K260209 · Karl Storz SE & CO. KG · HIF · Aug 19, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K260209
Device Name
ENDOFLATOR+; Smart Smoke Evacuation
Applicant
Karl Storz SE & CO. KG
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Aug 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ENDOFLATOR + is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with CO2 gas with the following operating modes:High Flow mode is indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure.Pediatric mode is indicated to fill and distend a peritoneal cavity with gas during a pediatric laparoscopic procedure.EVH mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.taTME mode is indicated to fill and distend the rectum and colon using CO2 gas during transanal minimally invasive surgery.The smart smoke evacuation software is an optional feature that triggers smoke evacuation of the compatible insufflator through analysis of the smoke volume in the endoscopic image.
Device Story
ENDOFLATOR+ is a CO2 insufflator for laparoscopic, EVH, and taTME procedures; creates/maintains body cavities via CO2 gas. Operates via touchscreen; displays surgical data; network-enabled for monitor routing. Features four modes: High Flow, Pediatric, EVH, taTME. Optional 'Smart Smoke Evacuation' feature uses IMAGE1 CCU to analyze endoscopic images for smoke volume; automatically triggers evacuation. Uses RFID-enabled tube sets (standard, heating, smoke evacuation, humidification) for automatic function recognition. Benefits include automated smoke management, precise pressure/flow control, and expanded pediatric utility. Used in OR settings by surgical staff.
Clinical Evidence
Bench testing only. Pediatric mode performance verified against K201361 for patients <2 years. Smart smoke evacuation verified for smoke identification accuracy, trigger reliability, and pressure stability (avoiding over/under-pressure). Software and cybersecurity validated per FDA guidance.
Technological Characteristics
Insufflator with touchscreen interface; 370x150x305mm; 10.7kg. Max pressure 30mmHg; max flow 50 L/min. RFID-based tube set recognition. Connectivity: Network-enabled for monitor display. Software-based smoke detection/evacuation. Sterilization: Not specified.
Indications for Use
Indicated for adults and pediatric patients (including those under 2 years) for cavity distension during laparoscopic, endoscopic vessel harvesting (EVH), and transanal minimally invasive surgery (taTME) procedures using CO2 gas.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
W.O.M. World of Medicine GmbH PNEUMOCLEAR (K201361)
Stryker Connected OR Hub with Voice and Device Control (K201434)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 19, 2026
Karl Storz SE & CO. KG
Emily Rhiel
Regulatory Affairs Senior Specialist
Dr. Karl-Storz-Straße 34
Tuttlingen, BW 78532
Germany
Re: K260209
Trade/Device Name: ENDOFLATOR+; Smart Smoke Evacuation
Regulation Number: 21 CFR§ 884.1730
Regulation Name: Laparoscopic Insufflator
Regulatory Class: II
Product Code: HIF, OSV
Dated: July 30, 2026
Received: July 30, 2026
Dear Emily Rhiel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260209 - Emily Rhiel
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260209 - Emily Rhiel
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JASON ROBERTS -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260209 | ? |
| Please provide the device trade name(s). | | ? |
| ENDOFLATOR+;Smart Smoke Evacuation | | |
| Please provide your Indications for Use below. | | ? |
| The ENDOFLATOR + is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with CO2 gas with the following operating modes:High Flow mode is indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure.Pediatric mode is indicated to fill and distend a peritoneal cavity with gas during a pediatric laparoscopic procedure.EVH mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery.taTME mode is indicated to fill and distend the rectum and colon using CO2 gas during transanal minimally invasive surgery.The smart smoke evacuation software is an optional feature that triggers smoke evacuation of the compatible insufflator through analysis of the smoke volume in the endoscopic image. | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
**STORZ**
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
ENDOFLATOR + with Smart Smoke Evacuation
K260209
Page 1 of 5
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92.
