The SmartGuide provides fixed guiding of the precise insertion of linear instruments, such as needles through mechanical means with the use of diagnostic ultrasound equipment. The Needle Guide and Grids are attached over the transducer/ probe/ scanning instruments. The device provides a fixed path for the imaging guidelines for visualizing guided instrument placement procedures. UC-CARE SmartGuide Needle Guides and Grids are furnished sterile; single use patient/ procedure, and disposable. The Needle Guides and Grids are intended for use by clinicians, in clinical and hospital settings, to guide linear instruments during Transperineal ultrasound procedure that require the precision use of a needle.
Device Story
Single-use, sterile, polycarbonate needle guide; attaches to ultrasound probes via reusable Navigo Workstation sensor fixator; provides mechanical fixed-path guidance for linear instruments (needles) during transperineal biopsy procedures. Used in clinical/hospital settings by physicians. Device maintains fixed spatial relationship between transducer and needle to ensure alignment within ultrasound imaging plane. Output is physical needle path; healthcare provider uses this to visualize and perform precise instrument placement under ultrasound guidance. Benefits include accurate, stable needle positioning during biopsy, reducing risk of procedural error.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including channel alignment, needle path/spacing verification, equidistance/parallelism, cover breach/probe damage, retention force (8N), and needle drag testing. Biocompatibility evaluated per ISO 10993-1 (cytotoxicity, irritation, sensitization, acute systemic toxicity, pyrogenicity).
Technological Characteristics
Material: Medical-grade polycarbonate. Principle: Mechanical guidance via fixed-path channels. Energy: None. Form factor: Disposable guide with entry cone, mounted via reusable Navigo fixator. Sterilization: Ethylene Oxide (EtO). Connectivity: None (mechanical accessory).
Indications for Use
Indicated for adult patients in need of a biopsy via needle during endocavity ultrasound-guided procedures. Assists physicians by providing fixed guiding for precise insertion of linear instruments, retaining the needle tip and barrel within the ultrasound beam.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}

UC-CARE, Ltd.
Hadas Sheynise
SQA and V&V Manager
Apollo Bldg., POB. 310,
Yokneam, 2069200
Israel
July 30, 2026
Re: K260161
Trade/Device Name: SmartGuide Disposable Guides and Grids
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: ITX
Dated: June 30, 2026
Received: June 30, 2026
Dear Hadas Sheynise:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260161 - Hadas Sheynise
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260161 - Hadas Sheynise
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D. O'hara -S
Date: 2026.07.30 13:47:03 -04'00'
For
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260161
Device Name
SmartGuide Disposable Needle Guides and Grids
Indications for Use (Describe)
SmartGuide Disposable Needle Guides and Grids are used to assist and aid physicians in performing an endocavity diagnosis ultrasound needle guided procedure using guided intervention by providing fixed guiding for the precise insertion of linear instruments, such as needles. The Needle Guides and Grids are designed to aid adult patient population, in need of a biopsy, via the use of a needle, during an ultrasound procedure by retaining the needle tip and barrel within the ultrasound beam.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
UC-CARE
510(k) Summary-K260161
Page 1 of 9
28 July 2026
### 510(k) Summary (K260161)
### UC-Care Ltd.'s SmartGuide
### 1. Submitter Information
UC-Care Ltd.
Apollo Building,
P.O. Box 310,
Yokneam 2069200, Israel
Phone: +972-4-909-7427
Facsimile: +972-4-993-7323
E-mail: support@uc-care.com
Contact Person: Hadas Sheynise
Phone: +972-54-4818648
Email: shadas@uc-care.com
Date of Preparation: 17 June, 2026
### Identification of Subject Device
Name of Device: SmartGuide Disposable Needle Guides and Grids
Common or Usual Name: Disposable Needle Guides and Grids
Proprietary/Trade name: SmartGuide Disposable Needle Guides and Grids
Classification Name: Diagnostic ultrasonic transducer accessories
Regulation Number: 21 CFR 892.1570
Product Code: ITX
Regulatory Class: II
### 2. Identification of Predicate
| Device Name | 510(k) Number | Type |
| --- | --- | --- |
| Disposable Needle Guides and Grids | K223689 | Primary |
| Disposable Guides KDNG00 | K180970 | Reference |
| GTK Disposable needle Guides | K170741 | Reference |
{5}
UC-CARE
510(k) Summary-K260161
Page 2 of 9
### 3. Device Description
The SmartGuide Disposable Needle Guides and Grids (hereinafter referred to as "SmartGuide") is a single-use, disposable polycarbonate needle guide, provided sterile (EtO) and individually packaged in Tyvek®/PET-PE sterile peel pouches.
