← Product Code [LSE](/productcode/LSE) · K260232

# Elecsys EBV VCA IgG (K260232)

_Roche Diagnostics · LSE · Oct 2, 2026 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K260232

## Device Facts

- **Applicant:** Roche Diagnostics
- **Product Code:** [LSE](/productcode/LSE.md)
- **Decision Date:** Oct 2, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3235
- **Device Class:** Class 1
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K260232 · Oct 2, 2026 | Elecsys EBV VCA IgG | Roche Diagnostics | Retrospective clinical laboratory samples | Retrospective clinical samples were used to supplement prospective clinical data to evaluate the Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) of the Elecsys EBV VCA IgG assay against a composite consensus comparator. | Retrospective study; Clinical performance; EBV serology |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Multisite clinical study (retrospective cohort); Retrospective clinical performance evaluation; Follow-up/Duration: Not applicable | 253 retrospective samples (including 192 pediatric) selected based on routine EBV test results to be presumed acute or presumed seronegative.; Sample Size: 253 | Composite consensus comparator (based on 3 FDA-cleared EBV assay panels) | Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) |

## Indications for Use

Immunoassay for the in vitro qualitative detection of IgG antibodies to Epstein-Barr virus, including viral capsid antigens (VCA), in human serum and plasma. When performed in conjunction with other Epstein-Barr virus markers, this assay is intended for use as an aid in the diagnosis of Epstein-Barr virus infection, such as infectious mononucleosis. The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.

## Device Story

Fully automated electrochemiluminescence immunoassay (ECLIA) for qualitative detection of IgG antibodies to EBV VCA; utilizes cobas e 801 analyzer. Input: human serum or plasma samples. Process: 18-minute sandwich immunoassay; biotinylated recombinant EBV-specific antigens and ruthenium-labeled antigens form complex with sample antibodies; streptavidin-coated microparticles capture complex; voltage application induces chemiluminescent emission measured by photomultiplier. Output: qualitative result (reactive/non-reactive) based on COI cutoff. Used in clinical laboratories; operated by trained personnel. Results aid clinicians in diagnosing EBV infection status (e.g., acute vs. past infection) when interpreted alongside other EBV markers.

## Clinical Evidence

Multisite clinical study of 974 specimens (721 prospective, 253 retrospective) comparing Elecsys EBV VCA IgG to composite consensus comparator. Prospective adult cohort: 100% PPA, 61.9% NPA. Prospective pediatric cohort: 99.0% PPA, 92.7% NPA. Retrospective adult cohort: 91.3% PPA, 81.6% NPA. Retrospective pediatric cohort: 92.3% PPA, 83.5% NPA. Overall performance across serological statuses demonstrated clinical utility as an aid in diagnosis.

## Technological Characteristics

Electrochemiluminescence immunoassay (ECLIA). Reagents: biotinylated recombinant EBV-specific antigen, ruthenium-labeled recombinant EBV-specific antigen, streptavidin-coated microparticles. Platform: cobas e 801 analyzer. Automated barcode identification. Calibration: lot-specific, 12-week or 28-day stability. Standards: CLSI EP05-A3, EP09c, EP07, EP37, EP25-A, EP12-A2, I/LA21-A2.

## Regulatory Identification

Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).

## Predicate Devices

- LIAISON VCA IgG ([K040120](/device/K040120.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

[LOGO]

FDA

U.S. FOOD & DRUG

ADMINISTRATION

### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY

ASSAY AND INSTRUMENT

### I Background Information:

A 510(k) Number

K260232

B Applicant

Roche Diagnostics

C Proprietary and Established Names

Elecsys EBV VCA IgG

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LSE | Class I | 21 CFR 866.3235 - Epstein-Barr Virus Serological Reagents | MI - Microbiology  |

### II Submission/Device Overview:

A Purpose for Submission:

To obtain clearance of a new device

B Measurand:

Epstein Barr virus IgG antibodies to Viral Capsid Antigen (VCA) IgG

C Type of Test:

Fully automated qualitative chemiluminescent immunoassay (CLIA)

### III Intended Use/Indications for Use:

A Intended Use(s):

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

See Indications for Use below.

### B Indication(s) for Use:

Immunoassay for the in vitro qualitative detection of IgG antibodies to Epstein-Barr virus, including viral capsid antigens (VCA), in human serum and plasma. When performed in conjunction with other Epstein-Barr virus markers, this assay is intended for use as an aid in the diagnosis of Epstein-Barr virus infection, such as infectious mononucleosis.
The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.

