K260208 · embecta Medical II, LLC · FMF · Jul 13, 2026 · General Hospital
Device Facts
Record ID
K260208
Device Name
Ultra-Fine™ SAFETY Insulin Syringe
Applicant
embecta Medical II, LLC
Product Code
FMF · General Hospital
Decision Date
Jul 13, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ultra-Fine SAFETY Insulin Syringes are intended to be used for subcutaneous administration of insulin for the treatment of diabetes mellitus at a concentration of 100U/mL (U-100), while reducing the occurrence of accidental needle sticks during normal handling and disposal of the used needle/syringe combination.
Device Story
Ultra-Fine SAFETY Insulin Syringe is a single-use, sterile, piston-style syringe for subcutaneous U-100 insulin delivery. Device assembly includes a barrel with graduated scale, plunger rod with stopper, needle, and integrated safety mechanism. Operated by patients or clinicians; safety mechanism activated after injection to cover the needle, preventing accidental sticks during disposal. Device provides accurate insulin dosing via graduated markings; safety feature reduces sharps injury risk. Performance verified against ISO 8537 and ISO 23908 standards.
Clinical Evidence
No clinical data. Evidence consists of bench testing, biocompatibility evaluation, and a sharps injury prevention simulated use study. Performance testing met all acceptance criteria per ISO 8537, ISO 9626, ISO 7864, and ISO 23908. Biocompatibility endpoints (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity) were evaluated per ISO 10993-1.
Technological Characteristics
Piston syringe; materials include polypropylene, polystyrene, polycarbonate, 304 stainless steel, and silicone lubricants. Features integrated safety mechanism. Sterilized via gamma irradiation (SAL 10^-6). Complies with ISO 8537 (syringe performance), ISO 9626/7864 (needle performance), and ISO 23908 (sharps protection).
Indications for Use
Indicated for subcutaneous administration of U-100 insulin for the treatment of diabetes mellitus in patients requiring insulin therapy.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 13, 2026
embecta Medical II, LLC
Ewelina Rog
Senior Manager, Regulatory Affairs
300 Kimball Dr.
Parsippany, New Jersey 07054
Re: K260208
Trade/Device Name: Ultra-Fine™ SAFETY Insulin Syringe
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF, MEG
Dated: June 16, 2026
Received: June 16, 2026
Dear Ewelina Rog:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260208 - Ewelina Rog
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260208 - Ewelina Rog
Page 3
Sincerely,
**SHRUTI N. MISTRY -S**
Shruti Mistry, M.S.
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260208 | ? |
| Please provide the device trade name(s). | | ? |
| Ultra-Fine™ SAFETY Insulin Syringe | | |
| Please provide your Indications for Use below. | | ? |
| Ultra-Fine SAFETY Insulin Syringes are intended to be used for subcutaneous administration of insulin for the treatment of diabetes mellitus at a concentration of 100U/mL (U-100), while reducing the occurrence of accidental needle sticks during normal handling and disposal of the used needle/syringe combination. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
emberta™
# 510(k) Summary
Submitter
emberta Medical II LLC
300 Kimball Drive
Parsippany, NJ 07054
Contact Person
Ewelina Rog
Senior Manager, Regulatory Affairs
Phone: (734) 433-8817
Date Prepared
July 10, 2026
Trade Name
Ultra-Fine™ SAFETY Insulin Syringe
Common Name
Piston Syringe
Device Class
II
Regulation Number
21 CFR 880.5860
Classification Panel
80; General Hospital
Product Code
FMF (Syringe, Piston)
MEG (Syringe, Antistick)
# Predicate Device
Legally marketed predicate device to which substantial equivalence is being claimed:
K190054 – BD Insulin Syringe
# Device Description
The Ultra-Fine SAFETY Insulin Syringe is a single-use, plastic, sterile, non-pyrogenic device designed for subcutaneous administration of 100U/mL (U-100) insulin. The Ultra-Fine SAFETY Insulin Syringe device assembly consists of a needle shield, needle (cannula), barrel with graduated scale, plunger rod with a plunger stopper, and a safety mechanism. When activated, the safety mechanism covers the needle (cannula) and guards against accidental needle sticks during normal handling and disposal of the used device. The device operates on the principles of a piston syringe.
# Indications For Use
Ultra-Fine SAFETY Insulin Syringes are intended to be used for subcutaneous administration of insulin for the treatment of diabetes mellitus at a concentration of 100U/mL (U-100), while reducing the occurrence of accidental needle sticks during normal handling and disposal of the used needle/syringe combination.
# Summary of Technological Characteristics
The technological characteristics of the subject Ultra-Fine SAFETY Insulin Syringe are similar to the predicate BD Insulin Syringe. A comparison between Ultra-Fine SAFETY Insulin Syringe
emberta.com
{5}
(subject device) and BD Insulin Syringe (predicate device), cleared under K190054, is outlined in Table 1 below.
