Aristotle Mini Guidewire

K260223 · Scientia Vascular, Inc. · MOF · Aug 6, 2026 · Cardiovascular

Device Facts

Record IDK260223
Device NameAristotle Mini Guidewire
ApplicantScientia Vascular, Inc.
Product CodeMOF · Cardiovascular
Decision DateAug 6, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The Aristotle Mini Guidewire is intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature.

Device Story

Aristotle Mini Guidewire is a steerable, radiopaque guidewire used by physicians to navigate neuro and peripheral vasculature. It facilitates the introduction and positioning of catheters and interventional devices. The device features a 0.011" distal/0.012" proximal diameter, a 304v stainless steel core, and a nitinol distal tip. It includes a hydrophilic coating on the distal portion and PTFE on the proximal portion to reduce friction. The device is supplied with a shaping mandrel, introducer, and torque device. During procedures, the physician manipulates the guidewire under fluoroscopic guidance; the radiopaque platinum coil allows for visualization. The device is intended for use in clinical settings by trained physicians. It benefits patients by enabling access to tortuous distal vasculature for minimally invasive interventions.

Clinical Evidence

No clinical testing was performed. Substantial equivalence is supported by bench performance testing, including visual inspection, dimensional verification, particulate testing, tensile testing, kink resistance, torqueability/strength, tip flexibility, flex/fracture, coating integrity, corrosion resistance, and simulated use in anatomical models. Usability and radiopacity were validated by physicians in a human cadaver model.

Technological Characteristics

Core: 304v stainless steel; Tip: Nitinol; Coatings: Hydrophilic (distal), PTFE (proximal); Radiopaque marker: Platinum coil; Dimensions: 0.011" distal/0.012" proximal OD, 200-300 cm lengths; Sterilization: 100% Ethylene Oxide (ISO 11135); Standards: ISO 11070, ISO 10993-1, ISO 10993-7.