### C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

### D Special Instrument Requirements:

cobas e 801 Analyzer

### IV Device/System Characteristics:

### A Device Description:

The Elecsys EBV VCA IgG is an electrochemiluminescence immunoassay (ECLIA) to detect IgG antibodies to EBV VCA.
Elecsys EBV VCA IgG kit is comprised of a cobas e pack and the lot-specific calibrators, EBVVCAG Cal1 and EBVVCAG Cal2. The cobas e pack consists of Reagent 1 (R1), Reagent 2 (R2) and Microparticles (M):

- R1 contains biotinylated EBV-VCA specific antigen
- R2 contains EBV VCA-specific antigen (recombinant, E. coli) labeled with ruthenium complex
- M contains streptavidin-coated microparticles.

Calibrators are a two-concentration level set, negative calibrator 1 and positive calibrator 2. The PreciControl EBV IgM/VCA IgG kits are required to perform the assay and are available separately from the cobas e pack. There are 2 levels of control material provided: Negative control, EBV MG1, and positive control EBV MG3.

### B Principle of Operation:

The Elecsys EBV VCA IgG is a qualitative, serological, sandwich immunoassay to be used on the cobas e 801 immunoassay analyzer with an overall 18-minute test time.

- 1st incubation: 21 μL of sample, biotinylated recombinant EBV-specific antigens and EBV-specific recombinant antigens labeled with a ruthenium complex form a sandwich complex.
- 2nd incubation: After addition of streptavidin-coated microparticles, the complex

K260232 - Page 2 of 15

{2}

becomes bound to the solid phase via interaction of biotin and streptavidin.

- microparticles bound to the sandwich complex are magnetically captured onto the surface of the electrode. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
- results are determined automatically by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value obtained by calibration.

### Interpretation of results

Results obtained with the Elecsys EBV VCA IgG test can be interpreted as follows:

|  Numeric result | Result message | Interpretation  |
| --- | --- | --- |
|  COI < 1.0 | Non-reactive | Negative for EBV-VCA-specific IgG  |
|  COI ≥ 1.0 | Reactive | Positive for EBV-VCA-specific IgG  |

### C Instrument Description Information:

1. Instrument Name:

cobas e 801 analyzer

2. Specimen Identification:

The system automatically reads barcode information from specimen tubes to identify and track samples throughout the testing process.

3. Specimen Sampling and Handling:

The cobas e 801 analyzer handles samples through an automated rack-based system, supporting continuous loading/unloading of up to 300 samples per hour for high-throughput immunoassays.

4. Calibration:

Calibration is performed as follows:

- once per reagent lot using EBVVCAG Cal1, EBVVCAG Cal2 and fresh reagent
• every 12 weeks when using the same reagent lot
• every 28 days when using the same cobas e pack on the analyzer
- as required, such as when quality control findings are outside the defined limits.

5. Quality Control:

The PreciControl EBV IgM/VCA IgG QC is run:

• every 24 hours when the test is in use
- ones per cobas e pack
• following a valid calibration.

K260232 - Page 3 of 15

{3}

Acceptable performance is achieved when the analyte values obtained are within the expected control interval for the system.

# V Substantial Equivalence Information:

A Predicate Device Name(s):

LIAISON VCA IgG

B Predicate 510(k) Number(s):

K040120

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K040120 Predicate Device | K260232 Candidate Device  |
| --- | --- | --- |
|  Device Trade Name | LIAISON VCA IgG | Elecsys EBV VCA IgG  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | The LIAISON VCA IgG assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON Analyzer family* for the qualitative determination of specific IgG antibodies to Epstein-Barr virus (EBV) viral capsid antigen (VCA) p18 synthetic peptide in human serum. When performed in conjunction with other EBV markers, this assay can be used as an aid in the clinical laboratory diagnosis of Epstein-Barr Viral Syndrome in patients with signs and symptoms of EBV infection such as infectious mononucleosis. Caution: U.S. Federal Law restricts this device to sale by or on the order of a licensed practitioner. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens, or infants. Assay performance characteristics have not been established for the diagnosis of nasopharyngeal carcinoma, Burkitt's lymphoma, and other EBV-associated lymphomas. | Immunoassay for the in vitro qualitative detection of IgG antibodies to Epstein-Barr virus, including viral capsid antigens (VCA), in human serum and plasma. When performed in conjunction with other Epstein-Barr virus markers, this assay is intended for use as an aid in the diagnosis of Epstein-Barr virus infection, such as infectious mononucleosis. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.  |