The SmartGuide is used in conjunction with reusable Navigo Workstation sensor fixators, which attach to compatible ultrasound probes and are designed to secure the SmartGuide and the Navigo system probe sensor during use.
The SmartGuide is designed to mechanically guide a needle or other linear insertion tool in Transperineal biopsy procedures in multiple available tracks located within the ultrasound US image plane (centralized to the ultrasound probe).
The SmartGuide Disposable Needle Guides and Grids are classified as a Class II medical device under 21 CFR 892.1570 (Diagnostic ultrasonic transducer accessories), Product Code ITX.
### 4. Intended Use and Indication for Use statement
#### Indication For Use
SmartGuide Disposable Needle Guides and Grids are used to assist and aid physicians in performing an endocavity diagnosis ultrasound needle guided procedure using guided intervention by providing fixed guiding for the precise insertion of linear instruments, such as needles. The Needle Guides and Grids are designed to aid adult patient population, in need of a biopsy, via the use of a needle, during an ultrasound procedure by retaining the needle tip and barrel within the ultrasound beam.
#### Intended Use
The SmartGuide provides fixed guiding of the precise insertion of linear instruments, such as needles through mechanical means with the use of diagnostic ultrasound equipment. The Needle Guide and Grids are attached over the transducer/ probe/ scanning instruments. The device provides a fixed path for the imaging guidelines for visualizing guided instrument placement procedures. UC-CARE SmartGuide Needle Guides and Grids are furnished sterile; single use patient/ procedure, and disposable. The Needle Guides and Grids are intended for use by clinicians, in clinical and hospital settings, to guide linear instruments during Transperineal ultrasound procedure that require the precision use of a needle.
{6}
UC-CARE
510(k) Summary-K260161
Page 3 of 9
## 5. Summary of Technological Characteristics
Subject, Predicate Device and Reference Devices' indications for use place the SmartGuide Needle Guides and Grids and AMD (510k# K223689) /KOELIS (510k# K180970)/ GeoTek (510k# K170741) Ultrasound Needle Guides in device body contact categories as follows:
a. Surface devices, intact skin / mucosal membranes/ breached surfaces, limited contact duration (<24 hours)
b. External communicating devices, blood path indirect, tissue communicating limited contact duration (<24 hours)
Subject and Predicate Device have a similar Intended Use and Subject, Predicate device, and Reference devices (K180970, K170741) provide mechanical means for performing needle/ instrument guided procedures with the use of diagnostic ultrasound transducers. These devices provide the same fixed path for the needle or instrument that when coupled by the ultrasound system software corresponds to the on-screen imaging guidelines for visualizing guided instrument placement procedures. Predicate device, Reference Devices, and Subject device are furnished sterile; and the entire guide is single use patient/ procedure, disposable.
The principles of operation of the SmartGuide are similar to those of the predicate and reference devices and include:
- Placement of a probe attachment bracket onto the ultrasound transducer using locating and fixation features
- Securing the disposable needle guide to the transducer using a mechanical locking mechanism
- Maintaining a fixed spatial relationship between the transducer and the needle guide to ensure accurate alignment of the needle within the ultrasound imaging plane
- Releasing the locking mechanism to remove the needle guide from the transducer after use
The primary predicate device and the SmartGuide both utilize a mechanical attachment bracket to secure the disposable needle guide to the ultrasound transducer and to maintain stable and repeatable needle guidance during use. In the primary predicate device, the mounting bracket and cannula are integrated into a single disposable component that attaches externally to the transducer. In the SmartGuide, the disposable needle guide is mounted onto a probe attachment bracket (Navigo workstation fixator), which is a reusable component that is part of the previously cleared Navigo System (K250664).
The use of the Navigo probe attachment bracket does not alter the intended use, principle of operation, or functional performance of the SmartGuide. The bracket does not introduce additional patient contact beyond that associated with the disposable needle guide and serves solely as a mechanical fixation interface between the transducer and the guide. Accordingly, this design configuration demonstrates substantial equivalence to the predicate device design.
