The ETHICON™ Circular Staplers have application throughout the alimentary tract for end-to-end, end-to-side, and side-to-side anastomoses.
Device Story
Sterile, single-use mechanical surgical stapler; creates anastomoses in alimentary tract. Device delivers two concentric rows of staples while simultaneously cutting tissue. Features detachable anvil for placement; available in four sizes (21mm, 25mm, 29mm, 33mm) and two shaft lengths (26cm, 37cm). Operated manually by surgeons in clinical settings. Knife component utilizes UNS SS304 stainless steel manufactured via Electrochemical Machining (ECM) for improved edge consistency. Device compresses tissue to specific staple height; ensures hemostasis and proper healing. Benefits include reliable tissue transection and anastomosis formation.
Clinical Evidence
Bench testing only. Performance data included force-to-fire, washer/test skin full cut, thin tissue full cut, design thickness tissue full cut, and staple crossing tests. Biocompatibility testing confirmed compliance with ISO 10993-1. No animal or clinical studies were required.
Technological Characteristics
Mechanical, non-powered surgical stapler. Materials: UNS SS304 stainless steel knife. Dimensions: 21mm, 25mm, 29mm, 33mm end-effectors; 26cm or 37cm shaft lengths. Sterilization: Gamma radiation. Biocompatibility: ISO 10993-1 compliant. Manufacturing: Electrochemical Machining (ECM) for knife edge.
Indications for Use
Indicated for patients requiring end-to-end, end-to-side, or side-to-side anastomoses throughout the alimentary tract. Contraindicated if combined tissue thickness cannot be comfortably compressed to the closed staple height or if the internal diameter of the structure is less than 21 mm.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 11, 2026
Ethicon Endo- Surgery., LLC
Aysenur Erucman
Senior Regulatory Affairs Specialist
475 Calle C
Guaynabo, PR 00969
Re: K260196
Trade/Device Name: ETHICON™ Circular Stapler (CDH21mm); ETHICON™ Circular Stapler (CDH25mm); ETHICON™ Circular Stapler (CDH29mm); ETHICON™ Circular Stapler (CDH33mm); ETHICON™ Circular Stapler - XL Sealed (ECS21mm); ETHICON™ Circular Stapler - XL Sealed (ECS25mm); ETHICON™ Circular Stapler - XL Sealed (ECS29mm); ETHICON™ Circular Stapler - XL Sealed (ECS33mm)
Regulation Number: 21 CFR 878.4750
Regulation Name: Implantable Staple
Regulatory Class: Class II
Product Code: GDW
Dated: January 22, 2026
Received: January 22, 2026
Dear Aysenur Erucman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260196 - Aysenur Erucman
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260196 - Aysenur Erucman
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260196 | |
| Device Name ETHICON™ Circular Stapler (CDH21mm, CDH25mm, CDH29mm, CDH33mm) ETHICON™ Circular Stapler-XL Sealed (ECS21mm, ECS25mm, ECS29mm, ECS33mm) | |
| Indications for Use (Describe) The ETHICON™ Circular Staplers have application throughout the alimentary tract for end-to-end, end-to side, and side-to-side anastomoses. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary K260196
I. SUBMITTER
Company: ETHICON Endo-Surgery, LLC
475 Calle C
Guaynabo, PR 00969
Contact: Aysenur Huseyinoglu Erucman
Senior Regulatory Affairs Specialist
ETHICON Endo-Surgery, Inc.
Phone: 201-725-4434
Email: ahuseyi1@its.jnj.com
Date Prepared: April 22, 2026
II. SUBJECT DEVICES
Trade Names:
- ETHICON™ Circular Stapler CDH 21mm
- ETHICON™ Circular Stapler CDH 25mm
- ETHICON™ Circular Stapler CDH 29mm
- ETHICON™ Circular Stapler CDH 33mm
- ETHICON™ Circular Stapler-XL Sealed ECS 21mm
- ETHICON™ Circular Stapler-XL Sealed ECS 25mm
- ETHICON™ Circular Stapler-XL Sealed ECS 29mm
- ETHICON™ Circular Stapler-XL Sealed ECS 33mm
Common or Usual Name: Surgical Stapler with Implantable Staples
Classification Name: Implantable staple (21 CFR 878.4750)
Regulatory Class: Class 2 -Staple, Implantable
Product Code: GDW
III. PREDICATE DEVICE
| Predicate Device 510(k) Number | Predicate Device Name | Predicate Device Model Numbers |
| --- | --- | --- |
| K201280 | ETHICON™ Circular Stapler – ETHICON™ XL Circular Stapler | CDH21mm, CDH25mm, CDH29mm, CDH33mm, ECS21mm, ECS25mm, ECS29mm, ECS33mm |
{5}
2
# IV. REFERENCE DEVICE
| Reference Device 510(k) Number | Reference Device Name | Reference Device Model Numbers |
| --- | --- | --- |
| K163523 | ECHELON CIRCULAR™ Powered Staplers | CDH21P,CDH25P, CDH29P, CDH33P, ECS21, ECS25P, CS29P,ECS33P |
# V. DEVICE DESCRIPTION
The ETHICON™ Circular Staplers are sterile, single-use surgical instruments designed to simultaneously staple and cut tissue to create an anastomosis. Each device delivers two concentric rows of staples on either side of the cut line and includes a detachable anvil to facilitate placement. The staplers are available in two configurations (CDH and ECS), two shaft lengths (26 cm standard and 37 cm XL), and four end-effector sizes (21 mm, 25 mm, 29 mm, and 33 mm). All configurations function identically in terms of tissue compression and staple formation, differing primarily in shaft length and pneumatic sealing features for laparoscopic use.