K260232 - Page 4 of 15

{4}

|  Measurand | IgG antibodies to the viral capsid antigen (VCA) of EBV | Same  |
| --- | --- | --- |
|  Assay Type | Qualitative | Same  |
|  Technology | Chemiluminescence immunoassay (CLIA) | Same  |
|  General Device Characteristic Differences |  |   |
|  Sample Matrix | Serum | Serum Li-heparin, K2-EDTA, K3-EDTA  |
|  Result Interpretation | • < 18.0 U/mL Negative • 18.0 to 21.9 U/mL Equivocal • ≥ 22.0 U/mL Positive | COI < 1.0 non-reactive COI ≥ 1.0 reactive  |
|  Application/Test Time | 30 minutes LIAISON Analyzer 15 minutes LIAISON XL Analyzer | 18 minutes  |
|  Reagent Composition | Coated Magnetic particles Conjugate: Mouse monoclonal antibodies to human IgG | Streptavidin-coated microparticles R1: Biotinylated EBV-specific antigen (recombinant) R2: EBV-specific antigen with ruthenium complex  |
|  Calibrator | Human serum/defibrinated plasma containing low and high EBV VCA IgG levels | Human serum lyophilized positive and negative for EBV VCA IgG  |
|  Instrument Platform | LIAISON Analyzer LIAISON XL Analyzer | Cobas e analytical unit  |

### VI Standards/Guidance Documents Referenced:

- CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition (Reaffirmed: September 2019).
- CLSI EP09c Measurement Procedure Comparison and Bias Estimation Using Patient Samples - 3rd Edition (2018).
- CLSI EP07 Interference Testing in Clinical Chemistry; Approved Guideline - 3rd Edition (2018).
- CLSI EP37 Supplemental Tables for Interference Testing in Clinical Chemistry - 1st Edition (2018).
- CLSI EP25-A (Replaces EP25-P) Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline (2009).
- CLSI EP12-A2 User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - 2nd Edition (2008).
- CLSI I/LA21-A2 Clinical Evaluation of Immunoassays; Approved Guideline - 2nd Edition.

### VII Performance Characteristics (if/when applicable):

#### A Analytical Performance:

K260232 - Page 5 of 15

{5}

## 1. Precision/Reproducibility:

### A. Within-Laboratory Precision Study

A precision study was conducted to establish within-laboratory precision of the Elecsys EBV VCA IgG on the cobas e 801 analyzer. Five human serum samples covering the analytical range and two controls were tested at: 1 run per day, in replicate of 4 each for 21 days (n = 84).

In addition, a study was conducted to demonstrate lot-to-lot variability between three lots of Elecsys EBV VCA IgG at 1 run per day for 5 days on one cobas e 801 analyzer. Precision study results on cobas e 801 analyzer are summarized in Tables 1 and 2.

Table 1. Elecsys EBV VCA IgG within-laboratory precision study results on the cobas e 801

|  Sample | N | Mean COI | Repeatability |   | Between-Run/Day |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | % CV | SD | % CV | SD | % CV  |
|  Sample 1 | 84 | 0.351 | 0.014 | 3.88 | 0.015 | 4.38 | 0.021 | 5.85  |
|  Sample 2 | 84 | 0.855 | 0.029 | 3.39 | 0.020 | 2.37 | 0.035 | 4.14  |
|  Sample 3 | 84 | 1.78 | 0.039 | 2.18 | 0.035 | 1.96 | 0.052 | 2.94  |
|  Sample 4 | 84 | 17.8 | 0.343 | 1.93 | 0.463 | 2.61 | 0.577 | 3.2  |
|  Sample 5 | 84 | 77.2 | 1.40 | 1.81 | 1.39 | 1.80 | 1.97 | 2.55  |
|  PC EBV-MG1 | 84 | 0.420 | 0.0134 | 3.2 | 0.012 | 2.91 | 0.019 | 4.50  |
|  PC EBV-MG3 | 84 | 6.68 | 0.0879 | 1.3 | 0.127 | 1.89 | 0.182 | 2.73  |

Table 2. Elecsys EBV VCA IgG 5-day Lot-to-Lot Precision on cobas e 801

|  Sample | N | Mean COI | Repeatability |   | Between-Run/Day |   | Between-Lot |   | Within-Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%)  |
|  Sample 1 | 75 | 0.379 | 0.00530 | 1.4 | 0.0132 | 3.5 | 0.0408 | 10.8 | 0.0432 | 11.4  |
|  Sample 2 | 75 | 0.386 | 0.00613 | 1.6 | 0.0148 | 3.8 | 0.0387 | 10.0 | 0.0419 | 10.9  |
|  Sample 3 | 75 | 0.667 | 0.00846 | 1.3 | 0.0265 | 4.0 | 0.000000 | 0.0 | 0.0278 | 4.2  |
|  Sample 4 | 75 | 0.708 | 0.00828 | 1.2 | 0.0260 | 3.7 | 0.0297 | 4.2 | 0.0404 | 5.7  |
|  Sample 5 | 75 | 0.870 | 0.00886 | 1.0 | 0.0318 | 3.7 | 0.0525 | 6.0 | 0.062 | 7.1  |