{7}
UC-CARE
510(k) Summary-K260161
Page 4 of 9
Substantial Equivalence (Comparison Table)
| | Subject Device | Primary Predicate Device K223689 | Reference Device K180970 | Reference Device K170741 |
| --- | --- | --- | --- | --- |
| Company | UC-Care Ltd. | Advance Medical Designs | KOELIS | Geotek |
| Device Name | SmartGuide Disposable Needle Guides and Grids | Disposable Needle Guides and Grids | Disposable Guides KDNG00 | GTK Disposable needle Guides |
| Regulation Number | 21 CFR 892.1570 | 21 CFR 892.1570 | 21 CFR 892.1570 | 21 CFR 892.1570 |
| Device Class | Class II | Class II | Class II | Class II |
| Product Code | ITX | ITX | ITX | ITX |
| Indications for Use | SmartGuide Disposable Needle Guides are used to assist and aid physicians in performing an endocavity diagnosis ultrasound needle guided procedure using guided intervention by providing fixed guiding for the precise insertion of linear instruments, such as needles. The Needle Guides and Grids are designed to aid adult patient population, in need of a biopsy, via the use of a needle, during an ultrasound procedure by retaining the needle tip and barrel within the ultrasound beam. | Disposable Needle Guides and Grids are used to assist and aid physicians in performing an endocavity diagnosis ultrasound needle guided procedure using guided intervention by providing fixed guiding for the precise insertion of linear instruments, such as needles. The Needle Guides and Grids are designed to aid adult patient population, in need of a biopsy of an internal organ, or internal delivery or removal of fluid within the body cavity, via the use of a needle, during an ultrasound procedure by retaining the needle tip and barrel within the ultrasound beam. | The disposable guide KDNG00 is dedicated for endocavity diagnosis ultrasound needle guided procedure. it is intended for clinicians and assistant clinicians, in clinic and hospital, to guide linear instrument with a transrectal approach. | GTK disposable Needle Guides when used in conjunction with ultrasound system transducer where configuration is suitable (e.g Aloka, Alpinion, BK ,Esaote, GE, Hitachi, Medison, Mindray, Philips, Toshiba, Siemens and Shimadzu ultrasound systems) and attached to the ultrasound system's transducers, is to facilitate proper needle placement to access anatomical structures. |
| Intended Use | The Subject Device provides fixed guiding of | The Subject Device provides fixed guiding of the | The disposable guide KDNG00 provides a fixed | NA |
{8}
UC-CARE
510(k) Summary-K260161
Page 5 of 9
| | the precise insertion of linear instruments, such as needles through mechanical means with the use of diagnostic ultrasound equipment. The Needle Guide and Grids are attached over the transducer/ probe/ scanning instruments. The device provides a fixed path for the imaging guidelines for visualizing guided instrument placement procedures. UC-CARE SmartGuide Needle Guides and Grids are furnished sterile; single use patient/ procedure, and disposable. The Needle Guides and Grids are intended for use by clinicians in clinical and hospital settings, to guide linear instruments during Transperineal ultrasound procedure that require the precision use of a needle. | precise insertion of linear instruments, such as needles through mechanical means with the use of diagnostic ultrasound equipment. The Needle Guide and Grid is attached over the transducer/ probe/ scanning instruments. The device provides a fixed path for the imaging guidelines for visualizing guided instrument placement procedures. Advance Medical Designs (AMD) Needle Guides and Grids are furnished sterile; single use patient/ procedure, and disposable. The single use, disposable feature helps prevent transfer of microorganisms, body fluid, and particulate material to the patient and healthcare worker during reuse of the transducer. The Needle Guides and Grids are intended to clinicians and assistant clinicians, in clinical and hospital settings, to guide linear instruments during any ultrasound | path for the needle that when coupled by the ultrasound system software corresponds to on-screen imaging guidelines for visualizing guided instrument placement procedures. | |
| --- | --- | --- | --- | --- |
{9}
UC-CARE
510(k) Summary-K260161
Page 6 of 9
| | | procedure that require the precision use of a needle. | | |
| --- | --- | --- | --- | --- |
