MiPACS V5 is a software-only medical and dental DICOM image viewing system with integrated image acquisition functionality. MiPACS V5 is intended for use by trained healthcare professionals for the acquisition, viewing, manipulation, and management of medical and dental images in DICOM format. MiPACS V5 is not intended to replace primary diagnostic workstations and does not provide automated diagnosis, image interpretation, or clinical decision support.
Device Story
MiPACS V5 is a web-based, software-only medical/dental DICOM image management system. It interfaces with external dental imaging hardware to acquire, view, manipulate, and manage digital images. Operated by trained healthcare professionals in clinical settings, the system functions via a client-server architecture using standard web browsers. It provides tools for image navigation and management but performs no automated diagnosis, image interpretation, or clinical decision support. The device facilitates clinical workflows by enabling secure, authorized access to patient imaging data within on-premises infrastructure. It benefits patients by streamlining the availability and management of diagnostic images for clinicians.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included software verification and validation (unit, integration, and system-level), performance testing for image rendering consistency, DICOM conformance, connectivity, database integrity, and end-to-end clinical workflow validation.
Technological Characteristics
Software-only, web-based DICOM image management system. Operates in a client-server model via standard web browsers. Supports DICOM/DICOM web standards (QIDO-RS, WADO-RS, STOW-RS). Implements role-based access control, session management, and encrypted communication. Deployed on customer-managed, on-premises infrastructure.
Indications for Use
Indicated for trained healthcare professionals for the acquisition, viewing, manipulation, and management of medical and dental DICOM images in clinical environments.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 25, 2026
Apryse Software Corp
Hilal Chami
Manager, Support and Implementation
2399 Blake St.
Denver, CO 80205-2186
Re: K260231
Trade/Device Name: MiPACS V5 (1.0.0)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: May 29, 2026
Received: May 29, 2026
Dear Hilal Chami:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260231 - Hilal Chami
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260231 - Hilal Chami
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260231
?
Please provide the device trade name(s).
?
MiPACS V5 (1.0.0.5)
Please provide your Indications for Use below.
?
MiPACS V5 is a software-only medical and dental DICOM image viewing system with integrated image acquisition functionality. MiPACS V5 is intended for use by trained healthcare professionals for the acquisition, viewing, manipulation, and management of medical and dental images in DICOM format. MiPACS V5 is not intended to replace primary diagnostic workstations and does not provide automated diagnosis, image interpretation, or clinical decision support.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
# 510(k) Summary
K260231
# General Summary of Submission / Executive Summary
Document Date: 2026-06-24
Device/Software Name: MiPACS V5
Version/Build: 1.0.0
Sponsor/Manufacturer: Apryse Software Corp
Submission Type: Traditional 510(k)
Product Code / Regulation: LLZ / 21 CFR 892.2050
Device Class: Class II
Predicate Device: MiPACS HTML5 Web Viewer (K212984)
Administrative Contact: Hilal Chami, Support, Integrations, & QA Manager,
hilal.chami@apryse.com, 919-208-5312
# 1. Administrative Information
- Common Name: Dental Picture Archiving and Communication System (PACS)
- Trade/Proprietary Name: MiPACS V5
- Classification Regulation: 21 CFR 892.2050
- Product Code: LLZ (System, Image Processing, Radiological)
- Classification Panel: Radiology
MiPACS V5 is regulated as a Class II medical device under the Radiology panel and is subject to premarket notification requirements.
# 2. Device Description
MiPACS V5 is a software-only, web-based medical device designed for the viewing, acquisition, manipulation, and management of digital dental images in DICOM format. The device is intended for use by trained healthcare professionals in clinical environments and operates within customer-managed, on-premises infrastructure.
The system consists of the following primary functional components:
- Web-Based User Interface:
- A browser-based interface providing image viewing, navigation, manipulation, and acquisition controls.
- Application Logic Layer:
- Manages clinical workflows, user sessions, image handling, validation, and coordination of acquisition and storage functions.
- Image Acquisition Interface:
{5}
○ Interfaces with compatible, vendor-supported dental imaging devices to initiate image acquisition and receive image data. Image generation and hardware performance remain the responsibility of the acquisition device manufacturer.
- DICOM Communication Layer:
○ Supports DICOM and DICOM web standards (including QIDO-RS, WADO-RS, and STOW-RS) for image storage, retrieval, and transfer.
- Security and Access Control Layer:
○ Implements authentication, role-based authorization, session management, audit logging, and encrypted communication.
MiPACS V5 is deployed in a client–server model using standard web browsers on supported operating systems. No automated diagnosis, image analysis, or clinical decision-support functionality is provided.
### 3. Indications for Use
MiPACS V5 is a software-only medical and dental DICOM image viewing system with integrated image acquisition functionality. MiPACS V5 is intended for use by trained healthcare professionals for the acquisition, viewing, manipulation, and management of medical and dental images in DICOM format. MiPACS V5 is not intended to replace primary diagnostic workstations and does not provide automated diagnosis, image interpretation, or clinical decision support.
### 4. Predicate Device Comparison (Substantial Equivalence)
- Primary Predicate: MiPACS HTML5 Web Viewer (K212984)
- Manufacturer: MiPACS by Apryse
- Intended Use:
○ MiPACS V5 has the same intended use as the predicate device:
▪ to support the acquisition, viewing, manipulation and management of digital dental images to aid clinical workflows.
- Technological Characteristics:
○ Both MiPACS V5 and the predicate device are web-based software systems that operate within clinical environments and utilize DICOM standards for image handling.
○ MiPACS V5 builds upon the predicate architecture by adding software-mediated image acquisition functionality via compatible acquisition devices. The fundamental viewing, navigation, storage, and security mechanisms remain consistent with the predicate.
- Addressing Differences:
{6}
○ The added acquisition functionality was evaluated through software verification and validation testing, risk management activities, and cybersecurity assessment. These differences do not raise new questions of safety or effectiveness.
## 5. Summary of Non-Clinical Testing
Non-clinical testing was conducted to demonstrate that MiPACS V5 meets its specified requirements and performs safely and effectively for its intended use.
- **Software Verification & Validation:**
- ○ Software V&V activities were conducted in accordance with FDA guidance on software contained in medical devices (June 2023) at the Basic Documentation Level. Testing included unit, integration, and system-level verification of functional, performance, and security requirements.
- **Performance/Interoperability:**
- ○ Bench testing verified image rendering consistency, DICOM conformance, connectivity with external systems, database integrity, and end-to-end clinical workflows including image acquisition, storage, retrieval, and cache management.
## 6. Cybersecurity
Cybersecurity controls were implemented under a Secure Product Development Framework and documented within a Cybersecurity Management Plan consistent with FDA cybersecurity guidance. Activities included threat modeling, SBOM-based dependency analysis, security requirement verification, and risk-based vulnerability assessment.
## 7. Conclusion
All testing demonstrated that MiPACS V5 meets defined acceptance criteria and supports safe and effective clinical use. Based on the intended use, technological characteristics, and results of testing, **MiPACS V5 is substantially equivalent to the predicate device**, MiPACS HTML5 Web Viewer (K212984) and does not raise new questions of safety or effectiveness.
Jun 24 2026
Purpose: Document Approval
Hilal Chami
Authenticated via SMS 2FA
hilal.chami@apryse.com
Support, Integrations, & Quality Assurance Manager
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.