K260232 - Page 6 of 15

{6}

|  Sample 6 | 75 | 0.884 | 0.00903 | 1.0 | 0.0336 | 3.8 | 0.00555 | 0.6 | 0.0352 | 4  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Sample 7 | 75 | 2.79 | 0.0401 | 1.4 | 0.132 | 4.7 | 0.0941 | 3.4 | 0.167 | 6  |
|  Sample 8 | 75 | 16.4 | 0.415 | 2.5 | 0.679 | 4.1 | 1.09 | 6.6 | 1.35 | 8.2  |
|  Sample 9 | 75 | 61.1 | 1.17 | 1.9 | 2.89 | 4.7 | 4.07 | 6.7 | 5.12 | 8.4  |
|  PC EBV-MG1 | 75 | 0.420 | 0.00707 | 1.7 | 0.0162 | 3.9 | 0.0503 | 12.0 | 0.0533 | 12.7  |
|  PC EBV-MG3 | 75 | 6.80 | 0.0587 | 0.9 | 0.293 | 4.3 | 0.326 | 4.8 | 0.442 | 6.5  |

### B. Reproducibility

Reproducibility of the Elecsys EBV VCA IgG on the cobas e 801 analyzer was determined with a panel of 5 human sera and 2 controls. Samples were measured in triplicate determination using 1 reagent lot, in 2 runs per day for 5 days at 3 sites (n = 90). Variance components were calculated absolutely, as well as in relative contribution to the total size variance.

The reproducibility study results on the cobas e 801 analyzer are summarized in Tables 3.

Table 3. Elecsys EBV VCA IgG Reproducibility on the cobas e 801

|  Sample | N | Mean COI | Repeatability |   | Between-Run |   | Between-Day |   | Between-Sites |   | Reproducibility  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD COI | CV % | SD COI | CV % | SD COI | CV % | SD COI | CV % | SD COI | CV %  |
|  Sample 1 | 90 | 0.448 | 0.012 | 2.76 | 0.000 | 0.000 | 0.005 | 1.03 | 0.018 | 4.02 | 0.022 | 4.99  |
|  Sample 2 | 90 | 0.856 | 0.017 | 1.96 | 0.006 | 0.738 | 0.005 | 0.574 | 0.019 | 2.23 | 0.027 | 3.11  |
|  Sample 3 | 90 | 1.24 | 0.021 | 1.67 | 0.013 | 1.07 | 0.000 | 0.000 | 0.009 | 0.692 | 0.026 | 2.10  |
|  Sample 4 | 90 | 25.6 | 0.332 | 1.30 | 0.333 | 1.30 | 0.000 | 0.000 | 0.000 | 0.000 | 0.470 | 1.84  |
|  Sample 5 | 90 | 85.6 | 1.47 | 1.71 | 1.51 | 1.76 | 0.000 | 0.000 | 0.000 | 0.000 | 2.10 | 2.46  |
|  PC EBV-MG1 | 90 | 0.478 | 0.009 | 1.98 | 0.000 | 0.000 | 0.002 | 0.393 | 0.019 | 3.94 | 0.021 | 4.43  |
|  PC EBV-MG3 | 90 | 5.78 | 0.077 | 1.32 | 0.049 | 0.852 | 0.032 | 0.559 | 0.034 | 0.584 | 0.102 | 1.77  |

### 2. Linearity:

Not applicable as the device is a qualitative assay

### 3. Analytical Specificity/Interference:

### A. Cross-Reactivity

K260232 - Page 7 of 15

{7}

Potentially cross-reacting samples, consisting in characterized positive for potentially cross-reacting analytes, were tested with the Elecsys EBV VCA IgG assay. A total of 670 samples were tested. The following results were obtained.