Table 1. Comparison of Technological Characteristics with the Predicate Device
| General Information Feature | Subject Device (Ultra-Fine SAFETY Insulin Syringe) | Predicate Device (BD Insulin Syringe) K190054 | Comparison |
| --- | --- | --- | --- |
| **General Information** | | | |
| Device Class | II | II | **Same** |
| Regulation Number | 21 CFR 880.5860 | 21 CFR 880.5860 | **Same** |
| Product Code | FMF (Syringe, Piston) MEG (Syringe, Antistick) | FMF (Syringe, Piston) | Different – FDA product code MEG was included to support the subject device's antistick claim. Although the device may be classified under two product codes, both are governed by the same FDA regulation number (21 CFR 880.5860); therefore, this difference does not impact the determination of substantial equivalence. |
| Specific Drug Use | U-100 Insulin | U-100 Insulin | **Same** |
| Indications for Use | Ultra-Fine SAFETY Insulin Syringes are intended to be used for subcutaneous administration of insulin for the treatment of diabetes mellitus at a concentration of 100U/mL (U-100), while reducing the occurrence of accidental needle sticks during normal handling and disposal of the used needle/syringe combination. | Insulin syringes are intended for subcutaneous injection of U-100 insulins. | **Same** – Both the predicate and subject devices are intended for subcutaneous injection of insulin. The subject device includes an additional statement to describe the presence of the safety mechanism, but this feature does not alter the overall intended use. |
| Single Use Only | Yes | Yes | **Same** |
| Volumetric Capacity | 0.3 mL, 0.5 mL, 1 mL | 0.3 mL, 0.5 mL, 1 mL | **Same** |
| Needle (Cannula) Gauge Size | 29G, 30G, 31G | 27G, 28G, 29G, 30G, 31G | Different – Predicate device has an additional 27G and 28G configuration. |
| Needle (Cannula) Length Size | 6mm, 8mm, 12.7mm | 6mm, 8mm, 12.7mm, 16mm | Different – Predicate device has an additional 16mm configuration. |
2 embecta.com
embecta
{6}
Table 1. Comparison of Technological Characteristics with the Predicate Device
| General Information Feature | Subject Device (Ultra-Fine SAFETY Insulin Syringe) | Predicate Device (BD Insulin Syringe) K190054 | Comparison |
| --- | --- | --- | --- |
| Nozzle Tip Type | Permanently attached needle | Permanently attached needle | **Same** |
| Mechanism of Connection – Hub | 0.3 mL: Snap-Fit 0.5 mL: Snap-Fit 1 mL: Snap-Fit | 0.3 mL: Snap-Fit 0.5 mL: Snap-Fit 1 mL: Integral | Different – Verification testing conducted on the subject device confirmed the snap-fit meets all applicable requirements and functions as intended in accordance with ISO 8537. The differences between the 1 mL configurations do not raise different questions of safety and effectiveness. |
| Graduated Scale – Unit Increments | 0.3 mL: ½-unit or 1-unit 0.5 mL: 1-unit 1 mL: 1-unit | 0.3 mL: ½-unit or 1-unit 0.5 mL: 1-unit 1 mL: 2-unit | Different – Verification testing conducted on the subject device confirmed the graduated scale unit increments meet all applicable requirements and function as intended in accordance with ISO 8537. The differences between the 1 mL configurations do not raise different questions of safety and effectiveness. |
| Safety Mechanism | Yes | No | Different – Specific to the safety mechanism of the subject device. *See comment #1* |
| Non-Pyrogenic | Yes | Yes | **Same** |
| Sterilization Method | Gamma Irradiation | Gamma Irradiation | **Same** |
| SAL 10^{-6} | Yes | Yes | **Same** |
| Packaging Configuration | Blister Package | Blister Package Self-Contained Polybag | **Same** – Predicate device has an additional self-contained polybag configuration. |
| Shelf-Life | 5 years | 5 years | **Same** |
| **Materials** | | | |
| Needle (Cannula) | 304 Stainless Steel | 304 Stainless Steel | **Same** |
3
embecta.com
embecta
{7}
Table 1. Comparison of Technological Characteristics with the Predicate Device
| General Information Feature | Subject Device (Ultra-Fine SAFETY Insulin Syringe) | Predicate Device (BD Insulin Syringe) K190054 | Comparison |
| --- | --- | --- | --- |
| **Needle Shield** | Polypropylene | Polyethylene | Different – The subject needle shield features a relief cutout to accommodate the safety mechanism. *See comment #2* |
| **Needle Shield Color** | Orange Colorant | Orange Colorant | Different – While the shield color is unchanged, the underlying color formulation differs. *See comment #2* |
| **Cannula Lubricant** | Silicone Fluid | Silicone Fluid | **Same** |
| **Adhesive** | UV Cured Adhesive | UV Cured Adhesive | **Same** |
| **Graduated Scale** | Solvent Based Markem Black Colored Ink | Solvent Based Markem Black Colored Ink | **Same** |
| **Plunger Rod** | Polystyrene | Polystyrene | **Same** |
| **Plunger Stopper** | Rubber Stopper | Rubber Stopper | **Same** |
| **Plunger Stopper Lubricant** | Medical Grade Silicone | Medical Grade Silicone | **Same** |
| **Barrel** | Polypropylene | Polypropylene | **Same** |
| **Barrel Lubricant** | Medical Grade Silicone | Medical Grade Silicone | **Same** |
| **Snap-Fit Hub** | Polypropylene | Polypropylene | **Same** |
