Performance validation dataset: >6,000 ultrasound data points from >1,800 patient examinations.
>1 (musculoskeletal ultrasound experts)
Indications for Use
DIANA is intended for use in patients undergoing ultrasound examination of the hand and wrist joints. The device automatically analyzes musculoskeletal structures on pre-captured ultrasound images and provides decision support for identification of Synovial hypertrophy, Doppler activity, and Osteophytes presence/absence. The user of the decision support shall be a healthcare professional trained and qualified in musculoskeletal (MSK) ultrasound. DIANA is limited to use with the system ARTHUR which performs automatic acquisition of ultrasound images.
Device Story
DIANA (Diagnosis Aid Network for Arthritis) is a software medical device for musculoskeletal ultrasound image analysis; inputs are pre-captured ultrasound images of hand and wrist joints acquired by the ARTHUR system. Device uses locked neural networks to segment anatomical structures; applies rule-based decision logic to segmentations to classify presence/absence of synovial hypertrophy, Doppler activity, and osteophytes based on EULAR-OMERACT scoring. Outputs include binary clinical decision support, visual overlays on original images, and historical longitudinal data. Used by trained healthcare professionals in clinical settings to support diagnosis and monitoring. Clinicians review outputs alongside original images to verify analysis; device serves as decision support and does not replace clinical judgment.
Clinical Evidence
Performance validation used >6,000 ultrasound data points from >1,800 patient examinations (72.4% female, 27.6% male) across Denmark, Germany, and the US. Reference standard established by up to five MSK ultrasound experts using EULAR-OMERACT scoring. Results: Synovial hypertrophy (Sens: 0.70, Spec: 0.91); Osteophytes (Sens: 0.78, Spec: 0.84); Doppler activity (Sens: 0.82, Spec: 0.97). Usability testing confirmed user comprehension of outputs and overlays.
Technological Characteristics
Software-based medical device; utilizes locked neural network algorithms for image segmentation and rule-based logic for binary classification. Inputs: B-mode images and Doppler videos. Outputs: Structured text and visual overlays. Connectivity: Limited to use with ARTHUR system. Evaluates synovial hypertrophy, Doppler activity, and osteophytes per EULAR-OMERACT scoring system.
Indications for Use
Indicated for adult patients undergoing ultrasound examination of hand and wrist joints for clinical investigation or monitoring of musculoskeletal diseases. Intended for use by healthcare professionals trained in musculoskeletal ultrasound and the EULAR-OMERACT scoring system.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
July 24, 2026
ROPCA ApS
Trine Straarup Winther
Head of Quality
Cortex Park 26 E
Odense M, 5230
Denmark
Re: K260192
Trade/Device Name: Diana
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: June 16, 2026
Received: June 16, 2026
Dear Trine Straarup Winther:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260192 - Trine Straarup Winther
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260192 - Trine Straarup Winther
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director, Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260192 | ? |
| Please provide the device trade name(s). | | ? |
| DIANA | | |
| Please provide your Indications for Use below. | | ? |
| DIANA is intended for use in patients undergoing ultrasound examination of the hand and wrist joints. The device automatically analyzes musculoskeletal structures on pre-captured ultrasound images and provides decision support for identification of Synovial hypertrophy, Doppler activity, and Osteophytes presence/absence. The user of the decision support shall be a healthcare professional trained and qualified in musculoskeletal (MSK) ultrasound. DIANA is limited to use with the system ARTHUR which performs automatic acquisition of ultrasound images. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) #: K260192
510(k) Summary
Prepared on: 2026-07-24
# Contact Details
21 CFR 807.92(a)(1)
Applicant Name
ROPCA ApS
Applicant Address
Cortex Park 26 E Odense M 5230 Denmark
Applicant Contact Telephone
+45 22 53 22 07
Applicant Contact
Ms. Trine Straarup Winther
Applicant Contact Email
tw@ropca.com
# Device Name
21 CFR 807.92(a)(2)
Device Trade Name
DIANA
Common Name
DIANA
Classification Name
Medical Image Management and Processing System
Regulation Number
892.2050
Product Code(s)
QIH
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K222406
Clarius AI
QIH
# Device Description Summary
21 CFR 807.92(a)(4)