Table 4. Cross-Reactivity Study Results

|  Cross-Reactants |  | Elecsys EBV VCA IgG  |   |
| --- | --- | --- | --- |
|   |  Number of samples tested | Number of Reactive | Number of Non-reactive  |
|  CMV | 61 | 2 | 59  |
|  Rubella | 215 | 3 | 212  |
|  Toxoplasmosis | 7 | 0 | 7  |
|  HAV | 111 | 1 | 110  |
|  HBV | 123 | 0 | 123  |
|  HIV^{A)} | 5 | 0 | 5  |
|  HCV | 7 | 1 | 6  |
|  HHV6 | 17 | 0 | 17  |
|  Treponema pallidum^{A)} | 5 | 0 | 5  |
|  Measles | 6 | 1 | 5  |
|  Mumps | 17 | 0 | 17  |
|  Borrelia burgdorferi | 6 | 1 | 5  |
|  VZV | 13 | 1 | 12  |
|  Parvovirus B19 | 5 | 0 | 5  |
|  SLE^{A)} | 5 | 0 | 5  |
|  HEV | 14 | 1 | 13  |
|  ANA^{A)} | 5 | 0 | 5  |
|  HSV-1 IgG | 27 | 1 | 26  |
|  HSV-2 IgG^{A)} | 8 | 0 | 8  |
|  EBV IgM | 8 | 0 | 8  |
|  EBV EBNA IgG^{A)} | 5 | 0 | 5  |

A) 4 samples spiked of HIV, Treponema, SLE, ANA, EBV EBNA IgG; 6 spiked samples of HSV-2

Cross-reactivity cannot be ruled out for CMV, Rubella, HAV, HCV, Borrelia burgdorferi, Measles, HSV-1, and HEV. No evidence for other cross-reactivity was observed.

### B. Interference

#### Endogenous Interference

The effect of the following endogenous substances and pharmaceutical compounds on assay performance was tested on the cobas analyzer. Interferences were tested up to the concentrations listed in Table 5. No significant interference with the assay was observed.

Table 5. Endogenous Substances tested Concentrations

|  Interferent Substance | Concentration tested  |
| --- | --- |
|  Bilirubin | 66 mg/dL  |
|  Hemoglobin | 1000 mg/dL  |

K260232 - Page 8 of 15

{8}

|  Intralipid | 2000 mg/dL  |
| --- | --- |
|  Biotin | 1200 ng/mL  |
|  Rheumatoid factors | 1200 IU/mL  |
|  Human serum albumin | 7 g/dL  |
|  Cholesterol | 400 mg/dL  |

### Exogenous interference

Testing was performed with 17 commonly used pharmaceuticals. Interferences were tested up to the concentrations listed in Table 6. No significant interference with the assay was found except for itraconazole. Itraconazole at 10 mg/L in the sample can lead to reduced COI values of up to 15%.

In rare cases, interference caused by extremely high titers of antibodies to analyte-specific antibodies, streptavidin, or ruthenium can occur. These effects are minimized by suitable test design.

Table 6. Exogenous Substances tested Concentrations

|  Active agent | Drug Conc [mg/L]  |
| --- | --- |
|  Acetylcystein | 150  |
|  Acetylsalicylic Acid | 30.0  |
|  Ampicillin-Na | 75.0  |
|  Ascorbic acid | 52.5  |
|  Cefoxitin | 750  |
|  Doxycyclin | 18.0  |
|  Heparin | 3300 IU/L  |
|  Levodopa | 7.50  |
|  Methyldopa +1.5 | 22.5  |
|  Metronidazole | 123  |
|  Rifampicin | 48.0  |
|  Acetaminophen | 156  |
|  Cyclosporine | 1.80  |
|  Ibuprofen | 219  |
|  Theophyllin | 60.0  |
|  Phenylbutazone | 321  |
|  Itraconazol | 50.0 and 10.0  |

### C. IgG Class Specificity

A VCA IgG class specificity study was performed by depleting EBV IgM with DTT treatment. The results showed no decreased signal with DTT treatment compared to untreated specimens demonstrating that the Elecsys EBV VCA IgG assay selectively detects EBV VCA antibodies.

### D. Hook Effect

High positive samples were serially diluted to determine if high-dose hook effect was found with the Elecsys EBV VCA IgG assay. No false negative result was observed.

### 4. Detection Limit (LoD):

K260232 - Page 9 of 15

{9}

Not applicable

5. Stability, Expected Values (Controls, Calibrators, or Methods):

### A. Reagent Stability Study

On-board reagent stability:

Five samples covering negative, around the cut-off, moderately positive and high positive were tested with the Elecsys EBV VCA IgG test on the cobas e 801 analyzer. The results showed that the reagent pack was stable on-board up to 16 weeks.

Calibration stability:

Five samples covering negative, around the cut-off, moderately positive and high positives were tested with the Elecsys EBV VCA IgG test on cobas e 801 analyzer. The following stability results were obtained:

- Lot calibration - up to 12 weeks
- Calibration when using the same cobas e pack on the analyzer – up to 28 days.