| **Pushrod** | Polycarbonate | N/A – Specific to the safety mechanism of the subject device. | Different – For the subject device, the safety mechanism is comprised of the pushrod and adapter, which is not present in the predicate device. *See comment #2* |
| | White Colorant | | |
| **Adapter** | Polycarbonate | N/A – Specific to the safety mechanism of the subject device. | |
| **Performance** | | | |
| **Syringe Performance** | ISO 8537 | ISO 8537 | **Same** |
| **Needle (Cannula) Performance** | ISO 8537 ISO 9626 ISO 7864 | ISO 8537 ISO 9626 ISO 7864 | **Same** |
4 embecta.com
embecta
{8}
Table 1. Comparison of Technological Characteristics with the Predicate Device
| General Information Feature | Subject Device (Ultra-Fine SAFETY Insulin Syringe) | Predicate Device (BD Insulin Syringe) K190054 | Comparison |
| --- | --- | --- | --- |
| Safety Performance | ISO 23908 | N/A – Specific to the safety mechanism of the subject device. | Different – Specific to the safety mechanism of the subject device. Design verification per ISO 23908 confirmed the safety mechanism meets all applicable requirements and functions as intended, supporting a determination of substantial equivalence. |
# Discussion of differences in technological characteristics:
- Comment #1: Design verification per ISO 23908, the sharps injury prevention simulated use study, and the human factors summative validation study demonstrated that the safety mechanism performs as intended and can be used safely and effectively by the intended users. These results show that the addition of the safety mechanism does not raise different questions of safety and effectiveness compared to the predicate device.
- Comment #2: A biocompatibility evaluation was conducted in accordance with ISO 10993-1 and supports that the identified differences do not raise different questions of safety and effectiveness, supporting a determination of substantial equivalence.
# Summary of Non-Clinical Performance Data
Performance testing was conducted per FDA recognized consensus standards and met all acceptance criteria. The results confirmed that the subject device is substantially equivalent to the predicate device. The following performance data was provided in support of the substantial equivalence determination:
- ASTM D4169-22 - Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F88/F88M-23 - Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F2096-11 - Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) (Reapproved 2019)
- ISO 8537:2016 - Sterile single-use syringes, with or without needle, for insulin
- ISO 9626:2016 - Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 7864:2016 - Sterile hypodermic needles for single use - Requirements and test methods
- ISO 23908:2024 - Sharps injury protection - Sharps protection mechanisms for single-use needles, introducers for catheters and needles used for blood testing, monitoring, sampling and medical substance administration - Requirements and test methods
- USP <161> - Medical Devices - Bacterial Endotoxin and Pyrogen Tests
5
embecta.com
embecta
{9}
## Biocompatibility Testing
A biocompatibility evaluation for the Ultra-Fine SAFETY Insulin Syringe was performed based on the requirements of ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The Ultra-Fine SAFETY Insulin Syringe is categorized as externally communicating with direct tissue contact and is used for a prolonged duration (more than 24 hours but less than 30 days). The following biological endpoints were evaluated and used to demonstrate substantial equivalence:
- Cytotoxicity
- Irritation Intracutaneous Reactivity
- Sensitization
- Acute Systemic Toxicity
- Material Mediated Pyrogenicity
- Genotoxicity
- Subacute Toxicity (Systemic Toxicity)
## Sterilization
The Ultra-Fine SAFETY Insulin Syringe is sterilized by gamma irradiation. The sterilization process has been validated in accordance with ISO 11137-1:2025, Sterilization of health care products - Radiation - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices, ISO 11137-2:2013[AMD1:2022], Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, and ANSI AAMI ISO 13004:2022, Sterilization of health care products - Radiation - Substantiation of selected sterilization dose: Method VDmaxSD. The sterilization parameters were chosen to assure the sterilization dose provides a minimum Sterility Assurance Level (SAL) of 10⁻⁶.
## Sharps Injury Prevention Simulated Use
In accordance with FDA guidance document, *Medical Devices with Sharps Injury Prevention Features*, a sharps injury prevention simulated use study was conducted using the Ultra-Fine SAFETY Insulin Syringe to evaluate the sharps injury prevention (SIP) feature. This study concluded with zero failures observed and was considered successful according to the criteria outlined within the FDA guidance.
## Summary of Clinical Testing
Not applicable.
## Conclusion
The differences between the predicate and the subject device do not raise different questions of safety or effectiveness. The information submitted within this Traditional 510(k) demonstrates the Ultra-Fine SAFETY Insulin Syringe is substantially equivalent to the predicate BD Insulin Syringe.
6 embecta.com
embecta
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.