Indications for Use

Indicated for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. Not intended for use in the coronary vasculature.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 6, 2026 Scientia Vascular, Inc. Jamila Abou-Bakr Regulatory Affairs Specialist 2460 South 3270 West West Valley City, Utah 84119 Re: K260223 Trade/Device Name: Aristotle Mini Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF, DQX Dated: July 3, 2026 Received: July 6, 2026 Dear Jamila Abou-Bakr: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260223 - Jamila Abou-Bakr Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K260223 - Jamila Abou-Bakr Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, NAIRA MURADYAN -S Naira Muradyan, PhD Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260223 | ? | | Please provide the device trade name(s). | | ? | | Aristotle Mini Guidewire | | | | Please provide your Indications for Use below. | | ? | | The Aristotle Mini Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} Scientia Vascular, Inc. Traditional 510(k) K260223 ![img-0.jpeg](img-0.jpeg) ## 510(k) SUMMARY (K260223, Per 21 CFR 807.92) 510(k) Sponsor: Scientia Vascular, Inc. 2460 South 3270 West West Valley City, UT 84119 Tel (888) 385-9016 Contact Person(s): Jamila Abou-Bakr Regulatory Affairs Specialist Tel: (888) 385-9016 Email: regulatory@scientiavascular.com Date Prepared: July 14, 2026 Page 1 of 7 {5} Scientia Vascular, Inc. Traditional 510(k) K260223 **Table 1. Subject Device Information** | Trade Name: | Aristotle Mini Guidewire | | --- | --- | | Common Name: | Guidewire | | Classification Name | Catheter Guide Wire per 21 CFR 870.1330 | | Primary Product Code: | MOF | | Subsequent Product Code | DQX | | Predicate Device: | Aristotle 14 Guidewire (K243938) | | Reference Device: | PVS 1600 Synchro 0.010” Guidewire (K032146) | ## DEVICE DESCRIPTION Scientia Vascular’s Aristotle Mini Guidewire has a 0.011” distal/0.012” proximal outer diameter (OD) and is compatible with catheters with a minimum inner diameter of 0.013”. The guidewire has a straight, shapeable tip and is supplied in lengths of 200 cm, 225 cm, and 300 cm. It is provided in one stiffness profile – standard. The Aristotle Mini Guidewire is intended to facilitate access to distal neuro and peripheral vasculature, and it can be used with guide catheters or intermediate catheters. The guidewire incorporates a radiopaque platinum coil to facilitate fluoroscopic visualization, and is designed with variable shaft stiffness from the flexible distal tip to the more rigid proximal end. The distal outer surface of the guidewire is coated with a hydrophilic coating, and the proximal outer surface is coated in polytetrafluoroethylene (PTFE). The coated surfaces are intended to reduce friction during manipulation of the guidewire within the vessel and/or catheter. A shaping mandrel, guidewire introducer, and torque device accessories are included with the Aristotle Mini Guidewire. ## INDICATIONS FOR USE The Aristotle Mini Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. ## INTENDED USE The Aristotle Mini Guidewire is intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS The subject device is substantially equivalent to the predicate device and has the same or similar: - Indications for use Page 2 of 7 {6} Scientia Vascular, Inc. Traditional 510(k) K260223 • Technological characteristics • Packaging and sterilization of devices Table 2 provides details on the technological characteristics of the subject device compared to the predicate device and the reference device. Table 2. Technological Characteristics Comparison | Characteristic | Subject Device Aristotle Mini Guidewire (K260223) | Predicate Device Aristotle 14 Guidewire (K243938) | Reference Device PVS 1600 Synchro 0.010" Guidewire (K032146)* | | --- | --- | --- | --- | | Indications for Use | The Aristotle Mini Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | The PVS 1600 Synchro 0.010" Guidewire series of products are intended for general intravascular use, including the neuro and peripheral vasculature. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. The device should be used only by physicians trained in percutaneous, intravascular techniques and procedures. | | Anatomical Location | Neuro and peripheral vasculature | Neuro and peripheral vasculature | Neuro and peripheral vasculature | | Dimensions | OD: 0.011" distal / 0.012" proximal | OD: 0.014" | OD: 0.010" distal / 0.012" proximal | | | Length: 200 cm, 225 cm, 300 cm | Length: 200 cm, 300 cm | Length: 180 – 300 cm | | Core Wire | 304v Stainless Steel | 304v Stainless Steel | 304 Stainless Steel | | Distal Tip | Length: 35 cm Material: Nitinol | Length: 35 cm Material: Nitinol | Length: 25-65 cm Material: Nitinol | | Tip Shape | Straight, Shapeable | Straight, Shapeable | Straight, Shapeable | Page 3 of 7 {7} Scientia Vascular, Inc. Traditional 510(k) K260223 | Characteristic | Subject Device Aristotle Mini Guidewire (K260223) | Predicate Device Aristotle 14 Guidewire (K243938) | Reference Device PVS 1600 Synchro 0.010" Guidewire (K032146)* | | --- | --- | --- | --- | | Stiffness Profiles | Standard | Soft, Standard | Support (Stiff) to Flex (Soft) | | Distal Coating | Hydrophilic | Hydrophilic | Hydrophilic | | Proximal Coating | PTFE | PTFE | PTFE | | Radiopaque Marker | 1 radiopaque marker at distal tip *Length:* 5 cm | 1 radiopaque marker at distal tip *Length:* 10 cm | 1 radiopaque marker at distal tip *Length:* 10 cm | | Centering Coil(s) | None | Stainless-steel | NA | | Introducer (Accessory) | Provided with each guidewire | Provided with each guidewire | Provided with each guidewire (combined with shaping mandrel) | | Torque Device (Accessory) | Provided with each guidewire | Provided with each guidewire | Provided with each guidewire | | Shaping Mandrel (Accessory) | Provided with each guidewire | Provided with each guidewire | Provided with each guidewire (combined with introducer) | | Packaging Configuration | Single HDPE tubing hoop sealed in Tyvek Pouch | Single HDPE tubing hoop sealed in Tyvek Pouch | Single tubing hoop sealed in pouch | | Sterilization Method | 100% Ethylene Oxide (EO) | 100% Ethylene Oxide (EO) | Irradiation | | Shelf Life | 1 year | 1 year | NA | *Completed based on 510(k) Summary, product labeling, or public information available on the device. Characteristics for which information could not be found have been labeled as not available (NA). When compared with the predicate device, the subject device has the technological characteristic differences shown in the table above. These differences do not raise new questions of safety and effectiveness for the subject device. To evaluate the effects of these differences, performance testing was conducted to demonstrate the subject device performs as intended. Page 4 of 7 {8} Scientia Vascular, Inc. Traditional 510(k) K260223 ## NON-CLINICAL PERFORMANCE TESTS ### Biocompatibility Biocompatibility testing was performed according to requirements set by FDA guidance document, “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process”, and ISO 10993-1, “Biological evaluation of medical devices — Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process.” The subject device is considered an externally communicating device with circulating blood contact for a limited duration (≤ 24 hours). All manufacturing processes used for the subject device are the same as for the predicate, as well as all materials with the exception of the hydrophilic coating (same coating components, new formulation). Therefore, the following tests were performed with representative devices focusing on biocompatibility of the hydrophilic coating: - Cytotoxicity - Sensitization - Irritation/Intracutaneous Reactivity - Acute Systemic Toxicity and Pyrogenicity - Hemocompatibility ### Sterilization The subject device was adopted into an existing validated sterilization cycle that uses 100% EO to achieve a sterilization assurance level (SAL) of 10⁻⁶, and testing for EO and ethylene chlorohydrin (ECH) residuals and bacterial endotoxin levels was performed. The subject device followed the requirements of AAMI TIR 28:2016, “Product adoption and process equivalence for ethylene oxide sterilization,” to ensure it continues to meet the requirements of the following standards: - ISO 11135:2014/A1:2018, “Sterilization of Health Care Products - Ethylene Oxide – Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices” - ANSI/AAMI/ISO 10993-7:2008, “Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals” ### Bench Performance Testing Following review of the risk assessments conducted in accordance with ISO 14971:2019 performance testing was performed on the subject device. Performance testing was conducted to ensure compliance with ISO 11070:2014, “Sterile single-use intravascular introducers, dilators and guidewires,” and per recommendations of the FDA guidance document, “Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling” (2019). Table 3 below summarizes these tests and their results: Page 5 of 7 {9} Scientia Vascular, Inc. Traditional 510(k) K260223 **Table 3. Summary of Bench Performance Tests** | Test | Test Method Summary | Results | | --- | --- | --- | | Visual Inspection and Dimensional Verification | Test per ISO 11070 | Acceptance criteria met | | Particulate | Particulate measured and counted after use in a simulated pathway model. | Acceptance criteria met | | Tensile Testing | Test per ISO 11070 | Acceptance criteria met | | Kink Resistance | Test per FDA guidance document, “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019) | Acceptance criteria met | | Torqueability | Test per FDA guidance document, “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019) | Acceptance criteria met | | Torque Strength | Test per FDA guidance document, “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019) | Acceptance criteria met | | Tip Flexibility – Column Buckling | Test per FDA guidance document, “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019) | Acceptance criteria met | | Flex and Fracture | Test per ISO 11070 | Acceptance criteria met | | Shapeability and Tip Shape Retention | Evaluated the distal tip shapeability and shape retention after simulated use. | Acceptance criteria met | | Coating Lubricity, Durability, and Integrity | Test per ISO 11070 and FDA guidance document, “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019) | Acceptance criteria met | | Corrosion Resistance | Test per ISO 11070 and FDA guidance document, “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019) | Acceptance criteria met | | Compatibility with Agents | Guidewires were tested for compatibility with IsoVue 300, IsoVue 370, DMSO, and heparinized saline. | Acceptance criteria met | | Simulated Use and Compatibility | Conducted simulated use in anatomical model designed to simulate the tortuous anatomy of the neurovasculature. | Acceptance criteria met | | Usability and Radiopacity Validation | Physicians evaluated subject and comparator guidewires for various performance characteristics in a human cadaver. | Acceptance criteria met | ## ANIMAL TESTING No animal testing was deemed necessary to support the substantial equivalence of the subject device. Page 6 of 7 {10} Scientia Vascular, Inc. Traditional 510(k) K260223 ## CLINICAL TESTING No clinical testing was deemed necessary to support the substantial equivalence of the subject device. ## CONCLUSION The subject device has the same intended use and indications for use statement as the predicate device. The identified technological differences do not raise new questions of safety or effectiveness regarding the use of the subject device. The testing demonstrates that the subject device is substantially equivalent to the predicate device. Page 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...