### B. Sample Stability Study

A study was conducted to evaluate the effect of sample handling and storage conditions at room temperature (20 °C), refrigerated (2 - 8°C), and frozen (≤ -20°C). In addition, sample stability was evaluated after multiple freeze/thaw cycles. Sample stability studies support the following stability claims:

|  Storage Condition | Claimed Stability  |
| --- | --- |
|  Room Temp Storage (20-25 °C) | Up to 5 Days  |
|  Refrigerated Storage (2-8 °C) | Up to 14 Days  |
|  Frozen Storage (≤-20 °C) | Up to 3 Months  |
|  Freeze/Thaw Cycles | No more than 3 cycles  |

### C. Fresh versus Frozen Sample Stability Study

A study was conducted to demonstrate that the Elecsys EBV VCA IgG detects VCA IgG in both fresh and frozen samples. Fifty samples covering low negative, high negative, around the cut-off, and moderate positives were tested as fresh and after freezing at -20 °C on the Elecsys EBV VCA IgG assay. The results demonstrated that EBV VCA IgG levels were comparable in fresh and frozen samples.

6. Assay Cut-Off:

A study to establish the Elecsys EBV IgG cut-off was performed using intended use samples. The results of this study established the cut-off value of the Elecsys EBV VCA IgG as 1.0 COI.

### B Comparison Studies:

K260232 - Page 10 of 15

{10}

### 1. Method Comparison:

Not applicable. See Section C - Clinical Studies

### 2. Matrix Comparison:

This study was conducted to determine whether the claimed matrices have equivalent performance on the Elecsys EBV VCA IgG test. The study evaluated 50 sets of matched samples collected in serum, lithium heparin, and EDTA plasma. The data supports comparable performance of samples collected in serum, lithium heparin, and EDTA plasma (Table 7).

Table 7. Summary of results for regression analysis for different plasma types

|  Matrix Tested | N Sample Tested | Slope | Intercept | Range tested (COI)  |
| --- | --- | --- | --- | --- |
|  Li Heparin vs Serum | 50 | 0.999 | 0.000 | 0.263-483  |
|  K2 EDTA vs Serum | 50 | 0.991 | 0.000 | 0.291-497  |
|  K3 EDTA vs Serum | 50 | 0.962 | 0.000 | 0.296-484  |

## C Clinical Studies:

### 1. Clinical Study:

A multisite study was conducted to evaluate clinical performance of the Elecsys EBV VCA IgG on the cobas e analyzer and to determine Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) with the predicate device. PPA and NPA to a composite consensus comparator were assessed using 974 specimens with routine laboratory requests for EBV testing. These specimens included 721 prospective samples not selected based on their routine EBV test results, and 253 retrospective samples selected based on their routine EBV test results to be presumed acute or presumed seronegative. Out of 721 prospective samples, 224 were from pediatric population. Out of 253 retrospective samples, 192 were from pediatric population.

All specimens were tested with the Elecsys EBV IgM assay, the Elecsys EBV VCA IgG assay, and the Elecsys EBV EBNA IgG assay (referred to as Elecsys EBV assay panel) and were also tested with 3 comparator FDA-cleared EBV assay panels.

Specimens were classified based on serological marker profile by the comparators. Table 8 presents the serological classification of the EBV disease status. Results for prospectively collected adults only and pediatrics are shown in Tables 9 and 10, respectively. Results for retrospectively collected adults only and pediatrics are shown in Tables 11 and 12, respectively. Prospective and retrospective percent agreement for Elecsys EBV VCA IgG in relation to the EBV serological status and the composite comparator consensus EBV VCA IgG results are shown in Table 13 and 14, respectively.

Table 8. Result interpretation Table used to determine the EBV serological status when using Elecsys EBV assays

|  EBV Status | Elecsys EBV Assay Results  |   |   |
| --- | --- | --- | --- |
|   |  EBV IgM | EBV VCA IgG | EBV EBNA IgG  |
|  EBV Seronegative | - | - | -  |

K260232 - Page 11 of 15

{11}

|  EBV Status | Elecsys EBV Assay Results  |   |   |
| --- | --- | --- | --- |
|   |  EBV IgM | EBV VCA IgG | EBV EBNA IgG  |
|  Acute Infection | + | - | -  |
|   |  + | + | -  |
|  Past Infection | - | + | +  |
|   |  - | + | -  |
|  Indeterminate EBV Status*  |   |   |   |
|  Isolated presence of EBNA IgG | - | - | +  |
|  Past or Recovery or Reactivation | + | + | +  |

* For indeterminate and/or any additional combination of results not shown in this table, further diagnostic testing is recommended