| Design | SmartGuide Designs Needle Guide and Grid is a plastic guide designed to be clipped on to an ultrasound probe, with an entry cone to easily introduce the needle into the channel. Fixation mechanism of the Guide on the probe: The guide is fixed on the probe with a reusable clip (Navigo Fixator) to allow the needle guide stability on the transducer and a lateral screw to fix the guide axial position within the clip. | Advance Medical Designs Needle Guide and Grid is a plastic guide designed to be clipped on to an ultrasound probe, with an entry cone to easily introduce the needle into the channel. Fixation mechanism A clip to allow the needle guide stability on the transducer and 2 pins for attachment in the notches for the probe. | Plastic guide designed to be clipped on an endocavity ultrasound probe. An entry cone to easily introduce the needle into the tube. Fixation mechanism A clip to allow the needle guide stability on the transducer and 2 pins for attachment in the notches for the probe. | The GTK Disposable Needle Guides are polymeric needle guides made from molded ABS/PC plastic and some models have a stainless cannula. The GTK disposable guides hold both the US probe and the needle guides in place relative to each other while the US guided biopsy is being performed. The GTK disposable needle guides is and assembly of 1 or 2 components, either just a carriage or a carriage with a rail (cannula). |
| Materials of Construction | Medical Grade Polycarbonate The material have met the requirements of ISO 10993-1 for biocompatibility. | Medical Grade Thermoplastic ABS Polypropylene 304 Stainless Steel All materials have met the requirements of ISO 10993-1 for biocompatibility. | Medical Grade Polycarbonate All materials have met the requirements of ISO 10993-1 for biocompatibility. | ABS or Polycarbonate |
| Safety/ Biocompatibility | Meets ISO 10993-1 biocompatibility requirements for | Meets ISO 10993-1 biocompatibility requirements | Meets ISO 10993-1 biocompatibility requirements | NA |
{10}
UC-CARE
510(k) Summary-K260161
Page 7 of 9
| | limited contact duration: • Surface device – intact skin contact • External communicating indirect tissue contact (via biopsy needle only) **Demonstrated to be:** • Non-cytotoxic (ISO 10993-5) • Non-irritating (ISO 10993-23) • Non-sensitizing (ISO 10993-10) • Non-pyrogenic (ISO 10993-11) and • Non-acute systemic toxic (ISO 10993-11) | for limited contact duration: • surface devices of breached or compromised surface • External communicating indirect blood path/tissue contact Demonstrated to be non-toxic, non-sensitizing, non-irritating, non-hemolytic, and nonpyrogenic | for limited contact duration: • surface devices of breached or compromised surface • Externally communicating tissue bone/dentin Demonstrated to be non-toxic, non-sensitizing, non-irritating, (DOES NOT INCLUDE PYROGEN/HEMO/ OR ACUTE SYSTEMIC) | |
| --- | --- | --- | --- | --- |
| **Effectiveness** | Subject, Predicate, and Reference Devices are designed for secure and aligned fit to the transducer or probe, while not altering the transducer or probe design integrity or function. When used with their intended fixation mechanisms, the design features ensure accurate and consistent needle guidance relative to the transducer. The exterior shapes of the guides are contoured for the patient comfort with no sharp edges. Based on comparison to the primary predicate and the reference devices, UC-CARE SmartGuide needle guides and grids devices have the same intended use and comparable technological characteristics. Therefore, the UC-CARE SmartGuide needle guides and grids are substantially equivalent to the legally marketed primary predicate device, the disposable endocavity needle guide marketed by Advance Medical Designs (K223689). | | | |
| **Sterilization** | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| **Shelf Life** | 6 Months | 3 years | 3 years | 5 years |
| **Accessories** | None provided. | Ultrasound gel packet and covers. | None provided. | Ultrasound gel packet and covers. |
| **Energy Type** | None | None | None | None |
| **Software** | None | None | None | None |
### Non-Clinical Performance Testing
The biological safety of the SmartGuide was evaluated based on its intended use, material composition, and the nature and duration of patient contact, and in comparison, to the predicate and reference devices. The evaluation was conducted using a risk-based approach
{11}
UC-CARE
510(k) Summary-K260161
Page 8 of 9
in accordance with ISO 10993-1:2018 / EN ISO 10993-1:2020 and FDA guidance Use of International Standard ISO 10993-1: Biological Evaluation of Medical Devices – Part 1 (September 2023).
