQ^{™}IVL System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, iliofemoral, popliteal, and infra-popliteal arteries. Not for use in the coronary, carotid or cerebral vasculature. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
# 510(k) Summary – K260246
Name, Address, Phone Number of Applicant
Bolt Medical, Inc.
2131 Faraday Ave
Carlsbad, CA 92008
Phone: 760-979-1890
Contact Person
Stephanie Onstot
Date Prepared
August 21, 2026
Device Name and Classification
| Trade Name: | SEISMIQ™ Intravascular Lithotripsy System |
| --- | --- |
| Common Name: | Catheter, lithotripsy, peripheral, transluminal |
| CFR Classification: | 21 CFR 870.1250 |
| Classification Name: | Percutaneous catheter |
| Product Code: | PPN |
Predicate Device
The primary predicate for the SEISMIQ Intravascular Lithotripsy (IVL) System is the SEISMIQ Intravascular Lithotripsy System (Previously named the Bolt IVL System, K250225) cleared by FDA on March 25, 2025.
Reference Device
The reference device for the Bolt IVL Lithotripsy System is the Shockwave Medical Lithoplasty System S⁴ (K191840) cleared by FDA on August 7, 2019.
Indications for Use / Intended Use
The SEISMIQ (IVL) System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, iliofemoral, popliteal, and infra-popliteal arteries. Not for use in the coronary, carotid or cerebral vasculature.
Device Description
The SEISMIQ Intravascular Lithotripsy System is a proprietary balloon catheter and console designed to enhance percutaneous transluminal angioplasty by delivering calcium disrupting lithotripsy prior to balloon dilatation at low pressures. The application of lithotripsy mechanical pulse waves alters the structure of an occlusive vascular deposit (stenosis) allowing low-pressure balloon dilatation of the stenosis and facilitates the passage of blood.
The SEISMIQ IVL Catheter is delivered through the peripheral arterial system of the lower extremities to the lesion site. The balloon is partially inflated, and the lithotripsy emitters generate pulsatile mechanical energy within the balloon at the target treatment site allowing subsequent dilatation of a peripheral artery stenosis using low balloon pressure. The SEISMIQ IVL Catheter is a single-use device supplied sterile to the customer.
K260246
Page - 1
{5}
Bolt Medical, Inc.
Premarket Notification, Traditional 510(k)
SEISMIQ™ IVL System
The SEISMIQ IVL Console delivers energy through the integrated catheter cabling to the emitters located inside the catheter balloon in the SEISMIQ IVL Catheter. The SEISMIQ IVL Console is a non-sterile, reusable device.
## Technological Comparison
This 510(k) Premarket Notification is being submitted for minor modifications to the SEISMIQ™ IVL System. The product line extension adds additional catheter sizes with a smaller ballon diameter and length to the SEISMIQ IVL System portfolio.
To increase the size range of the system, the additional catheter sizes include a smaller ballon diameter and length as compared to catheters cleared under K250225. Differences between the two catheters do not raise new questions of safety and effectiveness.
The SEISMIQ IVL System has the same intended use principles of operation and has substantially equivalent technological characteristics, including the same fundamental scientific technology, design, energy source, shelf life and sterilization as the predicate IVL System.
## Summary of Performance Data
Objective evidence demonstrating that the SEISMIQ IVL System design output meets the product design input requirements as well as that device performance characteristics conform to user needs and intended uses as defined in the product specification was provided. Testing was conducted in accordance with Bolt Medical's risk analysis procedures, and all applicable FDA guidance documents and relevant international standards. Testing included:
- Catheter diameter and balloon profile
- Tensile strength
- Kink resistance / flexibility
- Catheter torsional strength
- Balloon Inflation / deflation time
- Minimum RBP
- Balloon compliance
- Fatigue (multiple inflations)
- Pushability and trackability
- Fluoroscopic visibility
- Particulate evaluation
- Pulsing cycles
- Sonic output
Results demonstrate that the performance of the SEISMIQ IVL System meets its design specifications and demonstrates substantial equivalence for its intended use.
## Summary of Biocompatibility Testing
Biocompatibility testing was conducted on the predicate device (K250225) in accordance with ISO 10993-1:2020 and FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and leveraged for the subject device.
K260246
Page - 2
{6}
Bolt Medical, Inc.
Premarket Notification, Traditional 510(k)
SEISMIQ™ IVL System
## Summary of Animal Testing
A chronic GLP study was conducted in a porcine in vivo model to determine the safety of delivering pulsatile mechanical energy to peripheral arteries with the Bolt IVL System, and submitted as part of K250225. There have been no modifications made to the subject device that impact the delivered therapy of the SEISMIQ IVL System. Therefore, data from K250225 is leveraged for the subject device submission.
## Summary of Clinical Data
Clinical evidence which evaluated the safety and performance of the Bolt IVL System was submitted as part of K250225 for above the knee (RESTORE ATK) and below the knee (RESTORE BTK), and is leveraged for the subject device submission. There have been no modifications made to the subject device that impact the delivered therapy of the SEISMIQ IVL System.
## Basis for Substantial Equivalence
The SEISMIQ IVL System shares the same intended use and mechanism of action as well as similar technological characteristics as the predicate device. Any differences between the SEISMIQ IVL System and the predicate device were evaluated through bench testing which demonstrated acceptable device performance and confirmed that there are no new questions of safety or effectiveness. Therefore, the SEISMIQ IVL System is substantially equivalent to the predicate device.
K260246
Page - 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.