| **Submitter Information:** | KARL STORZ SE & Co. KG Dr.-Karl-Storz-Straße 34 78532 Tuttlingen, Germany |
| --- | --- |
| | **Contact:** Emily Rhiel Regulatory Affairs Senior Specialist Phone: (774) 318-2820 Email: emily.rhiel@karlstorz.com |
| | **Date of Preparation:** August 17, 2026 |
| **Device Information:** | Trade/Proprietary Name: ENDOFLATOR +; Smart Smoke Evacuation Common Name: Laparoscopic Insufflator Classification Name: Laparoscopic Insufflator Classification Number: 21 CFR 884.1730 Product Code: HIF (primary), OSV (secondary) Regulatory Class: Class II |
| **Predicate Device(s):** | **Insufflator predicates:** K250388 KARL STORZ ENDOFLATOR + K201361 W.O.M. World of Medicine GmbH PNEUMOCLEAR **Smart Smoke predicate:** K201434 Stryker Connected OR Hub with Voice and Device Control The predicate devices have not been subject to a design related recall. |
| **Device Description:** | The ENDOFLATOR + is an insufflation device with integrated smoke evacuation for laparoscopic examinations and surgery as well as transanal endoscopy and endoscopic vessel harvesting. Insufflation creates and maintains a cavity in the patient’s body. The subject device offers four operating modes: High Flow, Pediatric, EVH, and taTME. Smoke evacuation can be used with different flow rates for removal and filtration of CO_{2} and surgical smoke during operation. When used as a standalone insufflator, smoke evacuation can be used in HF, Continuous, or Manual Activation mode. When the insufflator is combined with the IMAGE1 CCU, Smart mode becomes available which enables the insufflator to conduct automatic smoke evacuation through analysis of the endoscopic image. The ENDOFLATOR + is operated directly on the touchscreen. All data required during the surgery is displayed simultaneously on the |
{5}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
ENDOFLATOR + with Smart Smoke Evacuation
K260209
Page 2 of 5
| | touchscreen. The data can also be routed to display the primary surgical monitor when the insufflator is network enabled. The subject device is to be used with one of the following tube sets depending on the selected mode and desired features (smoke evacuation, heating, and humidification) needed: UI610 (standard), UI611 (heating), UI612 (smoke evacuation), UI613 (heating + smoke evacuation), and UI614 (heating + smoke evacuation + humidification). The tube sets are designed with a radio frequency identification (RFID) transponder technology tube set recognition function, that recognizes the type of tube set that has been connected and which functions are available based on the selection. |
| --- | --- |
| Indications for Use: | ENDOFLATOR + The ENDOFLATOR + is a CO₂ insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with CO₂ gas with the following operating modes: - High Flow mode is indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. - Pediatric mode is indicated to fill and distend a peritoneal cavity with gas during a pediatric laparoscopic procedure. - EVH mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery. - taTME mode is indicated to fill and distend the rectum and colon using CO₂ gas during transanal minimally invasive surgery. Smart Smoke Evacuation The smart smoke evacuation software is an optional feature that triggers smoke evacuation of the compatible insufflator through analysis of the smoke volume in the endoscopic image. Comparison to Predicates The subject device's pediatric mode is expanded to include patients under 2 years old, which was excluded from the initial clearance (K250388). However, this patient population is included in the indications of the K201361 predicate. The smart smoke evacuation software indications differ from the K201434 predicate, as the subject device only accomplishes automatic smoke detection and evacuation, while the Stryker Connected OR |
{6}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
ENDOFLATOR + with Smart Smoke Evacuation
K260209
Page 3 of 5
| | system also includes voice and remote control of medical devices in addition to automatic smoke evacuation. Considering the indications of all three predicates together, the indications of the subject device, Endoflator + with Smart Smoke Evacuation, do not introduce any new indications. | | | |
| --- | --- | --- | --- | --- |
| Comparison of the Technological Characteristics: | Insufflator Comparison | | | |
| | | Subject Device: KARL STORZ ENDOFLATOR + with Smart Smoke Evacuation | Predicate Device: K250388 KARL STORZ ENDOFLATOR + | Predicate Device: K201361 W.O.M PNEUMOCLEAR |