DIANA stands for Diagnosis Aid Network for Arthritis and is a software medical device, based on neural networks used for musculoskeletal ultrasound image analysis, that classifies and segments anatomical features, contributing to the diagnosis and monitoring of musculoskeletal diseases with a binary score(presence/absence). DIANA is a decision support tool for evaluating and monitoring the possible disease within anatomical joints based upon the ultrasound images of hand and wrist joints captured by ARTHUR. DIANA provides clinical decision support based upon the following parameters: Synovial hypertrophy, Doppler activity, and Osteophytes presence. DIANA is intended for adult patient utilization who are under clinical investigation, or monitoring of, musculoskeletal diseases and is intended to be utilized by trained healthcare professionals with experience in musculoskeletal diseases and the Global EULAR-OMERACT scoring system.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
DIANA is intended for use in patients undergoing ultrasound examination of the hand and wrist joints. The device automatically analyzes musculoskeletal structures on pre-captured ultrasound images and provides decision support for identification of Synovial hypertrophy, Doppler activity, and Osteophytes presence/absence. The user of the decision support shall be a healthcare professional trained and qualified in musculoskeletal (MSK) ultrasound. DIANA is limited to use with the system ARTHUR which performs automatic acquisition of ultrasound images.
# Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device has the same intended use as the predicate device. Both devices are intended to assist healthcare professionals trained in musculoskeletal (MSK) ultrasound, by preforming non-invasive processing of musculoskeletal ultrasound images, using segmentation of anatomical structures utilizing artificial intelligence algorithms. Both devices provide information to support clinical assessment and are not intended to replace clinical judgment.
{5}
The subject device is intended for use with ultrasound images of the hand and wrist, whereas the predicate device is intended for use with ultrasound images of the foot. This difference in anatomical site does not raise different questions of safety or effectiveness because both devices analyze musculoskeletal ultrasound images with comparable imaging characteristics using the same technological approach. Both anatomical regions comprise bones, tendons, synovium, and other soft tissues that can be reliably visualized and assessed using musculoskeletal ultrasound.
The subject device and the predicate device utilize non-adaptive (locked) artificial intelligence image segmentation algorithms to identify and segment anatomical structures in pre-captured ultrasound images. The generated segmentations are displayed to the user as overlays on the original ultrasound images to support image interpretation. The subject device applies rule-based decision logic to these segmentation results to generate binary outputs indicating the presence or absence of synovial hypertrophy, Doppler activity, and osteophytes. In contrast, the predicate device derives measurements of anatomical structure thickness from the generated segmentations and does not apply classification or decision logic.
Although the subject device applies rule-based decision logic to generate binary clinical decision support outputs, whereas the predicate device provides anatomical measurements without classification or decision logic, these technological differences do not raise different questions of safety or effectiveness. Based on the similarities in intended use, underlying segmentation technology, performance validation, and overall technological characteristics, the subject device is substantially equivalent to the predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device applies a predefined and standardized methodology for the evaluation of ultrasound images to support the assessment of possible musculoskeletal disease activity. Binary disease activity classification is performed for synovial hypertrophy, Doppler activity, and osteophytes based on the EULAR OMERACT scoring system with the cut-off values:
- Synovial hypertrophy: grade 0 and 1 would classify as absence of disease. Grade 2 and 3 is classified as presence of disease
- Synovial Doppler activity: grade 0 would classify as absence of disease. Grade 1, 2 and 3 is classified as presence of disease
- Osteophytes: grade 0 and 1 would classify as absence of disease. Grade 2 and 3 is classified as clinical presence of the disease.
The presence or absence of disease parameters is provided to the clinician as structured textual information along with corresponding visual output. In addition, the subject device provides a historical view of disease presence to support longitudinal clinical assessment. These outputs are intended to support clinical review and decision-making by qualified healthcare professionals and do not replace clinical judgment. The subject device further provide the healthcare professional with presents the underlying ultrasound images, so that the clinician can independently verify each analysis against the ultrasound image. The clinician is advised to access the ultrasound images within the report and include other diagnostic parameters.
