Avenir® Müller Stem: The Avenir Müller Uncemented Hip Stems are intended for cementless use in total or hemi hip arthroplasty and are indicated for the following conditions: Noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g. rheumatoid arthritis. Acute traumatic fracture of the femoral head or neck. Avenir Complete™ Hip System: The Avenir Complete Hip System is intended for cementless use in total or hemi hip arthroplasty and is indicated in the following conditions: Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemi-arthroplasty or total hip replacement (THR) Acute traumatic fracture of the femoral head or neck Avascular necrosis of the femoral head
Device Story
Avenir Müller Stem and Avenir Complete Hip System are cementless femoral hip prostheses; used in total or hemi hip arthroplasty to replace damaged/diseased joints. Devices feature wedge-shaped titanium alloy (Ti-6Al-4V) stems with proximal-to-distal taper and proximal ribs for stability. Implantation performed by surgeons in clinical settings using system-specific instrumentation. Submission specifically addresses addition of a second production line for cleaning, porous CP-Titanium coating, hydroxyapatite (HA) coating, and packaging. Devices are provided sterile for single use. Clinical benefit derived from restoration of hip joint function and stability.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including distal fatigue testing per ISO 7206-4 and characterization of the dual CP-Ti/HA coating.
Technological Characteristics
Materials: Ti-6Al-4V (Protasul-64WF) alloy. Coatings: Commercially Pure-Titanium (CP-Ti) and hydroxyapatite (HA) applied via air plasma spray. Design: Wedge-shaped, proximal-to-distal taper, proximal ribs. Sterilization: Sterile, single-use. Connectivity: None.
Indications for Use
Indicated for patients requiring total or hemi hip arthroplasty due to noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis), inflammatory joint disease (e.g., rheumatoid arthritis), acute traumatic femoral head/neck fracture, advanced joint wear, or failed previous hip surgery (e.g., osteotomy, arthrodesis, prior arthroplasty).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 27, 2026
Zimmer, Inc.
Ana Montoya
Regulatory Affairs Specialist
1800 W Center St.
Warsaw, Indiana 46580
Re: K260182
Trade/Device Name: Avenir® Müller Stem; Avenir Complete™ Hip System
Regulation Number: 21 CFR 888.3353
Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis
Regulatory Class: Class II
Product Code: LZO, KWL, KWY, KWZ, LWJ, MEH
Dated: January 21, 2026
Received: January 21, 2026
Dear Ana Montoya:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260182 - Ana Montoya
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260182 - Ana Montoya
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LIMIN SUN -S
Limin Sun, Ph.D.
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-0740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260182 | |
| Device Name Avenir® Müller Stem | |
| Indications for Use (Describe) The Avenir Müller Uncemented Hip Stems are intended for cementless use in total or hemi hip arthroplasty and are indicated for the following conditions: • Noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g. rheumatoid arthritis. • Acute traumatic fracture of the femoral head or neck. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
FORM FDA 3881 (8/23)
Page 2 of 2
PSC Publishing Services (301) 443-0740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260182 | |
| Device Name Avenir Complete™ Hip System | |
| Indications for Use (Describe) The Avenir Complete Hip System is intended for cementless use in in total or hemi hip arthroplasty and is indicated in the following conditions: • Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases • Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemi-arthroplasty or total hip replacement (THR) • Acute traumatic fracture of the femoral head or neck • Avascular necrosis of the femoral head | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
ZIMMER BIOMET
K260182 Page 1 of 3
510(k)Summary
Prepared on: 2026-04-24
# Contact Details
Applicant Name: Zimmer, Inc
Applicant Address: 1800 W Center Street Warsaw IN 46580 United States
Applicant Contact Telephone: +57 3193346263
Applicant Contact: Ms. Ana Montoya
Applicant Contact Email: ana.montoya@zimmerbiomet.com
Device Name: Avenir® Müller Stem
Avenir Complete™ Hip System
Common Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
Classification Name: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Regulation Number: 888.3353
Product Code(s): LZO, KWL, KWY, KWZ, LWJ, MEH
# Legally Marketed Predicate Devices
The predicate devices for subject device of this submission are listed in Table 1
Table 1. Predicate Devices for Avenir Müller® Stem and Avenir Complete™ Hip System.
| | Device Name | Manufacturer | 510(k) Number |
| --- | --- | --- | --- |
| Primary Predicate | Avenir® Müller Stem | Zimmer Switzerland Manufacturing GmbH | K193030 |
| Additional Predicate | Avenir Complete™ Hip System | Zimmer, Inc. | K182048 |
| Additional Predicate | Avenir Complete™ Hip System - Size 0 Coxa Vara Stems | Zimmer, Inc. | K192189 |
| Reference Device(s) | Avenir® Müller Stem (MR Labeling) | Zimmer Switzerland Manufacturing GmbH | K200112 |
# Device Description Summary
# Reason for Submission
This submission requests the addition of a second production line (Line B) at supplier Medicoat S.A.S (formerly Zimmer Etapes, France) for cleaning, porous coating (Commercially Pure-Titanium), hydroxyapatite (HA) coating, and packaging of the subject Avenir Müller® Stem and Avenir Complete™ Hip System.
{6}
ZIMMER BIOMET
K260182 Page 2 of 3
# 510(k)Summary
The Avenir Stems are intended for use in total or hemi hip arthroplasty. These stems are designed for cementless implantation into the proximal femur to replace the damaged or diseased natural joint. The stems are manufactured from a titanium alloy Ti-6Al-4V (Protasul-64WF). They have a wedge-shaped design with a proximal-to-distal taper. The stem and neck are a single unit, and the stem features proximal ribs on the anterior and posterior surfaces which are designed to increase stability. The Avenir Stems are provided sterile and are for single use only. System specific instrumentation is available to prepare the femur for implantation of the Avenir femoral stems.
# Intended Use/Indications for Use
Avenir® Müller Stem:
The Avenir Müller Uncemented Hip Stems are intended for cementless use in total or hemi hip arthroplasty and are indicated for the following conditions:
- Noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g. rheumatoid arthritis.
- Acute traumatic fracture of the femoral head or neck.
Avenir Complete™ Hip System:
The Avenir Complete Hip System is intended for cementless use in total or hemi hip arthroplasty and is indicated in the following conditions:
- Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases
- Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemiarthroplasty or total hip replacement (THR)
- Acute traumatic fracture of the femoral head or neck
- Avascular necrosis of the femoral head
# Indications for Use Comparison
The subject devices have the same intended use and indications for use as the predicate devices.
# Technological Comparison
The subject device has similar technological characteristics as the identified predicate devices. The rationale for substantial equivalence is based on consideration of the following characteristics:
- Materials: Identical to predicate devices.
- Design Features: Identical to predicate devices.
- Sterilization: Identical to predicate devices.
- Air Plasma Spray Coating: Similar to primary predicate device.
Any minor differences are addressed via predicate device and / or performance testing, and do not raise new questions of safety or effectiveness.
# Non-Clinical and/or Clinical Tests Summary & Conclusions
- Distal Fatigue testing of Avenir® Müller Stem and Avenir Complete™ Hip System per ISO 7206-
{7}
ZIMMER BIOMET
K260182 Page 3 of 3
# 510(k)Summary
4; and
- Coating Characterization of the dual CP-Ti/HA) coating
- No clinical testing was required to support substantial equivalence.
# Conclusion
The subject Avenir Müller Stem and Avenir Complete Hip System have the same intended use and indications for use as the predicate devices. The differences in technological characteristics between the subject and the predicate devices do not raise new questions of safety and effectiveness. The performance testing and information provided demonstrate that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.