The SLIM Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices in the neurovasculature.
Device Story
SLIM Microcatheter is a single-lumen, variable-stiffness, composite catheter designed for neurovascular, peripheral, and coronary procedures. It features a hydrophilic coating for lubricity and a proximal luer fitting. Three radiopaque markers (two distal, one 3cm from tip) facilitate fluoroscopic visualization and coil delivery. Supplied sterile for single use, it includes a steam shaping mandrel and peel-away introducer. Used by clinicians in clinical settings, the device is navigated through the vasculature under fluoroscopy to deliver diagnostic agents or interventional devices (e.g., embolization coils). It is compatible with guidewires, DMSO, and liquid embolic agents. By providing a trackable pathway, it enables precise placement of therapeutic materials or diagnostic contrast, potentially improving procedural outcomes in neurovascular interventions.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), dimensional verification, mechanical performance (tensile strength, kink resistance, burst pressure, torque strength), particulate matter testing, coating integrity, and usability in clinically relevant anatomical models. Compatibility with interventional devices and embolic agents was verified.
Technological Characteristics
Single-lumen, variable-stiffness composite catheter. Materials: medical-grade plastics and metals. Coating: hydrophilic. Dimensions: 0.017" ID, 0.034" OD, 150-170 cm length. Markers: 3 radiopaque bands. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: None (mechanical device). Standards: ISO 10555-1 (mechanical), ISO 80369-7 (hub), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for patients requiring intravascular infusion of diagnostic agents (e.g., contrast media) or delivery of interventional devices within the peripheral, coronary, and neurovasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 13, 2026
Balt USA, LLC
Subha Elango
Manager, Regulatory Affairs
29 Parker
Irvine, California 92618
Re: K260199
Trade/Device Name: SLIM Microcatheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: QJP, DQY, KRA
Dated: June 12, 2026
Received: June 15, 2026
Dear Subha Elango:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260199 - Subha Elango
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260199 - Subha Elango
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260199
Device Name
SLIM Microcatheter
Indications for Use (Describe)
The SLIM Microcatheter is intended for general intravascular use, including the peripheral, coronary and neurovasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices in the neurovasculature.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
balt
## 510(k) Summary
510(k) Summary: K260199
| **Applicant:** | Balt USA, LLC 29 Parker Irvine, CA 92618 Registration No: 3014162263 |
| --- | --- |
| **Contact Person:** | Subha Elango Manager, Regulatory Affairs Email: subha.elango@baltgroup.com |
| **Date Summary Prepared:** | June 12, 2026 |
| **Trade Name:** | SLIM Microcatheter |
| **Common Name:** | Percutaneous Catheter |
| **Review Panel:** | Neurology, Cardiovascular |
| **Product Code:** | QJP, DQY, KRA |
| **Regulation Number:** | 21 CFR 870.1250, 21 CFR 870.1210 |
| **Device Classification:** | Class II |
| **Predicate Device:** | Phenom Catheters 510(k) #: K210230 |
| **Reference Device:** | NG Delivery Catheter 510(k) #: K230609 Headway Duo Microcatheter 510(k) #: K120917 |
### **Device Description:**
The SLIM Microcatheter is a single lumen, variable stiffness, composite microcatheter. The design facilitates the advancement of the microcatheter and is intended to assist in the delivery of interventional devices in the neurovasculature. The outer surface of the SLIM Microcatheter is coated with a hydrophilic coating to increase lubricity. The proximal end of the SLIM Microcatheter incorporates a luer fitting for the attachment of accessories. A total of three (3) radiopaque markers are built into the microcatheter - two (2) radiopaque markers at the distal end help to facilitate fluoroscopic visualization and one (1) marker 3cm from the distal tip to facilitate the delivery of embolization coils.
A Steam Shaping Mandrel and Peel-away Introducer Tube are included within the tray. The SLIM Microcatheter is provided sterile, non-pyrogenic, and is intended for single use only.
### **Indications for Use:**
The SLIM Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices in the neurovasculature.