Table 9. PPA and NPA results of prospective population for adult cohort

|  Elecsys EBV VCA IgG | Reference Method |   | Total  |   |
| --- | --- | --- | --- | --- |
|   |   |  Reactive |   | Non-reactive  |
|   | Reactive | 476 | 8 | 484  |
|   |  Non-reactive | 0 | 13 | 13  |
|  Total |   | 476 | 21 | 497  |
|  |   |   |   |   |
|   |   | Estimate | 95% CI  |   |
|  PPA |   | 100% | 99.2% - 100%  |   |
|  NPA |   | 61.9% | 38.4% - 81.9%  |   |

Table 10. PPA and NPA results of prospective population for pediatric cohort

|  Elecsys EBV VCA IgG | Reference Method |   | Total  |   |
| --- | --- | --- | --- | --- |
|   |   |  Reactive |   | Non-reactive  |
|   | Reactive | 100 | 9 | 109  |
|   |  Non-reactive | 1 | 114 | 115  |
|  Total |   | 101 | 123 | 224  |
|  |   |   |   |   |
|   |   | Estimate | 95% CI  |   |
|  PPA |   | 99.0% | 94.6% - 100%  |   |
|  NPA |   | 92.7% | 86.6% - 96.6%  |   |

Table 11. PPA and NPA results of Retrospective adult cohort

|  Elecsys EBV VCA IgG | Reference Method |   | Total  |   |
| --- | --- | --- | --- | --- |
|   |   |  Reactive |   | Non-reactive  |
|   | Reactive | 21 | 7 | 28  |
|   |  Non-reactive | 2 | 31 | 33  |
|  Total |   | 23 | 38 | 61  |
|  |   |   |   |   |
|   |   | Estimate | 95% CI  |   |
|  PPA |   | 91.3% | 72.0% - 98.9%  |   |
|  NPA |   | 81.6% | 65.7% - 92.3%  |   |

K260232 - Page 12 of 15

{12}

Table 12. PPA and NPA results of Retrospective pediatric cohort

|  Elecsys EBV VCA IgG | Reference Method |   | Total  |   |
| --- | --- | --- | --- | --- |
|   |   |  Reactive |   | Non-reactive  |
|   | Reactive | 60 | 21 | 81  |
|   |  Non-reactive | 5 | 106 | 111  |
|  Total |   | 65 | 127 | 192  |
|  |   |   |   |   |
|   |   | Estimate |   |   |
|  PPA |   | 92.3% | 83.0% - 97.5%  |   |
|  NPA |   | 83.5% | 75.8% - 89.5%  |   |

Table 13. Percent Agreement between Elecsys EBV VCA IgG and Composite Comparator EBV VCA IgG by EBV serological status in Prospective Samples (adults and pediatric) (n=721)

|  EBV Serological Status (IgM / VCA IgG / EBNA IgG) | PPA Estimate and 95% CI |   |   | NPA Estimate and 95% CI  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  n/n | % | (95% CI) | n/n | % | (95% CI)  |
|  Seronegative (-/-/-) | n/a | n/a | n/a | 124/134 | 92.5% | (86.7-96.4%)  |
|  Acute (+/-/-) | n/a | n/a | n/a | 2/6 | 33.3% | (4.3-77.7%)  |
|  Acute (+/+/−) | 9/9 | 100.0% | (66.4-100.0%) | n/a | n/a | n/a  |
|  Total Acute | 9/9 | 100.0% | (66.4-100.0%) | 2/6 | 33.3% | (4.3-77.7%)  |
|  Past Infection (-/+/+) | 532/532 | 100.0% | (99.3-100.0%) | n/a | n/a | n/a  |
|  Past Infection (-/+/−) | 32/33 | 97.0% | (84.2-99.9%) | n/a | n/a | n/a  |
|  Total Past Infection | 564/565 | 99.8% | (99.0-100.0%) | n/a | n/a | n/a  |
|  Isolated Presence of EBNA IgG (-/-/+) | n/a | n/a | n/a | 1/4 | 25.0% | (0.6-80.6%)  |
|  Transient (+/+/+) | 3/3 | 100.0% | (29.2-100.0%) | n/a | n/a | n/a  |
|  Other Indeterminate (+/-/+) | n/a | n/a | n/a | n/a | n/a | n/a  |
|  Total Indeterminate | 3/3 | 100.0% | (29.2-100.0%) | 1/4 | 25.0% | (0.6-80.6%)  |
|  Overall | 576/577 | 99.8% | (99.0-100.0%) | 127/144 | 88.2% | (81.8-93.0%)  |