# VI. INDICATIONS FOR USE
ETHICON CIRCULAR Staplers have application throughout the alimentary tract for end-to-end, end-to-side, and side-to-side anastomoses.
# VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Subject and Predicate Device are sterile, single-use, mechanical stapling devices designed to simultaneously cut and staple tissue to create an anastomosis in the alimentary tract. They share identical end-effector designs, staple configurations, device sizes, ergonomic features, packaging, and sterilization methods.
{6}
| Characteristic | Subject Device- ETHICON™ Circular Stapler – ETHICON™ XL Circular Stapler | Predicate Device (K201280) | Reference Device (K163523) |
| --- | --- | --- | --- |
| Indication for Use | The ETHICON™ Circular Staplers have application throughout the alimentary tract for end-to-end, end-to side, and side-to-side anastomoses. | Same. | Same. The ECHELON CIRCULAR™ Powered Staplers have application throughout the alimentary tract for end-to-end, end-to-side, and side-to-side Anastomoses. |
| Intended Use | Creation of anastomoses. | Same. | Same. |
| Contraindications | Do not use where the combined tissue thickness cannot be comfortably compressed to the closed staple height (Refer to the ETHICON™ Circular Stapler Product Code Table below) or where the internal diameter of the structure is less than 21 mm. If the device is used on tissue that cannot comfortably compress to the closed staple height, or can easily compress to less than the closed staple height, an inadequate anastomosis could be formed resulting in leakage, inadequate hemostasis, or improper healing. | Same. | Tissue thickness should be carefully evaluated before firing any stapler. Do not use where the combined tissue thickness cannot be comfortably compressed to the selected closed staple height, or too easily compresses to less than the selected closed staple height (Refer to the ECHELON CIRCULAR™ Powered Stapler Product Codes Table below for range of adjustable closed staple height) or where the internal diameter of the structure is less than 23 mm. If the device is used on tissue that cannot be comfortably compressed to the selected closed staple height, or too easily compresses to less than the selected closed staple height, an inadequate anastomosis could be formed resulting in leakage, inadequate hemostasis, or improper healing. |
| Sterile, Single Patient Use | Yes | Same. | Same. |
| Functional Use | Simultaneously transect and staple tissue | Same. | Same. |
| Biocompatibility of Materials | All tissue-contacting materials comply with ISO 10993-1. | Same. | Same. |
{7}
| Characteristic | Subject Device- ETHICON™ Circular Stapler – ETHICON™ XL Circular Stapler | Predicate Device (K201280) | Reference Device (K163523) |
| --- | --- | --- | --- |
| Packaging Method | Packaged in a clear PETG blister with a heat sealed preprinted Tyvek Lid. | Same. | Same. |
| Device Operation | Mechanical (hand) power used to staple and cut while creating an anastomosis closure. | Same. | Battery power used to staple and cut while creating an anastomosis closure |
| Sterilization Method | Gamma Radiation | Same. | EO sterilization |
| Shelf Life | 5 years | Same. | 3 years |
The key technological differences pertain to the knife component. The Subject Device incorporates a knife made from UNS SS304 stainless steel, replacing the UNS SS305 stainless steel used in the predicate device. Additionally, the knife edge in the subject device is formed using Electrochemical Machining (ECM)—a non-contact precision process that provides improved edge consistency and dimensional control. In contrast, the predicate device knife was manufactured through conventional grinding and buffing techniques.
## VIII. PERFORMANCE DATA:
The following performance data demonstrate that the Subject Device is substantially equivalent to the Predicate Device. The modifications made to the knife component were shown not to impact the safety, effectiveness, or overall performance of the device.
### Bench Testing
- Force to Fire
- Washer & Test Skin Full Cut
- Thin Tissue Full Cut
- Design Thickness Tissue Full Cut
- Fully cut indicated tissue and Crossing Staples
**Animal Testing**: Animal testing was not required for this submission, as substantial equivalence was demonstrated through non-clinical performance testing.
**Clinical studies**: Clinical studies were not required for this submission.
{8}
Biocompatibility: Studies were performed and confirmed that the Subject Device is biocompatible for the intended patient contact. No new risks were introduced as a result of the material change. All tissue contacting materials comply with ISO 10993-1.
Electrical Safety and Electromagnetic Compatibility: The device is not powered and therefore this does not apply.
IX. CONCLUSIONS
In conclusion, Design verification testing confirmed that did not introduce new issues of safety and effectiveness. The changes do not introduce a new indication for use or change the intended use, and the Subject Device is substantially equivalent to the Predicate Device.
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.