Page 1 of 5
{5}
balt
# **Device Comparison:**
| Device Attribute | Predicate Device Phenom Catheters **K210230** | Reference Device Headway Duo Microcatheter **K120917** | Reference Device NG Delivery Catheter **K230609** | Subject Device SLIM Microcatheter **K260199** |
| --- | --- | --- | --- | --- |
| Indications for Use | Phenom™ Catheters are intended for the introduction of interventional devices or diagnostic agents into the neuro, peripheral, and coronary vasculatures. | The Headway Duo Microcatheter is intended for general intravascular use, including the peripheral and coronary vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as embolization materials. The Headway Duo Microcatheter is intended for neurovascular use, for the infusion of diagnostic agents, such as contrast media, and therapeutic agents that have been cleared or approved for use in the neurovasculature and are compatible with the inner diameter of the Headway Duo Microcatheter. | The NG Delivery Catheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices, such as distal access catheters, in the neurovasculature. | The SLIM Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices in the neurovasculature. |
| FDA Product Code | DQY; KRA; QJP | DQO | QJP; DQY | QJP; DQY; KRA |
| **Dimensional Attributes** | | | | |
| Inner Diameter | 0.017" – 0.0445" | 0.0165" | 0.021" | 0.017" |
| Shaft Outer Diameter | Proximal: 2.2 F – 4.7 F (0.029" – 0.061") Distal: 1.8 F – 4.2 F (0.024" – 0.055") | 2.1 F (0.028") | 0.059" – 0.069" | 0.034" |
| Effective Length | 120 cm – 160 cm | 157 cm and 168 cm | 152 cm | 150 cm, 160 cm, 170 cm |
| Coating Length | 90 cm – 100 cm | 115 cm | 60 cm | 60 cm |
| Materials | Commonly used medical grade plastics & metals with hydrophilic coating | Same as K210230 | Same as K210230 | Same as K210230 |
| Marker band | 1 or 2 tip markers | 2 tip markers | 2 marker bands | 3 marker bands |
| **Packaging/Condition Supplied** | | | | |
| Condition | Sterile and Single Use | Same as K210230 | Same as K210230 | Same as K210230 |
Page 2 of 5
{6}
balt
| Device Attribute | Predicate Device Phenom Catheters K210230 | Reference Device Headway Duo Microcatheter K120917 | Reference Device NG Delivery Catheter K230609 | Subject Device SLIM Microcatheter K260199 |
| --- | --- | --- | --- | --- |
| Supplied | | | | |
| Sterilization Method | Ethylene Oxide (EO), SAL 10^{-6} | Same as K210230 | Same as K210230 | Same as K210230 |
| Packaging Configuration | For the hoop configuration, the catheter is placed in a dispenser hoop and packaged in a pouch. For the tray configuration, the dispenser-hooped catheter and its accessories are placed in a tray, which is then sealed in the pouch. The pouch is then placed inside a carton box. | Catheter is placed in a dispenser hoop and accessories on a card that is then inserted into the pouch. The pouch is then placed inside a carton box. | Catheter is placed in a dispenser hoop and accessories on a mounting card that is then inserted into the pouch. The pouch is then placed inside a carton box. | Same as K230609 |
| Shelf-Life | 12 months | Not available | 12 months | Same as K210230 |
### Non-Clinical Performance Testing:
#### Biocompatibility:
A biocompatibility assessment was conducted for the SLIM Microcatheter in accordance with the FDA biocompatibility guidance, “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”,” issued on September 8, 2023.
The subject device is an externally communicating device directly contacting circulating blood for a limited (≤ 24 hours) duration and the device was assessed for cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, direct and indirect hemolysis, complement activation, and thrombogenicity.
#### Performance Data – Bench Testing:
The following performance bench tests were conducted to assess the performance of the SLIM Microcatheter.
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Dimensional Verification | Measurements taken to verify that the device meets dimensional specifications. | Pass |
| Visual Inspection | Visual inspection was conducted for surface defects. | Pass |
| Tensile Strength | Peak tensile force was evaluated per ISO 10555-1 after preconditioning in a simulated use model. | Pass |
| Kink Resistance | Kink resistance was evaluated after preconditioning in a simulated use model. | Pass |
Page 3 of 5
{7}
balt
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Liquid Leak | The device was exposed to a liquid pressure for 30 seconds. The device was inspected for leakage per ISO 10555-1. | Pass |
| Air Leakage | The device was tested for air leakage into the hub per ISO 10555-1. | Pass |
| Dynamic Burst | Tested to verify the device can withstand internal liquid pressure under dynamic flow conditions with the distal end open. | Pass |
| Static Burst | The distal tip of the catheter was blocked, and fluid was injected into the lumen at increasing pressure until the catheter burst per ISO 10555-1 and the static burst pressure was compared with the maximum pressure generated with manual syringe injection. | Pass |
| Turns to Failure (Torque Strength) | The device was evaluated for torque strength by measuring the number of catheter rotations until failure after preconditioning in a simulated use model. | Pass |
| Hub Verification | The device shall meet the functional requirements with established acceptance criteria per ISO 80369-7. | Pass |
| Particulate Matter | The catheter underwent simulated use testing and particulate testing was conducted including the predicate device for comparison. | Pass |
| Tip Buckling | The maximum force to cause catheter tip buckling while constrained at varying distances from the tip was measured. | The tip stiffness was comparable to the predicate and other cleared catheters. |
| Coating Integrity | The coating integrity was inspected before and after preconditioning in a simulated use model. | No evidence of surface damage or coating defects. |
| Contrast and Saline Exposure | After the device was used to deliver saline and contrast media, the device was inspected for damage, and dimensional attributes were measured. | No visual evidence of damage or dimensional changes. |
| Radiopacity | Tested to verify that radiopaque markers are visible using standard fluoroscopic equipment. | Pass |
| Flow Rate | The flow rate of contrast media and saline and associated injection pressures were evaluated. | The flow rate was characterized. |
Page 4 of 5
{8}
balt
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Design Validation/Usability | The subject and predicate devices were prepared in accordance with their respective instructions for use and tested for device usability in a clinically relevant anatomical model. | Device preparation, introduction, trackability, and retrieval were comparable to the predicate. |
| Compatibility – Interventional Devices | The subject device was evaluated for compatibility with interventional devices such as guidewires, neurovascular embolization coils, DMSO, and liquid embolic agents. Compatibility with DMSO and liquid embolic agents was established through risk assessment and testing with DMSO and representative liquid embolic agents, including the SQUID™ Liquid Embolic Agent and TRUFILL™ n-Butyl Cyanoacrylate (n-BCA) Liquid Embolic System. | The subject device is compatible for use with guidewires, neurovascular embolization coils, DMSO and liquid embolic agents. |
# Performance Data – Animal:
No animal study was deemed necessary. A determination of substantial equivalence is based on successful completion of non-clinical bench testing.
# Clinical Testing:
No clinical study was deemed necessary. A determination of substantial equivalence is based on successful completion of non-clinical bench testing.
# Conclusion:
Based on the same intended use, similar technological characteristics, and the results of the nonclinical testing, the subject device is found to be substantially equivalent to the predicate device. The differences in technological characteristics do not raise new questions of safety and effectiveness. Non-clinical testing was conducted to demonstrate that the subject device meets the specifications and performs as intended.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.