Consistent with the predicate and reference devices in terms of limited contact duration, the SmartGuide is intended for use of less than 24 hours. Based on its intended use, the SmartGuide contacts directly the patient's skin and has potential indirect contact with internal non-circulatory tissues through minimal and brief contact mediated by a biopsy or interventional needle. The overall surface area of contact is small and the duration of exposure is short.
Based on this contact characterization and the use of well-characterized medical-grade polycarbonate materials that are also used in the predicate and reference devices, a risk-based biological evaluation determined that cytotoxicity (ISO 10993-5) and irritation (ISO 10993-23) represent the most relevant and sensitive biological endpoints for the SmartGuide, additionally sensitization (ISO 10993-10), acute systemic toxicity and pyrogenicity (ISO 10993-11) tests were preformed. These endpoints were evaluated on the finished device (See ATT.6.3 Cytotoxicity report_25T_55198_03, ATT.6.4 Irritation report_25T_55198_04, ATT.6.5 Sensitization Test Report_26T_32356_02_03, ATT.6.6 ISO Acute Systemic Toxicity Study Report_26T_32356_05, ATT.6.7 Pyrogen Study_Report_26T_32356_06).
### Bench Testing:
According to the FDA guidance titled "Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions, December 20, 2019" bench performance tests were performed to evaluate the performance, safety, and effectiveness of a medical device under controlled, non-clinical conditions.
The SmartGuide submitted in this premarket notification [510(k)] is a modification of its predicate device Advance Medical Designs (K223689) Disposable Needle Guides and Grids device.
The non-clinical bench performance testing for the SmartGuide utilize similar test methods, standards, and, where applicable, acceptance criteria to those used for the cleared predicate.
1) Channel alignment testing - Needle guides were tested on test fixtures to ensure needle path falls within the design tolerances specified for the design.
2) Needle path and spacing verification testing- The Needle guide was mounted on the US probe (BK 8848). Needles were inserted through adjacent channels; spacing between the needles was measured.
3) Equidistance and parallelism test - By using side view image. Equidistance was demonstrated by assessing the distance between the exit openings (center-to-center). This distance between exit openings was measured (in pixels) by counting the number of pixels between the gray markings of the exit openings (center-to-center).
{12}
UC-CARE
510(k) Summary-K260161
Page 9 of 9
4) Cover breach and probe damage testing - Water leak testing was performed to demonstrate material attachment of needle guide over a cover did not cause damage to cover or probe.
5) Retention and movement testing - Force testing was performed on needle guide attachment to ensure a force of 8N would not cause the guide to dislodge.
6) Needle drag testing Force testing- was performed by passing a cannula through the needle guide to ensure binding would not occur.
7) Clinical simulation and Design requirements testing - A Bench test designed to test the clinical setting during a Navigation procedure via the Navigo system and US system was performed. The design requirements were tested.
# Clinical Tests
Clinical tests were not required to demonstrate substantial equivalence.
# Conclusions
The UC-Care SmartGuide Disposable Needle Guides and Grids are substantially equivalent to the legally marketed primary predicate device, the Advance Medical Designs Disposable Needle Guides and Grids (K223689). The SmartGuide has the same intended use and indications for use as its primary predicate device, as well as similar technological characteristics and principles of operation.
The technological characteristics of the SmartGuide are similar to those of the primary predicate device, while reference devices (K180970 and K170741) were used to support the evaluation of materials of construction and biological safety. The SmartGuide is sterilized using ethylene oxide (EtO), consistent with the predicate and reference devices, with no differences in the sterilization method or sterility assurance level.
The identified technological and design differences between the SmartGuide and its primary predicate device do not alter the intended use, principle of operation, or functional performance, and do not raise new risks. This was supported by the non-clinical bench performance testing for the SmartGuide utilized similar test methods to those used for the predicate device, and by a risk-based biological evaluation supporting the biological safety of the device for its intended use.
Thus, the UC-Care SmartGuide Disposable Needle Guides and Grids are substantially equivalent to the legally marketed primary predicate device, the Advance Medical Designs Disposable Needle Guides and Grids (K223689).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.