The technological characteristics, mode of operation, and output of the subject device are comparable to those of the predicate device and do not raise any new or different questions of safety or effectiveness.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Performance testing was conducted to demonstrate that DIANA performs as specified in the intended use.
Usability testing was performed to evaluate user comprehension and interpretation of the device outputs. The results demonstrated that intended users were able to understand and interpret the clinical decision support results, including the anatomical segmentations displayed as overlays on the original ultrasound images. The usability evaluation demonstrated that the presentation of the results supports transparent interpretation and clinical review by healthcare professionals trained and qualified in musculoskeletal (MSK) ultrasound.
Validation testing was performed using independent datasets collected from multiple clinical sites, representative patient populations, and different ultrasound systems to evaluate the generalizability of the algorithms across representative clinical settings.
The performance validation dataset consisted of more than 6,000 ultrasound data points (B-mode images and Doppler videos) obtained from more than 1,800 patient examinations (72.4% female and 27.6% male). The dataset included examinations acquired at hospitals in Denmark (46%), Germany (39%), and the United States (15%) using GE LOGIQ E10, GE LOGIQ Totus, and Canon Aplio A ultrasound systems. The validation dataset included metacarpophalangeal (MCP), proximal interphalangeal (PIP), distal interphalangeal (DIP), and radiocarpal/intercarpal (RCIC) joints.
Device performance was evaluated by comparison with a reference standard established by up to five musculoskeletal ultrasound experts, each with more than 10 years of clinical experience. The reference standard was established using the EULAR-OMERACT ultrasound scoring system. For performance evaluation, the expert OMERACT scores were converted to binary classifications (presence or absence) for synovial hypertrophy, Doppler activity, and osteophytes. DIANA provides binary clinical decision support indicating the presence or absence of these findings.
The clinical performance of the device is defined by the mean sensitivity and specificity for the detection of synovial hypertrophy, osteophytes, and Doppler activity. DIANA demonstrated a sensitivity of 0.70 (95% CI: 0.64–0.77) and a specificity of 0.91 (95% CI: 0.88–
{6}
0.93) for synovial hypertrophy. For osteophyte detection, the sensitivity was 0.78 (95% CI: 0.73–0.83) and the specificity was 0.84 (95% CI: 0.79–0.91). For Doppler activity, DIANA achieved a sensitivity of 0.82 (95% CI: 0.62–0.94) and a specificity of 0.97 (95% CI: 0.94–1.00).
Performance testing demonstrated that DIANA performs as intended by providing binary clinical decision support for the assessment of synovial hypertrophy, Doppler activity, and osteophytes from pre-captured ultrasound images of the hand and wrist. The device achieved high specificity across all three assessment categories 0.84–0.97, while maintaining sensitivities between 0.70 and 0.82.
The subject device and the predicate device are software-based medical devices intended to provide analysis of pre-captured musculoskeletal ultrasound images using non-adaptive (locked) artificial intelligence algorithms. Both devices employ image segmentation to identify musculoskeletal anatomical structures. The resulting segmentations are used to generate device-specific outputs that support interpretation by healthcare professionals trained in musculoskeletal (MSK) ultrasound.
The subject device is intended for use with ultrasound images of the hand and wrist, whereas the predicate device is intended for a different musculoskeletal anatomical region. This difference in anatomical site does not raise different questions of safety or effectiveness because both devices analyze musculoskeletal ultrasound images with comparable imaging characteristics using the same technological approach. Both devices utilize locked AI algorithms and provide anatomical segmentations for user review. The predicate device derives anatomical measurements from the segmentations, whereas the subject device applies rule-based decision logic to generate binary clinical decision support outputs.
Performance validation of the subject device was conducted using independent multicenter datasets representative of the intended patient population. The validation results demonstrated that the subject device performs as intended for its intended use and support that the differences in anatomical site do not adversely affect the safety or effectiveness of the device.
Although the subject device differs from the predicate by applying rule-based decision logic to segmentation results to generate binary clinical decision support outputs, whereas the predicate provides anatomical measurements derived from segmentation, these technological differences do not raise new questions of safety or effectiveness.
Based on the similarities in intended use, underlying segmentation technology, mode of operation, and performance testing, the subject device is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.