| | Target Population | Adults and Pediatrics (all sub-populations) | Adults, Children, Adolescents | Same as subject |
| | User Interface | Touchscreen | Same as subject | Same as subject |
| | Dimensions and Weight | 370 x 150 x 305mm 10.7kg | Same as subject | Same as subject |
| | Design | High/low pressure units Safety valve assembly Gas heater Control hardware Smoke evacuation unit | Same as subject | Same as subject |
| | User Modes | • Pediatric: ○ <2 years ○ ≥2 years • High Flow • Endoscopic Vessel Harvesting (EVH) • Transanal Total Mesorectal Excision (taTME) | • Pediatric ≥2 years • High Flow • Endoscopic Vessel Harvesting (EVH) • Transanal Total Mesorectal Excision (taTME) | • Pediatric • Standard High Flow/Bariatric • Advanced Flow • Endoscopic Vessel Harvesting (EVH) • Transanal Minimally Invasive Surgery (TAMIS) |
| | Maximum specifications | Pressure: 30mmHg Insufflation Gas Flow: 50 L/min | Same as subject | Same as subject |
{7}
STORZ
KARL STORZ — ENDOSKOPE
KARL STORZ Premarket Notification
ENDOFLATOR + with Smart Smoke Evacuation
K260209
Page 4 of 5
| | Smoke Evacuation Flow: 12 L/min | | |
| --- | --- | --- | --- |
| **Tubing Set Features** | Heating Humidification Smoke Evacuation | Same as subject | Same as subject |
| **Smoke Evacuation Modes** | HF Continuous Smart Manual Activation | HF Continuous Manual Activation | Information not available |
| **Insufflation Tubing Sets** | Standard (UI610) Heating (UI611) Smoke Evacuation (UI612) Heating + Smoke Evacuation (UI613) Heating + Smoke Evaluation + Humidification (UI614) | Same as subject | Insufflation tube set with integrated filter (ST295) Insufflation tube set with integrated filter and heating wire (ST296) Insufflation tube set with integrated filter, heating wire and humidification (ST297) Insufflation and smoke evacuation tube set with integrated filter, heating wire and humidification (ST298) Insufflation and smoke evacuation tube set with integrated filter (ST299) |
| **Smart Smoke Evacuation Comparison** | | | |
| | **Subject Device:** KARL STORZ ENDOFLATOR + with Smart Smoke Evacuation | **Predicate Device: K201434** Stryker Connected OR Hub with Device and Voice Control | |
| **System Components** | ENDOFLATOR + and insufflation tubing IMAGE1 CCU with Smart Smoke Evacuation Software | PNEUMOCLEAR and insufflation tubing Connected OR Hub console Device control and voice control software | |
| **Smoke Detection** | Automated | Same as subject | |
| **Smoke Evacuation** | Automated | Same as subject | |
{8}
STORZ
KARL STORZ—ENDOSKOPE
KARL STORZ Premarket Notification
ENDOFLATOR + with Smart Smoke Evacuation
K260209
Page 5 of 5
| | Digital Smoke Removal | No | Yes |
| --- | --- | --- | --- |
| | The subject device has the same intended use as the predicates. The differences between the subject device and predicates do not raise different questions of safety and effectiveness and have been confirmed through testing. | | |
| Non-Clinical Performance Data: | The following non-clinical performance data were provided in support of the substantial equivalence determination. Software and Cybersecurity Validation • Software verification and validation testing was conducted and documentation was provided as recommended by the FDA's guidance, "Content of Premarket Submissions for Device Software Functions." • The cybersecurity was evaluated according to the FDA guidance "Cybersecurity in Medical Device: Quality System Considerations and Content of Premarket Submissions." Bench Testing • Pediatric Mode: Testing of the subject device's pediatric mode compared to the K201361 PNEUOMOCLEAR device was conducted to verify that the subject device's mode was as safe & effective as the predicate for pediatrics under 2 years old. • Smart Smoke: Verification testing to demonstrate correct identification of smoke level and smoke evacuation trigger was conducted. Additionally, rapid activation/deactivation of smoke evacuation was simulated to demonstrate that smart smoke does not increase the risk of over or under pressure during use. All bench testing passed, thus demonstrating that the expansion of the pediatric patient population and smart smoke evacuation are safe & effective to use as indicated. | | |
| Conclusion: | The conclusions drawn from the non-clinical performance data demonstrate that the subject device is as safe and effective as the predicate devices. As such, the subject device is substantially equivalent to the predicate devices. | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.