Table 14. Percent Agreement between Elecsys EBV VCA IgG and Composite Comparator EBV VCA IgG by EBV serological status in Retrospective Samples (adults and pediatric) (n=253)

|  EBV Serological Status (IgM / VCA IgG / EBNA IgG) | PPA Estimate and 95% CI |   |   | NPA Estimate and 95% CI  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  n/n | % | (95% CI) | n/n | % | (95% CI)  |
|  Seronegative (-/-/-) | n/a | n/a | n/a | 118/118 | 100.0% | (96.9-100.0%)  |
|  Acute (+/-/-) | n/a | n/a | n/a | 19/47 | 40.4% | (26.4-55.7%)  |
|  Acute (+/+/−) | 71/76 | 93.4% | (85.3-97.8%) | n/a | n/a | n/a  |

K260232 - Page 13 of 15

{13}

|  Total Acute | 71/76 | 93.4% | (85.3-97.8%) | 19/47 | 40.4% | (26.4-55.7%)  |
| --- | --- | --- | --- | --- | --- | --- |
|  Past Infection (-/+/+) | 2/2 | 100.0% | (15.8-100.0%) | n/a | n/a | n/a  |
|  Past Infection (-/+/−) | 7/9 | 77.8% | (40.0-97.2%) | n/a | n/a | n/a  |
|  Total Past Infection | 9/11 | 81.8% | (48.2-97.7%) | n/a | n/a | n/a  |
|  Isolated Presence of EBNA IgG (-/-/+) | n/a | n/a | n/a | n/a | n/a | n/a  |
|  Transient (+/+/+) | 1/1 | 100.0% | 2.5-100.0 | n/a | n/a | n/a  |
|  Other Indeterminate (+/-/+) | n/a | n/a | n/a | n/a | n/a | n/a  |
|  Total Indeterminate | 1/1 | 100.0% | 2.5-100.0 | n/a | n/a | n/a  |
|  Overall | 81/88 | 92.0% | (84.3-96.7%) | 137/165 | 83.0% | (76.4-88.4%)  |

### 2. Clinical Cut-Off

See Section A8

### D Expected Values/Reference Range:

Table 15. Summary for Elecsys EBV VCA IgG Results by Age Group and Sex

|  Age Group (Years) | Sex | Reactive |   | Non-reactive |   | Total  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  n | % | n | % | n  |
|  2 to <6 years | Female | 3 | 15.8 | 16 | 84.2 | 19  |
|   |  Male | 18 | 30.0 | 42 | 70.0 | 60  |
|   |  Not Reported | 0 | 0.0 | 1 | 100.0 | 1  |
|   |  Total | 21 | 26.3 | 59 | 73.8 | 80  |
|  6 to <11 years | Female | 13 | 37.1 | 22 | 62.9 | 35  |
|   |  Male | 18 | 35.3 | 33 | 64.7 | 51  |
|   |  Total | 31 | 36.0 | 55 | 64.0 | 86  |
|  11 to <16 years | Female | 28 | 52.8 | 25 | 47.2 | 53  |
|   |  Male | 19 | 38.8 | 30 | 61.2 | 49  |
|   |  Total | 47 | 46.1 | 55 | 53.9 | 102  |
|  16 to <22 years | Female | 57 | 62.6 | 34 | 37.4 | 91  |
|   |  Male | 34 | 59.6 | 23 | 40.4 | 57  |
|   |  Total | 91 | 61.5 | 57 | 38.5 | 148  |
|  22-39 years | Female | 82 | 82.8 | 17 | 17.2 | 99  |
|   |  Male | 67 | 85.9 | 11 | 14.1 | 78  |
|   |  Total | 149 | 84.2 | 28 | 15.8 | 177  |
|  40-59 years | Female | 86 | 91.5 | 8 | 8.5 | 94  |
|   |  Male | 103 | 95.4 | 5 | 4.6 | 108  |
|   |  Total | 189 | 93.6 | 13 | 6.4 | 202  |
|  ≥60 years | Female | 70 | 95.9 | 3 | 4.1 | 73  |

K260232 - Page 14 of 15

{14}

|   | Male | 104 | 98.1 | 2 | 1.9 | 106  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Total | 174 | 97.2 | 5 | 2.8 | 179  |
|  Total | Female | 339 | 73.1 | 125 | 26.9 | 464  |
|   |  Male | 363 | 71.3 | 146 | 28.7 | 509  |
|   |  Not Reported | 0 | 0.0 | 1 | 100.0 | 1  |
|   |  Overall | 702 | 72.1 | 272 | 27.9 | 974  |

### E Other Supportive Instrument Performance Characteristics Data:

Carryover Study: A carryover study was conducted. No carryover was detected.

Electrical safety and electromagnetic compatibility (EMC) testing were performed, and the system was found to be acceptable.

Software and cybersecurity documentation were acceptable.

### VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

### IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K260232 - Page 15 of 15

---

**Source:** [https://fda-staging.innolitics.com/device/K260232](https://fda-staging.innolitics.com/device/K260232)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
